K123955 · Life Technologies Corporation · OOI · Mar 8, 2013 · Clinical Chemistry
Device Facts
Record ID
K123955
Device Name
QUANTSTUDIO DX REAL-TIME PCR INSTRUMENT
Applicant
Life Technologies Corporation
Product Code
OOI · Clinical Chemistry
Decision Date
Mar 8, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.2570
Device Class
Class 2
Indications for Use
The QuantStudio™ Dx Real-Time PCR Instrument with QuantStudio™ Dx Software is intended to perform fluorescence-based PCR to provide detection of FDA cleared and approved nucleic acid sequences in human-derived specimens. The QuantStudio™ Dx Real-Time PCR Instrument with QuantStudio™ Dx Software is intended for in vitro diagnostic use by trained laboratory technologists in combination with nucleic acid reagent kits/tests manufactured and labeled for diagnostic purposes on this instrument.
Device Story
Benchtop real-time PCR instrument; utilizes fluorescent-based PCR reagents for qualitative/quantitative detection of target nucleic acid sequences. System includes instrument with eGUI touchscreen, thermal block (Peltier-based), heated cover, plate adaptor, and computer workstation with QuantStudio™ Dx software. Operated by trained laboratory technologists in clinical settings. Workflow involves manual transfer of amplification mixture to 96-well plates (10-30 µL volume), followed by automated thermal cycling, optical detection of stimulated fluorescence, and data analysis. Output provides diagnostic results for nucleic acid targets. Benefits include standardized, automated amplification and detection for clinical molecular diagnostics.
Clinical Evidence
Bench testing and clinical performance evaluation conducted using Quidel Molecular Direct C. difficile Assay. Clinical study (n=792) compared device against tissue culture cytotoxin assay and enhanced toxigenic culture. Sensitivity 93.3% (vs cytotoxin) and 87.3% (vs culture); Specificity 93.4% (vs cytotoxin) and 98.7% (vs culture). Precision/reproducibility studies confirmed performance across sites and operators.
Technological Characteristics
Benchtop real-time PCR instrument; Peltier-based thermal cycling; microvolume fluorimeter for optical detection of stimulated fluorescence. Supports 96-well Fast PCR plates (10-30 µL reaction volume). Connectivity via computer workstation. Software-controlled automated amplification and detection.
Indications for Use
Indicated for in vitro diagnostic use by trained laboratory technologists to detect FDA-cleared/approved nucleic acid sequences in human-derived specimens using fluorescence-based PCR.
Regulatory Classification
Identification
Instrumentation for clinical multiplex test systems is a device intended to measure and sort multiple signals generated by an assay from a clinical sample. This instrumentation is used with a specific assay to measure multiple similar analytes that establish a single indicator to aid in diagnosis. Such instrumentation may be compatible with more than one specific assay. The device includes a signal reader unit, and may also integrate reagent handling, hybridization, washing, dedicated instrument control, and other hardware components, as well as raw data storage mechanisms, data acquisition software, and software to process detected signals.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems.” See § 862.1(d) for the availability of this guidance document.
The special control is FDA's guidance document entitled "Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems."
{0}------------------------------------------------
K123955
### MAR 0 8 2013
### 510(k) Summary
### Summary of Safety and Effectiveness
### Submitter Information - 21 CFR 807.92(a)(1):
- Life Technologies Corporation Submitter: 5791 Van Allen Way Carlsbad, CA 92008
- Manufacturer: Life Technologies Holdings Pte Ltd Blk 33, #07-06, Marsiling Industrial Estate, Road 3 Singapore 739256
Establishment Registration No: 3003673482
- Contact: Deanna Vella, Regulatory Affairs Manager
Phone: 760-918-3000
760-476-6934 Fax:
E-mail: deanna.vella@lifetech.com
Alternate Contact: Nikki Arora, Engineer, Regulatory Affairs
Phone: 650-554-2268
Fax: 650-638-6786
nikki.arora@lifetech.com E-mail:
Date Prepared: December 20, 2012
### Name of Device and Classification - 21 CFR 807.92(a)(2):
QuantStudio™ Dx Real-Time PCR Instrument Product Name:
Device Classification: Class II
Product Code: OOI, Real-Time Nucleic Acid Amplification System for Real Time Instruments.
{1}------------------------------------------------
# ذابة ( ) - ) - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
Predicate: Abbott m2000™ System consisting of m2000sp and m2000rt instruments, K092705
### Device Description - 21 CFR 807.92(a)(4):
The QuantStudio™ Dx Real-Time PCR Instrument is a bench top Real-Time PCR instrument that uses fluorescent-based polymerase chain reaction (PCR) reagents to provide qualitative or quantitative detection of target nucleic acid sequences (targets) using real-time analysis.
