K124051 · Somersault Orthopedics, Inc. · LLZ · May 17, 2013 · Radiology
Device Facts
Record ID
K124051
Device Name
THE VAULT SYSTEM
Applicant
Somersault Orthopedics, Inc.
Product Code
LLZ · Radiology
Decision Date
May 17, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
The VAULT® System is intended for use as a software interface and image manipulation system for the transfer of imaging information from a medical scanner such as Computerized Axial Tomography (CT) or Magnetic Resonance Imaging (MRI). It is also intended as pre-operative software for simulating/evaluating implant placement and surgical treatment options. The physician chooses the out-put data file for printing and/or subsequent use in CAD modeling or CNC/Rapid-prototyping.
Device Story
VAULT® System is a web-based surgical planning software for orthopedic and trauma surgeons. Input: DICOM medical images (CT/MRI). Operation: System converts DICOM images to graphical formats (JPEG, BMP, PNG, TIFF) prior to planning; provides tools for image manipulation, measurement of anatomical features/landmarks, and overlaying of implant templates. Output: Surgical plan data files for printing, CAD modeling, or CNC/rapid-prototyping. Usage: Surgeon-directed; accessed via secure web interface. Clinical impact: Facilitates preoperative simulation and implant sizing, aiding surgical decision-making and potentially improving procedural outcomes.
Clinical Evidence
Bench testing only. Validation included objective and subjective testing of functional requirements, anatomical/phantom model digital file testing, DICOM process verification, and image calculation/measurement accuracy. Safety risk/hazard analysis performed per ISO 14971:2007. No clinical data presented.
Technological Characteristics
Web-based software interface; PC compatible. Supports DICOM standard for image input. Performs image manipulation, measurement, and template overlay. Software validation per FDA guidance; risk management per ISO 14971:2007. Connectivity: Web-accessed via secure login.
Indications for Use
Indicated for use by physicians as a preoperative planning tool for orthopedic and trauma surgery. Assists in simulating and evaluating implant placement and surgical treatment options using CT or MRI imaging data.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K124051
## Section 510(k) Releasable Summary
VAULT® System Surgery Planning Software
# MAY 1 7 2013
Proprietary Name:
Surgery Planning Software
Classification Name
& CFR Reference:
Common Name:
Sec. 892.2050 System, image processing, radiological
(a) Identification. A picture archiving and communications system is a device that provides one or more capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images. Its' hardware components may include workstations, digitizers, communications devices, computers, video monitors, magnetic, optical disk, or other digital data storage devices, and hardcopy devices. The software components may provide functions for performing operations related to image manipulation, enhancement, compression or quantification.
(b) Classification. Class II (special controls; voluntary standards-Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern). [63 FR 23387. Apr. 29. 1998]
Regulatory Class: Class II
Device Product Code: LLZ
Company Contact:
Mr. David W. Schlerf Somersault Orthopedics, Inc. 142 N. Milpitas Blvd., Suite 262 Milpitas, CA 95035-4401 925.768.0247 - Direct
April 12. 2013 Date Prepared:
Predicate Device(s): 1) Mimics Software - K073468 · 2) TraumaCAD Software- K073714
Summary Device Description: The WAULT® System software described here was developed in conformance with reference to the FDA Guidance Document for Industry "Guidance for the Submission of Premarket Notifications for Medical Image Management Devices, July 27, 2000". Based on the information contained in Section G of that document, a final determination of submission content was developed. A secondary reference entitled "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. May 11, 2005 was also used and resulted in a determination of a "MODERATE" level of concern for the software.
187
{1}------------------------------------------------
The UAULT® System software is made available to the user via a web-accessed software interface. The program is a surgeon directed surgical planning package primarily but not exclusively directed at trauma and orthopedic indications. After secure log-in the user requests, creates, reviews and finally authorizes their desired surgical plan. When authorizing, the surgeon/user may choose additional options such as implant sizing and/or various file output options.
Discussion of Similarities and Differences: The primary and most immediately noticeable difference is that the VAULT® Sustem software is Web-based while the predicate devices are local (PC/Laptop/Notebook) or local server based (hospital/clinic/office). Locally installed software presents minor to major complicating factors related to version, hardware, speed, operating system conflicts, etc. Those complicating factors are eliminated or reduced by direct company control of single edition server based web-portal program access. All users access the same optimized revision level software.
A second apparent difference in the basic operation of the Vault System and the predicate(s) software is the initial mass conversion and use of graphical images such as JPEG, BMP, PNG, of TIFF by the Vault System for planning. Both Mimics and TraumaCAD perform their planning on the DICOM images and then save/export files to JPEG file format(s) for presentation and export/import purposes.
The difference between the systems is the fact that the predicate software packages perform DICOM to JPEG conversion as a post-plan individual step. The WAULT® Sustem software performs this activity initially, converting all image files at once before planning and saving or exporting, Functionally there is no significant difference between the program capabilities other than improved performance on the part of the VAULT® System software.
