← Product Code [NLQ](/productcode/NLQ) · K124033

# MEDISISS REPROCESSED ULTRASONIC INSTRUMENTS (K124033)

_Surgical Instruments Service and Savings, Inc. · NLQ · Jan 7, 2014 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K124033

## Device Facts

- **Applicant:** Surgical Instruments Service and Savings, Inc.
- **Product Code:** [NLQ](/productcode/NLQ.md)
- **Decision Date:** Jan 7, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The MEDISISS Reprocessed Harmonic FOCUS Curved Shears FCS9 is indicated for soft-tissue incisions when bleeding control and minimal thermal injury are desired. The instrument can be used as an adjunct to, or substitute for, electrosurgery, lasers, and steel scalpels in general, otorhinolaryngologic (ear, nose and throat [ENT]), plastic, pediatric, gynecologic, urologic, exposure to orthopedic structures (such as spine and joint space), and other open procedures.

## Device Story

Reprocessed ultrasonic surgical instrument; originally manufactured by Ethicon Endo-Surgery. Device performs soft-tissue incision and coagulation. Used in clinical settings by surgeons for general, ENT, plastic, pediatric, gynecologic, urologic, and orthopedic open procedures. MEDISISS cleans, refurbishes, tests, and re-sterilizes used devices for additional clinical use. Output is mechanical ultrasonic energy for cutting and hemostasis; replaces or supplements electrosurgery, lasers, or steel scalpels. Benefits include reduced bleeding and thermal tissue damage.

## Clinical Evidence

Bench testing only. Evaluated grasping/pull force, cutting effectiveness, device integrity, thermal analysis, tissue sticking, and simulated use. Biocompatibility testing included cytotoxicity, sensitization, irritation, and acute systemic toxicity. Cleaning validation confirmed removal of protein, carbohydrate, hemoglobin, and endotoxins. Sterilization and stability were verified.

## Technological Characteristics

Ultrasonic scalpel; reprocessed device. Functionality identical to predicate. Sterilization via ethylene oxide. Complies with IEC 60601-1. No changes to materials or energy source from original predicate design.

## Regulatory Identification

Single-use reprocessed versions of the ultrasonic devices used in surgical procedures either for fragmentation, emulsification and aspiration of soft tissue and hard tissue or for ligation of vessels (under product code “LFL”). Reprocessing validation data for this device type must be included in a 510(k) submission.. (70 FR 56911, available at https://www.govinfo.gov/content/pkg/FR-2005-09-29/pdf/05-19510.pdf).

## Predicate Devices

- Ethicon Endo-Surgery Harmonic FOCUS Shears ([K100597](/device/K100597.md))
- Ethicon Endo-Surgery Harmonic FOCUS Shears ([K063192](/device/K063192.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the word "MEDISISS" in black font. To the left of the word is a graphic of lines radiating outward, resembling a sun or starburst. A trademark symbol is located to the upper right of the word.

(i

it

510(k) 124033 MEDISISS Reprocessed Harmonic FOCUS Curved Shears FCS9

### K124033

# Section 5: 510(K) Summary

| Submitter/<br>Owner              | MEDISISS<br>2747 SW 6th St.<br>Redmond, OR 97756                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |  |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Contact Name                     | Brandi Panteleon<br>Director of Product Development<br>P: 541-923-3310<br>F: 541-923-3375<br>E: bpanteleon@medisiss.com                                                                                                                                                                                                                                                                                                                                                                                 |  |  |
| Date Prepared                    | December 26, 2012                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |  |  |
| Device Names                     | Proprietary Name: MEDISISS Reprocessed Harmonic FOCUS<br>Curved Shears FCS9<br>Common Name: scalpel, ultrasonic, reprocessed                                                                                                                                                                                                                                                                                                                                                                            |  |  |
| Classification                   | Scalpel, Ultrasonic, Reprocessed, Unclassified, product code NLQ                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |  |
| Predicate<br>Devices             | K100597   Ethicon Endo-Surgery Harmonic FOCUS Shears<br>K063192   Ethicon Endo-Surgery Harmonic FOCUS Shears                                                                                                                                                                                                                                                                                                                                                                                            |  |  |
| Device<br>Description            | MEDISISS Reprocessed Harmonic FOCUS Curved Shears FCS9<br>(originally manufactured by Ethicon Endo-Surgery). Following<br>clinical use, the instruments are cleaned, refurbished, tested,<br>inspected, packaged, sterilized with ethylene oxide and returned to<br>the user facility by MEDISISS for an additional clinical use.                                                                                                                                                                       |  |  |
| Intended Use                     | The MEDISISS Reprocessed Harmonic FOCUS Curved Shears<br>FCS9 is indicated for soft-tissue incisions when bleeding control<br>and minimal thermal injury are desired. The instrument can be used<br>as an adjunct to, or substitute for, electrosurgery, lasers, and steel<br>scalpels in general, otorhinolaryngologic (ear, nose and throat<br>[ENT]), plastic, pediatric, gynecologic, urologic, exposure to<br>orthopedic structures (such as spine and joint space), and other<br>open procedures. |  |  |
| Technological<br>Characteristics | The technological characteristics of the subject device are<br>substantially equivalent to the predicate device listed in this<br>submission. The subject device has the same functionality and<br>indications as the predicate device.                                                                                                                                                                                                                                                                 |  |  |

