SONICONE PLUS ULTRASOUND WOUND CARE SYSTEM AND ACCESSORIES
K123980 · Misonix, Inc. · NRB · Mar 5, 2013 · General, Plastic Surgery
Device Facts
Record ID
K123980
Device Name
SONICONE PLUS ULTRASOUND WOUND CARE SYSTEM AND ACCESSORIES
Applicant
Misonix, Inc.
Product Code
NRB · General, Plastic Surgery
Decision Date
Mar 5, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4410
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Misonix SonicOne Ultrasonic Would Care System and Accessories are indicated for use in the debridement of wounds, such as, but not limited to, burn wounds, diabetic ulcers. bedsores and vaginal ulcers. soft tissue debridement and cleansing of the surgical site in applications in which, in the physician's judgment would require the use of an ultrasonic aspirator with sharp debridement.
Device Story
System comprises generator, handpiece with piezoelectric crystals, and titanium tip. Generator feeds 22.5 kHz electrical signal to crystals; crystals vibrate; titanium tip amplifies vibration for tissue ablation. Irrigation unit provides solution to operative site. Used in clinical settings by physicians for wound debridement and surgical site cleansing. Accessories include probe tips, wrenches, tube sets, and handpiece sheaths. Output allows for precise tissue removal; benefits patient through effective wound bed preparation and debridement.
Clinical Evidence
No clinical data obtained. Substantial equivalence supported by bench testing, including output frequency, power, tip displacement, irrigation flow, life tests, and electrical/EMC safety testing. Compliance with voluntary standards (UL 60601-1, ISO 10993, ISO 14971, etc.) demonstrated.
Technological Characteristics
System uses 22.5 kHz ultrasonic energy via piezoelectric handpiece and titanium tip. Includes irrigation unit. Software-controlled generator. Materials evaluated per ISO 10993-1. Sterilization via ethylene oxide (ISO 11135). Complies with IEC 60601-1, IEC 62304, and IEC 60601-1-4.
Indications for Use
Indicated for debridement of wounds (burns, diabetic ulcers, bedsores, vaginal ulcers) and soft tissue debridement/cleansing of surgical sites where physician determines ultrasonic aspiration with sharp debridement is required.
Regulatory Classification
Identification
A low energy ultrasound wound cleaner is a device that uses ultrasound energy to vaporize a solution and generate a mist that is used for the cleaning and maintenance debridement of wounds. Low levels of ultrasound energy may be carried to the wound by the saline mist.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Low Energy Ultrasound Wound Cleaner.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
SonicOne Ultrasonic Wound Care System and Accessories (K112782)
Alliger Ultrasonic Surgical System AUSS-7 (K070313)
{0}------------------------------------------------
KI2-3980
#### Exhibit E 510(k) SUMMARY
# Misonix SonicOne Plus Ultrasonic Wound Care System and Accessories
This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21CFR 807.92.
#### Submitter's Identification 1 .
| Submitter's Name: | MISONIX INCORPORATED |
|-------------------|-----------------------------------------|
| Address: | 1938 New Highway, Farmingdale, NY 11735 |
| Telephone Number: | 516-694-9555 |
| Contact Person: | Ronald R. Manna |
| Date Prepared: | December 20, 2012 |
#### 2. Name of Device
4.
| Proprietary Name: | Misonix SonicOne Plus Ultrasonic Wound Care System<br>and Accessories |
|--------------------|-----------------------------------------------------------------------|
| Common/Usual Name: | Ultrasonic Surgical System<br>Ultrasonic Surgical Aspirator |
| Name: | Low Energy Ultrasonic Wound Cleaner |
| C.F.R. Section | 21 CFR 878.4410 |
| Product Codes: | NRB, FQH |
#### 3. Predicate Device Information
Predicate Devices SonicOne Ultrasonic Wound Care System and Accessories K112782 Misonix Inc. Alliger Ultrasonic Surgical System AUSS-7 K070313 Microtek Medical Inc. Equipment Drapes K050322 Device Description The SonicOne Plus Ultrasonic Wound Care System and Accessories is comprised of a generator, which feeds a 22.5 kHz electrical signal to a piezoelectric crystals mounted in a hand-held handpiece; the crystals then vibrate at the same frequency. The titanium tip attached to the handpiece amplifies the vibration. An irrigation unit is provided to introduce irrigation solution to the operative site. Accessories include Probe tips, Wrenches, Sterile and non
{1}------------------------------------------------
****************************************************************************************************************************************************************************
sterile Tube sets and sterile Surgical Procedure bags and sterile Handpiece Sheaths.
