K124016 · Djo, LLC · IPF · May 6, 2013 · Physical Medicine
Device Facts
Record ID
K124016
Device Name
EMPI PHOENIX
Applicant
Djo, LLC
Product Code
IPF · Physical Medicine
Decision Date
May 6, 2013
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Empi Phoenix is a multifunctional electrotherapy device with various treatment modes that allow for neuromuscular electrical stimulation (NMES) and transcutaneous electrical nerve stimulation (TENS) and pulsed current stimulation (PCS). As an NMES device, indications are for the following conditions: - Retarding or preventing disuse atrophy - Maintaining or increasing range of motion - Re-educating muscles - Relaxation of muscle spasms - Increasing local blood circulation As a TENS device, indications are for the following conditions: - Symptomatic relief and management of chronic, intractable pain - Adjunctive treatment for post-surgical and post-trauma acute pain - Relief of pain associated with arthritis - As a pulsed current device, indications are for the following conditions: - Reduction of edema (under negative electrode) - Reduction of muscle spasm - Influencing local blood circulation (under negative electrode) - Retardation or prevention of disuse atrophy - Facilitation of voluntary motor function - Maintenance of increase of range of motion
Device Story
Empi Phoenix is a 2-channel, hand-held electrotherapy device for home use; provides NMES, TENS, and pulsed DC (PDC) stimulation. Inputs: user-selected programs and intensity levels via device interface. Operation: microprocessor-controlled transformer generates symmetrical square biphasic waveforms (NMES/TENS) or net positive DC waveforms (PDC). Patient places electrodes (optionally using Thigh Garment accessory) and adjusts intensity as prescribed by clinician. Output: electrical stimulation to target muscles/nerves. Benefits: pain management, edema reduction, muscle rehabilitation, and improved range of motion. Clinician sets program parameters; patient manages daily intensity adjustments. Device includes LCD for status monitoring (time, intensity, battery, load).
Clinical Evidence
No clinical data. Evidence consists of bench testing, software validation per FDA guidance, and a Summative Human Factors and Usability Study confirming device safety and usability in the home environment. Compliance with IEC 60601-1, 60601-1-2, 60601-1-11, and 60601-2-10 standards demonstrated.
Technological Characteristics
Hand-held, 2-channel electrotherapy device. Powered by two removable AA batteries. Microprocessor-controlled transformer. Symmetrical square biphasic waveforms (NMES/TENS); net positive DC waveforms (PDC). Fixed pulse duration of 300 us for TENS. Frequency modulation for TENS. LCD interface. Connectivity: optional remote switch. Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home use), IEC 60601-2-10 (stimulator performance).
Indications for Use
Indicated for patients requiring NMES for disuse atrophy, range of motion maintenance, muscle re-education, spasm relaxation, or blood circulation; TENS for chronic intractable pain, post-surgical/post-trauma acute pain, or arthritis pain; and pulsed current for edema reduction, muscle spasm reduction, blood circulation, disuse atrophy, or motor function facilitation.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for DJO Global. The logo consists of a stylized globe on the left and the text "DJO GLOBAL" on the right. The word "DJO" is in a larger font than the word "GLOBAL". The globe is black and white and has a textured surface.
#### Section 5 510(k) Summary
## MAY 0 6 2013
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
The assigned 510(k) number is: K124016
Submitted by:
DJO, LLC 1430 Decision Street Vista, CA 92081 .
Contact Person:
Maria Pronina Regulatory Affairs Specialist Phone: 760-734-3047 Fax: 760-734-5694
Date Summary Prepared: December 21, 2012
Trade Name:
Empi Phoenix
NMES; TENS
Class II
Common/Usual Name:
Classification Name:
Powered muscle stimulator (21 CFR 890.5850); Transcutaneous electrical nerve stimulator for pain relief (21 CFR 882.5890)
Product Code:
IPF, Powered muscle stimulator GZJ, Transcutaneous electrical nerve stimulator for pain relief NYN, Stimulator, electrical, transcutaneous, for arthritis
Regulatory Class:
Predicate Device(s):
Compex® Rehab (K090632) EMPI Continuum (K093324)
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for DJO Global. The logo consists of a stylized globe on the left and the text "DJO GLOBAL" on the right. The word "GLOBAL" is smaller and located below "DJO".
### Device Description:
The Empi Phoenix is a multifunctional electrotherapy device indicated for retarding or preventing disuse atrophy, maintaining or increasing range of motion, and re-educating muscles. The Empi Phoenix device provides two channels of neuromuscular electrical stimulation (NMES), transcutaneous electrical stimulation (TENS) for pain management, and a pulsed DC Edema program to increase local blood circulation and reduce edema (swelling). These programs allow the patient to receive electrotherapy throughout the recovery cycle using a single device.
Its simplified programming makes the Empi Phoenix device convenient for home use: after placing the electrodes and selecting the program as prescribed by a healthcare professional, the patient only needs to increase the intensity to begin therapy. The device may be used with conventional leadwires and electrodes. For NMES treatment, it also may be paired with the Empi Phoenix™ Thigh Garment (provided as an accessory), which is designed to make electrode placement and treatment of the knee/quadriceps easier for the patient and clinician.
#### Intended Use:
As an NMES device, indications are for the following conditions:
- Retarding or preventing disuse atrophy -
- Maintaining or increasing range of motion -
- Re-educating muscles -
- Relaxation of muscle spasms ・
- Increasing local blood circulation -
As a TENS device, indications are for the following conditions:
- Symptomatic relief and management of chronic, intractable pain -
- Adjunctive treatment for post-surgical and post-trauma acute pain -
- Relief of pain associated with arthritis -
As a pulsed current device, indications are for the following conditions:
- -Reduction of edema (under negative electrodes)
- Reduction of muscle spasm -
- Influencing local blood circulation (under negative electrode) -
- Retardation or prevention of disuse atrophy ー
- Facilitation of voluntary motor function -
- Maintenance of increase of range of motion -
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for DJO Global. The logo consists of a stylized globe on the left and the text "DJO GLOBAL" on the right. The word "GLOBAL" is placed below "DJO" in a smaller font size. There is a trademark symbol to the right of the letters DJO.
