K130039 · Gambro Renal Products, Inc. · KDI · May 2, 2013 · Gastroenterology, Urology
Device Facts
Record ID
K130039
Device Name
REVACLEAR 300 DIALYZER, REVACLEAR 400 DIALYZER
Applicant
Gambro Renal Products, Inc.
Product Code
KDI · Gastroenterology, Urology
Decision Date
May 2, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The dialyzer is intended for the treatment of chronic and acute renal failure by hemodialysis.
Device Story
The Revaclear 300 and 400 are single-use hemodialyzers used for treating chronic and acute renal failure. The device consists of hollow fibers made of a polyarylethersulfone (PAES) and polyvinylpyrrolidone (PVP) blend. During operation, patient blood flows through the inside of the hollow fibers while dialysis solution flows countercurrently in the outer compartment. Waste products and uremic toxins are removed from the blood via diffusion and convection across the membrane, driven by hydrostatic and transmembrane pressure. The device is used in a clinical hemodialysis setting. Healthcare providers use the device to facilitate the removal of solutes and plasma water from the patient's blood, thereby managing renal failure.
Clinical Evidence
No clinical data. Evidence consists of bench testing only. In vitro performance testing conducted per ISO 8637 included urea, creatinine, and vitamin B12 clearance, flow resistance, and ultrafiltration coefficient. Biocompatibility evaluated per ISO 10993-1 and FDA G-95-1 (cytotoxicity, sensitization, systemic toxicity, genotoxicity, hemocompatibility). Mechanically induced hemolysis testing compared free hemoglobin levels against predicates, showing no systemic increase in hemolysis.
Technological Characteristics
Hollow fiber dialyzer; membrane material: polyarylethersulfone (PAES) and polyvinylpyrrolidone (PVP) blend. Performance testing per ISO 8637; biocompatibility per ISO 10993-1 and ISO 10993-18. External communicating device, prolonged contact duration (>24h to ≤ 30 days).
Indications for Use
Indicated for treatment of chronic and acute renal failure by hemodialysis in patients requiring such therapy.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
K130039 page 1 of 3
### GAMBRO.Renal Products
#### 5.0 510(k) SUMMARY
### MAY 0 2 2013
#### 510(k) Summary
| Submitter: | Gambro Renal Products, Inc.<br>14143 Denver West Parkway, Suite 400.<br>Lakewood, Colorado 80401 |
|--------------------------------|--------------------------------------------------------------------------------------------------|
| Contact: | Kae Miller, Regulatory Affairs, Americas<br>Phone: 303.222.6724<br>Fax: 303.222.6916 |
| Date Prepared: | February 4, 2013 |
| Device Name: | Gambro Revaclear™ 300 and 400 Dialyzer |
| Common/Classification<br>Name: | Dialyzer, high permeability with or without sealed dialysate<br>system |
| Regulation Number: | 21 CFR 876.5860 |
| Product Code: | KDI |
| Review Panel: | Gastroenterology / Urology |
### Predicate Devices:
| Polyflux HD-C4 (BIG) (Revaclear Max) | Cleared - April 24, 2006 | K060195 |
|--------------------------------------|-----------------------------|---------|
| Polyflux HD-C4 (Small) (Revaclear) | Cleared - September 7, 2007 | K072232 |
### Device Description:
This device is intended for the treatment of chronic and acute renal failure by hemodialysis.
The intended population of this device is identical to those of the Polyflux HD-C4 (BIG), cleared for marketing in the United States under 510(k) notification K060195 and Polyflux HD-C4 (Small), cleared for marketing in the United States under 510(k) notification K072232.
The membrane used in this device is a blend of polyarylethersulfone (PAES) and polyvinylpyrrolidone (PVP), which is equivalent to the membrane utilized in the Gambro Polyflux HD-C4 (BIG) and Polyflux HD-C4 (Small) single use hemodialyzers cleared for marketing in the United States under 510(k) Notifications (K060195 and K072232).
Blood enters a blood inlet port where it is distributed to the hollow fibers. The patient's blood traverses the inside of the hollow fibers and exits the device via a blood exit port. By means of a hydrostatic pressure or transmembrane pressure which is created by a combination of positive and negative pressures across the membrane, plasma water along
Page 16 of 53
{1}------------------------------------------------
## GAMBRO.Renal Products
with certain lower and middle molecular weight solutes pass through the membrane and into the dialysate or filtrate compartment of the device. Uremic toxins and waste products are removed from the patient's blood in this device by means of both diffusion and convection through the membrane and into the countercurrent flowing dialysis solution during hemodialysis. The dialysate exits the devices via a dialysate outlet port.
