PASSEO-35

K130161 · Biotronik, Inc. · LIT · Apr 23, 2013 · Cardiovascular

Device Facts

Record IDK130161
Device NamePASSEO-35
ApplicantBiotronik, Inc.
Product CodeLIT · Cardiovascular
Decision DateApr 23, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Passeo-35 peripheral dilatation catheter is indicated to dilate stenosis in the renal, iliac, femoral, popliteal and infrapopliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

Device Story

Passeo-35 PTA Catheter is a dual-lumen balloon catheter used for percutaneous transluminal angioplasty. Device features a tapered soft tip, radiopaque markers for fluoroscopic positioning, and a silicone-coated shaft to enhance trackability. Operates via inflation device connected to a proximal Luer port; compatible with 0.035" guidewires. Used by physicians in clinical settings to dilate stenotic peripheral vessels and arteriovenous dialysis fistulae. Mechanical expansion of the balloon restores vessel patency, improving blood flow. No software or electronic components.

Clinical Evidence

No clinical data submitted.

Technological Characteristics

Dual-lumen balloon catheter; silicone-coated shaft; radiopaque markers; compatible with 0.035" guidewires. Mechanical dilatation principle. No software, electronics, or energy source.

Indications for Use

Indicated for patients requiring dilation of stenotic segments in renal, iliac, femoral, popliteal, and infrapopliteal arteries, and treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K130161 p. 1 of 2 APR 2 3 2013 # Passeo-35 PTA Catheter Special 510(k) Premarket Notification 510(k) Summary BIOTRONIK, Inc. Name and Address of Sponsor: 6024 Jean Road Lake Oswego, OR 97035 510(k) Contact Person and Phone Number: Jon Brumbaugh Vice President Regulatory Affairs and Compliance BIOTRONIK, Inc. 6024 Jean Road Lake Oswego, OR 97035 Phone: (888) 345-0374 Fax: (503) 635-9936 jon.brumbaugh@biotronik.com Date Prepared: Device Name: Proprietary Name: Common Name: Catheter Classification: Classification Name: Product Code: January 21, 2013 Passeo-35 Percutaneous Transluminal Angioplasty (PTA) Class II (21 CFR 870.1250) Catheter, angioplasty, peripheral, transluminal LIT #### Predicate Device: BIOTRONIK indicates Passeo-35 was previously cleared through 510(k) notification and it is its own predicate device for this Special 510(k) changes being effected labeling changes: - BIOTRONIK's Passeo-35 PTA Catheter (K082933, cleared November 3, 2008). . #### General Description: The Passeo-35 peripheral transluminal angioplasty (PTA) balloon catheter is indicated for dilatation of stenotic segments in peripheral vessels. One radiopague marker is located at either end of the balloon to facilitate fluoroscopic visualization and positioning of the balloon catheter towards and across the lesion. The dilatation balloon is designed to inflate to a known diameter at a specific inflation pressure consistent with the compliance chart on the label. The balloon catheter includes a tapered soft tip to facilitate advancement of the catheter. The balloon catheter shaft has two Luer ports at the proximal end. One port (inflation port) serves for connecting an inflation device to inflate/deflate the balloon. The other port (guidewire port) enables insertion of the guide wire. The balloon catheter is a dual lumen design with both lumens contained within one tube. The smaller lumen is the balloon inflation lumen. The larger lumen permits the use of guide wires with a maximum diameter of 0.035" to facilitate advancement of the Passeo-35 catheter towards and through the lesion(s) to be dilated. The balloon catheter is compatible with introducer sheath (introducer) sizes according to the recommendations on the label. The balloon catheter has a silicone coating to improve the trackability and pushability characteristics. Page 1 of 2 {1}------------------------------------------------ #### Indication for Use: The indications for use statement may be found in Appendix 01. This device is prescription only device as stated on the IFU, "Caution: Federal (USA) law restricts this device to sale by or on the order of a physician". The Passeo-35 peripheral dilatation catheter was cleared with the indication for use of, "The Passeo-35 peripheral dilatation catheter is indicated to dilate stenosis in the renal, iliac, femoral, popliteal and infrapopliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae." #### Technological Characteristics The subject device as the predicate device, Passeo-35 PTA Catheter (K082933, cleared November 3, 2008). The only differences pertain to the changes in the IFU and labeling, the materials, manufacture and technology have remained unchanged. #### Nonclinical Test Data The determination of substantial equivalency on this subject device does not rely upon the nonclinical test data. There is no nonclinical test data submitted in this application. #### Clinical Test Data The determination of substantial equivalency on this subject device does not rely upon the clinical data. There is no clinical data submitted in this application. #### Conclusion Based on the minor nature of the labeling changes, the subject Passeo-35 catheter is substantially equivalent to the predicate Passeo-35 catheter. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans. April 23, 2013 BIOTRONIK. Inc. c/o Mr. Jon Brumbaugh Vice President, Regulatory Affairs & Compliance 6024 Jean Road Lake Oswego, OR 97224 Re: K130161 Trade/Device Name: Passeo-35 PTA Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II (two) Product Code: LIT Dated: January 21, 2013 Received: January 23, 2013 Dear Mr. Brumbaugh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 {3}------------------------------------------------ Page 2 - Mr. Jon Brumbaugh device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.goy/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ## Matthew G. Hillebrenner for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K130161 ### Indications for Use 510(k) Number (if known): K130161 Device Name: Passeo-35 PTA Catheter Indications For Use: The Passeo-35 peripheral dilatation catheter is indicated to dilate stenosis in the renal, iliac, femoral, popliteal and infrapopliteal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) # Matthew G割訓llebrenner Page 1 of ___1________________________________________________________________________________________________________________________________________________________________ Page 33 of 86
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...