#### Technological Comparison to Predicate Devices:
The Empi Phoenix device is technologically equivalent to Compex® Rehab (K090632) and EMPI Continuum (K093324) with respect to technical and performance characteristics.
The Empi Phoenix and the Compex Rehab have the same NMES indications for use. Both are hand-held devices. The Empi Phoenix uses two AA batteries that must be removed from the device for recharging. The Compex Rehab uses one 4.6 V NiMH battery that is charged in the device. The Empi Phoenix is a simpler form of the Compex Rehab. The Phoenix provides two channels of stimulation instead of four. For the Phoenix, the user only needs to select the desired program and adjust the intensity. On the Compex Rehab, the user has several display options and must choose the appropriate body part before selecting the program. The progression of warm-up, work (contraction alternating with active rest), and cool down is the same on the two devices.
The Empi Phoenix and Empi Continuum devices both provide TENS and Pulsed DC treatments. Both are 2-channel devices using two removable AA batteries. Both have an optional remote switch. The Phoenix is more convenient for home use because the user only needs to select the program and the intensity. The Continuum can be locked by the clinician so that the user only has these options, but if the device is not locked, the Continuum has more display and parameter choices for the user.
For TENS, the Continuum offers five body-part specific programs and one custom program. All programs have adjustable treatment time and frequency, and the custom program also allows the selection of type and degree of modulation and cycling time. Both devices use a fixed pulse duration of 300 us. The type of modulation is different between the two devices. Continuum uses either intensity modulation or a simple modulated pulse (SMP) which modulates both frequency and intensity. The Phoenix uses frequency modulation. The purpose of modulation in a TENS treatment is to reduce the body's acclimation to the signal, and either type of modulation can do this effectively.
For PDC, the Continuum and the Phoenix both provide 266 uA of dc current regardless of the amplitude. The Continuum does this by mixing triphasic (2 positive 60 usec pulses and one negative one) and biphasic (one positive and one negative pulse) waveforms. The Phoenix achieves the same result by adjusting the pulse width of the positive waveform compared to the set 60 usec negative pulse width. On the Continuum, the user selects between acute or chronic edema and the software provides net positive or net negative dc waveforms. The Phoenix always generates a net positive waveform, and the user connects the leadwires to the electrodes differently for acute or chronic edema.
{3}------------------------------------------------
The Phoenix and both predicate devices use a microprocessor and a transformer to generate symmetrical square biphasic waveforms for NMES and TENS. All devices use an LCD to display the selected program, remaining time, and channel intensity, and to show open load and low battery icons.
#### Performance Testing:
- . The device's software has been validated in accordance with the requirements set forth in the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005). The software validation tests demonstrated that the software version meets its design requirements.
- 하 A Human Factors and Usability Study was conducted to validate the usability of the Empi Phonix device in the home environment. The results of the Summative Validation support the instructions for successfully using the device as intended. The result of the Human Factors and Usability Study substantiates the acceptability of the risks identified during the risk assessment activities.
- The Empi Phoenix was tested and found to comply with the following standards: .
- . IEC 60601-1 for basic safety and essential performance
- IEC 60601-1-2 for electromagnetic compatibility
- 루 IEC 60601-1-11 for use in a home healthcare environment
- IEC 60601-2-10 for performance of nerve and muscle stimulators .
#### Conclusion:
Based on the performance testing and the supporting documentation, it can be concluded that the Empi Phoenix device is safe, effective and substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three abstract shapes resembling birds in flight, stacked vertically.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 6, 2013
DJO, LLC Attention: Ken Fisher SVP Quality Assurance and Regulatory Affairs 1430 Decision Street Vista, CA 92081
Re: K124016
Trade/Device Name: Empi Phoenix Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF, GZJ, NYN Dated: March 13, 2013 Received: March 27, 2013
Dear Mr. Fisher:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
Page 2 - Mr. Ken Fisher
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address:
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to:
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address:
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
## Joyce M. Whang -S
for V
for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### Section 4 Indications for Use
K124016 510(k) Number (if known):
Device Name: _________________________________________________________________________________________________________________________________________________________________
The Empi Phoenix is a multifunctional electrotherapy device with various treatment modes that allow for neuromuscular electrical stimulation (NMES) and transcutaneous electrical nerve stimulation (TENS) and pulsed current stimulation (PCS).
As an NMES device, indications are for the following conditions:
- Retarding or preventing disuse atrophy -
- Maintaining or increasing range of motion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
- Re-educating muscles
- Relaxation of muscle spasms -
- -Increasing local blood circulation
As a TENS device, indications are for the following conditions:
- -Symptomatic relief and management of chronic, intractable pain
- -Adjunctive treatment for post-surgical and post-trauma acute pain
- Relief of pain associated with arthritis -
As a pulsed current device, indications are for the following conditions:
- -Reduction of edema (under negative electrode)
- Reduction of muscle spasm -
- Influencing local blood circulation (under negative electrode) -
- Retardation or prevention of disuse atrophy -
- Facilitation of voluntary motor function -
- -Maintenance of increase of range of motion
Prescription Use X AND/OR (Part 21 CFR 801 Subpart D)
Over-the-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER
#### PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Joyce M. Whang -S
Page 1 of 1
K124016 Empi Phoenix Abbreviated 510(k): New Device
Page 14 of 714
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.