The intended use of the Spacelabs Healthcare CardioExpress (CardioExpress) is to acquire ECG signals from adult and pediatric patients using ECG electrodes. CardioExpress is only intended to be used in hospitals or healthcare facilities by doctors and trained healthcare professionals. The cardiogram recorded by CardioExpress can help users to analyze and diagnose heart disease. However, the interpreted ECG with measurements and interpretive statements are only intended to be used on adult patients, and is offered to clinicians on an advisory basis only.
Device Story
CardioExpress (models SL3, SL6, SL12) is an electrocardiograph for acquiring, analyzing, displaying, and recording ECG signals. Device consists of a signal acquisition module and a central processing unit with a rechargeable battery. Inputs are ECG signals from patient electrodes; device amplifies, filters, and analyzes signals using peak-to-peak detection. Outputs include ECG waveforms and analysis results presented on an LCD display, recorded via thermal or USB printer, or stored in memory. Data can be transmitted to a PC via UART or Ethernet. Used in hospitals/healthcare facilities by doctors and trained professionals. Auto-analysis software provides measurements and interpretive statements for adult patients on an advisory basis to assist clinical diagnosis.
Clinical Evidence
No clinical data. Performance established via bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), usability (IEC 60601-1-6), alarm systems (IEC 60601-1-8), and specific electrocardiograph performance standards (IEC 60601-2-25, IEC 60601-2-51).
Technological Characteristics
Electrocardiograph with 3, 6, or 12-channel configurations. Features 12-lead simultaneous acquisition, thermal dot-matrix recorder, LCD display, and rechargeable battery. Connectivity via UART or Ethernet. Complies with IEC 60601-1, 60601-1-2, 60601-1-6, 60601-1-8, 60601-2-25, and 60601-2-51. Software-based auto-analysis for measurements and interpretation.
Indications for Use
Indicated for adult and pediatric patients requiring ECG signal acquisition for heart disease analysis and diagnosis. Auto-interpretation and measurement features are indicated for adult patients only. Intended for use in hospitals or healthcare facilities by trained professionals.
Regulatory Classification
Identification
An electrocardiograph is a device used to process the electrical signal transmitted through two or more electrocardiograph electrodes and to produce a visual display of the electrical signal produced by the heart.
Predicate Devices
Edan Instruments, Inc. / SE-601 Series Electrocardiograph (K090367)
Edan Instruments, Inc. / SE-3, SE-300A, SE-300B, SE-6, SE-600, SE-12, SE-12 Express, SE-1200 AND SE-1200 Express (K091513)
Submission Summary (Full Text)
{0}------------------------------------------------
| Submission Date: | 25 January 2013 |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Spacelabs Healthcare<br>1 Harforde Court<br>John Tate Road<br>Hertford<br>SG13 7NW<br>United Kingdom |
| Submitter Contact: | Mr. Al Van Houdt<br>Spacelabs Healthcare<br>Phone: +1 (425) 657-7200, ext 5970<br>Fax: +1 (425) 657-7210<br>Email: al.vanhoudt@spacelabs.com |
| Application<br>Correspondent: | Thomas Kroenke<br>Principal Consultant<br>Speed To Market, Inc.<br>PO Box 3018<br>Nederland, CO 80466 USA<br>tkroenke@speedtomarket.net<br>303 956 4232 |
| Manufacturing Site: | Edan Instruments, Inc.<br>2/F And 5/F, Block A/B, Unit 8, Xing Hua Building,<br>Nanhai Rd, Nanshan<br>Shenzhen CHINA 518067 |
| Trade Name: | Spacelabs Healthcare CardioExpress, Models 98400-SL3-IEC,<br>98400-SL3-AHA, 98400-SL6-IEC, 98400-SL6-AHA, 98400-SL12-<br>IEC, and 98400-SL12-AHA |
| Common Name: | Electrocardiograph |
| Classification Name: | Electrocardiograph |
| Classification<br>Regulation: | 21 CFR §870.2340 |
| Product Code: | DPS |
-
{1}------------------------------------------------
·
:
.
