The Merz Dental GmbH Denture Base Resins and Prosthetic Color System are for use in: - Relining a denture surface that contacts tissue - Repairing a fractured denture, or - Forming a new denture base
Device Story
Merz Dental Denture Base Resins (Promolux, Promolux Hi, Weropress, PremEco PCS) are PMMA-based materials used by dental technicians for denture fabrication. Inputs include PMMA powder and MMA liquid; these are mixed into a dough or applied via brush-on technique. The system uses standard heat or cold-curing methods to polymerize the resin. Promolux Hi is formulated for increased impact resistance; PremEco PCS includes tints for gingival aesthetic customization. The final output is a hardened denture base used for relining, repair, or new construction. The device is used in dental laboratory settings. The resulting denture base provides structural support and aesthetic restoration for the patient.
Clinical Evidence
Bench testing only. Materials tested for in vitro cytotoxicity (ISO 10993-5), guinea pig sensitization (ISO 10993-10.2), physical properties (ISO 1567:2000 for cold-curing, ISO 20795-1 for hot-curing), and residual methyl methacrylate monomer requirements. Results confirm compliance with standards and similarity to predicate performance.
Technological Characteristics
Materials: Polymethylmethacrylate (PMMA) and methacrylate monomer. Curing: Heat-cured (Promolux, Promolux Hi) or cold-cured (Weropress, PremEco PCS). Form factor: Two-part powder/liquid system. Standards: ISO 1567:2000 (cold-curing), ISO 20795-1 (hot-curing), ISO 10993-5/10.2 (biocompatibility).
Indications for Use
Indicated for patients requiring denture relining, repair of fractured dentures, or fabrication of new denture bases.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
Vertex cold-curing denture base material (K102640)
Submission Summary (Full Text)
{0}------------------------------------------------
مسمسلمسلمسلسل
## 510(k) Summary
Merz Dental GmbH Denture Base Resins with Prosthetic Color System
## Submitter
| Company Name: | Merz Dental GmbH |
|------------------|-------------------------------------------|
| Address: | Eetzweg 20<br>Lutjenburg, Germany D-24321 |
| Contact Name: | Dr. Med. Claudia Bobrowski |
| Telephone No. | (011) 43 81/4 03-4 11 |
| Fax No. | (011) 43 81/4 09-1 07 |
| e-mail: | Claudia.bobrowski@merz-dental.de |
| Date of Summary: | July 17, 2013 |
AUG 2 2 2013## Device Name
| Proprietary name: | Promolux, Promolux Hi, Weropress, PremEco PCS |
|----------------------|---------------------------------------------------------------------------------------------------------|
| Common name: | Denture Base Resin |
| Classification name: | Resin, Denture, Relining, Repairing, Rebasing<br>(21CFR 872.3760 Product Code EBI) |
| Predicate Device: | Vertex hot-curing denture base material (K102654)<br>Vertex cold-curing denture base material (K102640) |
# Device Description:
Merz Dental Denture Base Resins with Prosthetic Color System are composed of the following products:
| • Promolux | A heat processed PMMA denture base resin |
|---------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| • Promolux Hi | A heat-processed PMMA denture base resin formulated for<br>greater strength |
| · Weropress | A two part cold curing PMMA denture base resin |
| : PremEco PCS | A two part cold curing PMMA denture base resin with<br>individual tints for aesthetic coloration of gingival portion of<br>a denture base |
These products are intended for the preparation of denture bases. They are composed of PMMA and cured by standard heat or cold setting methods. The Prosthetic Color System
{1}------------------------------------------------
is a three part system composed of the following: (1) PMMA polymer with various tints in a powder form, (2) MMA liquid for mixing with the powder, and (3) a cold-casting resin (PMMA and MMA) for final preparation of the denture base. The hot and cold curing denture materials contain standard dental colorants and catalyst systems. One of the hot cured dental base materials, Promolux Hi, is formulated to increase impact resistance.
