K130152 · Andrew Technologies, LLC · MUU · Apr 18, 2013 · General, Plastic Surgery
Device Facts
Record ID
K130152
Device Name
HYDRASOLVE LIPOPLASTY SYSTEM
Applicant
Andrew Technologies, LLC
Product Code
MUU · General, Plastic Surgery
Decision Date
Apr 18, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The HydraSolve™ Lipoplasty System is intended to be used for liquefaction and aspiration of localized subcutaneous fatty deposits for the purpose of aesthetic body contouring. Adipose tissue harvested with the HydraSolve™ Lipoplasty System may be used for autologous transfer following the method provided in the manual.
Device Story
HydraSolve™ Lipoplasty System performs body contouring and adipose tissue harvesting. System inputs include pressurized, heated, and pulsed saline solution delivered via a specialized cannula. The Phaser™ energy transfer system liquefies subcutaneous fatty deposits while simultaneous aspiration removes the material. Operated by trained medical professionals in a clinical setting. Output is aspirated adipose tissue, which can be processed via centrifuge for autologous transfer. Device includes a console (controller, power supply, Phaser™ unit), tumescent infusion component, and a sterile treatment kit (pumping mechanism, heat exchanger, handpiece with integral cannula). Benefits include selective tissue extraction and preparation of fat for grafting. Healthcare providers use the system to facilitate fat transfer procedures.
Clinical Evidence
No clinical data. Non-clinical bench testing only. Testing evaluated aspirated adipocyte cell viability, fat graft survivability, and fat graft quality to support the safety and effectiveness of the device for autologous fat transfer.
Technological Characteristics
System utilizes Phaser™ energy transfer technology (heat and pressure) to deliver pulsed, heated saline through a cannula for tissue liquefaction. Components include a console, wetting infusion unit, aspiration unit, and sterile treatment kit with handpiece. Compatible with COTS cannulae and waste canisters. Class II device (21 CFR 878.5040).
Indications for Use
Indicated for aesthetic body contouring via liquefaction and aspiration of localized subcutaneous fatty deposits in patients requiring adipose tissue harvesting for autologous transfer.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{0}------------------------------------------------
#### 510(k) Summary A.4.
# APR 1 8 2013
| Submitted by: | Herbert Crane<br>Vice President Regulatory Affairs and Quality Assurance |
|---------------------------------------------|----------------------------------------------------------------------------------------------|
| Address: | Andrew Technologies<br>3 Haddon Avenue<br>Haddonfield, NJ 08033 |
| Telephone: | (959) 502-1907 |
| Facsimile: | (856) 433-8092 |
| Date of Submission: | 17 January 2013 |
| Classification Name:<br>Device Product Code | Suction Lipoplasty System (21 CFR 878.5040)<br>MUU |
| Trade or Proprietary<br>or Model Name: | HydraSolve™ Lipoplasty System |
| Legally Marketed Devices: | HydraSolve™ Lipoplasty System (K121218)<br>Harvest AdiPrep Adipose Transfer System (K121005) |
### Device Description:
The Andrew Technologies HydraSolve "" Lipoplasty System is a liposuction system used to perform body contouring. It is designed to perform selective tissue extraction through a cannula that utilizes pressurized, heated, and pulsed saline solution to suction. Adipose tissue aspirated with the HydraSolve™ Lipoplasty System can be harvested for subsequent autologous transfer.
The device includes the HydraSolve™ Console which contains within its durable case: a user interface, controller, power supply, and Phaser™ energy transfer systems (both heat and pressure). The system also includes a tumescent infusion component which is compatible with commercial off-the-shelf (COTS) turnescent cannulae and tubing. The device also includes the Sterile Treatment Kit consisting of a pumping mechanism, heat exchanger and tubing and a reusable limited-use handpiece with integral cannula. The device interfaces with COTS waste canisters and suction tubing.
#### Indications for Use:
The HydraSolve™ Lipoplasty System is intended to be used for liquefaction and aspiration of localized subcutaneous fatty deposits for the purpose of aesthetic body contouring. Adipose tissue harvested with the HydraSolve™ Lipoplasty System may be used for autologous transfer following the method provided in the manual.
### Comparison to Predicate Device
A table comparing the technological characteristics of the subject device and predicate devices is included as a third page of this Summary of Safety and Effectiveness. With the exception of the Indications for Use statement, the subject device is identical to the predicate device cleared under K121218.
Andrew Technologies Traditional 510(k) Notification HydraSolve™ Lipoplasty System (AFT Indication) January 2013
{1}------------------------------------------------
K130152 Page 2 of 3
## Summary of testing to demonstrate safety and effectiveness
Nonclinical test data was used to support the decision of safety and effectiveness. Clinical testing was not necessary. Non-clinical testing consisted of testing aspirated adipocyte cell viability fat graft survivability, and fat graft quality.