The QuantStudio™ Dx Real-Time PCR Instrument system includes the following components:
- QuantStudio™ Dx Real-Time PCR instrument with embedded graphical user . interface (eGUI) Touchscreen
- Thermal Block, also referred to as the sample block, with associated Heated Cover . and Plate Adaptor
- Calibration and verification materials for instrument qualification .
- . Computer workstation with a monitor, keyboard and mouse
- QuantStudio™ Dx instrument software .
### Intended Use/Indications for Use - 21 CFR 807.92(a)(5):
The QuantStudio™ Dx Real-Time PCR Instrument with QuantStudio™ Dx Software is intended to perform fluorescence-based PCR to provide detection of FDA cleared and approved nucleic acid sequences in human-derived specimens. The QuantStudio™ Dx Real-Time PCR Instrument with QuantStudio™ Dx Software is intended for in vitro diagnostic use by trained laboratory technologists in combination with nucleic acid reagent kits/tests manufactured and labeled for diagnostic purposes on this instrument.
### Summary of technological characteristics of the device compared to the predicate devices- 21 CFR 807.92(a)(6):
The Life Technologies QuantStudio™ Dx Real-Time PCR Instrument ("Subject Device") and the legally marketed devices, Abbott m2000™ System ("Predicate Device") is described in the table below:
{2}------------------------------------------------
Predicates Comparison – Life Technologies QuantStudio™ Dx Real-Time PCR
Instrument vs. Abbott m2000™ System
| Item | Subject Device<br>QuantStudio™ Dx<br>Real-Time PCR<br>Instrument | Predicate Device<br>Abbott m2000™ System | |
|------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Similarities | | | |
| 510(k) | N/A | K092705 | |
| Regulation | 862.2570<br>Instrumentation for<br>clinical multiplex test<br>systems. | Same | |
| Product Code | OOI: Real-Time<br>Nucleic Acid<br>Amplification System<br>for Real Time<br>Instruments. | OOI: Real-Time Nucleic<br>Acid Amplification<br>System for Real Time<br>Instruments. | |
| Device Class | Class II | Same | |
| Intended Use | The QuantStudio™ Dx Real-Time PCR<br>Instrument with<br>QuantStudio™ Dx<br>Software is intended to<br>perform fluorescence-<br>based PCR to provide<br>detection of FDA<br>cleared and approved<br>nucleic acid sequences<br>in human-derived<br>specimens. The<br>QuantStudio™ Dx<br>Real-Time PCR<br>Instrument with<br>QuantStudio™ Dx<br>Software is intended<br>for in vitro diagnostic<br>use by trained<br>laboratory<br>technologists in<br>combination with<br>nucleic acid reagent<br>kits/tests manufactured<br>and labeled for<br>diagnostic purposes on | The Abbott m2000™<br>System is intended for in<br>vitro diagnostic use in<br>performing FDA cleared<br>and approved nucleic acid<br>testing in clinical<br>laboratories. It comprises<br>the Abbott m2000sp and<br>the Abbott m2000rt<br>instruments. The Abbott<br>m2000sp is an automated<br>system for performing<br>sample preparation for<br>nucleic acid testing. The<br>Abbott m2000rt is an<br>automated system for<br>performing fluorescence-<br>based PCR to provide<br>quantitative and qualitative<br>detection of nucleic acid<br>sequences. | |
| Item | Subject Device<br>QuantStudio™ Dx<br>Real-Time PCR<br>Instrument | Predicate Device<br>Abbott m2000™ System | |
| Similarities | | | |
| | this instrument. | | |
| Technology/<br>Detection | Real-Time PCR | Same | |
| | Specimen Types | Nucleic acid | Same |
| Assay Format | Homogeneous, closed<br>tube PCR | Same | |
| Degree of<br>Automation | Requires manual<br>transfer of<br>amplification mixture<br>to<br>amplification/detection<br>instrument.<br>Automated control of<br>amplification,<br>detection, and data<br>analysis. | Same | |
| Primary<br>Operational<br>Amplification<br>and Detection<br>Components | Integrated thermal<br>cycler and<br>microvolume<br>fluorimeter for walk<br>away PCR<br>amplification and<br>detection | Same | |
| | Heating Method<br>for Amplification | Peltier device with<br>sample block | Same |
| Detection<br>Procedure | Optical detection of<br>stimulated<br>fluorescence | Same | |
| Detection<br>Chemistries | Fluorescence labeled<br>target-specific probes | Same | |
| Item | Subject Device<br>QuantStudioTM Dx Real-Time PCR Instrument | Predicate Device<br>Abbott m2000TM System | |
| Differences | | | |
| Product Code | OOI: Real-Time Nucleic Acid Amplification System for Real Time Instruments. | OOI: Real-Time Nucleic Acid Amplification System for Real Time Instruments. | |
| User Interface | PC with instrument-specific software. Instrument has touchscreen console. | PC with instrument-specific software | |
| Amplification Reaction Volume | 10-30 $\mu$ L in 96-well Fast PCR plates | 25-100 $\mu$ L in 96-well PCR plates | |
| Sample Preparation | No automated sample processing instrument offered in conjunction with the QuantStudioTM Dx Instrument. | Pairing with the m2000sp instrument provides automated sample processing. | |
・
·
:
.
