K130153 · Philips Medical Systems · MKJ · Mar 12, 2013 · Cardiovascular
Device Facts
Record ID
K130153
Device Name
HEARSTART MRX WITH AIRWAY CONFIMRATION ASSIST
Applicant
Philips Medical Systems
Product Code
MKJ · Cardiovascular
Decision Date
Mar 12, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic, Pediatric
Indications for Use
The HeartStart MRx is intended for use in the hospital and pre-hospital settings by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced life support, advanced cardiac life support or defibrillation. When operating as a semi-automatic external defibrillator in AED Mode, the HeartStart MRx is suitable for use by medical personnel trained in basic life support that includes the use of an AED. When operating in Monitor, Manual Defib or Pacer Mode, the HeartStart MRx is suitable for use by healthcare professionals trained in advanced cardiac life support.
Device Story
HeartStart MRx with Airway Confirmation Assist (ACA) is a modified defibrillator/monitor; adds airway verification functionality to existing platform. Input: CO2 waveform data from capnography monitor. Operation: ACA continuously analyzes CO2 waveforms against predetermined criteria to identify valid breaths; assesses airway establishment based on breath characteristics; outputs Pass/Fail decision to display. Used in hospital/pre-hospital settings by trained medical personnel. Provides adjunct information to clinical examination for verifying endotracheal/supraglottic airway placement and assessing ventilatory status. Benefits patient by aiding in airway management during resuscitation.
Clinical Evidence
Bench testing only. No clinical data presented. Substantial equivalence supported by performance and usability testing of the ACA feature and demonstration of technological equivalence to predicate devices.
Technological Characteristics
Defibrillator/monitor with integrated capnography and ACA software. Utilizes SMART Biphasic waveform. Connectivity includes ECG, SpO2, NBP, EtCO2, invasive pressure, and temperature monitoring. Software-based airway analysis algorithm. Manufactured using identical materials and processes as the predicate HeartStart MRx.
Indications for Use
Indicated for termination of ventricular tachycardia and ventricular fibrillation in patients. AED therapy for unresponsive, pulseless, non-breathing patients >= 8 years. Manual defibrillation for pulseless ventricular fibrillation/tachycardia or synchronous termination of atrial/ventricular arrhythmias. Pacing for symptomatic bradycardia or asystole. Monitoring functions include SpO2, NBP, EtCO2, 12-lead ECG, invasive pressures, temperature, and O-CPR feedback. Must be used by or on order of a physician.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
Zoll E Series Defibrillator/Monitor with Intubation Assist Option (K080903)
Submission Summary (Full Text)
{0}------------------------------------------------
| | 510(k) Summary | MAR 1 2 2013 |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Submitter: | Philips Medical Systems<br>3000 Minuteman Road<br>Andover MA, 01810<br>USA | |
| Contact Person: | Mark Puopolo<br>Sr. Manager, Regulatory Affairs<br>Phone: 978.659.4244<br>Fax: 978.659.3690<br>mark.puopolo@philips.com | |
| Date Prepared: | January 15, 2013 | |
| Trade Name: | HeartStart MRx with Airway Confirmation Assist<br>Defibrillator/Monitor | |
| Common Name: | Automatic External Defibrillator | |
| Classification<br>Name: | Automatic External Defibrillator | |
| Predicate Device: | Philips HeartStart MRx Defibrillator/Monitor (K061707)<br>Oridion Capnostream 20 (K082268)<br>Zoll E Series Defibrillator/Monitor with Intubation Assist<br>Option (K080903) | |
| Device<br>Description: | The Philips HeartStart MRx with Airway Confirmation Assist<br>(ACA) Defibrillator/Monitor is a modification of the FDA<br>cleared HeartStart MRx Defibrillator/Monitor. This function<br>of the MRx used to verify airway placement and aid in<br>assessing ventilatory status when an adult or pediatric patient<br>is being ventilated using an endotracheal or supraglottic<br>airway and the airway is being monitored using capnography. | |
| | Based on predetermined criteria, Airway Confirmation Assist<br>continuously analyzes the CO2 waveform coming from the<br>capnography monitor to decide if the waveform represents a<br>valid breath. If the Airway Confirmation Assist is enabled in<br>the MRx, and based on characteristics of the valid breaths,<br>ACA will decide if the airway is established. It will then<br>report a Pass/Fail decision to the MRx. The<br>defibrillator/monitor will display the result to the user in the<br>configured format (graphic and text). As an adjunct to clinical | |
:
: 上一篇:
{1}------------------------------------------------
## K130153 - page 2 of 3
examination, ACA provides additional information along with capnography to help in verifying airway placement and aiding in assessment of ventilation status.
