ACCESS TOTAL BHCG (5TH IS)

K130020 · Beckman Coulter, Inc. · DHA · Oct 1, 2013 · Clinical Chemistry

Device Facts

Record IDK130020
Device NameACCESS TOTAL BHCG (5TH IS)
ApplicantBeckman Coulter, Inc.
Product CodeDHA · Clinical Chemistry
Decision DateOct 1, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1155
Device ClassClass 2

Indications for Use

The Access Total βhCG (5th IS) assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of total βhCG levels in human serum and plasma using the Access Immunoassay Systems. The assay is intended for use as an aid in the early detection of pregnancy. The Access Total βhCG (5th IS) calibrators are intended to calibrate the Access Total βhCG (5th IS) assay for the quantitative determination of total βhCG levels in human serum and plasma using the Access Immunoassay Systems.

Device Story

The Access Total βhCG (5th IS) assay is an automated, paramagnetic particle, chemiluminescent immunoassay used on Access Immunoassay Systems to measure total βhCG (intact hCG and free β-subunit) in human serum and plasma. The device functions as a sandwich immunoassay; it captures hCG molecules from patient samples using specific antibodies. The system produces a quantitative result, which clinicians use as an aid in the early detection of pregnancy. The assay is intended for use in clinical laboratory settings by trained personnel. By providing accurate, standardized quantitative measurements of βhCG, the device assists healthcare providers in confirming pregnancy status, facilitating timely clinical management.

Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP5-A2 (total CV 3.1–21.7% across ranges). Linearity confirmed 0.6–1350 mIU/mL. LoB/LoD ≤ 0.5 mIU/mL; LoQ 0.6 mIU/mL. Method comparison (N=224) vs. predicate showed correlation (r=0.99). No cross-reactivity with FSH, LH, TSH, or hCG α-subunit. No high-dose hook effect up to 1,000,000 mIU/mL. Reference intervals established for non-pregnant females (18–40y, >40y, post-menopausal).

Technological Characteristics

Paramagnetic particle, chemiluminescent sandwich immunoassay. Reagents: rabbit anti-βhCG alkaline phosphatase conjugate, goat anti-mouse IgG, mouse monoclonal anti-βhCG. Substrate: Lumi-Phos 530. Measuring range: 0.6–1350 mIU/mL. Standardized to WHO 5th IS (07/364). Automated on UniCel DxI 800. Calibrators: 6 levels (0, 6, 35, 195, 620, 1350 mIU/mL) in BSA matrix.

Indications for Use

Indicated for the quantitative determination of total βhCG in human serum and plasma as an aid in the early detection of pregnancy. For prescription use.