| Substantially<br>Equivalent Devices: | New Spacelabs Models | Predicate<br>510(k) Number | Predicate<br>Manufacturer / Model |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| | Spacelabs Healthcare<br>CardioExpress, Models<br>98400-SL3-IEC,<br>98400-SL3-AHA, | K090367 | Edan Instruments, Inc. /<br>SE-601 Series<br>Electrocardiograph |
| | 98400-SL6-IEC,<br>98400-SL6-AHA,<br>98400-SL12-IEC, and<br>98400-SL12-AHA | K091513 | Edan Instruments, Inc. /<br>SE-3, SE-300A, SE-300B,<br>SE-6, SE-600, SE-12, SE-<br>12 Express, SE-1200<br>AND SE-1200 Express |
| Device Description: | The Spacelabs Healthcare (Spacelabs) CardioExpress, Models 98400-<br>SL3-IEC, 98400-SL3-AHA, 98400-SL6-IEC, 98400-SL6-AHA,<br>98400-SL12-IEC, and 98400-SL12-AHA, are a series of<br>electrocardiographs (ECGs) designed to acquire, analyze, display, and<br>record ECG signals from ECG electrodes connected to a patient.<br><br>After been amplified, filtered and analyzed, the ECG signal<br>waveforms and analysis results are presented on a liquid crystal diode<br>(LCD) display, and recorded on the paper through either a thermal or<br>USB printer. ECG data, result and patient information may be stored<br>in the memory file. The file can be transmitted to a personal computer<br>(PC) through either an UART or Ethernet connection.<br><br>The CardioExpress can optionally contain auto analysis software<br>which assists in performing auto measurement and auto interpretation<br>of data. The CardioExpress consists of two (2) basic components: (1)<br>the signal acquisition module, and (2) the central processing unit. All<br>models contain a rechargeable battery.<br><br>The CardioExpress series of ECG monitors are provided in three (3)<br>primary models: CardioExpress SL-3; CardioExpress SL-6; and<br>CardioExpress SL-12. | | |
| Intended Use: | The intended use of the Spacelabs Healthcare CardioExpress<br>(CardioExpress) is to acquire ECG signals from adult and pediatric<br>patients using ECG electrodes. CardioExpress is only intended to be<br>used in hospitals or healthcare facilities by doctors and trained<br>healthcare professionals. The cardiogram recorded by CardioExpress<br>can help users to analyze and diagnose heart disease. However, the<br>interpreted ECG with measurements and interpretive statements are<br>only intended to be used on adult patients, and is offered to clinicians on an advisory basis only. | | |
{2}------------------------------------------------
| Technology<br>Comparison: | Characteristic | Predicate Devices | Proposed Devices |
|---------------------------|-------------------------------------|------------------------------------|------------------|
| CardioExpress SL-3 | Channels | 3 | Same |
| | Lead Color Coded Option | AHA and IEC | Same |
| | Leads | 12 standard | Same |
| | Acquisition Mode | 12 simultaneous leads | Same |
| | Heart Rate Recognition<br>Technique | Peak-to-peak detection | Same |
| | Heart Rate Range | 30 - 300 beats per minute<br>(BPM) | Same |
| | Recorder | Thermal dot-matrix<br>recorder | Same |
| CardioExpress SL-6 | Channels | 6 | Same |
| | Lead Color Coded Option | AHA and IEC | Same |
| | Leads | 12 standard | Same |
| | Acquisition Mode | 12 simultaneous leads | Same |
| | Heart Rate Recognition<br>Technique | Peak-to-peak detection | Same |
| | Heart Rate Range | 30 - 300 beats per minute<br>(BPM) | Same |
| | Recorder | Thermal dot-matrix<br>recorder | Same |
| CardioExpress SL-12 | Channels | 12 | Same |
| | Lead Color Coded Option | AHA and IEC | Same |
| | Leads | 12 standard | Same |
| | Acquisition Mode | 12 simultaneous leads | Same |
| | Heart Rate Recognition<br>Technique | Peak-to-peak detection | Same |
| | Heart Rate Range | 30 - 300 beats per minute<br>(BPM) | Same |
| | Recorder | Thermal dot-matrix<br>recorder | Same |
p
{3}------------------------------------------------
