ImageQube is intended for use by a physician or other medical professionals in the display and interpretation of medical images and demographic detail from all institutional imaging modalities, including, but not limited to, CT, MRI, NM, DR, US, PET Fusion, Angio and MG (including display of DICOM overlays and 3D Mammography images), along with secondary capture devices, such as film digitizers or other imaging sources. The ImageQube is designed for display, interpretation, storage and distribution of all modalities. Only pre-processed DICOM For Presentation images can be interpreted for primary diagnosis in mammography. Lossy compressed mammographic images and digitized film screen images must not be viewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor meeting all the technical specifications required by the FDA for the Performance of Screening and Diagnostic Mammography. Images that are printed to film must be done using an FDA-approved printer for the diagnosis of digital mammography images. Efficient mammography screening makes toolbars and thumbnails available on each monitor, while also handling DICOM overlay display. Acquired medical images may be displayed and manipulated further utilizing Multi-Planar Reconstruction (MPR), Anatomic Triangulation (AT), Dynamic Cross-Referencing, Maximum Intensity Projection (MIP), Slab and 3-D display, sent to and retrieved by radiologists in-house at facilities or at remote sites, stored, archived or printed. The ImageQube can operate as an independent device, or can also be interfaced with Rational Imaging PACS systems. Annotated print pages, transcribed reports and Key Image Summaries can also be accessed.
Device Story
ImageQube is a medical imaging software platform for acquisition, display, processing, and distribution of DICOM images from various modalities (CT, MRI, etc.). Operated by radiologists and medical professionals in clinical settings, it enables image manipulation via tools like Multi-Planar Reconstruction (MPR), Anatomic Triangulation (AT), Dynamic Cross-Referencing, Maximum Intensity Projection (MIP), and 3D display. The device functions as a standalone workstation or interfaces with existing PACS systems. It facilitates clinical decision-making by providing tools for image review, interpretation, and reporting, including access to annotated print pages and Key Image Summaries. Benefits include improved workflow efficiency and diagnostic support through advanced visualization capabilities.
Clinical Evidence
No clinical data. Substantial equivalence supported by software validation confirming all device features were compliant with software requirements.
Technological Characteristics
Software-based imaging processing system. Supports DICOM data exchange and JPEG image compression. Features include MIP, MPR, DICOM overlay, PET/CT fusion, slab, critical results, mammography, and peer review. Operates as an independent device or interfaces with Rational Imaging PACS.
Indications for Use
Indicated for use by physicians and medical professionals for display, interpretation, storage, and distribution of medical images from institutional modalities (CT, MRI, NM, DR, US, PET, Angio, MG). Includes support for 3D mammography and secondary capture devices. Contraindicated for primary diagnosis using lossy compressed mammographic images or digitized film screen images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Section 6
510(k) Summary
# 6. 510(k) Summary
This 510(k) summary information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT: | Intuitive Imaging Informatics, LLC |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| DATE PREPARED: | 17 July 2012 |
| CONTACT PERSON: | Cindy Simoni<br>Intuitive Imaging Informatics, LLC<br>30 Hackamore Lane, STE 6<br>Bell Canyon, CA 91307<br>Phone: 818.347.8909, x110<br>Fax: 888.908.6242 |
| TRADE NAME: | ImageQube |
| CLASSIFICATION NAME: | System, Imaging Processing |
| DEVICE CLASSIFICATION: | Class II |
| REGULATION NUMBER | 892.2050 |
| PRODUCT CODE | LLZ |
| PREDICATE DEVICES: | ImageQube (K051037)<br>Amicas PACS 6.0 (K082144)<br>Candelis ImageGrid (K080333)<br>Synapse Obliquus MIP/MPR/Fusion (K100881) |
# Substantially Equivalent To:
The ImageQube is substantially equivalent in intended use, principal of operation and technological characteristics to the current marketed ImageQube, Amicas PACS 6.0, Candelis ImageGrid and Synapse Obliquus MIP/MPR Fusion devices.
#### Description of the Device Subject to Premarket Notification:
ImageQube is designed for use by a physician or other medical professionals in the acquisition of medical images and demographic detail from all institutional imaging modalities, including, but not limited to CT, MRI, NM, DR, US, PET Fusion, Angio and MG (including display of DICOM overlay and 3D Mammography images), along with secondary capture devices, such as film digitizers or other imaging sources. The acquired medical images and demographic information may be displayed, processed, reviewed, sent to and retrieved by radiologists at remote sites, stored, archived or printed. Multiplanar Reconstruction (MPR), Anatomic Triangulation (AT), Dynamic Cross-
Intuitive Imaging Informatics, Inc. Modified ImageQube
Page 11 of 65 Premarket Notification
# JAN 2 9 2013
{1}------------------------------------------------
Referencing, Maximum Intensity Projection (MIP), Slab and 3D display are also available for optional use.
## Indication for Use:
ImageQube is intended for use by a physician or other medical professionals in the display and interpretation of medical images and demographic detail from all institutional imaging modalities, including, but not limited to, CT, MRI, NM, DR, US, PET Fusion, Angio and MG (including display of DICOM overlay and 3D Mammography images), along with secondary capture devices, such as film digitizers or other imaging sources. The ImageQube is designed for display, interpretation, storage and distribution of all modalities.
Only pre-processed DICOM For Presentation images can be interpreted for primary diagnosis in mammography. Lossy compressed mammographic images and digitized film screen images must not be viewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor meeting all the technical specifications required by the FDA for the Performance of Screening and Diagnostic Mammography. Images that are printed to film must be done using an FDAapproved printer for the diagnosis of digital mammography images. Efficient mammography screening makes toolbars and thumbnails available on each monitor, while also handling DICOM overlay display.