#### Intended Use
The Merz Dental GmbH Denture Base Resins and Prosthetic Color System are for use in:
- · Relining a denture surface that contacts tissue
- · Repairing a fractured denture, or
- · Forming a new denture base
#### Technological characteristics and substantial equivalence
The technological characteristics of the Merz Dental GmbH Denture Base Resins with Prosthetic Color System, and those of the predicate devices, are shown in Tables 1 and 2.
The Merz Dental GmbH Denture Base Resins and the predicate devices are composed of the same acrylic material (PMMA), utilize the same curing methods (heat or cold cured), and have similar physical properties. They also have the same indications for use (fabrication of denture bases). They differ with regard to impact resistance (Promolux Hi has a higher impact resistance than the heat-cured predicate). The Merz Dental GmbH Prostheic Color System also adds flexibility in the technician's ability to vary the coloring of the gingival portion of the denture base.
{2}------------------------------------------------
# Table 1: Comparison of Promolux and Promolux Hi to the Predicate Vertex Rapid Simplified
-
| Device | Vertex Rapid | Promolux | Promolux Hi |
|---------------|-------------------------|-------------------------|-------------------------|
| Heat-Cured | Simplified | | |
| Technology | PMMA Acrylic Plastic | PMMA Acrylic Plastic | PMMA Acrylic Plastic |
| Features | Heat curable | Heat curable | Heat curable |
| Materials | Two part mix of | Two part mix of | Two part mix of |
| | polymethylmethacrylate | polymethylmethacrylate | polymethylmethacrylate |
| | powder and | powder and | powder and |
| | methacrylate monomer | methacrylate monomer | methacrylate monomer |
| | liquid | liquid | liquid |
| Operations | Mixed into dough and | Mixed into dough and | Mixed into dough and |
| | pressed into flask | pressed into flask | pressed into flask |
| Indications | · Relining a denture | • Relining a denture | · Relining a denture |
| | surface that contacts | surface that contacts | surface that contacts |
| | tissue | tissue | tissue |
| | · Repairing a fractured | · Repairing a fractured | · Repairing a fractured |
| | denture, or | denture, or | denture, or |
| | · Forming a new | · Forming a new | · Forming a new |
| | denture base | denture base | denture base |
| | | | |
| Physical | Impact-resistance | Impact-resistance | Impact-resistance |
| Properties | Flexural strength | Flexural strength | Flexural strength |
| | Flexural modulus | Flexural modulus | Flexural modulus |
| | Water sorption | Water sorption | Water sorption |
| | Solubility | Solubility | Solubility |
| Bio- | Composed of | Composed of | Composed of |
| compatibility | biocompatible PMMA | biocompatible PMMA | biocompatible PMMA |
| | with colorants. | with colorants. | with colorants. |
.
ﺘ
{3}------------------------------------------------
| Device Cold<br>-Cured | Vertex Castavaria | Weropress | PremEco PCS |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Technology | PMMA Acrylic plastic | PMMA Acrylic plastic | PMMA Acrylic plastic |
| Features | Cold Curing | PMMA Acrylic plastic | PMMA Acrylic plastic |
| Materials | Two part mix of<br>polymethylmethacrylate<br>powder and<br>methacrylate monomer<br>liquid | Two part mix of<br>polymethylmethacrylate<br>powder and<br>methacrylate monomer<br>liquid | Two part mix of<br>polymethylmethacrylate<br>powder and<br>methacrylate monomer<br>liquid |
| Operations | Mixed dough and<br>pressed into flask | Mixed dough and<br>pressed into flask | Mix powder and liquid<br>and brush on |
| Indications | • Relining a denture<br>surface that contacts<br>tissue<br>• Repairing a fractured<br>denture,<br>• Forming a new<br>denture base | • Relining a denture<br>surface that contacts<br>tissue<br>• Repairing a fractured<br>denture,<br>• Forming a new<br>denture base | • Relining a denture<br>surface that contacts<br>tissue<br>• Repairing a fractured<br>denture,<br>• Forming a new<br>denture base |
| Physical<br>Properties | Impact-resistance<br>Flexural strength<br>Flexural modulus<br>Water<br>Solubility | Impact-resistance<br>Flexural strength<br>Flexural modulus<br>Water<br>Solubility | Impact-resistance<br>Flexural strength<br>Flexural modulus<br>Water<br>Solubility |
| Bio-<br>compatibility | Composed of<br>biocompatible PMMA<br>with colorants. | Composed of<br>biocompatible PMMA<br>with colorants. | Composed of<br>biocompatible PMMA<br>with colorants. |
## Table 2: Comparison of Weropress and PremEco PCS to Vertex Castavaria
## Performance Testing
Tests of the cured polymer base materials was performed according to ISO 1567:2000 "Dentistry -Base polymers - Part 1: Denture base polymers" (for cold curing -- Type II: Class 1), and ISO 20795- (for hot curing-Type II, Class 2) The materials met the physical and residual methyl methacrylate monomer requirements.
