← Product Code [NGX](/productcode/NGX) · K130074

# GYMFORM ABS-A-ROUND (K130074)

_Well Brain International , Ltd. · NGX · Aug 15, 2013 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K130074

## Device Facts

- **Applicant:** Well Brain International , Ltd.
- **Product Code:** [NGX](/productcode/NGX.md)
- **Decision Date:** Aug 15, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5850
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The GYMFORM® ABS-A-ROUND is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The ABS-A-ROUND may be considered a technique or method for muscle training. The 3-area belt is intended for use on the muscles in abdomen, left waist and right waist alternately. The Mini belt accessory is intended for use on the muscles in arms, legs (lower extremities), thighs and buttocks areas separately.

## Device Story

Battery-operated, three-channel muscle stimulation system; includes electronic stimulator module, fixation belts, and replaceable skin-contact electrodes. Device delivers electrical current to targeted muscle groups; user-controlled via buttons to adjust intensity levels and mode programs. Operates in a pre-set sequence: F channel (180s), 10s idle; L channel (180s), 10s idle; R channel (180s), 10s idle; cycle repeats. Used by healthy individuals for muscle conditioning/training. Output parameters (intensity, mode) are displayed on an LCD. Benefits include improved muscle performance through electrical stimulation.

## Clinical Evidence

Bench testing only. Compliance verified with electrical safety standards IEC 60601-1 and IEC 60601-2-10, and electromagnetic compatibility standard IEC 60601-1-2. Waveform testing performed to verify device parameters.

## Technological Characteristics

Battery-powered (3 x 1.5V AAA) muscle stimulator. 3-channel output. Features LCD for power, channel, intensity, and mode status. Materials contacting skin comply with ISO 10993-5 and ISO 10993-10. Output: 0-118V (0-1000 Ohm). Frequency range: 2, 10, 50, 90, 120 Hz. Pulse width: 100/120 us. Operates via pre-programmed stimulation sequences.

## Regulatory Identification

A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.

## Predicate Devices

- X2ABS Dual Channel Fitness Belt ([K102295](/device/K102295.md))
- Gymform Dual Shape flex belt ([K111781](/device/K111781.md))
- Contour Technology Muscle Stimulator, Model: MX9 ([K111476](/device/K111476.md))
- Prospera OTC TENS Electronic Pulse Massager ([K122744](/device/K122744.md))
- Body Control System '4M' ([K092476](/device/K092476.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Well Brain International Ltd. Sponsor: Subject Device: GYMFORM® ABS-A-ROUND, Model: VDPGYCIND0016 File No.: 510(k) submission report (V1.0), Chapter 6

Chapter 6. 510(k) Summary

## K130074

# 510(k) Summary

This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.

### 1. Submitter's Information

- � 510(k) Owner's Name: Well Brain International Ltd.
- � Establishment Registration Number: 3004950644
- Address: Room 1212, Harbour Crystal Centre, No. 100 Granville Road, Tsim Sha Tsui East, Kowloon, � Hong Kong SAR, China
- � Phone: (852) 2619-0833
- Fax: (852) 2429-0960 �
- ◆ Contact Person: Victor K Wai
- � Email: victor@wellbrain-intl.com

### 2. Subject Device Information

- GYMFORM® ABS-A-ROUND, Model: VDPGYCIND0016 � Trade Name:
- Powered muscle stimulator Common Name: �
- Stimulator, Muscle, Powered, For muscle conditioning Classification name: �
- Physical Medicine � Review Panel:
- NGX Product Code: �
- 2 � Regulation Class:
- 890.5850 Regulation Number: �

#### 3. Predicate Device Information

| Sponsor        | Leto<br>Enterprises<br>Incorporation  | Well Brain<br>International<br>Ltd. | Contour<br>Technology                                           | Prospera<br>Corporation                                 | SPORT-<br>ELEC S.A.               |
|----------------|---------------------------------------|-------------------------------------|-----------------------------------------------------------------|---------------------------------------------------------|-----------------------------------|
| Device<br>Name | X2ABS Dual<br>Channel<br>Fitness Belt | Gymform<br>Dual Shape<br>flex belt  | Contour<br>Technology<br>Muscle<br>Stimulator,<br>Model:<br>MX9 | Prospera<br>OTC TENS<br>Electronic<br>Pulse<br>Massager | Body<br>Control<br>System<br>"4M" |

Report by Well Brain International Ltd.