### Conclusion
The information provided in this submission demonstrates that the device is substantially equivalent to the predicate devices.
Andrew Technologies Traditional 510(K) Notification
HydraSolve™ Lipoplasty System (AFT Indication) January 2013
{2}------------------------------------------------
Substantial Equivalence Comparison to Predicate Devices
| ATTRIBUTE | SUBJECT DEVICE | PREDICATE | PREDICATE |
|--------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | HydraSolveTM Lipoplasty System<br>(AFT Indication) | Harvest AdiPrep<br>(K121005) | HydraSolveTM Lipoplasty System<br>(K121218) |
| User Population | Trained Medical Professionals | Trained Medical Professionals | Trained Medical Professionals |
| Major Components | - System Console<br>- Wetting Infusion Unit<br>- Aspiration Unit<br>- Phaser Stream Heating and Pumping<br>Unit | - Syringes with removable plunger<br>- Centrifuge tubes with filter<br>- Aspiration & fat injection cannula<br>- Fat injection syringes<br>- Skin puncture needles<br>- Oil extraction syringe & needle | - System Console<br>- Wetting Infusion Unit<br>- Aspiration Unit<br>- Phaser Stream Heating and Pumping<br>Unit |
| Technological<br>Characteristics | - Phaser Cannula<br>- Aspirate Collection Unit<br>- Pressurized, heated, and pulsed saline<br>to liquefy fatty deposits while aspiration<br>is engaged to remove liquefied material | N/A | - Phaser Cannula<br>- Aspirate Collection Unit<br>- Pressurized, heated, and pulsed saline to<br>liquefy fatty deposits while aspiration is<br>engaged to remove liquefied material |
| Intended Use | Aesthetic Body Contouring | Aesthetic Body Contouring | Aesthetic Body Contouring |
| Tissue Processing &<br>Implantation<br>Technique | Concentration of aspirated adipose<br>tissue using a centrifuge followed by<br>injection using a syringe | Concentration of aspirated adipose tissue<br>using a centrifuge followed by injection using<br>a syringe | N/A |
| Indications for Use | The HydraSolveTM Lipoplasty System is<br>intended to be used for liquefaction and<br>aspiration of localized subcutaneous<br>fatty deposits for the purpose of<br>aesthetic body contouring. Adipose<br>tissue harvested with the HydraSolveTM<br>Lipoplasty System may be used for<br>autologous transfer following the<br>method provided in the manual. | The AdiPrep Adipose Transfer System is<br>used in medical procedures involving the<br>harvesting and transferring of autologous<br>adipose tissue. The AdiPrep system is used<br>for concentrating adipose tissue harvested<br>with a legally marketed lipoplasty system.<br>The AdiPrep Adipose Transfer System is<br>intended for use in the following surgical<br>specialties when the concentration of<br>harvested adipose tissue is desired.<br><br>Neurosurgery, Gastrointestinal Surgery,<br>Urological Surgery, Plastic and<br>Reconstructive Surgery, General Surgery,<br>Orthopedic Surgery, Gynecological Surgery,<br>Thoracic Surgery, Laparoscopic Surgery,<br>Arthroscopic Surgery | The HydraSolveTM Lipoplasty System is<br>intended to be used for liquefaction and<br>aspiration of localized subcutaneous fatty<br>deposits for the purpose of aesthetic body<br>contouring.<br><br>The HydraSolveTM Lipoplasty System is<br>indicated for use in aesthetic body<br>contouring. |
Andrew Technologies Traditional 510(k) Notification
HydraSolve™ Lipoplasty System (AFT Indication)
January 2013
K130152 page 3 of
3
. .
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three horizontal lines extending from its body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Andrew Technologies % Mr. Herbert Crane Vice President, Regulatory Affairs and Quality Assurance 3 Haddon Avenue Haddonfield, New Jersey 08033
April 18, 2013
Re: K130152
Trade/Device Name: HydraSolve™ Lipoplasty System Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: MUU Dated: March 29, 2013 Received: April 01, 2013
Dear Mr. Crane:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{4}------------------------------------------------
Page 2 - Mr. Herbert Crane
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, FOR
Peter D Rümm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K130152
Device Name: HydraSolve™ Lipoplasty System
Indications For Use:
The HydraSolve™ Lipoplasty System is intended to be used for liquefaction and aspiration of localized subcutaneous fatty deposits for the purpose of aesthetic body contouring. Adipose tissue harvested with the HydraSolve™ Lipoplasty System may be used for autologous transfer following the method provided in the manual.
× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# David Krause
(Division Sign-Off) Division of Surgical Devices 510(k) Number: K130152
Page 1 of 1
A.3.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.