Acquired medical images may be displayed and manipulated further utilizing Multi-Planar Reconstruction (MPR), Anatomic Triangulation (AT), Dynamic Cross-Referencing, Maximum Intensity Projection (MIP), Slab and 3-D display, sent to and retrieved by radiologists in-house at facilities or at remote sites, stored, archived or printed. The ImageQube can operate as an independent device, or can also be interfaced with Rational Imaging PACS systems. Annotated print pages, transcribed reports and Key Image Summaries can also be accessed.
# Technological Characteristics:
The ImageQube has the same technological characteristics and is similar in overall. design, principal of operation and configuration compared to the predicates. The table below illustrates the similarities and differences in Technological Characteristics between the devices.
| | Modified ImageQube | ImageQube<br>(K051037) | Amicas PACS 6.0<br>(K082144) | Candelis ImageGrid<br>(K080333) | Synapse Obliquus<br>MIP/MPR/Fusion<br>(K100881) |
|------------------------------------------|--------------------|------------------------|------------------------------|---------------------------------|-------------------------------------------------|
| Use | | | | | |
| Storing,<br>distributing,<br>displaying, | Yes | Yes | Yes | Yes | Yes |
Intuitive Imaging Informatics, Inc. Modified ImageQube
Page 12 of 65 Premarket Notification
{2}------------------------------------------------
# Section 6
# 510(k) Summary
| | Modified<br>ImageQube | ImageQube<br>(K051037) | Amicas PACS<br>6.0<br>(K082144) | Candelis<br>ImageGrid<br>(K080333) | Synapse<br>Obliquus<br>MIP/MPR/<br>Fusion<br>(K100881) |
|----------------------------------------------------------|-----------------------|------------------------|---------------------------------|------------------------------------|--------------------------------------------------------|
| analyzing,<br>manipulating<br>and enhancing<br>of images | | | | | |
| Standards | | | | | |
| DICOM (data<br>exchange) | Yes | Yes | Yes | Yes | Yes |
| JPEG (image<br>compression) | Yes | Yes | Yes | Yes | Yes |
| Features | | | | | |
| Maximum<br>Intensity<br>Projection<br>(MIP) | Yes | No | Yes | Yes | Yes |
| Multiplanar<br>Resolution<br>(MPR) | Yes | No | Yes | Yes | Yes |
| DICOM<br>Overlay | Yes | No | Yes | Yes | No |
| PET/CT Fusion | Yes | No | No | Yes | Yes |
| Slab | Yes | No | Yes | Unknown | Unknown |
| Critical Results | Yes | No | Yes | Unknown | Yes |
| Mammography | Yes | No | Yes | Yes | No |
| Peer Review | Yes | No | Unknown | Unknown | Yes |
# Performance
Support of the substantial equivalence of the ImageQube device was provided as a result of software validation, which confirms all features of the ImageQube device were compliant with the software requirements.
# Basis for Determination of Substantial Equivalence:
Upon reviewing and comparing intended use, design, principle of operation and overall technological characteristics, the ImageQube is determined by Intuitive Imaging Informatics, LLC to be substantially equivalent to existing legally marketed devices.
Intuitive Imaging Informatics, Inc. Modified ImageQube
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services-USA. The logo features a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and the USA. The department's name is written in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
# January 29, 2013
Intuitive Imaging Informatics, LLC c/o Mr. Jeff D Rongero Senior Project Manager 12 Laboratory Drive RESEARCH TRIANGLE PARK NC 27709 .
Re: K124048
.
Trade/Device Name: ImageQube Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 21, 2012 Received: December 31, 2012
Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2-Mr. Rongero
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Sean M. Boyd -S for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 5. Indications for Use Statement
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
## Device Name: ImageQube
#### Indications for Use:
ImageQube is intended for use by a physician or other medical professionals in the display and interpretation of medical images and demographic detail from all institutional imaging modalities, including, but not limited to, CT, MRI, NM, DR, US, PET Fusion, Angio and MG (including display of DICOM overlays and 3D Mammography images), along with secondary capture devices, such as film digitizers or other imaging sources. The ImageQube is designed for display, interpretation, storage and distribution of all modalities.
Only pre-processed DICOM For Presentation images can be interpreted for primary diagnosis in mammography. Lossy compressed mammographic images and digitized film screen images must not be viewed for primary image interpretations. Mammographic images may only be interpreted using an FDA approved monitor meeting all the technical specifications required by the FDA for the Performance of Screening and Diagnostic Mammography. Images that are printed to film must be done using an FDA-approved printer for the diagnosis of digital mammography images. Efficient mammography screening makes toolbars and thumbnails available on each monitor, while also handling DICOM overlay display.
Acquired medical images may be displayed and manipulated further utilizing Multi-Planar Reconstruction (MPR), Anatomic Triangulation (AT), Dynamic Cross-Referencing, Maximum Intensity Projection (MIP), Slab and 3-D display, sent to and retrieved by radiologists in-house at facilities or at remote sites, stored, archived or printed. The ImageQube can operate as an independent device, or can also be interfaced with Rational Imaging PACS systems. Annotated print pages, transcribed reports and Key Image Summaries can also be accessed.
#### AND/OR
Prescription Use X_ (21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices and Radiological Health (OIR)
Division Sign-off
Office of In Vitro Diagnostic Device Evaluation and Safety 510k ¥124048
Page of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.