AUG 1 5 2013

{1}------------------------------------------------

| Sponsor:        | Well Brain International Ltd.              |
|-----------------|--------------------------------------------|
| Subject Device: | GYMFORM® ABS-A-ROUND, Model: VDPGYCIND0016 |
| File No.:       | 510(k) submission report (V1.0), Chapter 6 |

| 510(k)<br>Number     | K102295  | K111781  | K111476  | K122744  | K092476  |
|----------------------|----------|----------|----------|----------|----------|
| Product<br>Code      | NGX      | NGX      | NGX      | NGX      | NGX      |
| Regulation<br>Number | 890.5850 | 890.5850 | 890.5850 | 890.5850 | 890.5850 |
| Regulation<br>Class  | 2        | 2        | 2        | 2        | 2        |

#### 4. Device Description

The ABS-A-ROUND is a three-channel battery operated muscle stimulation system specifically designed to training muscles.

The 3-area belt is intended for use on the muscles in abdomen, left waist and right waist alternately. The stimulation of each channel is active atternately for 2 cycles with its set output intensity level and mode program in below default activation sequence: F channel is active for 180 seconds solely, stop for 10 seconds as idle interval; then L channel is active for 180 seconds solely, stop for 10 seconds as idle interval; then R channel is active for 180 seconds solely, stop for 10 seconds as idle interval; repeat one cycle with above activation sequence.

The Mini belt accessory is intended for use on the muscles in arms, legs (lower extremities), thighs and buttocks areas separately.

It is comprised of an electronic stimulator module for signal generation, two belts for fixation, and electrodes for signal connection to skin. The electrodes are replaceable.

Power is derived from 3 batteries located in a compartment protected by a removable battery cover for the Fitness Belt. There is no current passed from side to side.

The stimulator sends gentle electrical current to targeted muscle group through the electrodes placed on the skin. The parameters of the unit are controlled by the buttons. Its intensity level can be adjustable by user.

#### 5. Intended Use / Indications for Use

The GYMFORM® ABS-A-ROUND is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The ABS-A-ROUND may be considered a technique or method for muscle training. The 3-area belt is intended for use on the muscles in abdomen, left waist and right waist alternately. The Mini belt accessory is intended for use on the muscles in arms, legs (lower extremities), thighs and buttocks areas separately.

#### 6. Performance Summary

Testing has been carried out to assure compliance with recognized electrical safety standards: - IEC 60601-1 and IEC 60601-2-10 standards for electrical safety

Report by Well Brain International Ltd.

{2}------------------------------------------------

| Sponsor:        | Well Brain International Ltd.               |
|-----------------|---------------------------------------------|
| Subject Device: | GYMFORM ® ABS-A-ROUND, Model: VDPGYCIND0016 |
| File No.:       | 510(k) submission report (V1.0), Chapter 6  |

- IEC 60601-1-2 standard for electromagnetic compatibility

Performance data has been verified according to the requirements of the FDA "Guidance for Pre Market Submissions and for Software Contained in Medical Devices".

The waveform test report has also been provided to verify the parameters of the device.

#### 7. Comparison to predicate device and conclusion

The technological characteristics, features, specifications, materials, mode of operation, and intended use of GYMFORM® ABS-A-ROUND is substantially equivalent to the predicate devices quoted above. The differences between the subject device and predicate devices do not raise new issues of safety or effectiveness.

