S TEST REAGENT CARTRIDGE ALP (ALKALINE PHOSPHATASE)
K130141 · Hitachi Chemical Diagnostics, Inc. · CJE · Mar 22, 2013 · Clinical Chemistry
Device Facts
Record ID
K130141
Device Name
S TEST REAGENT CARTRIDGE ALP (ALKALINE PHOSPHATASE)
Applicant
Hitachi Chemical Diagnostics, Inc.
Product Code
CJE · Clinical Chemistry
Decision Date
Mar 22, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1050
Device Class
Class 2
Indications for Use
The S TEST Reagent Cartridge Alkaline Phosphatase (ALP) is intended for the quantitative measurement of alkaline phosphatase activity in serum, lithium heparinized plasma, or sodium citrate plasma using the HITACHI Clinical Analyzer. The S TEST Reagent Cartridge Alkaline Phosphatase (ALP) is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. Measurements of alkaline phosphatase are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.
Device Story
The Hitachi Clinical Analyzer is an automated, bench-top, wet chemistry system for clinical or physician office laboratories. It utilizes single-use plastic reagent cartridges containing two reservoirs (R1, R2) and a reaction cell/photometric cuvette. The system processes serum or plasma samples by pipetting sample and reagent into the cuvette, mixing, and incubating. A multi-wavelength photometer measures absorbance to calculate analyte concentration. The reagent cartridge includes a dot code label with chemistry parameters and calibration factors. Results are available in approximately 15 minutes. Healthcare providers use the quantitative ALP results to assist in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.
Clinical Evidence
Clinical performance was evaluated at three physician office laboratory (POL) sites. Precision studies (n=30 replicates per level) showed total precision %CVs ranging from 3.0% to 9.2%. Method comparison studies (n=72-77 per site) against a reference method yielded correlation coefficients (r) of 0.99, with slopes ranging from 0.942 to 0.980, demonstrating accuracy across the dynamic range of 11 to 745 U/L.
Indicated for quantitative measurement of alkaline phosphatase activity in serum, lithium heparinized plasma, or sodium citrate plasma for diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases. For prescription and point-of-care use in clinical or physician office laboratories.
Regulatory Classification
Identification
An alkaline phosphatase or isoenzymes test system is a device intended to measure alkaline phosphatase or its isoenzymes (a group of enzymes with similar biological activity) in serum or plasma. Measurements of alkaline phosphatase or its isoenzymes are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.
Predicate Devices
Roche Cobas c systems ALP2 (k100853)
Submission Summary (Full Text)
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k130141
B. Purpose for Submission:
New Device
C. Measurand:
Alkaline phosphatase (ALP)
D. Type of Test:
Quantitative, enzymatic activity
E. Applicant:
Hitachi Chemical Diagnostics
F. Proprietary and Established Names:
S TEST Reagent Cartridge Alkaline Phosphatase (ALP)
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1050 Alkaline Phosphatase (ALP) or isoenzymes test system
2. Classification:
Class II
3. Product code:
CJE
4. Panel:
Clinical Chemistry (75)
{1}
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
The S TEST Reagent Cartridge Alkaline Phosphatase (ALP) is intended for the quantitative measurement of alkaline phosphatase activity in serum, lithium heparinized plasma, or sodium citrate plasma using the HITACHI Clinical Analyzer. The S TEST Reagent Cartridge Alkaline Phosphatase (ALP) is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
Measurements of alkaline phosphatase are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.
3. Special conditions for use statement(s):
For prescription and point-of-care use.
4. Special instrument requirements:
Hitachi Clinical Analyzer E40
I. Device Description:
The S TEST reagent cartridges for the Hitachi Clinical Analyzer E40 are made of plastic and include two small reservoirs capable of holding two separate reagents R1 (2-Ethylaminoethanol Buffer and Magnesium Chloride) and R2 (p-Nitrophenyl phosphate disodium salt and 2-Amino-2-hydroxymethyl-1,3-propanediol Buffer), separated by a reaction cell/photometric cuvette. The cartridges also include a dot code label that contains all chemistry parameters, calibration factors, and other production-related information, e.g., expiration dating.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Roche Cobas c systems ALP2
2. Predicate 510(k) number(s):
k100853
{2}
3. Comparison with predicate:
| | Hitachi S Test Reagent Cartridge ALP (Candidate Device) | Roche Cobas c systems ALP2 (Predicate Device, k100853) |
| --- | --- | --- |
| Similarities | | |
| Intended Use/Indications for Use | For the quantitative measurement of alkaline phosphatase activity in serum, lithium heparinized plasma, or sodium citrate plasma Measurements of alkaline phosphatase are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases. | Same |
| Specimen Type | Human serum or (sodium citrate or lithium heparin) plasma | Same |
| Differences | | |
| Test Principle | In the presence of magnesium ions, ALP reacts with p-NPP to release p-nitrophenol. | In the presence of magnesium and zinc ions, p-NPP is cleaved by phosphatases into phosphate and p-nitrophenol |
| Testing Environment | Physician office or clinical lab | Clinical lab |
| Detection Wavelength | 405/508 nm | 480/450 nm |
| Claimed measuring range | 10 to 1,000 U/L | 5 to 1,200 U/L |
| Detection Limit | 1.8 U/L | 5 U/L |
K. Standard/Guidance Document Referenced (if applicable):
CLSI/NCCLS EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; 2004
CLSI/NCCLS EP6-A: Evaluation of Linearity of Quantitative Measurement Procedures, A Statistical Approach; 2003
{3}
CLSI EP7-A2: Interference Testing in Clinical Chemistry; Approved Guideline, 2005
CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation; 2004
## L. Test Principle:
Alkaline phosphatase (ALP) in the sample reacts with its substrate, p-nitrophenyl phosphate (p-NPP), in ethylaminoethanol (EAE) buffer, to release p-nitrophenol (yellow). The ALP activity is determined by measuring the rate of p-nitrophenol production. This rate of change in absorbance is directly proportional to the ALP activity in the serum.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
20-day In-house Precision: The studies followed CLSI EP5-A2, where three levels of archived (stored frozen) patient serum samples were each tested in two runs, twice a day, for 20 days on the Hitachi E40 Clinical Analyzer. The results were as follows:
ALP- Low, Level 1, Summary
| ALP | Within-Run | Total |
| --- | --- | --- |
| Mean (U/L) | 42.7 | 42.7 |
| SD (U/L) | 1.3 | 2.25 |
| %CV | 3.1% | 5.3% |
ALP- Middle, Level 2, Summary
| ALP | Within-Run | Total |
| --- | --- | --- |
| Mean (U/L) | 80.5 | 80.5 |
| SD (U/L) | 3.07 | 4.65 |
| %CV | 3.8% | 5.8% |
ALP- High, Level 3, Summary
| ALP | Within-Run | Total |
| --- | --- | --- |
| Mean (U/L) | 555.6 | 555.6 |
| SD (U/L) | 13.99 | 24.69 |
| %CV | 2.5% | 4.4% |
{4}
Point-of-Care Precision:
Three levels of samples A (low), B (middle), and C (high) were tested by three POL sites, six times a day for five days on the Hitachi E40 Clinical Analyzer. The samples were native (neat) serum specimens (stored frozen). The precision estimates are described below:
| Site # | Sample | Mean (U/L) | Within-run Precision | | Total Precision | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD (U/L) | %CV | SD (U/L) | %CV |
| 1 | A | 13.2 | 0.80 | 6.1 | 0.96 | 7.3 |
| 2 | A | 12.8 | 0.89 | 6.9 | 1.00 | 7.8 |
| 3 | A | 13.4 | 1.18 | 8.8 | 1.23 | 9.2 |
| 1 | B | 76.2 | 1.22 | 1.6 | 2.26 | 3.0 |
| 2 | B | 68.8 | 4.39 | 6.4 | 5.24 | 7.6 |
| 3 | B | 70.0 | 4.56 | 6.5 | 6.28 | 9.0 |
| 1 | C | 408.6 | 4.22 | 1.0 | 26.01 | 6.4 |
| 2 | C | 390.1 | 23.21 | 6.0 | 23.21 | 6.0 |
| 3 | C | 402.9 | 13.21 | 3.3 | 12.33 | 3.1 |
b. Linearity/assay reportable range:
A linearity study was conducted based on the CLSI EP6-A guideline by comparing observed versus expected values for 10 samples. Ten (10) serial dilutions (4 to 1168 U/L were prepared and tested. The dilutions were prepared using a commercial linearity/calibration set. The calibration samples were assigned their reference values arithmetically from the labeled values and were tested in duplicate by the Hitachi E40 Clinical Analyzer. The mean Hitachi results (y-axis) were plotted against the expected values (x-axis). The parameters of linear regression are as follows:
$$
y = 0.9651x + 3.6328
$$
$$
R^2 = 0.999
$$
The ALP assay is linear between $4\mathrm{U/L}$ and $1106\mathrm{U/L}$. The results of the study support the sponsor's claim that the Hitachi S Test ALP Reagent Cartridge is linear across the measuring range of 10 to $1,000\mathrm{U/L}$.
{5}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Each lot of S TEST ALP cartridges is calibrated by the manufacturer prior to shipment using material traceable to Japanese Enzyme Reference Material (JC-ERM). The barcode printed on each cartridge provides the analyzer with lot-specific calibration data. Alkaline Phosphatase concentration is directly determined by multiplying the change in absorbance of the unknown sample by the calibrator factor on the barcode. No user calibration is needed.