These well known denture base materials were tested and found to comply with in vitro cytotoxicity (ISO 10993-5), guinea pig sensitization (ISO 10993-10.2), physical properties (ISO 1567:2000 for cold-curing polymers; and ISO 20795-1 for hot-curing polymers), and residual monomer requirements. These materials have a long history of safe use and have been extensively tested for physical and biological properties.
{4}------------------------------------------------
#### Conclusions:
Merz Dental GmbH denture base resins (Promolux Hi, Weropress and PremEco PCS) have similar technological and performance characteristics as the currently marketed predicate denture base resins (Vertex Rapid Simplified and Vertex Castavaria). The products are composed of the same biocompatible acrylic plastic (PMMA), utilize the same preparation and casting techniques, and have identical indications for use. Their physical properties are similar for each type of curing system (hot or cold) with the exception that the hot-curing Merz Dental Promolux Hi has a higher impact resistance than the hot-curing Vertex Rapid Simplified. In addition, the Merz Dental PremEco PCS product allows greater flexibility in the coloring of the gingival portion of the denture than the predicate device. Based on the above similarities and differences, it is concluded that the Merz Dental GmbH denture base resin products (Promolux, Promolux Hi, Weropress and PremEco PCS) are substantially equivalent to the predicate denture base resin products (Vertex Rapid Simplified and Vertex Castavaria).
{5}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight, composed of three curved lines.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 22, 2013
Merz Dental GmbH C/O Richard G. Hunter, M.S., RAC Principal Washington Regulatory Consultants 5616 Mariola Place NE ALBUQUERQUE, New Mexico 87111
Re: K130076
Trade/Device Name: Merz Denture Base Resins with Prosthetic Color System Regulation Number: 21 CFR 872.3760 Regulation Name: Resin Denture. Relining, Repairing, Rebasing Regulatory Class: II Product Code: EBI Dated: July 17, 2013 Received: July 18, 2013
Dear Mr. Hunter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Oosmetie for (110) the device. subject to the general controls provisions of the Act. The r ou may, therefore, many of the Act include requirements for annual registration, listing of general controls proficioning practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it n your device is elaboriise (boo assns). Existing major regulations affecting your device can be may be subject to additions. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA *s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address hup://www.fda.gov/MedicalDevices/Resources/orYou/Industry/default.htm. Also. please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou may of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincercly yours.
Mary S. Runner -S
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### 4. Indications for Use
# 510(k) Number (if known): ¥13007 ص
Device Name: Merz Denture Base Resins with Prosthetic Color System
#### Indications for Use:
The Merz Dental GmbH Denture Base Resins with Prosthetic Color System (Promolux, Promolux Hi, Weropress and PremEco PCS) are indicated for:
- Relining a denture surface that contacts tissue ●
- Repairing a fractured denture, or
- Forming a new denture base
Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR .
Over-The-Counter Use
(21 CFR Part 801 Subpart D)
(21 CFR Part 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Andrew I. Steen -S. پياز -2013.08.21 11:08:54 -04'00'
(Division Sign-Off) Division of Anesthesiology, General Hospital infection Control, Dental Devices
Page 1 of __ 1_
510(k) Number:_K | 30076
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.