Regulatory Classification

Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k130020 B. Purpose for Submission: New device C. Measurand: Total beta human chorionic gonadotropin (βhCG) in human serum and plasma D. Type of Test: Quantitative chemiluminescent immunoassay E. Applicant: Beckman Coulter, Inc. F. Proprietary and Established Names: Access Total βhCG (5th IS) Assay Access Total βhCG (5th IS) Calibrators G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | DHA | II | 21 CFR 862.1155, Human chorionic gonadotropin (HCG) test system | Chemistry (75) | | JIX | II | 21 CFR 862.1150, Calibrator | Chemistry (75) | {1} H. Intended Use: 1. Intended use(s): See Indications for Use below 2. Indication(s) for use: The Access Total βhCG (5th IS) assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of total βhCG levels in human serum and plasma using the Access Immunoassay Systems. The assay is intended for use as an aid in the early detection of pregnancy. The Access Total βhCG (5th IS) calibrators are intended to calibrate the Access Total βhCG (5th IS) assay for the quantitative determination of total βhCG levels in human serum and plasma using the Access Immunoassay Systems. 3. Special conditions for use statement(s): For prescription use only. 4. Special instrument requirements: Beckman UniCel DxI 800. I. Device Description: The Access Total βhCG (5th IS) assay consists of the reagent pack and calibrators. Other items needed to run the assay include substrate and wash buffers. The Access Total βhCG (5th IS) assay reagent pack, Access Total βhCG (5th IS) Calibrators, along with the Access wash buffer and substrate, are designed for use with the UniCel DxI 800 Analyzer in a clinical laboratory setting. J. Substantial Equivalence Information: 1. Predicate device name(s): Siemens ADVIA Centaur Total hCG Assay 2. Predicate 510(k) number(s): k925277 {2} 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Access Total βhCG (5thIS) Assay and Access Total βhCG (5thIS) Calibrators | Predicate ADVIA Centaur Total hCG k925277 | | Intended use | The Access Total βhCG (5thIS) assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of total βhCG levels in human serum and plasma using the Access Immunoassay Systems. This assay is intended for use as an aid in the early detection of pregnancy.The Access Total βhCG (5thIS) Calibrators are intended to calibrate the Access Total βhCG (5thIS) assay for the quantitative determination of total βhCG levels in human serum and plasma using the Access Immunoassay Systems | Same | | Technology | Sandwich Immunoassay | Same | | Format | Chemiluminescent | Same | | Method | Automated | Same | | Measurand | Total βhCG (intact hCG and free beta-subunit) | same | | Calibration | Utilizes a stored calibration curve | Same | | Calibration Interval | 28 days | Same | | Differences | | | | Item | Access Total βhCG (5thIS) Assay and Access Total βhCG (5thIS) Calibrators | Predicate ADVIA Centaur Total hCG k925277 | | Measuring Range | 0.6 – 1350 mIU/mL | 2.0 – 1000 mIU/mL | | Standardization | WHO 5thInternational | WHO 4thInternational | {3} | Similarities | | | | --- | --- | --- | | Item | Access Total βhCG (5^{th} IS) Assay and Access Total βhCG (5^{th} IS) Calibrators | Predicate ADVIA Centaur Total hCG k925277 | | | Reference Standard, 07/364 | Reference Standard, 75/589 | | Sample Type | Serum or plasma | Serum | | Calibrator Levels | 6 levels (0, 6, 35, 195, 620, and 1350 mIU/mL) | 2 levels (7 and 300 mIU/mL) | | Matrix | Bovine serum albumin | Equine serum | ## K. Standard/Guidance Document Referenced (if applicable): - CLSI EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Method - CLSI EP6-A: Evaluation of Linearity of Quantitative Measurement Procedure: A Statistical Approach - CLSI EP7-A2: Interference Testing in a Clinical Laboratory - CLSI EP9-A2: Method Comparison and Bias Estimation using Patient Samples - CLSI EP14-A2: Evaluation of Matrix Effects - CLSI EP17-A2: A Protocol for Determining Limit of Detection and Limit of Quantitation - CLSI EPC28-A3: Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory ## L. Test Principle: The Access Total βhCG (5th IS) assay is a sequential two-step immunoenzymatic (“sandwich”) assay. A sample is added to a reaction vessel along with a citrate buffer. After an initial incubation, a rabbit anti-βhCG alkaline phosphatase conjugate and paramagnetic particles coated with goat anti-mouse IgG: mouse monoclonal anti-βhCG