### Indications for Use:
The dialyzer is intended for the treatment of chronic and acute renal failure by hemodialysis.
### Technological Characteristics:
The proposed device configurations have the same technological characteristics and are similar in design, function, composition, and operation, to the currently marketed configurations.
### Summary of Non-Clinical Tests:
In vitro testing was conducted to compare the performance of the proposed device configurations to the predicate configurations.
### In Vitro Performance Testing:
The performance characteristics of the submitted Revaclear 300 and 400 were determined according to requirements of ISO 8637" Cardiovascular implants and extracorporeal systems - Haemodialysers, haemodiafilters, haemofilters and haemoconcentrators".
The following characteristics were determined:
- . Clearances of urea, creatinine and vitamin B12
- Flow resistance of blood and dialysate compartments .
- . Ultrafiltration coefficient
Detailed measurement results are provided in Section 17 of this submission.
### Biocompatibility Evaluation:
Biocompatibility of the submitted devices Revaclear 300 and 400 was evaluated according to current ISO 10993-1 and FDA Memorandum G-95-1 the device is classified as external communicating device, body contact with circulating blood and prolonged (>24h to ≤ 30 days) contact duration.
Revaclear 400 was chosen as a master product for testing as it is the biggest product in the product family.
The biological end-points studied for the device were cytotoxicity, sensitization/ intracutaneous reactivity, systemic toxicity (acute), systemic toxicity (subacute), genotoxicity and hemocompatibility. These endpoints are recommended to be evaluated by the ISO 10993-1 and FDA Memorandum G-95-1.
{2}------------------------------------------------
### GAMBRO.Renal Products
In addition, chemical analyses were performed to characterize the materials and to identify and quantify potential leachables as recommended by ISO 10993-1 and ISO 10993-18. According to the results of the biological in vivo testing and of the chemical analyses performed in these studies, Revaclear 300 and 400 dialyzers did not induce any adverse biological effects.
A vailable data covers the requirements according to ISO 10993 and FDA Memorandum G-95-1. Based on chemical and biological test results, Revaclear 300 and 400 is approved for its intended use in haemodialysis for chronic and acute treatment of renal disease.
The complete biological and chemical report is provided in Attachment 4 of the submission.
### Mechanical Hemolysis:
Mechanically induced hemolysis was tested in three experiments. In each experiment three filters were tested in parallel, whereas Revaclear 300 and Revaclear 300 filters with alternative materials were compared to Revaclear predicate devices.
Samples were taken every 15 minutes starting at the beginning of recirculation and free hemoglobin (fHb) levels were detected in duplicate according to 405 nm method.
Using the above described conditions for the test set-up no systemic increase in free hemoglobin as marker for hemolysis could be detected for the Revaclear 300 filters compared to the predicate device. These results indicate that the filters are comparable regarding hemolysis.
The complete report is enclosed in Attachment 5 of the submission.
### Summary of Clinical Tests: N/A
#### Conclusion:
Testing performed on the Gambro Revaclear Dialyzers indicates that they are safe, effective and perform as well as the predicate devices, when used in accordance with the instructions for use.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 2, 2013
Gambro Renal Products, Inc. % Ms. Kae Miller Regulatory Affairs Manager, Americas 14143 Denver West Parkway, Suite 400 LAKEWOOD CO 80401
Re: K130039
Trade/Device Name: Revaclear™ 300 and 400 Dialyzer Regulation Number: 21 CFR§ 876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: February 4, 2013 Received: February 5, 2013
Dear Ms. Miller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 – Ms. Kae Miller
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## GAMBRO.Renal Products
### Indications for Use
510(k) Number (if known): ___K130039
Device Name: Revaclear™ 300 and 400 Dialyzer
Indications for Use:
Revaclear™ 300 and 400 dialyzers are indicated for treatment of chronic and acute renal failure by hemodialysis.
Prescription Use _ Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
### Herbert - Berner - S
RA 12-035 Revised 02/04/13
(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices K130039 510(k) Number
Page 15 of 53
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.