K130001 · Foshan Care Medical Technology Co., Ltd. · JCX · Jul 25, 2013 · General, Plastic Surgery
Device Facts
Record ID
K130001
Device Name
CARE SU01A SUCTION ASPIRATOR
Applicant
Foshan Care Medical Technology Co., Ltd.
Product Code
JCX · General, Plastic Surgery
Decision Date
Jul 25, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic, Pediatric, 3rd-Party Reviewed
Indications for Use
The CARE SU01A Suction aspirator is intended for professional use to remove infectious materials from wounds or fluids from a patient airway or respiratory support unit. The intended large population for this device is both adult and pediatric patients.
Device Story
AC-powered suction aspirator; provides general suction in hospitals or clinics. Device generates negative pressure via vacuum pump; pressure manually adjusted via regulating valve and vacuum gauge. Negative pressure transmitted through tubing to collection bottle; aspirates secretions or liquids from patient body. Includes 850ml collection bottle; bacteria filter isolates pump from aspirated fluids; mechanical overflow protection in bottle lid prevents fluid ingress to pump. Operated by healthcare professionals. Benefits patient by clearing airway/respiratory support systems or removing infectious wound materials.
Clinical Evidence
No clinical data. Bench testing only. Compliance with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and ISO 10079-1 (performance requirements for electrically powered suction equipment) confirmed.
Technological Characteristics
AC-powered (120 VAC, 60 Hz) vacuum pump. Max vacuum 560 mmHg. 850ml collection bottle with mechanical overflow protection and bacteria filter. Manual vacuum regulation and gauge. Complies with IEC 60601-1, IEC 60601-1-2, and ISO 10079-1. Non-software controlled.
Indications for Use
Indicated for professional use to remove infectious materials from wounds or fluids from a patient airway or respiratory support unit in adult and pediatric patients.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
K130001
# JUL 2 5 2013
# 510(K) SUMMARY
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 807.92
#### Submitter's Information 1.0
### Establishment Registration Name:
Foshan Care Medical Technology Co., Ltd.
The 2F and 3F, G Building, No.1 Huabao Road Southern, Zhangcha, Chancheng District, Foshan, Guangdong, P. R. China 528000
#### Contact Person of applicant
Mr. Gang Wu - Chairman of directors
The 2F and 3F, G Building, No.1 Huabao Road Southern, Zhangcha, Chancheng District, Foshan, Guangdong, P. R. China 528000
TEL: +86-757-8802 3265 FAX: +86-757-8382 8966 Email: wugang@gse.cn
### Contact Person of the Submission:
Ms. Kathy Guo
Foshan Care Medical Technology Co., Ltd.
The 2F and 3F, G Building, No.1 Huabao Road Southern, Zhangcha, Chancheng District, Foshan, Guangdong, P. R. China 528000
Email: medicaldevicefda@163.com
#### Device Information 2.0
| Type of 510(k) submission: | Traditional |
|----------------------------|--------------------------------------------------------|
| Device Common Name: | Power Suction Unit |
| Trade Name: | Model SU01A Suction aspirator |
| Model: | SU01A |
| Classification name: | Apparatus, Suction, Ward Use, Portable, AC-<br>Powered |
| Review Panel: | General & Plastic Surgery |
Page 1 of 4
{1}------------------------------------------------
| Product Code: | JCX |
|--------------------|----------|
| Regulation Class: | II |
| Regulation Number: | 878.4780 |
#### Predicate Device Information 3.0
| Sponsor: | EMG Technology Company, Ltd |
|-----------------------|-----------------------------|
| Device: | Suction Unit, Model:SUA01 |
| 510(K) Number: | K042349 |
| Product Code: | JCX |
| Regulation Class: | II |
| Regulation<br>Number: | 878.4780 |
#### Device description 4.0
Powered suction pumps are described in FDA regulations, 21 CFR 878.4780, as:
"A powered suction pump is an AC-powered device intended to be used to remove infectious materials from wounds or fluids from patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter. The FDA classified the device as a class II medical device".
The SU01 A Suction aspirator is designed to provide general suction for use in hospitals or clinics. This Suction aspirator is a professional medical suctioning device that produces a maximum vacuum of 560 mmHg. It is used with a power supply allowing operation on AC voltage (120 VAC, 60 Hz). The device is equipped with 850ml collection bottle and has been designed for convenient and reliable and reliable usages.
The vacuum pump of the SU01A transmits negative pressure throughout the tubing-system and the collection bottle to patient tubing that aspirates secretions or liquid form the body. The vacuum pump can be manually adjusted using the regulating value and the vacuum gage. The aspirated fluids are collected in the collection bottle that is isolated from the pump by a bacteria filter. It also has a mechanical overflow protection in the lid of the collection bottle to further prevent the fluids from being sucked into the pump.
#### Intended Use 5.0
The CARE SU01A Suction aspirator is intended for professional use to remove infectious materials from wounds or fluids from a patient airway or respiratory support unit. The intended large population for this device is both adult and pediatric patients.