| Elements of<br>Comparison                   | Subject<br>Device                                                                                                                                                                                                                                                                                                                                                                                                     | Predicate Device                                                                                                                                                                                                                                                                                                                           |                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                  | Remark                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                          |              |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Device<br>Name and<br>Model                 | GYMFORM®<br>ABS-A-<br>ROUND                                                                                                                                                                                                                                                                                                                                                                                           | X2ABS Dual<br>Channel<br>Fitness Belt                                                                                                                                                                                                                                                                                                      | Gymform<br>Dual Shape<br>flex belt                                                                                                         | Contour<br>Technology<br>Muscle<br>Stimulator,<br>Model: MX9                                                                                                                                                                                                                                                                                                                                                     | Prospera<br>OTC TENS<br>Electronic<br>Pulse<br>Massager                                                                                                                                                                                                                                                  | Body Control<br>System '4M'                                                                                                                                                                                                                                                                                                                              |              |
| 510 (K)<br>Number                           | Applying                                                                                                                                                                                                                                                                                                                                                                                                              | K102295                                                                                                                                                                                                                                                                                                                                    | K111781                                                                                                                                    | K111476                                                                                                                                                                                                                                                                                                                                                                                                          | K122744                                                                                                                                                                                                                                                                                                  | K092476                                                                                                                                                                                                                                                                                                                                                  | --           |
| Intended<br>Use &<br>Indications<br>for Use | The<br>GYMFORM®<br>ABS-A-<br>ROUND is<br>intended to<br>stimulate<br>healthy<br>muscles in<br>order to<br>improve or<br>facilitate<br>muscle<br>performance.<br>The ABS-A-<br>ROUND may<br>be<br>considered a<br>technique or<br>method for<br>muscle<br>training.<br>The 3-area<br>belt is<br>intended for<br>use on the<br>muscles in<br>abdomen,<br>left waist and<br>right waist<br>alternately.<br>The Mini belt | The X2ABS<br>Dual<br>Channel<br>Fitness Belt<br>is intended<br>for use by<br>healthy<br>persons to<br>apply trans-<br>coetaneous<br>electrical<br>muscle<br>stimulation<br>(EMS)<br>through skin<br>contact<br>electrodes<br>for the<br>following<br>purposes:<br>Improvement<br>of muscle<br>tone of the<br>muscles in<br>the<br>abdomen. | The<br>Gymform<br>Dual Shaper<br>flex belt is a<br>powered<br>muscle<br>stimulator<br>intended for<br>abdominal<br>muscle<br>conditioning. | The Contour<br>Technology<br>Muscle<br>Stimulator is<br>intended to<br>stimulate<br>healthy<br>muscles in<br>order to<br>improve or<br>facilitate<br>muscle<br>performance<br>. The<br>Contour<br>Technology<br>Muscle<br>Stimulator<br>may<br>therefore be<br>considered a<br>technique or<br>method for<br>muscle<br>training.<br>The Contour<br>Technology<br>Muscle<br>Stimulator<br>Ab Belt<br>accessory is | To be used<br>for<br>temporary<br>relief of pain<br>associated<br>with sore<br>and aching<br>muscles in<br>the shoulder,<br>waist, back,<br>neck, upper<br>extremities<br>(arm), and<br>lower<br>extremities<br>(leg) due to<br>strain from<br>exercise or<br>normal<br>household<br>work<br>activities. | The Body<br>Control -4M'<br>is intended<br>for use bv<br>healthy<br>persons to<br>apply trans-<br>coetaneous<br>electrical<br>muscle<br>stimulation<br>(EMS)<br>through skin<br>contact<br>electrodes<br>for the<br>following<br>purposes.<br>Improvement<br>of muscle<br>tone and<br>firmness, for<br>strengthenin<br>g muscles in<br>arms,<br>abdomen, | SE<br>Note 1 |

Report by Well Brain International Ltd.