### Statement of Intended Use:
The HeartStart MRx is intended for use in the hospital and pre-hospital settings by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced life support, advanced cardiac life support or defibrillation.
When operating as a semi-automatic external defibrillator in AED Mode, the HeartStart MRx is suitable for use by medical personnel trained in basic life support that includes the use of an AED.
When operating in Monitor, Manual Defib or Pacer Mode, the HeartStart MRx is suitable for use by healthcare professionals trained in advanced cardiac life support.
The SMART Biphasic waveform utilized in the HeartStart MRx has previously undergone clinical testing in adults. These trials support the waveform's effectiveness for defibrillation of ventricular tachyarrhythmias at 150J.
### Summary of Technological Characteristics:
In addition to being technologically equivalent to the predicate devices, the HeartStart MRx with Airway Confirmation Assist Defibrillator/Monitor has been subjected to performance and usability testing and it has been determined that the HeartStart MRx with Advanced Airway Confirmation Defibrillator/Monitor is suitable for its intended use.
81
{2}------------------------------------------------
# K130153 - page 3 of 3
### Summary of Nonclinical Data:
HeartStart MRx with Airway Confirmation Assist Defibrillator/Monitor is manufactured under the same conditions, using the similar processes and identical materials, as the Philips HeartStart MRx Defibrillator/Monitor, the legally marketed Philips Medical Systems predicate device. In addition to being technologically equivalent, the indications for use have not changed.
.
.
.
.. . .
・
11
82
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, composed of three curved lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 12, 2013
Mr. Mark Puopolo Sr. Manager, Regulatory Affairs Philips Medical Systems 3000 Minuteman Road Andover, MA 01810
Re: K130153
> HeartStartMRx Defibrillator/Monitor Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulation Class: Class III Product Code: MKJ (LDD, DRO, DPS, DXN, CCK, DQA, MWI, LIX) Dated: January 23, 2013 Received: January 24, 2013
Dear Mr. Mark Puopolo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Mr. Mark Puopolo
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Owen Faris -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The letters are black and appear to be slightly distressed or textured. The word is presented in all capital letters and is centered in the image.
#### Indications for Use
510(k) Number (if known):
K130153
Device Name: The HearStart MRx with Advanced Airway Confirmation Defibrillator/Monitor
Indications for Use: The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation.
The device is for use by qualified medical personnel trained in the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.
AED Therapy: To be used in the presence of a suspected cardiac arrest on patients of at Jeast 8 years of age that are unresponsive, not breathing and pulseless.
Manual Defibrillation: Asynchronous defibrillation is the initial treatment for ventricular fibrillation and ventricular tachycardia in patients that are pulseless and unresponsive. Synchronous defibrillation is indicated for the termination of certain atrial and ventricular arrhythmias.
Noninvasive External Pacing Therapy: The pacing option is intended for treating patients with symptomatic bradycardia. It can also be helpful in patients with asystole, if performed early.
Pulse Oximetry: The SpO2 option is intended for use when it is beneficial to assess the patient's oxygen saturation level.
Noninvasive Blood Pressure Monitoring: The NBP option is intended for noninvasive measurement of a patient's arterial blood pressure.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C) .
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 2
n P. Faris -S
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The letters are black against a white background. The word is the brand name of a multinational conglomerate corporation.
510(k) Number (if known):
K130153
Device Name: The HearStart MRx with Advanced Airway Confirmation Defibrillator/Monitor
Indications for Use: [continued ... ]
End-Tidal CO2: The EtCO2 option is intended for noninvasive monitoring of a patient's exhaled carbon dioxide and to provide a respiration rate.
12-Lead ECG: The 12-Lead ECG function is to provide a conventional diagnostic 12-Lead-ECG report, which may include measurement and interpretative statements.
O-CPR: The O-CPR option provides feedback designed to encourage rescuers to perform resuscitation in accordance with AHA/ERC guidelines for chest compression rate depth, and duty cycle and ventilation rate, volume and flow rate (inflation time).
Invasive Pressures: The Invasive Pressures option is indicated for measuring arterial, venous, intracranial, and other physiological pressures on patients.
Temperature: The Temperature option is indicated for measure temperature in patients.
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 2 of_ 2
Owen P. Faris -S
2013.03.12
13:22:19 -04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.