LIGHTLAS FAMILY OF MULTI-WAVELENGTH MEDICAL LASER SYSTEM
Applicant
Lightmed Corp.
Product Code
HQF · Ophthalmic
Decision Date
Sep 26, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4390
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
LightLas577/670 Multi-Wavelength Medical Laser System is an ophthalmic laser instrument intended for use by ophthalmologists. The 577 nm wavelength laser is indicated for retinal and pan-retinal photocoagulation, endophotocoagulation, macular photocoagulation to treat leaking vessels, and laser trabeculoplasty. The 670 nm wavelength laser is indicated for retinal and pan-retinal photocoagulation.
Device Story
Ophthalmic laser system consisting of console, electronic control, power supplies, and laser delivery units (LDU). LDUs include slitlamp attachments (CSO SL980, SL990, Zeiss SL30), laser indirect ophthalmoscopes (Heine Omega 500, Keeler All pupil II), and endophotocoagulation handpieces. Operated by ophthalmologists in clinical settings. User selects wavelength (577nm or 670nm) and parameters via LCD touch screen. System software controls wavelength selection and output ports. Laser energy delivered to ocular tissue for photocoagulation or trabeculoplasty. Provides therapeutic benefit by treating retinal conditions and leaking vessels. Safety features include door interlocks, constant monitoring of safety systems, and displays for user intervention.
Clinical Evidence
Bench testing only. Evaluated per IEC60601-1, IEC60601-1-1, IEC60601-1-2, IEC60825-1, IEC60601-1-4, IEC60601-2-22, and ISO14971. Testing included assembly, calibration, and alignment of console and delivery units. Results confirmed compliance with specifications.
Technological Characteristics
Semiconductor/Diode laser; 577nm/670nm wavelengths; continuous wave and micro-pulse (10% duty cycle) modes; thermoelectric (Peltier) cooling; coaxial red laser diode aiming beam (635-650nm, max 1.0mW); 100-240 Vac power; Class IV laser safety.
Indications for Use
Indicated for ophthalmologists to perform retinal/pan-retinal photocoagulation, endophotocoagulation, macular photocoagulation for leaking vessels, and laser trabeculoplasty using 577nm or 670nm wavelengths.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
{0}------------------------------------------------
### Section 11 510(k) Summary or Statement
# 510(K) Summary .
ﺬ
This is 510(K) summary in accordance with CFR807.82(c).
| | Submitter Information: | | | | |
|---------------------------------------------------------------------------------------------------------------------------------|------------------------|---------------------------------------------|--|--|--|
| | Submitter: | LightMed Corporation | | | |
| | Address: | NO.1-1, Lane1, Pao-An St. Sec. 3, | | | |
| | | Shulin Dist., New Taipei City 23861, Taiwan | | | |
| Owner/ Operation Number: Mr. Gary Lee, President / 9040850<br>Contact person in Taiwan: Jocelyn Liu, Regulatory Affairs Officer | | | | | |
| | | | | | |
| | FAX: +886-2-2676-4920 | | | | |
| | | | | | |
- Device Name B.
Product Name: LightLas577/670 Multi-Wavelength Medical Laser System Common Name: Ophthalmic Laser, Surgical Laser Classification Name: 86 HQF, Laser, Ophthalmic
Regulation Name: 21 CFR 886.4390, Ophthalmic Laser Regulatory Class: II Performance standards: 21 CFR 1040.10 & 1040.11
#### C. Device Technological Characteristics to Predicate Device:
| Manufacturer | LightMed Corporation<br>LightLas532/670 | LightMed Corporation<br>LightLas577/670 |
|-------------------|-----------------------------------------|----------------------------------------------------------|
| 510K Number | K103547 | - |
| Lasing Medium | Frequency doubled<br>Nd:YAG/Diode | Semiconductor/Diode |
| Pulsing System | Continuous | Continuous |
| Output Wavelength | 532nm/670nm | 577nm/670nm |
| Micro Pulse | - | 10% Duty Cycle<br>(Power"On" 200μs,<br>Power"Off" 1800μs |
{1}------------------------------------------------
| | | from 0.01s-Continuous<br>Wave for 577 nm only |
|-----------------------------|-------------------------------------------------------------------|-------------------------------------------------------------------|
| Average Power(AP) | 2W/0.7W | 2W/0.7W |
| Laser Safety Class | 4/IV | 4/IV |
| Exposure Sections (t) | 0.01s to continuous | 0.01s to continuous |
| Cooling Method | Solid State Cooling with<br>thermoelectric (Peltier)<br>heat pump | Solid State Cooling<br>with thermoelectric<br>(Peltier) heat pump |
| Aiming Beam: | Coaxial with treatment beam | |
| Aiming Beam: Type | Red Laser Diode | Red Laser Diode |
| Aiming Beam: Wavelength | 635-650nm (Red) | 635-650nm (Red) |
| Aiming Laser power | Maximum of 1.0mW | Maximum of 1.0mW |
| Laser Safety Class | 2/II | 2/II |
| Power requirements: Voltage | 100 to 240 Vac | 100 to 240 Vac |
### D. Device Description
LightLas577/670 Multi-Wavelength Medical Laser System consists of a Laser Console where the Laser is housed along with the Electronic Control system and Power Supplies and various Laser Delivery Units (LDU's).