{3}------------------------------------------------

| Sponsor: | Well Brain International Ltd.                                 |
|----------|---------------------------------------------------------------|
|          | Subject Device:    GYMFORM® ABS-A-ROUND, Model: VDPGYCIND0016 |
|          | File No .:    510(k) submission report (V1.0), Chapter 6      |

|                                                   | accessory is<br>intended for<br>use on the<br>muscles in<br>arms, legs<br>(lower<br>extremities),<br>thighs and<br>buttocks<br>areas<br>separately. |                                               |                                                | intended for<br>use on<br>abdominal<br>muscles only<br>for<br>strengthenin<br>g and toning<br>of abdominal<br>muscles.<br>The Contour<br>Technology<br>Muscle<br>Stimulator<br>BackPad<br>accessory is<br>intended for<br>use on the<br>lower back<br>muscles<br>only. |                                                                                                        | thighs and<br>buttocks<br>areas.                              |                | Power                                               | 3 x 1.5V<br>AAA<br>batteries                                                                                                      | 2 x 1.5V<br>AAA<br>batteries                                                                                                                                  | 12V AC<br>Adapter                                                                                                                          | 4 x 1.5V<br>AAA<br>batteries                                                                                                      | 6V battery                         | 3 x 1.5V<br>AAA<br>batteries           | SE<br>Note 4 |
|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|----------------|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|------------------------------------|----------------------------------------|--------------|
| Specification                                     |                                                                                                                                                     |                                               |                                                |                                                                                                                                                                                                                                                                        |                                                                                                        |                                                               |                | Electrode<br>Size (cm²)                             | Small: 33.0 x<br>1 piece<br>Big: 34.5 x 2<br>pieces                                                                               | 32                                                                                                                                                            | --                                                                                                                                         | 40.5                                                                                                                              | --                                 | 128                                    | SE<br>Note 5 |
| Channel                                           | 3                                                                                                                                                   | 2                                             | 1                                              | 2                                                                                                                                                                                                                                                                      | 2                                                                                                      | 2                                                             | SE<br>Note 2   | Maximum<br>Current<br>Density(mA/<br>cm² @<br>500Ω) | 0.057<br>(for the<br>smallest size<br>electrode<br>33.0 cm²)                                                                      | 0.032<br>(for the<br>smallest size<br>electrode<br>32.0 cm²)                                                                                                  | --                                                                                                                                         | 0.55 (for the<br>smallest size<br>electrode<br>40.5 cm²)                                                                          | 1.4                                | --                                     | SE<br>Note 5 |
| Synchronous<br>/Alternating<br>channels           | Alternating                                                                                                                                         | Synchronous                                   | (1 Channel)                                    | --                                                                                                                                                                                                                                                                     | Alternating                                                                                            | …                                                             | SE             | Timer Range                                         | Default time<br>(maximum)<br>is 19<br>minutes.                                                                                    | Default time<br>is 10<br>minutes,<br>minimum<br>time is 5<br>minutes                                                                                          | Default time<br>is 10<br>minutes                                                                                                           | --                                                                                                                                | 15 minutes                         | Default time<br>is 4 minutes<br>30 sec | SE           |
| Stimulated<br>muscles                             | Abdomen,<br>left waist and<br>right waist,<br>arms, legs,<br>thighs and<br>buttocks                                                                 | Abdomen                                       | Abdomen                                        | Abdomen<br>and lower<br>back                                                                                                                                                                                                                                           | Shoulder,<br>waist, back,<br>neck, upper<br>extremities<br>(arm), and<br>lower<br>extremities<br>(leg) | Arms,<br>abdomen,<br>thighs and<br>buttocks<br>areas          | SE             | LCD display                                         | Indicate the<br>following<br>information:<br>Power on/off,<br>Channel<br>indication,<br>Intensity<br>level, Mode<br>selection.    | Indicate the<br>following<br>information:<br>Sound<br>on/off,<br>Keylock,<br>Low battery,<br>Channel<br>indication,<br>Intensity<br>level, Mode<br>selection. | Indicate the<br>following<br>information:<br>Sound<br>on/off,<br>Keylock,<br>Low battery,<br>Channel<br>indication,<br>Intensity<br>level. | --                                                                                                                                | Indicate<br>On/Off<br>status only. | --                                     | SE           |
| Number of<br>programs                             | 6                                                                                                                                                   | 8                                             | 5                                              |                                                                                                                                                                                                                                                                        | 3                                                                                                      | 4                                                             | SE<br>Note 2   | Environment<br>for operating                        | Temperature<br>: 5 ~ 40°C<br>Humidity: 20<br>~ 65% RH                                                                             | Temperature<br>: 5 ~ 40°C<br>Humidity: 20<br>~ 65% RH                                                                                                         | Temperature<br>: 35 °F to<br>95 °F<br>Humidity: 20<br>to 80% RH                                                                            | --                                                                                                                                | --                                 | --                                     | SE<br>Note 4 |