#### Determination of Substantial Equivalence 6.0
Page 2 of 4
{2}------------------------------------------------
## Summary of Technological Characteristics of the Device Compared to the Predicate Device
The CARE SU01 A Suction aspirator uses the same fundamental technology as the EMG SUA01 Suction Unit for most features. The indications for use of CARE SU01A Suction aspirator are the same as EMG SUA01 Suction Unit. CARE SU01A Suction aspirator is similar to the predicate devices in its indications for use and patient population. The user interface is also similar to the predicated devices. The main difference is in the volume of the collection bottle, which is increased but still sufficient for the expected daily volumes. The smaller size improves mobility.
### Summary of Non-clinical Tests:
The CARE SU01 A Suction aspirator complies with voluntary standards for electrical safety, electromagnetic compatibility, and performance. The following quality assurance measures were applied to the development of the system:
- Electrical safety and electromagnetic compatibility testing per IEC 60601-1 ● and IEC 60601-1-2 standards, respectively
- Medical suction equipment Part 1 : Electrically powered suction equipment -● Safety requirements per ISO 10079-1
### Summary of Clinical Tests:
No clinical studies were performed
### CONCLUSION:
Foshan Care considers the SU01 A Suction aspirator to be as safe, as effective, and substantially equivalent to the predicate devices.
#### 7.0 Effectiveness and Safety Considerations
#### Effectiveness:
The SU01 A Suction aspirator complies with ISO10079-1:1999.
### Safety Considerations:
The applicant devices comply with IEC60601-1, Medical electrical equipment - Part l : General requirements for safety and IEC60601-1-2, Medical electrical equipment - Part 1-2: General requirements for safety -Collateral standard: Electromagnetic compatibility - Requirements and tests.
The SU01A Suction aspirator is not a software controlled medical device, the software test according the software guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" is not applicable.
The accessories which contact the patients directly are provided by the hospital but not the manufacture. All the accessories of the SU01 A Suction aspirator are not contact the patients, the biocompatibility test is not applicable.
{3}------------------------------------------------
The clinical performance test is not applicable
#### Comparison to predicate device and conclusion 8.0
### Comparison Analysis
The applicant device has same classification information, same indications and intended use, similar product design, similar technical specification and safety specification.
| | SU01A Suction | SUA01 Suction | Results |
|-------------------------------|---------------|---------------|---------|
| Attribute | Pump | Unit | |
| The pump shall be compliant | Meet the | Meet the | Pass |
| with IEC 60601-1 | requirements | requirements | |
| requirements | | | |
| The pump shall be compliant | Meet the | Meet the | Pass |
| with IEC 60601-1-2 | requirements | requirements | |
| requirements | | | |
| The pump shall be compliant | Meet the | Meet the | Pass |
| with ISO 10079-1 | requirements | requirements | |
| requirements | | | |
| The pump controls shall be | Meet the | Meet the | Pass |
| easily identifiable by user | requirements | requirements | |
| The Collection Bottle should | 850ml | 800ml | Pass |
| have volume reference | | | |
| markings | | | |
| The Collection Bottle should | Meet the | Meet the | Pass |
| be able to withstand | requirements | requirements | |
| maximum pressure delivered | | | |
| by the pump | | | |
| The degree of collapse of the | Less than 0,5 | Less than 0,5 | Pass |
| tubing shall be less than 0,5 | | | |
| throughout its entire length | | | |
| when subjected to the | | | |
| maximum vacuum stated | | | |
| The resistance to implosion | Meet the | Meet the | Pass |
| for collection container | requirements | requirements | |
| The flow rate of maximum | Meet the | Meet the | Pass |
| pumping power | requirements | requirements | |
#### Conclusion:
The applicant device is Substantially Equivalent (SE) to the predicate device which is US legally marketed device. Therefore, the applicant device is determined as safe and effectiveness.
Compare with predicate device, they are very similar in design principle, intended use, functions, material and the adopting applicable standards. The differences between applicant device and predicate device do not raise any new questions of safety or effectiveness.
Page 4 of 4
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white and appears to be a scanned image.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Foshan Care Medical Technology Co., Ltd. % Regulatory Technology Services, LLC Mr. Mark Job 1394 25th Street, NW Buffalo, Minnesota 55313
Re: K130001
Trade/Device Name: Care SU01A Suction Aspirator Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: JCX Dated: July 16, 2013 Received: July 17, 2013
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
July 25, 2013
{5}------------------------------------------------
Page 2 - Mr. Mark Job
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 1 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
FOR
Sincerely yours,
Peter D. Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Statement of Indications for Use
510(k) Number (if known): JS known): Suction aspirator Suction aspirator Device Name: SU01A Model:
Indications for Use:
The CARE SU01A Suction aspirator is intended for professional use to remove infectious The CARD DOGTT Caction abplaison a patient airway or respiratory support unit. The materials from Woulds of for this device is both adult and pediatric patients.
X Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1 =
| DSD-DIVISION SIGN-OFF | |
|------------------------------|---------------------|
| Division of Surgical Devices | |
| 510(k) Number: | K130001 |
| | Joshua C. Nipper -S |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.