The LDU's include:
- Slitlamp Integrated into CSO model SL980. (K992836) .
- Sirilamp Attachment for CSO model SL990 (K992836) and other Haag Streit clones. .
- Sittlamp Attachment for Zeiss model SL30 Stitlamp (K862004) .
- Laser Indirect Ophthalmoscope (LIO) using a Heine Omega 500. (K123316) ●
- Laser Indirect Ophthalmoscope (LIO) using a Keeler All pupil II (K854244) 0
- . Endophotocoagulation Handpieces (K122905)
The LightLas 577/670 Multi-Wavelength Medical Laser System may be used at either the 577 nm wavelength or the 670 nm wavelength. The 577 nm wavelength source can be delivered through both output ports (Ports 1 and 2) of the laser console, while the 670 mm wavelength source is delivered through Port 1 only. Users may choose either wavelength from a LCD touch screen, and only one output port and one wavelength may be selected for use at a time. Laser wavelength selection of the system is indicated on the panel and controlled by system software.
- E. Intended Use:
LightLas577/670 Multi-Wavelength Medical Laser System is an ophthalmic laser
{2}------------------------------------------------
instrument intended for use by ophthalmologists. The 577 nm wavelength laser is indicated for retinal and pan-retinal photocoagulation, endophotocoagulation, macular photocoagulation to treat leaking vessels, and laser trabeculoplasty. The 670 nm wavelength laser is indicated for retinal and pan-retinal photocoagulation. The intended use has not changed from the predicate device (K103547)
### Substantial Equivalence F.
LightLas577/670 Multi-Wavelength Medical Laser System has the same intend use, general design and fundamental scientific technology as the predicate device (K103547) Also the operating controls and functions are equivalent to the product. They have the same functional elements such as treatment wavelengths, treatment power, spot size and cooling system. Control systems such as the door interlock, and the safety systems and displays are constantly monitored in these systems for user intervention during a procedure or maintenance.
The difference in specifications of LightLas577/670 and LightLas532/670 (K103547) is that LightLas577/670 is based upon LightLas532/670 (K103547) with a modification of laser sources and corresponded control boards which causes different laser outputs. LightMed Corporation has demonstrated by evaluation of LightLas577/670 that modifications to current predicate devices are effective and safe.
There are no new hazards introduced by the Laser System as compared with the predicate devices.
#### Performance Data: ও
LightLas577/670 Multi-Wavelength Medical Laser System was evaluated according to the requirements of FDA recognized consensus standards (IEC60601-1, IEC60601-1-1, IEC60601-1-2, IEC60825-1, IEC60601-1-4, IEC60601-2-22 and ISO14971).These tests and evaluations were performed at LightMed facility and at outside test facilities as required.
Bench testing performance was completed, including the assembly, testing calibration and alignment of LightLasS77/670 laser console and laser delivery units. demonstrated that performance of LightLas577/670 test results All Multi-Wavelength Medical Laser System complies with specifications and requirements.
The conclusions drawn from the performance testing of the LightLas 577/670 Multi-Wavelength Medical Laser System demonstrate that the device is as safe, as effective, and performs as well as the legally marketed device.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609
Silver Spring, MD 20993-0002
September 26, 2013
Lightmed Corporation Ms. Jocelyn Liu Regulatory Affairs Coordinator No. 1-1. Lane 1. Pao-An St. Sec. 3. Shulin District, New Taipei City 23861, Taiwan
Re: K124043
Trade/Device Name: LightLas 577/670 Multi-Wavelength Medical Laser System Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HQF Dated: August 16, 2013 Received: August 23, 2013
Dear Ms. Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Deborah Falls -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement as Requested by FDA
510(K) Number (if Known): Product Name: LightLas577/670 Multi-Wavelength Medical Laser System Trade Name: LightLas 577/670
Indications for Use:
LightLas 577/670 is an ophthalmic laser instrument intended for use by ophthalmologists. The 577 mm wavelength laser is indicated for retinal and pan-retinal photocoagulation. endophotocoagulation, macular photocoagulation to treat leaking vessels. and laser trabeculoplasty. The 670 mm wavelength laser is indicated for retinal and pan-retinal photocoagulation.
The intended use has not changed from the predicate devices (K103547)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDR, Office of Device Evaluation (ODE)
Prescription Use: __ X (Per 21 CFR 801.109)
OR Over-The-Counter Use:
Charles Chilano 2013.09 26 14:53
(Division Sign-Off) Division of Ophthalmic and Ear, Nose and Throat Devices 510(k) Number K124043
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.