TRANS1LATERAL INTERBODY FUSION SYSTEM

K123997 · Trans1, Inc. · MAX · Apr 1, 2013 · Orthopedic

Device Facts

Record IDK123997
Device NameTRANS1LATERAL INTERBODY FUSION SYSTEM
ApplicantTrans1, Inc.
Product CodeMAX · Orthopedic
Decision DateApr 1, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

VEO is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. VEO is designed to be used with autogenous graft and supplemental spinal fixation that is cleared for use in the lumbar spine.

Device Story

VEO is a radiolucent PEEK interbody fusion implant; used in spinal fusion procedures to provide structural stability. Implanted via lateral access technique; accommodates patient anatomy and graft material size. Used in conjunction with autogenous graft and supplemental spinal fixation cleared for lumbar spine. Intended for use in skeletally mature patients with degenerative disc disease. Device interfaces with Class I access and disc preparation instruments.

Clinical Evidence

Bench testing only. Confirmatory mechanical testing performed per ASTM F2077-11 (static/dynamic compression, static compression shear, static torsion), ASTM F2267-04 (subsidence), and ASTM Draft F04.25.02.02 (expulsion). All results met or exceeded requirements.

Technological Characteristics

Radiolucent PEEK interbody fusion implant. Various heights available. Interfaces with Class I access and disc preparation instruments. Mechanical testing conducted per ASTM F2077-11, ASTM F2267-04, and ASTM Draft F04.25.02.02.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1, including those with up to Grade 1 spondylolisthesis or retrolisthesis, who have failed six months of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K123997 1 2013 APR Premarket Notification [510(K)] Summary (per 21 CFR 807.92) ## December 2012 #### Submitter: Contact Person: TranS1. Inc. 301 Government Center Drive Wilmington, NC 28403 Cheryl L Wagoner Director of Global Regulatory 910-332-1703 (phone), 910-233-1701 (fax) Proprietary Name: Classification: 888.3080: Intervertebral Fusion Device (MAX) TranS1® Lateral Interbody Fusion Device or TranSI VEO Legally Marketed Equivalent Device: TranS1 Lateral Interbody Fusion Device: K100210 ## Indications and Intended use: VEO is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from 1.2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. VEO is designed to be used with autogenous graff and supplemental spinal fixation that is cleared for use in the lumbar spine. # Device Description VEO is a radiolucent PEEK interbody fusion implant comprised of various heights to accommodate individual patient anatomy and graft material size. It also interfaces with Class I access instruments and disc preparation instruments. The lateral access technique allows a large cage to be implanted similar to marketed lateral lumbar interbody fusion sized cages. It is designed for use to provide structural stability in skeletally mature individuals. ## Technological Characteristics and Substantial Equivalence Documentation was provided to demonstrate that the modified TranS1® Lateral Interbody Fusion Device (brand name VEO) is substantially equivalent to the TranS1® Lateral Interbody Fusion Device as cleared The Subject device is substantially equivalent in intended use, level of attachment, in K100210. materials, labeling, sterilization, and technological characteristics. These devices have the same intended use and indications and rely on the same fundamental scientific technology, therefore the Subject device is substantially equivalent to the Predicate device. heet I of 2 ## Summary of Testing {1}------------------------------------------------ Confirmatory mechanical testing for the TranSI Interbody Fusion was performed per ASTM standards and included: | Test | Standard | |--------------------------|-------------------------| | Static Compression | ASTM F2077-11 | | Static Compression Shear | ASTM F2077-11 | | Static Torsion | ASTM F2077-11 | | Subsidence | ASTM F2267-04 | | Expulsion | ASTM Draft F04.25.02.02 | | Dynamic Compression | ASTM F2077-11 | All static and dynamic testing met or exceeded the requirements as established by the test protocol and applicable ASTM standards. The results demonstrated that the subject VEO device presents no new worst case for performance testing and the Subject device was therefore found to be substantially equivalent to the Predicate. ### Conclusion Based on the indications for use, technological characteristics, and comparison to predicate device, the Subject VEO (TranS1 Interbody Fusion System) has been shown to be substantially equivalent to legally marketed predicate devices, and safe and effective for its intended use. sheet 2 of 2 {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three overlapping human figures. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 Letter dated: April 1, 2013 TranS1, Incorporated % Ms. Cheryl L. Wagoner Director of Global Regulatory 301 Government Center Drive Wilmington, North Carolina 28403 Re: K123997 Trade/Device Name: TranS1® Lateral Interbody Fusion Device or TranS1 VEO Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: February 11, 2013 Received: February 14, 2013 Dear Ms. Wagoner: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Ms. Cheryl L. Wagoner forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mark NMelkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ 510(k) Number: K123997 Device Name: TranS1® Lateral Interbody Fusion Device or TranS1 VEO Indications for Use: VEO is indicated for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These patients should have had six months of non-operative treatment. VEO is designed to be used with autogenous graft and supplemental spinal fixation that is cleared for use in the lumbar spine. Prescription Use X (Part 21 CFR 801 Subpart D) . AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) > Anton E. Dmitriev, PhD Division of Orthopedic Devices
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...