complexes are added. The hCG binds to the immobilized monoclonal anti-βhCG on the solid phase while, at the same time, the rabbit anti-βhCG alkaline phosphatase conjugate reacts with different antigenic sites on the hCG. After incubation in a reaction vessel, materials bound to the solid phase are held in a magnetic field while unbound materials are washed away. Then, the chemiluminescent substrate Lumi-Phos® 530 is added to the vessel and light generated by the reaction is measured with a luminometer. The light production is directly proportional to the concentration of hCG in the sample. The amount of analyte in the sample is determined from a stored, multi-point calibration curve. ## M. Performance Characteristics (if/when applicable): 1. Analytical performance: {4} # a. Precision/Reproducibility: Seven serum and four plasma patient samples were tested following CLSI EP5-A2. One serum sample with a hCG concentration approximating 200,000 mIU/mL was tested using the onboard dilution feature. Three commercial controls were tested with each run. Each patient and control sample was run in duplicate. Testing occurred over twenty days (2 runs per day) for a total of 40 runs and 80 replicates. The study was run at an internal site on three instruments, using three reagent pack lots, one calibrator lot, and one lot of reagent pack per instrument. Results from multiple lots were similar. Results from one representative lot for serum and plasma samples are provided in the tables below. Precision Study - Serum | Sample ID | Grand Mean (mIU/mL) | Within Run SD (mIU/mL) | Within Run CV (%) | Between Run SD (mIU/mL) | Between Run CV (%) | Total SD (mIU/mL) | Total CV (%) | | --- | --- | --- | --- | --- | --- | --- | --- | | Serum 1 | 0.6 | 0.07 | 11.7 | 0.11 | 18.3 | 0.13 | 21.7 | | Serum2 | 4.1 | 0.11 | 2.7 | 0.24 | 6.0 | 0.27 | 6.6 | | Serum 3 | 24.0 | 0.68 | 2.8 | 0.30 | 1.2 | 0.74 | 3.1 | | Serum 4 | 106.7 | 1.97 | 1.8 | 2.60 | 2.4 | 3.26 | 3.1 | | Serum 5 | 791.3 | 24.24 | 3.1 | 40.81 | 5.2 | 47.46 | 6.0 | | Serum 6 | 1116.1 | 38.19 | 3.4 | 53.29 | 4.8 | 65.56 | 5.9 | | Serum 7 | 208,778 | 7,482 | 3.6 | 9,400 | 4.5 | 12,014 | 5.8 | Precision Study - Plasma | Sample ID | Grand Mean (mIU/mL) | Within Run SD (mIU/mL) | Within Run CV (%) | Between Run SD (mIU/mL) | Between Run CV (%) | Total SD (mIU/mL) | Total CV (%) | | --- | --- | --- | --- | --- | --- | --- | --- | | Plasma 1 | 1.4 | 0.06 | 4.3 | 0.06 | 4.3 | 0.08 | 5.7 | | Plasma 2 | 4.7 | 0.23 | 4.8 | 0.2 | 4.2 | 0.30 | 6.4 | | Plasma 3 | 675.0 | 19.81 | 2.9 | 21.45 | 3.2 | 29.2 | 4.3 | | Plasma 4 | 1112.3 | 43.35 | 3.9 | 58.09 | 5.2 | 72.48 | 6.5 | # b. Linearity/assay reportable range: # i) Linearity: Fifteen sample dilutions at concentrations across the measuring range of the Total $\beta \mathrm{hCG}$ ( $5^{\text{th}}$ IS) assay (0.6 to approximately $1350\mathrm{mIU / mL}$ ) were prepared by mixing different proportions of a high and a low serum sample. The hCG concentration of the low serum sample was determined using three lots of Total $\beta \mathrm{hCG}$ ( $5^{\text{th}}$ IS) assay and two lots of Total $\beta \mathrm{hCG}$ ( $5^{\text{th}}$ IS) Calibrators. The low sample was tested in replicates of eight and concentration was determined from the mean of the eight results to be approximately 0.03 {5} mIU/mL. The hCG concentration of the high sample was tested in eight replicates with three lots of Total $\beta$ hCG ( $5^{\text{th}}$ IS) assay and one lot of Total $\beta$ hCG ( $5^{\text{th}}$ IS) secondary calibrators, where the highest calibrator level was approximately $1700 \, \text{mIU/mL}$ . The high sample concentrations were determined from the mean of the eight Total $\beta$ hCG ( $5^{\text{th}}$ IS) replicates for each reagent lot. Linearity results from one representative lot are presented below in terms of expected versus observed. Results from other lots were similar. The results support linearity across the claimed measuring range of $0.6 - 1350\mathrm{mIU / mL}$ . | Sample ID | Expected (mIU/mL) | Observed (mIU/mL) | % Recovery | | --- | --- | --- | --- | | Sample 1 | 0.03 | 0 | 0 | | Sample 2 | 0.71 | 0.73 | 102.8 | | Sample 3 | 1.43 | 1.77 | 123.8 | | Sample 4 | 14.25 | 14.38 | 100.9 | | Sample 5 | 42.74 | 41.62 | 97.4 | | Sample 6 | 85.49 | 81.85 | 95.7 | | Sample 7 | 128.23 | 127.56 | 99.5 | | Sample 8 | 178.1 | 171.97 | 96.6 | | Sample 9 | 356.19 | 337.26 | 94.7 | | Sample 10 | 534.29 | 508.57 | 95.2 | | Sample 11 | 