### 5. Intended Use:
The Misonix SonicOne Ultrasonic Would Care System and Accessories are indicated for use in the debridement of wounds, such as, but not limited to, burn wounds, diabetic ulcers. bedsores and vaginal ulcers. soft tissue debridement and cleansing of the surgical site in applications in which, in the physician's judgment would require the use of an ultrasonic aspirator with sharp debridement.
### 6. Comparison to Predicate Device
SonicOne Plus Ultrasonic Wound Care System and Accessories are similar in design, material and operating parameters to the Misonix Inc. AUSS-7 Ultrasonic Surgical Aspirator and the Misonix Inc. SonicOne Ultrasonic Wound Care System
#### 7. Safety and Performance Data
The Misonix SonicOne Plus Ultrasonic Wound Care System and Accessories have been designed and tested to pass the following Voluntary Standards:
| UL 60601-1 2nd Edition | Medical Electrical Equipment, Part 1: General Requirements for Safety |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1* 2nd Edition | Medical Electrical Equipment, Part 1: General Requirements for Safety |
| ISO 10993-1:2009 | Biological evaluation of medical devices —Part 1: Evaluation and testing |
| ISO 10993-7:2008 | Biological evaluation of medical devices Part 7: Ethylene oxide sterilization residuals |
| ISO 14971:2007 | Medical devices -- Application of risk management to medical devices |
| ISO 15223-1:2007/A1:2008 | Medical devices—Symbols to be used with medical device labels, labeling, and information to be<br>supplied—Part 1: General requirements |
| ISO 11607:2006 | Packaging for Terminally Sterilized Medical Devices |
| ISO 11135:2007 | Sterilization of health care products -- Ethylene oxide -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices |
| IEC 62304: 2006 | Medical Device Software-Software Life Cycle Processes |
| IEC 60601-1-4 | General Requirements for safety-Programmable electrical medical systems |
| FCC Part 18 | EMC Requirements |
- 8. This device contains software. Software has been cleared under Software Validation FDA 510k #K070313. The software and its documentation meets all of the requirements of the FDA guideline "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and ISO 60601-1-4.
- 9. Sterilization Validations Items supplied sterile have been validated for sterility to ANSI/AAMI/ISO 11135:1994 or latest edition and ISO 11135:2007. For items to be reprocessed, cleaning/disinfecting/sterilization instructions are validated to insure appropriate level of cleaning/disinfection/sterilization when instructions are followed.
{2}------------------------------------------------
### 10. Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
Output Frequency Measurements Output Power Measurements (No Load to Maximum Load) Tip Displacement Measurements Irrigation Flowrate Measurements Life Tests Input Power Measurements EMI Tests Dielectric Tests on Mains Circuits Patient Current Leakage and Patient Sink Current Measurements Power Line Ground Leakage Measurements Dielectric Tests on Patient Circuits
#### Discussions of Clinical Tests Performed | | .
The FDA has cleared all indications for use in the predicates. As such, no additional clinic data was obtained in anticipation of this submission.
#### l 2. Conclusions
Misonix Inc. can state that the SonicOne Plus Ultrasonic Wound Care System and Accessories is substantially equivalent in Mode of Operation, Hardware Design and Output Parameters to the Misonix Inc. AUSS-7 and the SonicOne Ultrasonic Wound Care System. Based upon the system and hardware validations described herein, the Misonix Inc. believes the SonicOne Plus Ultrasonic Wound Care System and Accessories poses no new issues of safety or efficacy when used for tissue ablation during wound debridement.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings outstretched, symbolizing protection and care.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center- WO66-G609 Silver Spring, MD 209930002
Letter dated: March 5, 2013
Misonix, Incorporated % Mr. Ronald Manna Vice President, Regulatory Affairs 1938 New Highway Farmingdale, New York
Re: K123980
Trade/Device Name: SonicOne Plus Ultrasonic Wound Care System and Accessories Regulation Number: 21 CFR 878.4410 Regulation Name: Low energy ultrasound wound cleaner Regulatory Class: II Product Code: NRB, FQH Dated: February 11, 2013 Received: February 14, 2013
Dear Mr. Manna:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic · product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## K123980
### Exhibit C INDICATIONS FOR USE STATEMENT
Device Name: SonicOne Plus Ultrasonic Wound Care System and Accessories
The Misonix SonicOne Ultrasonic Would Care System and Accessories are indicated for use in the debridement of wounds, such as, but not limited to, burn wounds, diabetic ulcers, bedsores and vaginal ulcers, soft tissue debridement and cleansing of the surgical site in applications in which, in the physician's judgment would require the use of an ultrasonic aspirator with sharp debridement.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Long H. Chen = -A
for MXM
(Division Sign-off) Division of Surgical Devices 510(k) Number K123980
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.