{3}------------------------------------------------
and the control control control of the control of the control of
and the state of the state
{4}------------------------------------------------
Our analysis of the differences in user interface, amplification reaction volumes, and sample preparation between the Subject Device and the Predicate Device indicates that these differences do not impact performance or raise new/different questions of safety and effectiveness, and therefore render the Subject Device as Substantially Equivalent.
### Special Control/Guidance Document Referenced (if applicable):
Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems:
http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocum ents/ucm077819.htm
### Performance Data - 21 CFR 807.92(b):
As noted in the Cover Letter in this 510(k), Quidel® Corporation will submit a traditional 510(k) for the Molecular Real-Time PCR Direct C. Difficile Tox A/B that will be used with the QuantStudio™ Dx Real-Time PCR Instrument. To that end, testing to demonstrate non-clinical performance of the OuantStudio™ Dx Real-Time PCR Instrument was led by Quidel® Corporation as part of a collaboration agreement between
{5}------------------------------------------------
the two companies. This section provides a brief summary of the non-clinical performance studies and conclusions that demonstrate instrument performance when testing the Quidel Molecular Real-Time PCR Direct C. Difficile Tox A/B..
Complete non-clinical performance data can be found in Quidel's Molecular Real-Time PCR Direct C. Difficile Tox A/B traditional 510(k) submission.
### Non-Clinical Performance Data- 21 CFR 807.92(b)(1):
### Analytical performance:
#### Precision/Reproducibility: a.
Precision: For the Precision/Within Laboratory Repeatability study, a blinded fourmember panel consisting of C. difficile positive and negative sample was tested by two operators, twice a day using a single assay lot of Quidel Molecular Direct C. difficile Assay for twelve (12) days.
| QuantStudio™ Dx Real-Time PCR Instrument | | | | |
|------------------------------------------|--------|--------|----------|----------|
| <i>C. difficile</i> | 5X LoD | 2X LoD | 0.3X LoD | Negative |
| % Detection | 100% | 100% | 88% | 0% |
| Average Ct | 16.51 | 17.70 | 21.13 | N/A |
| STDEV | 0.42 | 0.76 | 1.37 | N/A |
| %CV | 2.6% | 4.3% | 6.5% | N/A |
Reproducibility: In order to confirm the reproducibility of the Quidel Molecular Direct C. difficile Assay a blinded and randomized study panel containing Clostridium difficile negative and positive samples were tested at three (3) test sites, two of which were clinical sites. Each site tested a reproducibility panel and assay controls for five (5) days in triplicate on each instrument. The testing was done by two operators at each site. Each operator ran the panel once a day using one lot of Quidel Molecular Direct C. difficile Assay.
{6}------------------------------------------------
| Reproducibility Results – QuantStudio™ Dx Instrument | | | | | | | | | | |
|------------------------------------------------------|---------|-----------|-----|---------|-----------|-----|---------|-----------|-----|------------------|
| Panel<br>Member<br>ID | Site 1 | | | Site 2 | | | Site 3 | | | Total<br>Results |
| | Results | AVE<br>Ct | %CV | Results | AVE<br>Ct | %CV | Results | AVE<br>Ct | %CV | |
| High<br>Negative<br>0.3x LoD | 8/30 | 22.9 | 5.0 | 15/30 | 22.5 | 1.3 | 15/30 | 22.5 | 1.5 | 38/90 |
| Low<br>Positive<br>2x LoD | 30/30 | 20.4 | 5.9 | 30/30 | 19.0 | 5.1 | 30/30 | 19.2 | 0.8 | 90/90 |
| Med<br>Positive<br>5x LoD | 30/30 | 18.4 | 4.2 | 30/30 | 17.5 | 0.4 | 30/30 | 17.9 | 0.7 | 90/90 |
| Negative<br>Specimen | 0/30 | N/A | N/A | 0/30 | N/A | N/A | 0/30 | N/A | N/A | 0/90 |
| Negative<br>Control | 0/30 | N/A | N/A | 0/30 | N/A | N/A | 0/30 | N/A | N/A | 0/90 |
| Positive<br>Control | 30/30 | 15.7 | 0.6 | 30/30 | 15.7 | 0.1 | 30/30 | 15.5 | 0.1 | 90/90 |
#### b. Detection limit:
The analytical sensitivity (limit of detection or LoD) of the Quidel Molecular Direct C. difficile Assay was determined on QuantStudio™ Dx instrument using quantified (CFU/mL) cultures of two C. difficile strains (ATCC BAA-1870 and ATCC BAA-1872) serially diluted in a negative fecal matrix. Analytical sensitivity (LoD) is defined as the lowest concentration at which 95% of all replicates tested positive.
| Instrument | Strain | |
|--------------------------------------|--------------------------------------|---------|
| ATCC BAA-1870 LoD<br>(CFU per assay) | ATCC BAA-1872 LoD<br>(CFU per assay) | |
| QuantStudio™<br>Dx | 4.2 E-01 | 4.0E-02 |
The final assay LoD is defined as the higher of the two strain concentrations where 95% positivity was observed. The final assay LoD is 4.2E-01 CFU/assay.