712.38 | 664.39 | 93.3 | | Sample 12 | 890.48 | 836.4 | 93.9 | | Sample 13 | 1068.58 | 1003.22 | 93.9 | | Sample 14 | 1246.67 | 1188.67 | 95.3 | | Sample 15 | 1432 | 1424.77 | 99.5 | # ii) Dilution Recovery # (a) Manual Dilution: Dilution recovery was tested on five serum samples and five plasma samples at hCG concentrations between 100 and $120\%$ of $1350~\mathrm{mIU / mL}$ . Six independent dilutions for each sample were made with wash buffer as the diluent at dilutions of 1/5, 1/10, 1/101, 1/200 and 1/250, resulting in sample concentrations of approximately $6 - 1500~\mathrm{mIU / mL}$ . This study was run on two instruments, using one reagent pack lot, two calibrator lots, and two lots of wash buffer. Neat samples were run in replicates of eight. Each sample dilution was measured in replicates of four. Three quality controls were run in replicates of two on each day of each run. {6} A 1/200 dilution is recommended in the Instructions for Use for over-range samples ( $&gt;1350\mathrm{mIU/mL}$ ). When both sample matrices were diluted 1/200, recovery ranged from $97\%$ to $115\%$ . The mean recoveries from manual dilution at 1/200 were $110\%$ for serum samples and $104\%$ for plasma samples. # (b) On-board Dilution: Verification studies were performed to determine the sample dilution recovery of the Access Total $\beta \mathrm{hCG}$ ( $5^{\text{th}}$ IS) assay. Sample dilution recovery was tested on five serum samples and five plasma samples at hCG concentrations between 100 and $120\%$ of $1350 \mathrm{mIU} / \mathrm{mL}$ . Each sample was diluted with wash buffer as the diluent at a dilution of 1/200 on the instrument using two different assay protocol files: dil-hCG or onboard dilution (OBD). This study was run on three instruments, using one reagent pack lot, two calibrator lots, and two lots of wash buffer. All samples were run in replicates of eight. Three quality controls were run in replicates of two on each day of each run. The results are summarized in the tables below. | Sample | Dil-hCG Average % Recovery | Dil-hCG Recovery Range (%) | OBD Average % Recovery | OBD Recovery Range (%) | | --- | --- | --- | --- | --- | | Serum 1 | 109 | 99-118 | 114 | 106-123 | | Serum 2 | 109 | 96-119 | 112 | 106-119 | | Serum 3 | 97 | 88-108 | 100 | 92-108 | | Serum 4 | 95 | 84-102 | 100 | 93-107 | | Serum 5 | 96 | 84-103 | 102 | 93-108 | | Sample | Dil-hCG Average % Recovery | Dil-hCG Recovery Range (%) | OBD Average % Recovery | OBD Recovery Range (%) | | --- | --- | --- | --- | --- | | Plasma 1 | 106 | 94-113 | 109 | 104-123 | | Plasma 2 | 104 | 96-124 | 107 | 101-113 | | Plasma 3 | 101 | 88-114 | 106 | 97-116 | | Plasma 4 | 99 | 87-109 | 106 | 98-113 | | Plasma 5 | 104 | 86-110 | 108 | 102-113 | Individual percent recoveries for serum and plasma generally demonstrated a bias of $\leq 15\%$ . The sponsor has a limitation in the labeling that states that there is potential for end-users to obtain dilution results with a bias $&gt;15\%$ from dil-hCG or OBD assay protocol dilution of serum samples (but not plasma). {7} c. Traceability, Stability, Expected values (controls, calibrators, or methods): **Traceability:** The measurand in the Access Total $\beta\mathrm{hCG}$ (5th IS) Calibrators is traceable to the WHO 5th International Standard (NIBSC Code 07/364). Traceability process is based on EN ISO17511. **Stability:** All stability study protocols were based on CLSI EP25-A. Stability protocols and acceptance criteria for real-time, opened and closed calibrators were reviewed and found to be acceptable. **Value Assignment:** Concentrations of calibrators are assigned through an internal procedure. Primary working calibrators are prepared from WHO 5th IS and secondary working calibrators are value assigned using the primary working calibrators and the Access Immunoassay System. Primary working calibrators and the secondary working calibrators are run together on the Access Immunoassay System in a matched, multi-replicate process. Product (commercial) calibrators are value assigned using the secondary working calibrators on the Access Immunoassay System following a similar value assignment process. The product calibrators are prepared at six levels; 0, 6, 35, 195, 620, and 1,350 mIU/mL. d. Detection limit: LoB, LoD, and LoQ below were determined according to CLSI EP17-A2. i) Limit of Blank (LoB) and Limit of Detection (LoD) One hundred fifty six (156) replicates of the S0 (zero) calibrator were