#### Summary of Performance Testing:
| Software Testing | Software was designed and developed according to a robust software<br>development process, and was rigorously verified and validated.<br>Software information is provided in accordance with internal<br>requirements and the following standards and guidance documents:<br>FDA guidance: The content of premarket submissions for software contained in medical devices, 11.May 05; FDA guidance: Off-the-shelf software use in medical devices. 09 Sep 99; FDA guidance: General principles of software validation; Final guidance for industry and FDA staff, 11 Jan 02; and FDA guidance: Cybersecurity for networked medical devices containing off-the-shelf (OTS) software, 14 January 2005. Test results indicate that the CardioExpress complies with its predetermined specifications and the applicable standards and guidance documents. |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical Safety<br>Testing | CardioExpress was tested for performance in accordance with the following Standard:<br>IEC 60601-1: 1988, Am1: 1991, and Am2: 1995. Medical electrical equipment. Part 1: Particular requirements for safety. Test results indicated that CardioExpress complies with the Standard. |
| Electromagnetic<br>Compatibility<br>Testing | CardioExpress was tested for performance in accordance with the following Standard:<br>IEC 60601-1-2: 2001, Medical electrical equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests. |
Test results indicated that CardioExpress complies with the Standards.
.
{4}------------------------------------------------
Performance Testing CardioExpress was tested for performance in accordance with internal requirements and the following standards:
- IEC 60601-1-6: 2004, Medical electrical equipment -- Part 1-6: . General requirements for safety - Colluteral standard: Usabilitv:
- IEC 60601-1-8: 2003, Am1: 2006, Medical electrical equipment -. Part 1-8: General requirements for safety - Collateral standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical svstems;
- IEC 60601-2-25: 1993. Am: 1999. Medical electrical equipment -. Part 2-25: Particular requirements for the safety of electrocardiographs; and
- IEC 60601-2-51: 2003, Medical electrical equipment Part 2-51: . Particular requirements for safety, including essential performance, of recording and analysing single channel and multichannel electrocardiographs.
Test results indicate that CardioExpress complies with its predetermined specifications and the applicable standards.
Conclusion
Verification and validation activities were conducted to establish the performance and safety characteristics of CardioExpress. The results of these activities demonstrate that CardioExpress is safe and effective when used in accordance with its intended use and labeling.
Therefore, CardioExpress is considered substantially equivalent to the predicate device.
{5}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 23, 2013
Spacelabs Healthcare Ltd. c/o Mr. Thomas Kroenke PO Box 3018 Nederland, CO 80466
Re: K130207
Trade/Device Name: CardioExpress Electrocardiographs (SL3. SL6. and SL12) Regulation Number: 21 CFR 870.2340 Regulation Name: Electrocardiograph Regulatory Class: Class II Product Code: DPS Dated: June 24, 2013 Received: June 24, 2013
Dear Mr. Thomas Kroenke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{6}------------------------------------------------
Page 2 - Mr. Thomas Kroenke
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Owen P. Faris -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Indications for Use
| 510(k) Number (if known): | K130207 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Spacelabs Healthcare CardioExpress, Models 98400-SL3-IEC,<br>98400-SL3-AHA, 98400-SL6-IEC, 98400-SL6-AHA,<br>98400-SL12-IEC, and 98400-SL12-AHA |
| Indications for Use: | The intended use of the Spacelabs Healthcare CardioExpress<br>(CardioExpress) is to acquire ECG signals from adult and<br>pediatric patients using ECG electrodes. CardioExpress is only<br>intended to be used in hospitals or healthcare facilities by doctors<br>and trained healthcare professionals. The cardiogram recorded<br>by CardioExpress can help users to analyze and diagnose heart<br>disease. However, the interpreted ECG with measurements and<br>interpretive statements are only intended to be used on adult<br>patients, and is offered to clinicians on an advisory basis only. |
X Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Digitally signed by
Owen P. Faris -S
Date: 2013.07.23
09:17:50-04'00'
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.