← Product Code [NUH](/productcode/NUH) · K124055

# T1040 AKA AURAWAVE (K124055)

_Endurance Therapeutics · NUH · Jul 19, 2013 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K124055

## Device Facts

- **Applicant:** Endurance Therapeutics
- **Product Code:** [NUH](/productcode/NUH.md)
- **Decision Date:** Jul 19, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5890
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

To be used for temporary relief of pain associated with sore and aching muscles in the upper and lower back due to strain from exercise or normal household work activities choose Manual modes 1, 2, 3, 4, 5, 6 or Auto 4. To be used for temporary relief of pain associated with sore and aching muscles in the upper extremities (arm) due to strain from exercise or normal household work activities choose Manual modes 1, 2, 3, 4, 5, 6 or Auto 1 or Auto 3. To be used for temporary relief of pain associated with sore and aching muscles in the lower extremities (leg) due to strain from exercise or normal household work activities choose Manual modes 1, 2, 6 or Auto1 or Auto1 or Auto 4. To be used for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis, choose Manual modes 1 and 3 or Auto 1, Auto 2, Auto 3, or Auto 4. Used to stimulate healthy muscles in order to improve and facilitate muscle performance choose Manual Mode 1 or Auto Mode 2.

## Device Story

T1040 (Aurawave) is a transcutaneous electrical nerve stimulator (TENS) for OTC use. Device delivers electrical pulses via skin electrodes to provide pain relief and muscle stimulation. User selects specific manual or automatic modes based on target body area (back, arm, leg) and clinical goal (pain relief vs. muscle performance). Device operates as a portable stimulator; intended for patient self-administration. Output consists of electrical stimulation patterns designed to manage chronic pain or facilitate muscle function. Clinical benefit includes symptomatic pain management and muscle performance improvement.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on technological characteristics and intended use consistent with predicate TENS devices.

## Technological Characteristics

Transcutaneous Electrical Nerve Stimulator (TENS). Operates via electrical pulse delivery to skin electrodes. Features multiple manual and automatic stimulation modes. Class II device (21 CFR 882.5890).

## Regulatory Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three diagonal lines that curve and merge into a single line at the bottom. The symbol is positioned to the right of the department's name, which is written in a circular arrangement around the symbol. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

## July 19, 2013

Endurance Therapeutics C/O Brad Brezden 122B First Street, NE Dauphin. Manitoba Canada, R7N IB5

Re: K124055

Trade/Device Name: T1040 aka Aurawave Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: NUH, NYN, NYN, NGX, GZJ Dated: July 8, 2013 Received: July 15, 2013

Dear Mr. Brad Brezden

We have reviewed your Section 510(k) promarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA is issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-342 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

## Page 2 - Mr. Brad Brezden

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address:

http://www.fda.gov/MedicalDevices/ResourcesforYow/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to:

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address: http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

## Victor Krauthamer -S

Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known): K124055

Device Name: T1040

Indications For Use:

To be used for temporary relief of pain associated with sore and aching muscles in the upper and lower back due to strain from exercise or normal household work activities choose Manual modes 1, 2, 3, 4, 5, 6 or Auto 4.

To be used for temporary relief of pain associated with sore and aching muscles in the upper extremities (arm) due to strain from exercise or normal household work activities choose Manual modes 1, 2, 3, 4, 5, 6 or Auto 1 or Auto 3. 

To be used for temporary relief of pain associated with sore and aching muscles in the lower extremities (leg) due to strain from exercise or normal household work activities choose Manual modes 1, 2, 6 or Auto1 or Auto1 or Auto 4.

To be used for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis, choose Manual modes 1 and 3 or Auto 1, Auto 2, Auto 3, or Auto 4.

Used to stimulate healthy muscles in order to improve and facilitate muscle performance choose Manual Mode 1 or Auto Mode 2.

AND/OR Over-The-Counter Use Prescription Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ので、この時には、その他の人気の人気の人気の人気の人気の人気の人気の人気の出来ないと思います。 この時は、 この時には、 この日に、 この

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Victor Krauthamer -S                                           |         |
|----------------------------------------------------------------|---------|
| 2013.07.18 18:05:41 -04'00'                                    |         |
| (Division Sign Off)                                            |         |
| Division of Neurological and Physical Medicine Devices (DNPMD) |         |
| 510(k) Number                                                  | K124055 |

Page 1 of 1

---

**Source:** [https://fda-staging.innolitics.com/device/K124055](https://fda-staging.innolitics.com/device/K124055)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