Commercially available controls are required but not provided. Users should follow federal, state, and local regulatory requirements regarding quality control practices.
d. Detection limit:
Detection limit studies were performed according to CLSI EP17-A Protocols for Determination of Limits of Detection and Limits of Quantitation. The blank sample for the reagent system was assayed on the Hitachi Clinical Analyzer E40, 20 times per day for three days for a total of 60 replicate results to determine LOB. Five low samples were assayed on the Hitachi Clinical Analyzer E40, 4 times per day for three days for a total of 60 replicate results with the specific reagent cartridges to determine LOD. Each of three samples targeted at 10 U/L were assayed 6 times per day for three days on three different analyzers for a total of 54 replicate results to determine LoQ. The LoQ was determined based on inter-assay precision of < 15% CV.
| Analyte | LoB (U/L) | LoD (U/L) | LoQ (U/L) |
| --- | --- | --- | --- |
| Alkaline Phosphatase | 1.02 | 1.8 | 10 |
The measuring range of the assay was 10 to 1000 U/L.
e. Analytical specificity:
Interference studies were performed according to CLSI EP7-A2, Interference Testing in Clinical Chemistry; Approved Guideline) were performed to determine the effects from potential interferents. Two levels of commercial control sera (ALP low and high, approximately 35 U/L and 100 U/L) were spiked to six levels with each interferent, and all seven samples (the 6 spiked samples and the neat, zero baseline sample) were tested in replicates of three on the Hitachi E40 Clinical Analyzer. In each case, the spiked sample result mean was compared to its neat control mean result, and recoveries were calculated. The sponsor claims non significant interference (≤10% difference) for the substances and concentrations listed in the table below.
{6}
Alkaline Phosphatase
| Interferent Compound | Highest Concentration Showing No Interference |
| --- | --- |
| Hemoglobin | 500 mg/dL |
| Unconjugated bilirubin | 50 mg/dL |
| Lipemia | 2,000 mg/dL |
| Ascorbic acid | 50 mg/dL |
f. Assay cut-off:
Not Applicable
2. Comparison studies:
a. Method comparison with predicate device:
In-house method comparison study was conducted using a total of 97 clinical specimens (4 diluted and 7 spiked) spanning the dynamic range (11 to 926 U/L), samples were assayed on the Hitachi Clinical Analyzer E40 in singleton on both the Hitachi S Test Reagent and the Roche Cobas c systems ALP2 (predicate device). The comparative data were analyzed by linear regression and are shown below.
Internal study summary – ALP (U/L)
| n | Hitachi Range | Regression Equation | “r” | 95% CI Slope | 95% CI Intercept |
| --- | --- | --- | --- | --- | --- |
| 97 | 11 to 926 U/L | y = 0.926x + 4.8075 | 0.996 | (0.909 to 0.943) | (-0.2 to 9.8) |
External site method comparison study: A series of approximately 70 serum specimens (4 diluted) with ALP values ranging from 11 to 745 U/L were assayed on the Hitachi Clinical Analyzer E40 at three sites using S TEST Reagent Cartridge ALP (y) and Roche Cobas c systems ALP2 (predicate device) as the reference method (x). The samples were native serum specimens. Linear regression analysis yielded the following results:
POL study summary- ALP (U/L)
| Site # | n | Sample Range | Regression Equation | “r” | 95% CI Slope | 95% CI Intercept |
| --- | --- | --- | --- | --- | --- | --- |
| 1 | 77 | 13 to 745 U/L | y=0.967x +2.9 | 0.99 | 0.949 to 0.984 | -0.6 to 6.4 |
| 2 | 72 | 11 to 688 U/L | y=0.942x -0.7 | 0.99 | 0.925 to 0.960 | -4.4 to 2.9 |
| 3 | 72 | 12 to 736 U/L | y=0.980x +0.8 | 0.99 | 0.961 to 0.999 | -3.2 to 4.7 |
{7}
# b. Matrix comparison
A study was performed to validate the use of two plasma types as an alternative to serum for the Hitachi Clinical Analyzer E40 with S TEST Reagent Cartridge ALP. The plasma types were sodium citrate and lithium heparin. Thirty-eight (38) matched serum/plasma samples that spanned the range of the assay (13 to 967 U/L) were assayed in singleton. The study set included five diluted samples and six spiked samples. The results were compared using linear regression (plasma = y-axis, each type). The performance characteristics were as follows.
$\mathrm{N} = 38$
Range (serum) = 13 to 967 U/L
| | Sodium Citrate Plasma | Heparinized Plasma |
| --- | --- | --- |
| Slope (95% CIs) | 1.03 (1.01 to 1.05) | 1.01 (1.00 to 1.02) |
| y-intercept (95% CIs) | -11.2 (-15.5 to -7.0) | -5.4 (-8.3 to -2.6) |
| r | 0.999 | 0.999 |
The sponsor claimed that lithium heparin and sodium citrate are acceptable anti-coagulants to be used with the ALP assay.
# 3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
# 4. Clinical cut-off:
Not applicable. Clinical studies are not typically submitted for this device type.
# 5. Expected values/Reference range:
| | Males | Females |
| --- | --- | --- |
| ALP Reference Range1 | 53 to 128 U/L | 42 to 141 U/L |
{8}
9
¹ Tietz Fundamentals of Clinical Chemistry, 4th Edition, WB Saunders Company, (1996)
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.