| Number of<br>phases<br>variations                 | 99 steps                                                                                                                                            | 28 steps                                      | 99 steps                                       | ==                                                                                                                                                                                                                                                                     | --                                                                                                     | --                                                            | SE<br>Note 2   | Environment<br>for storage                          | Temperature<br>: 0 ~ 40°C<br>Humidity: 10<br>~ 90% RH                                                                             | Temperature<br>: 0 ~ 40°C<br>Humidity: 10<br>~ 90% RH                                                                                                         | Temperature<br>: 32 °F to<br>131 °F<br>Humidity: 10<br>to 90% RH                                                                           | --                                                                                                                                | --                                 | --                                     | SE<br>Note 4 |
| Number of<br>adjustable<br>independent<br>outputs | 3                                                                                                                                                   | 2                                             | 1                                              | --                                                                                                                                                                                                                                                                     | 3                                                                                                      | 2                                                             | SE             | Standards                                           |                                                                                                                                   |                                                                                                                                                               |                                                                                                                                            |                                                                                                                                   |                                    |                                        |              |
| Output                                            | From 0 to<br>118 V (From<br>0 to 1000<br>Ohm)                                                                                                       | From 0 to 60<br>mA (From O<br>to 1000<br>Ohm) | From 5.6 to<br>40 mA<br>(From 0 to<br>500 Ohm) | From 0 to 22<br>mA (From 0<br>to 500 Ohm)                                                                                                                                                                                                                              |                                                                                                        | From 0 to<br>70mA from 0<br>to 60V- from<br>0 to 1000<br>Ohm) | SE<br>Note 3   | Biocompatibi<br>lity                                | All user<br>directly<br>contacting<br>materials are<br>compliance<br>with<br>ISO10993-5<br>and<br>ISO10993-<br>10<br>requirements | All user<br>directly<br>contacting<br>materials are<br>compliance<br>with<br>ISO10993-5<br>and<br>ISO10993-<br>10<br>requirements                             | All user<br>directly<br>contacting<br>materials are<br>compliance<br>with<br>ISO10993-5<br>and<br>ISO10993-<br>10<br>requirements          | All user<br>directly<br>contacting<br>materials are<br>compliance<br>with<br>ISO10993-5<br>and<br>ISO10993-<br>10<br>requirements | --                                 | --                                     | SE           |
| Frequency<br>range(Hz)                            | 2, 10, 50, 90,<br>120                                                                                                                               | 8.5 to 64                                     | 15,18, 25,30,<br>35                            | 1 to 120                                                                                                                                                                                                                                                               | 1 to 100                                                                                               | 50 to 70                                                      | SE<br>Note 3   | Electrical<br>Safety                                | Comply with<br>IEC 60601-1<br>and IEC<br>60601-2-10                                                                               | Comply with<br>IEC 60601-1<br>and IEC<br>60601-2-10                                                                                                           | Comply with<br>IEC 60601-1<br>and IEC<br>60601-2-10                                                                                        | Comply with<br>IEC 60601-1<br>and IEC<br>60601-2-10                                                                               | --                                 | --                                     | SE           |
| Pulse width<br>range(us)                          | 100 / 120                                                                                                                                           | 220                                           | 250                                            | 340                                                                                                                                                                                                                                                                    | 120 to 6800                                                                                            | 200                                                           | ട്ട്<br>Note 3 |                                                     |                                                                                                                                   |                                                                                                                                                               |                                                                                                                                            |                                                                                                                                   |                                    |                                        |              |
| Contraction<br>and<br>Relaxation<br>time          | Adjustable,<br>due to<br>different<br>modes.                                                                                                        | Adjustable,<br>due to<br>different<br>modes.  | Adjustable,<br>due to<br>different<br>modes.   | --                                                                                                                                                                                                                                                                     | Adjustable,<br>due to<br>different<br>modes.                                                           | Adjustable,<br>due to<br>different<br>modes.                  | SE             |                                                     |                                                                                                                                   |                                                                                                                                                               |                                                                                                                                            |                                                                                                                                   |                                    |                                        |              |

·

.

