← Product Code [OCY](/productcode/OCY) · K124052

# OSTAVI HYDROPHILIC GUIDEWIRE (K124052)

_Lake Region Medical · OCY · Feb 26, 2013 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K124052

## Device Facts

- **Applicant:** Lake Region Medical
- **Product Code:** [OCY](/productcode/OCY.md)
- **Decision Date:** Feb 26, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The endoscopic guidewire is intended for use in selective cannulation of the biliary ducts including, but not limited to the common bile, pancreatic, cystic, right and left hepatic ducts. The endoscopic guidewire is designed to be used during endoscopic pancreatic -biliary procedures for catheter introduction and exchanges of catheters, cannulas and sphicterotomes.

## Device Story

NaviPro™ Guidewire is a steerable, metallic (Nitinol) core guidewire with a polymer jacket and hydrophilic coating. Used by physicians during endoscopic pancreatic-biliary procedures to facilitate selective duct cannulation and the introduction or exchange of catheters, cannulas, and sphincterotomes. The device provides a track for interventional tools; it is available in various lengths and diameters with straight or pre-shaped distal tips. It is a manual, non-powered instrument.

## Clinical Evidence

Bench testing only. No clinical data provided. Non-clinical testing included visual, tactile, dimensional, and mechanical attribute inspections to demonstrate functionality and performance comparable to legally marketed devices. Biocompatibility was addressed via risk analysis, concluding no additional testing was required.

## Technological Characteristics

Materials: Nitinol core, polymer (polyurethane) jacket, hydrophilic coating. Dimensions: 260cm length; 0.018", 0.025", 0.035" diameters. Features: Steerable, straight or pre-shaped distal tip. Standards: ISO 10993 (biocompatibility), ISO 11070 (sterile guidewires), ISO 15223, EN 980, ISO 594.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Taxi Endoscopic Guidewire ([K081708](/device/K081708.md))
- Hydrophilic Coated Guidewire (ZIPwire) ([K000011](/device/K000011.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K124052
pg. 1 of 2

## SECTION 1.0: 510(k) SUMMARY

#### 1:1 MANUFACTURER / REGISTRATION INFORMATION CONTRACTION Lake Region Medical Contact Person: Tracy Mever 340 Lake Hazeltine Drive Title: Regulatory Specialist Chaska, MN 55318-1029 USA Telephone: 952-641-8510 FDA REGISTRATION NUMBER: 2126666 Fax: 952-448-3441

1:24 TRADE (RAME (PROPRIETATY NAME) ... (PA) ... (2) ... ) ... (2012) ... (2012) ... (2012) ... (2012) ... (2012) ... (2012) ... (2012) ... (2012) ... (2012) ... (2012) ... ( NAVIPRO™ Guidewire

1:3 DEVICE COMMON NAMES/USUAL NAMES/CLASSIFICATION NAMES CLASSIFICATION NAMES CLASS CLASS CLASS CLASS CLASS CLASS CLASS CLASS CLASS CLASS CLASS CLASS C These devices are commonly known as guides, Guidewires, or spring Guidewires. The current classification names and product codes are Gastroenterology/Urology (OCY).

11:40 CLASS OF DEVICE (1) 1979 - 1 (3) 2008 - 12 (300 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 This type of Guidewire was originally listed as a Class II device by the Gastroenterology/Urology (OCY) review panel.

1:5 {{CATIDENTIFICATION OF PREDICATE'DEVICE(s){ } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } } }

| 510(k) Number | Manufacturer        | Device Name                            |
|---------------|---------------------|----------------------------------------|
| K081708       | Lake Region Medical | Taxi Endoscopic Guidewire              |
| K000011       | Lake Region Medical | Hydrophilic Coated Guidewire (ZIPwire) |

4.6 KE COEVICE DESCRIPTION PARTICLE FOR CONTROLLERS COLOR COLOR COLOR COLOR COLOR COLLECTION CONSECTION CONSECTION Utilizing proprietary processes, these guides are constructed from a steerable, metallic core (Witinol) with a polymer jacket (polyurethane) coating. A hydrophilic coating is applied over the radiopaque polymer jacket. Guidewires are available in 260cm length and in diameters 0.018", 0.025" depending on specific design requirements. Guidewires may have a straight or a pre-shaped distal tip and are available in different tip flexibilities, which are dependent on the grind configuration of the core.

