SPEED SHIFT

K124022 · Biomedical Enterprises, Inc. · JDR · May 14, 2013 · Orthopedic

Device Facts

Record IDK124022
Device NameSPEED SHIFT
ApplicantBiomedical Enterprises, Inc.
Product CodeJDR · Orthopedic
Decision DateMay 14, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Speed Shift™ is indicated for: Fracture and osteotomy fixation and joint arthrodesis of the hand and foot.

Device Story

Speed Shift™ is a nitinol bone staple for extremity bone fragment fixation, osteotomy fixation, and joint arthrodesis. Delivered in an 'open' martensitic state, the implant undergoes a phase transformation to a 'closed' austenitic state upon exposure to ambient and body heat after insertion, providing compression. The device features a unique 'step bend' in the staple back, aligned with the legs, to reduce prominence during procedures like calcaneal slides. It is used in an operating room setting by orthopedic surgeons. The device provides mechanical fixation to stabilize bone fragments, potentially improving healing outcomes. It does not require external heating.

Clinical Evidence

Bench testing only. Pull-out strength and bending strength were evaluated per ASTM F564-10, demonstrating higher pull-out resistance and greater bending strength for Speed Shift™ compared to the predicate OSStaple™. Corrosion resistance was evaluated per ASTM F2129-08, demonstrating equivalence to the OSStaple™ Chill.

Technological Characteristics

Nitinol (ASTM F2063-05) bone staple; 1.8 x 1.8 mm square cross-section; features barbs on legs; step-bend geometry; thermally activated (martensitic to austenitic phase change via body heat); surface finish via mechanical tumbling, acid cleaning, and chemical passivation; sterile packaged.

Indications for Use

Indicated for fracture and osteotomy fixation and joint arthrodesis of the hand and foot in patients requiring bone fragment stabilization.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### (510(k) Summary) #### Product: Speed Shift™ Submitter Information BioMedical Enterprises, Inc. 14785 Omicron Drive, Ste. 205 San Antonio, Texas 78245 Telephone: (210) 677-0354 Fax: (210) 677-0355 Contact: Joe W. Soward Date Prepared: January 17, 2013 Classification name: Smooth or Threaded Metallic Bone Fastener (21 CFR 888.3040) | Classification: | Class II | |--------------------|--------------| | Product Code: | JDR | | Common/Usual Name: | Bone Staple | | Proprietary Name: | Speed Shift™ | #### Intended Use: The Speed Shift™ is indicated for: Fracture and osteotomy fixation and joint arthrodesis of the hand and foot. ### Substantial Equivalence: The Speed Shift™ is substantially equivalent to the predicate BME OSS1aple™ cleared in K993714 and the OSStaple™ Chill cleared in K102107. The predicates also include an intended use of fixation of proximal tibial metaphysis osteotomy which is not included in this submission. #### Device Description. The Speed Shift™ is a nitinol implant that comes in a range of sizes and models for use in extremity bone fragment fixation, osteotomy fixation, and joint arthrodesis. The implant is delivered to the operating room in an "open" martensitic state. The implant is then transformed by ambient and body heat after insertion, and contracts to a "closed" austenitic state. The implants do not require any external heating; they are completely transformed by body heat. This configuration change for the Speed Shift™ consists of a step bend of the staple back where the bend is in line with the legs of the staple rather than perpendicular to the legs as in the predicate devices. This in line bend allows Speed Shift™ to be useful for step osteotomies such as in the calcaneal slide procedure and other various mid-foot procedures reducing the prominence of the staple back after implantation as compared to the predicate devices. 2-1 {1}------------------------------------------------ | Product Name: | Speed Shift™ | Predicate<br>OSStaple™<br>(K993714) | Predicate<br>OSStaple Chill<br>(K102107) | |-----------------------------------|---------------------------------------------------------------------|---------------------------------------------------------------------|-----------------------------------------------------------------------| | Raw Material: | Nitinol, per ASTM<br>F2063-05 | Nitinol, per ASTM<br>F2063-05 | Nitinol, per ASTM<br>F2063-05 | | Bridge Lengths<br>(mm): | 15 and 20 | 9, 11, 13, 15, 18, 20,<br>and 30 | 9, 11, 13, 15, 18, and<br>20 | | Leg Lengths (mm): | 20 | 7, 8, 10, 12, 15, 18,<br>20, and 30 | 7, 8, 10, 12, 15, 18,<br>and 20 | | Cross-section<br>Dimensions (mm): | 1.8 x 1.8 square | Min 1.2 x 1.2<br>Max 2.0 x 3.0 | Min 1.2 x 1.2<br>Max 2.0 x 3.0 | | Barbs: | Barbs on the legs | Smooth legs | Barbs on the legs | | Heat Source: | Fully transformed<br>at room temperature | OSSforce™ electrical<br>heating unit | Body temperature | | Surface Finish: | Mechanical tumbling,<br>acid cleaning, and<br>chemical passivation. | Mechanical tumbling,<br>acid cleaning, and<br>chemical passivation. | Mechanical tumbling,<br>acid cleaning, and<br>chemical passivation. | | Storage: | Sterile packaged<br>stored at room temp<br>until used. | Sterile packaged,<br>stored at room temp<br>until used. | Sterile packaged, and<br>requires storage in<br>freezer prior to use. | Technological Characteristics Comparison to the Predicates ### Performance Bench Testing: Standard ASTM F564-10 (2010) was used to compare the pull-out strength of the new Speed Shift™ to the predicatc OSStaple™. Specimens of the largest and smallest sizes of the Speed Shift"M were used and compared to a comparably sized predicate OSStaple™M. The results showed that the Specd Shift™ has higher pull-out resistance than the predicate OSStaple™ Standard ASTM F564-10 (2010) was used to compare the mechanical strength ™. The results showed that the Speed Shift™ achieve greater bending strength when compared to the predicate OSStaple™. Standard ASTM F2129-08 was used to compare the corrosion resistance of representative samples of the new Speed Shift™ to the predicate OSStaple™. Test results demonstrate the corrosion resistance is equivalent to the predicate OSStaple Chill™ {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002 Letter dated: May 14, 2013 BioMedical Enterprises, Incorporated % Mr. Joe Soward Director, Ouality Assurance and Regulatory Affairs 14785 Omicron Drive, Suite 205 San Antonio, Texas 78245 Re: K124022 Trade/Device Name: Speed Shift™ Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: JDR Dated: April 12, 2013 Received: April 16, 2013 #### Dear Mr. Soward: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device.complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ ### Page 2 - Mr. Joe Soward forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Mark N.Mielkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K124022 Device Name: Speed Shift™ Indications For Use: The Speed Shift™ Fixation system is indicated for fracture and osteotomy fixation and joint arthrodesis of the hand and foot. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Casey L. Hanley, Ph.D. | |--------------------------------| | Division of Orthopedic Devices | Page 1 of *_*_
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...