measured for the LoB determination. Six very low hCG concentration serum samples between the anticipated LoB and ten times the anticipated LoB were measured in triplicates over 12 runs (36 replicates in total) to determine the LoD. The study was run on three instruments, using three reagent pack lots and two calibrator lots. Three quality controls were run in replicates of two on each day of each run. The sponsor claims that LoB and LoD is ≤ 0.5 mIU/mL for the Access Total $\beta\mathrm{hCG}$ (5th IS) Assay are supported. ii) Limit of Quantitation (LoQ) The WHO 5th IS was diluted with hCG-negative serum at nominal concentrations of 0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 4.0 and 6.0 mIU/mL to determine 8 {8} the LoQ. The study was run on three instruments, using three reagent pack lots and one calibrator lot. Each sample dilution was run in replicates of ten. Three quality controls were run in duplicates on each day. The sponsor claimed that the LoQ is $0.6\mathrm{mIU / mL}$ and demonstrated that recovery is within $\pm 0.1\mathrm{mIU / mL}$ and $\% \mathrm{CV}$ is $&lt; 20\%$ at the LoQ. # e. Analytical specificity: # i) Cross-reactivity: A cross-reactivity study was performed on pituitary hormones FSH, LH, TSH and $\alpha$ -subunit of hCG. The potential cross-reactants were added to the Access Total $\beta$ hCG ( $5^{\text{th}}$ IS) Calibrator S0 (zero calibrator) and one patient sample with an hCG concentration of $\sim 2.9$ mIU/mL. Results from these cross-reactant spiked samples were evaluated against that of the unspiked sample alone, defined as the control. No cross-reactivity was observed at the concentrations that were tested for LH, FSH, TSH and hCG $\alpha$ -subunit as shown below. | Substance | concentration (mIU/mL) | | --- | --- | | FSH | 1 | | LH | 103 | | TSH | 1000 | | hCG α-subunit | 500 | # ii) Isoform Recognition: Known concentrations of commercially available WHO reference standards (isoforms) were diluted in normal human serum (negative for hCG) for isoform recognition studies. The results (as shown in the table below) demonstrated that the Access Total $\beta \mathrm{hCG}$ ( $5^{\text{th}}$ IS) assay measures four hCG isoforms: intact hCG, $\beta$ subunit of hCG, nicked intact hCG and nicked $\beta \mathrm{hCG}$ , while free $\alpha$ subunit and $\beta$ -core fragment yield no detectable response. | hCG Isoform | WHO Code | Mean % Recovery | | --- | --- | --- | | Intact hCG | 99/688 | 109% | | hCG-β, subunit, highly purified | 99/650 | 193% | | hCGn, nicked dimer | 99/642 | 127% | | hCG-βn, nicked β subunit | 99/692 | 127% | | hCG-β, subunit, less pure | 75/551 | 140% | | hCG-βcf, beta core fragment* | 99/708 | 0.5% | | hCG-α, subunit, highly purified | 99/720 | 0.1% | *Note: high levels of $\beta$ -core fragment were not tested. # iii) Interference Study: {9} An interference study was performed by assessing common sample abnormalities (including hemolysis, icterus, and lipemia), common prescription drugs, over-the-counter drugs, and medications most often prescribed in the patient population for which the test is ordered. Two patient serum samples at hCG concentrations of approximately $2.9\mathrm{mIU/mL}$ and 500 mIU/mL were used for the study. The potential interferents were spiked individually into each patient sample with a low and high concentration of potential interferent. Results from these spiked patient samples were compared to results of the unspiked sample alone, defined as the control. No interference was observed at the concentrations tested as shown below. | Potential Interferent | Low Level | High Level | | --- | --- | --- | | Acetaminophen | 3 mg/dL | 20 mg/dL | | Acetylsalicylic acid | 39 mg/dL | 65 mg/dL | | Bilirubin (conjugated) | 0.3 mg/dL | 40 mg/dL | | Bilirubin (unconjugated) | 1.22 mg/dL | 40 mg/dL | | Hemoglobin | 200 mg/dL | 500 mg/dL | | Heparin (low molecular weight) | 100 U/dL | 7200 U/dL | | Human serum albumin | 3.5 g/dL | 6 g/dL | | Ibuprofen | 7 mg/dL | 50 mg/dL | | Intralipid | 0.3 g/dL | 3 g/dL | | Multi-vitamin | 0.3% (v/v) dilution | 0.9% (v/v) dilution | iv) High Dose Hook Effect: To determine any hook effect, $1,000,000\mathrm{mIU/mL}$ of commercially available WHO reference standards (intact hCG, highly purified hCG $\beta$-subunit, nicked hCG dimer, and nicked hCG $\beta$-subunit) were spiked into one human serum patient sample (negative for hCG). The difference between the different concentrations of hCG isoforms and the Access Total $\beta\mathrm{hCG}$ ($5^{\mathrm{th}}$ IS) S5 calibrator ($\sim 1350\mathrm{mIU/mL}$) was calculated. No high dose hook effect was observed at the concentrations tested. f. Assay cut-off: See detection limit above. 