OUTSIDE DIAMETER: 0.018", 025" and .035"

| LENGTHS:<br>Status and any and the country of the first of the country of                                                                                                                                           | come of the revelopment of the relations of the program more of the many of the many of the many of the many of the many of the many of the many<br>260 cn<br>Ciri<br>Company of the program and the first of the first of the first of the first of the first of the first of the first of the first of the first of the first of the first of the<br> |  | and the country of the production of the first of the first of the first | (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) = (1) =<br>the control control control control control and the program and consistence of the supportunity of the may be any and |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Comments of Children Children<br> | and the minister of market in the successful and the best of the country of the                                                                                                                                                                                                                                                                         |  |                                                                          | below the country of the country of the provide the provide the provide ware support want any are and                                                                                                                                                                                                  |

1:7 COMPLIANCE WITH APPLICABLE STANDARDS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . The NaviPro™ Guidewire is in compliance with ISO 10993, ISO 11070, ISO 15223, EN 980, ISO 594

## FEB 2 6 2013

{1}------------------------------------------------

#### SECTION 1.0: 510(k) SUMMARY

## 1.8 INTENDED USE STATEMENT, CARDEN CONSELLER

The endoscopic guidewire is intended for use in selective cannulation of the biliary ducts including, but not limited to the common bile, pancreatic, cystic, right and left hepatic ducts. The endoscopic guidewire is designed to be used during endoscopic pancreatic -biliary procedures for catheter introduction and exchanges of catheters, cannulas and sphicterotomes.

1:9 A CONTRAINDICATIONS CONTRAINDICATIONS CONTRACTOR CONTRACTORS CONTRACTOR None Known

1.10 COMPARISON COMPARISON COMMENT OF COMPARISON COMPARISON r 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - The NaviPro™ Guidewire is substantially equivalent to the Taxi Endoscopic Guidewire with 510(k) number K081708 with desgin changes.

The design is of the NaviPro™ is substantially equivalent to the Hydrophilic Coated (ZIPwire) Guidewire with 510(k) K000011.

2.1110 QUALIFICATION TESTING: 14.5 2017 10:16 10.00 10.00 10.00 10.00 10.00 10.00 The conclusions drawn from non-clinical and biocompatibility testing demonstrate the device is as safe, as effective and performs as safely and effectively as the legally marketed device.

NON-CLINICAL TESTING CONSULT CONTRACT CONTRACT CONSULT CONTRACT CONTRACT 1000 1000 2000 2000 2000 2000 1000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 2000 In order to demonstrate the safety and effectiveness of the NaviPro™ Guidewire, Lake Region Medical performed testing to establish requirements. Test pieces were tested and inspected according to established specific inspection criteria requirements for visual/tactile, dimensional and mechanical attributes. The results of these tests demonstrated the functionality and performance characteristics of these guidewire are comparable to the similar currently marketed devices. See section 13.0 for the results.

BIOCOMPATIBILITY TESTING . 00 Million Collection Collection Conce Risk analysis concluded that there was no additional biocompatibility testing required.

# 1:12 SUBSTANTIAL EQUIVALENCE DATA LA LA LA LEGENCE PORTER PO

The NaviPro™ Guidewire has a similar intended use as Taxi Endoscopic Guidewires legally marketed by Lake Region Medical and cleared by 510(k) K081708.

The NaviPro™ Guidewire is the exact same design as the ZIPwire, with a different intended use, legally marketed by Lake Region Medical and cleared by 510(k) K000011.

The NaviPro™ Guidewire has the exact same physical characteristics as the secondary predicate device (ZIPwire). The safety and effectiveness of these differences have been proven in the aforementioned Qualification Testing.

The NaviPro ""Guidewire is substantially equivalent to the Taxi Endoscopic Guidewires cleared under 510(k) K081708 and to the ZIPwire cleared under 510(k) K000011. All test results support the claim of substantial equivalence.

{2}------------------------------------------------

### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

## February 26, 2013

Lake Region Medical % Ms. Tracy Meyer Regulatory Specialist 340 Lake Hazeltine Drive CHASKA MN 55318

Re: K124052

Trade/Device Name: NaviProTM Guidewire Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: OCY Dated: December 28, 2012 Received: December 31, 2012

Dear Ms. Meyer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{3}------------------------------------------------

Page 2 - Ms. Tracy Meyer

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/3/Picture/7 description: The image shows the text "Benjamin R. Fisher - S". The text is in a simple, sans-serif font and is black. The letters "R.F" are stylized with a geometric design, with the letters overlapping and outlined with multiple lines. The overall impression is a simple logo or nameplate.

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## INDICATIONS FOR USE

K124052 510(k) NUMBER (IF KNOWN):

DEVICE NAME: NaviPro™ Guidewire

INDICATIONS FOR USE:

The endoscopic guidewire is intended for use in selective cannulation of the billary ducts including, but not limited to the common bile, pancreatic, cystic, right and left hepatic ducts. The endoscopic guidewire is designed to be used during endoscopic pancreatic-biliary procedures for catheter introduction and exchanges of catheters, cannulas and sphicterotomes.

PRESCRIPTION USE x (Part 21 CFR 801 Subpart D) AND/OR

OVER-THE-COUNTER USE (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

PAGE 1 of

# Benjam 2013.02 2

(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K124052

7 | P a g e

---

**Source:** [https://fda-staging.innolitics.com/device/K124052](https://fda-staging.innolitics.com/device/K124052)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).