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the word "MEDISISS" in bold, black letters. To the left of the word, there is a graphic of lines radiating outward, resembling a sun or starburst. The word appears to be a logo or brand name.

.

510(k) 124033 Reprocessed Harmonic FOCUS Curved Shears FCS9

|                        | The functional characteristics of the subject device have been<br>evaluated and found to be substantially equivalent to the predicate<br>device based on the below tests:                                                                                                                                                                                                                                                                               |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance<br>Testing | grasping/pull force; cutting effectiveness /functionality; drop test; device integrity; cutting/coagulation evaluation for a prolonged period of time; thermal analysis characterization; tissue sticking: simulated use; IEC 60601-1; cleaning; o protein, carbohydrate, hemoglobin, and endotoxins; biocompatibility; o cytotoxicity, sensitization, irritation, and acute systemic toxicity; performance qualification; sterilization and stability. |
| Conclusion             | Based on a comparison of the Indications for Use, technological<br>characteristics, and performance data to the predicate device, the<br>MEDISISS Reprocessed Harmonic FOCUS Curved Shears FCS9<br>are substantially equivalent to the predicate devices.                                                                                                                                                                                               |

.

ﺗﻬ

ﺗ

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an eagle or bird-like figure with three wing-like extensions.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

#### MEDISISS

Ms. Brandi Panteleon Director, Product Development 2747 Southwest 6th Street Redmond, Oregon 97756

#### January 7, 2014

Re: K124033

Trade/Device Name: MEDISISS Reprocessed Harmonic FOCUS Curved Shears FCS9 Regulatory Class: Unclassified Product Code: NLQ Dated: December 20, 2013 Received: December 26, 2013

Dear Ms. Panteleon:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Brandi Panteleon

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Joshua C. Nipper -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director For Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "MEDISISS" in bold, black letters. There are small lines radiating outwards from the left side of the word, giving the impression of light or energy. The font is simple and sans-serif, and the overall design is clean and modern. The word appears to be a brand name or logo.

510(k) 124033 MEDISISS Reprocessed Harmonic FOCUS Curved Shears FCS9

.

## Section 4: Indications for Use

510(k) Number: FBD K124033

Device Name: MEDISISS Reprocessed Harmonic FOCUS Curved Shears FCS9

Indications for Use:

The MEDISISS Reprocessed Harmonic FOCUS Curved Shears FCS9 is indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. The instrument can be used as an adjunct to or substitute for electrosurgery, lasers, and steel scalpels in general, otorhinolaryngologic (ENT), plastic, pediatric, gynecologic, urologic, exposure to orthopedic structures (such as spine and joint space), and other open procedures.

Prescription Use _ X

AND/OR

Over-The-Counter Use

(Part 21 CFR 801.109)

(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Long H.<br>Chen -A           | <div style="font-size: smaller;">Digitally signed by Long H. Chen-A</div> <div style="font-size: smaller;">ON: c=US, o=U.S. Government ou=HHS,</div> <div style="font-size: smaller;">ou=FDA ou=People, cn=Long H Chen-A</div> <div style="font-size: smaller;">0.9.2342.19200300.100.1.1-1.300369056</div> <div style="font-size: smaller;">Date: 2014.01.01 15:02:18 -05'00'</div> | for BSA |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| (Division Sign-off)          |                                                                                                                                                                                                                                                                                                                                                                                      |         |
| Division of Surgical Devices |                                                                                                                                                                                                                                                                                                                                                                                      |         |
| 510(k) Number:               | K124033                                                                                                                                                                                                                                                                                                                                                                              |         |

---

**Source:** [https://fda-staging.innolitics.com/device/K124033](https://fda-staging.innolitics.com/device/K124033)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