188
{2}------------------------------------------------
| Feature | VAULT® System | Mimics<br>K073468 | TraumaCAD<br>K073714 |
|-----------------|-------------------------------------------------------------------------|---------------------------------------------------------|----------------------------------------------------------|
| Indication for | The VAULT® System is intended for | Mimics software is intended | The TraumaCAD program is |
| Use: | use as a software interface and | for use as a software interface | indicated for assisting healthcare |
| | image manipulation system for the | and image manipulation | professionals in preoperative |
| | transfer of imaging information | system for the transfer of | planning of orthopedic surgery. |
| | from a medical scanner such as | imaging information from a | The device allows for overlaying |
| | Computerized Axial Tomography | medical scanner such as a CT | of prosthesis templates on |
| | (CT) or Magnetic Resonance | scanner or Magnetic | radiological images, and |
| | Imaging (MRI). It is also intended | Resonance Imaging scanner | includes tools for performing |
| | as pre-operative software for | to an output file. It is also | measurements on the image and |
| | simulating/evaluating implant | intended as preoperative | for positioning the template. |
| | placement and surgical treatment | software for simulating / | Clinical judgments and |
| | options. The physician chooses the | evaluating surgical treatment<br>options. Mimics is not | experience are required to<br>properly use the software. |
| | out-put data file for printing and/or<br>subsequent use in CAD modeling | intended for mammography | |
| | or CNC/Rapid-prototyping | imaging. | |
| Classification: | Class II | Class II | Class II |
| ProCode: | LLZ | 1.1.7. | 1.1.7. |
| Computer: | PC Compatible | PC & Workstation | PC Compatible |
| | | Compatible | |
| # of on-Screen | J | ()ne | Onc |
| . Images | | | |
| Runs on | Yes | ا es | Yes |
| Server | | | |
| Elective | | | |
| Procedures | Yes | Yes | Yes |
| Overlav & | | | |
| Templates | Yes | Yes | Yes |
| Preoperative | | | |
| Planning | Yes | Yes | Yes |
| Controls Life | | | |
| Saving Devices | No | No | No |
| Uses DICOM | | | |
| images | Yes | Yes | Yes |
| Expandability | Yes | Yes | Yes |
| | | | |
Summary Non-Clinical Performance Data: Performance data were evaluated by conducting and repeating a defined series of objective and subjective test conditions. Results of this testing demonstrates that the UAULT® System provides an equivalent level of performance and clinical utility to the cited equivalent comparison software systems. Functional requirements as defined by the VAULT® System Software Requirements Specification (SRS) were tested and traceability was performed and documented using FDA's General Principles of Software Validation guidance document.
Safety requirements were developed using a safety risk/hazard analysis based on ISO 14971:2007 approach. Validation included boundary values and stress testing as defined by the FDA's Guidance for the Content of Premarket Submission for Software Contained in Medical Devices guidance document. Software testing addressed the following areas of functionally.
- . Anatomical and phantom model digital file testing demonstrated the required level of accuracy and functionality.
- . Digital file image upload controlled by DICOM process met specifications.
{3}------------------------------------------------
- Image file integrity, accuracy and suitability following required conversion, save and . transfer operations met all specifications.
- . Image calculations & measurement of anatomic features and landmarks meets specifications.
Indication for Use: The VAULT® System is intended for use as a software interface and image manipulation system for the transfer of imaging information from a medical scanner such as Computerized Axial Tomography (CT) or Magnetic Resonance Imaging (MRI). It is also intended as pre-operative software for simulating/evaluating implant placement and surgical treatment options. The physician chooses the out-put data file for printing and/or subsequent use in CAD modeling or CNC/Rapid-prototyping.
Conclusion: Substantial equivalence is represented by comparison with the published capabilities of the predicate software devices. The referenced devices have equivalent uses, restrictions, clinical utility, safety and effectiveness to the VAULT® System. It is concluded that the VAULT® System may be reasonably expected to perform in a manner equivalent to the predicate devices. No new issues of either safety or effectiveness are raised.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol resembling an abstract eagle or bird-like figure with three wing-like shapes.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 17, 2013
Somersault Orthopedics, Inc. % Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street NW BUFFALO MN 55313
Re: K124051
Trade/Device Name: Vault® System Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Il Product Code: LLZ Dated: April 15, 2013 Received: April 19, 2013
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Mr. Job
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.
Sincerely yours,
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K124051
Device Name:
VAULT® System
Indications for Use:
The VAULT® System is intended for use as a software interface and image manipulation system for the transfer of imaging information from a medical scanner such as Computerized Axial Tomography (CT) or Magnetic Resonance Imaging (MRI). It is also intended as preoperative software for simulating implant placement and surgical treatment options. The physician chooses the out-put data file for printing and/or subsequent use in CAD modeling or CNC/Rapid-prototyping.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Smh7).
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k): K124051
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.