:

{4}------------------------------------------------

#### Well Brain International Ltd. Sponsor: GYMFORM® ABS-A-ROUND, Model: VDPGYCIND0016 Subject Device: 510(k) submission report (V1.0), Chapter 6 File No.:

{5}------------------------------------------------

| Sponsor:        | Well Brain International Ltd.              |
|-----------------|--------------------------------------------|
| Subject Device: | GYMFORM® ABS-A-ROUND, Model: VDPGYCIND0016 |
| File No.:       | 510(k) submission report (V1.0), Chapter 6 |

| EMC | Comply with<br>IEC 60601-1-2 | Comply with<br>IEC 60601-1-2 | Comply with<br>IEC 60601-1-2 | Comply with<br>IEC 60601-1-2 | -- | -- | SE |
|-----|------------------------------|------------------------------|------------------------------|------------------------------|----|----|----|
|-----|------------------------------|------------------------------|------------------------------|------------------------------|----|----|----|

#### Comparison in Detail(s):

#### Note 1:

Although the muscles for intended use of subject device are different from predicate devices, we can find that the predicate device Contour Technology Muscle Stimulator (K111476), Prospera OTC TENS Electronic Pulse Massager (K122744) and Body Control System '4M' (K092476) can also be used on the same muscles according to the device description of 510(k) summary. So the subject device and predicate device K111476 have the same intended use aspect.

### Note 2:

Although the number of channels, programs, and phase variations of subject device are different from the predicate devices, they are all compliance with IEC 60601-2-10 requirements. So the differences of the function specifications will not raise any safety or effectiveness issue.

#### Note 3:

Although the "Output", "Frequency range" and "Pulse width range" of subject device are a little different from the predicate devices, they are all compliance with IEC 60601-2-10 requirement and FDA guidance requirement for the EMS. So the differences of function will not raise any safety or effectiveness issue.

#### Note 4:

Although the power, operating and storage environment are a little different from the predicate devices, they are all compliance with IEC 60601-1 requirements. So the differences will not raise any safety or effectiveness issue.

#### Note 5:

Although the electrode size and maximum current density of subject device are a little different from the predicate devices, they are all compliance with IEC 60601-2-10 requirement and FDA guidance requirement for the EMS. So the differences of function will not raise any safety or effectiveness issue.

8. Date of the summary prepared: August 12, 2013

Report by Well Brain International Ltd.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 15, 2013

Well Brain International Ltd. c/o Victor K Wai Managing Director Rm. 1212, 12 Floor, Harbour Crystal Centre No. 100 Granville Road, Tsim Sha Tsui East Kowloon, Hong Kong China 999077

Re: K130074

Trade/Device Name: Gymform® ABS-A-ROUND, model VDPGYCIND0016 Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: NGX Dated: July 8, 2013 Received: July 11, 2013

Dear Mr. Wai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{7}------------------------------------------------

Page 2 - Mr. Victor K. Wai

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (2) CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Joyce M. Whang -S

for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

## Indications for Use

510(k) Number (if known): _ K130074

Device Name: GYMFORM® ABS-A-ROUND, Model: VDPGYCIND0016

Indications For Use:

The GYMFORM® ABS-A-ROUND is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The ABS-A-ROUND may be considered a technique or method for muscle training. The 3-area belt is intended for use on the muscles in abdomen, left waist and right waist alternately. The Mini belt accessory is intended for use on the muscles in arms, legs (lower extremities), thighs and buttocks areas separately.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

# Joyce M. Whang -S

(Division Sign Off) Division of Neurological and Physical Medicine Devices (DNPMD)

510(k) Number

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K130074](https://fda-staging.innolitics.com/device/K130074)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