2. Comparison studies: a. Method comparison with predicate device: Patient serum samples with hCG concentrations falling within the measuring range of the ADVIA Centaur Total hCG Assay (2-1000 mIU/mL) were evaluated. A total of 224 serum samples, including 85 fresh and 139 frozen samples were used in the analysis. {10} The study was run at two external sites, one instrument per site, with two reagent pack lots and one calibrator lot for the Total $\beta \mathrm{hCG}$ ( $5^{\text{th}}$ IS) assay. All samples were run in singleton. The correlations between paired measurements for fresh, frozen, and all samples were analyzed. Each analysis was completed by fitting the observed Access Total $\beta \mathrm{hCG}$ ( $5^{\text{th}}$ IS) values (dependant variable, y) into a Passing-Bablok linear regression model, with the observed ADVIA Centaur Total hCG values as the only independent variable (x, standard). Method Comparison vs. ADVIA Centaur Total hCG (Fresh Samples) | N | Concentration Range (mIU/mL) | Intercept (95% CI) | Slope (95% CI) | Correlation Coefficient (r) | | --- | --- | --- | --- | --- | | 85 | 4.9 – 1000 | 1.37 (0.56 – 2.67) | 1.00 (0.97 – 1.02) | 0.99 | Method Comparison vs. ADVIA Centaur Total hCG (Frozen Samples) | N | Concentration Range (mIU/mL) | Intercept (95% CI) | Slope (95% CI) | Correlation Coefficient (r) | | --- | --- | --- | --- | --- | | 139 | 3.2 – 1000 | 3.42 (2.65 – 4.42) | 1.09 (1.07 – 1.12) | 0.99 | Method Comparison vs. ADVIA Centaur Total hCG (All Samples) | N | Concentration Range (mIU/mL) | Intercept (95% CI) | Slope (95% CI) | Correlation Coefficient (r) | | --- | --- | --- | --- | --- | | 224 | 3.2 – 1000 | 2.87 (2.25 – 3.79) | 1.04 (1.02 – 1.06) | 0.99 | # b. Matrix comparison: The sample types for the study were serum (no gel), serum (gel) and lithium-heparin plasma. 42 matched sets of serum and plasma (lithium-heparin) samples were used. The sample (neat or spiked) concentrations spanned the reportable range of the assay (0.5 -1350 mIU/mL). Spiked samples were prepared using commercially available recombinant hCG antigen spiked into the Total $\beta \mathrm{hCG}$ ( $5^{\text{th}}$ IS) calibrator S0 (zero) matrix (BSA). The study was run on one instrument using one reagent pack lot and one calibrator lot. One replicate per sample type was analyzed. Linear regression results are shown below. {11} | Sample Type | N | Intercept | | Slope | | | --- | --- | --- | --- | --- | --- | | | | Estimate | 95% CI | Estimate | 95% CI | | Serum (no gel) vs. serum (gel) | 42 | -0.05 | -0.15 – 0.10 | 0.99 | 0.98 – 1.01 | | Serum (no gel) vs. lithium heparin plasma | 42 | -0.08 | -0.28 – 0.05 | 1.05 | 1.02 - 1.07 | 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: Not applicable. 5. Expected values/Reference range: Expected values for apparently healthy non-pregnant adult females, from the U.S. population, with no known serious chronic or underlying conditions were used in the study. Separate upper 95th percentiles were determined for: a) women $\geq 18$ and $&lt; 40$ years of age; b) women $\geq 40$ years of age; and, c) post-menopausal women (female subjects who had not had a menstrual period for 12 months or more). Samples for the study were prospectively procured at one internal site and one external site. The study was run at one external site on one instrument, using one reagent pack lot and one calibrator lot. The data was analyzed separately for each population using the non-parametric method and following CLSI C28-A3. Additionally, a two-sided non-parametric $95\%$ confidence interval (CI) was used for the analysis. The upper $95^{\text{th}}$ percentile is reported as an estimate of the upper reference limit (URL) of serum hCG levels in apparently healthy non-pregnant female populations. {12} | Subject Category (non-pregnant females) | N | Median (mIU/mL) | Range (mIU/mL) | 95thPercentile (mIU/mL), 95% CI | | --- | --- | --- | --- | --- | | ≥ 18 and < 40 years | 132 | 0 | 0 – 0.6 | 0.3, [0.2 – 0.4] | | ≥ 40 years | 141 | 0 | 0 – 3.1 | 1.5, [1.1 – 2.9] | | Post-menopausal | 134 | 2.8 | 0.1 – 11.6 | 7.7, [6.4 – 10.4] | # N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. # O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...