{7}------------------------------------------------
### Clinical Performance Data- 21 CFR 807.92(b)(2):
Comparison studies:
Method comparison with predicate device: a.
Performance characteristics of the Quidel Molecular Direct C. difficile Assay were established during a prospective study conducted August to November 2012. Seven hundred and ninety-two (792) samples used for this study were collected from patients suspected of having Clostridium difficile-associated disease (CDAD) at four (4) distinct geographical sites across the United States. These specimens were tested with the Quidel Assay on the Life Technologies QuantStudio™ Dx Real-Time PCR Instrument at one of three (3) facilities.
| Combined Sites – Age and Gender Distribution | | | | |
|----------------------------------------------|--------|--------|---------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Age | Gender | | Total prevalence of the<br>Quidel Molecular<br>Direct C. difficile<br>Assay on the<br>QuantStudioTM Dx<br>Real-Time PCR<br>Instrument | |
| | Male | Female | Total | |
| Unknown | | | 2 | 50% (1/2) |
| ≤ 2 years | 5 | 5 | 10 | 10% (1/10) |
| 2 to <12 years | 28 | 21 | 49 | 24% (12/49) |
| 12 to <18 years | 10 | 14 | 24 | 21% (5/24) |
| 18 to 21 years | 6 | 7 | 13 | 8% (1/13) |
| >21 to 59 years | 158 | 170 | 328 | 18% (60/328) |
| ≥ 60 years | 163 | 203 | 366 | 18% (65/366) |
| Total | 370 | 420 | 792 | 18% (145/792) |
Patient age and gender for the combined sites are presented below.
* includes two (2) patient samples with unknown age and gender.
### Tissue Culture Cytotoxicity Assay Comparison
Seven hundred and ninety-two (792) samples were tested by both the Quidel Molecular Direct C. difficile Assay and the tissue culture cytotoxin assay. Three (3) specimens (0.4%) were indeterminate in the cytotoxin assay due to toxicity in the antitoxin well. One (1) specimen (0.1%) was invalid in the Quidel Molecular Direct C. difficile Assay
{8}------------------------------------------------
when initially tested. The specimen yielded a valid result (it was negative) when retested according to the Quidel Molecular Direct C. difficile Assay draft instructions for use. We elected to calculate clinical performance based on the initial test result obtained for each specimen. Therefore, the data below is for the remaining seven hundred and eighty-eight (788) specimens.
| Combined Sites – Combined Ages | | | | | | | | |
|-------------------------------------------------------------------------------------------|-------|-----|-----|-------|-------------|--------|-------|-------|
| Tissue Culture Cytotoxin | | | | | | 95% CI | | |
| | | POS | NEG | Total | Sensitivity | | | |
| Quidel<br>Molecular<br>Real-<br>Time<br>PCR<br>Direct C.<br>difficile<br>Tox A/B<br>Assay | POS | 98 | 45* | 143 | 93.3% | 86.9% | 96.7% | |
| | NEG | 7** | 638 | 645 | Specificity | 93.4% | 91.3% | 95.0% |
| | | | | | | | | |
| | | | | | | | | |
| | | | | | | | | |
| | Total | 105 | 683 | 788 | | | | |
* Of the forty-five (45) discordant specimens (Quidel Molecular Positive/Tissue Culture Cytotoxin Negative) reported, forty-four (44) were tested with a FDA-cleared molecular device. Thirty-five (35) of these specimens were positive for C. difficile, and nine (9) were negative. The remaining specimen was unavailable for testing.
**Seven (7) discordant specimens (Quidel Negative/Tissue Culture Cytotoxin Positive) reported were tested with a FDA-cleared molecular device. Two (2) of these specimens were found positive for C. difficile, and five (5) were negative.
{9}------------------------------------------------
### Enhanced Toxigenic Culture Comparison
Seven hundred and ninety-two (792) samples were tested by both the Ouidel Molecular Direct C. difficile Assay and enhanced toxigenic culture. One (1) specimen (0.1%) was invalid in the Quidel Molecular Direct C. difficile Assay when initially tested. The specimen yielded a valid result (it was negative) when retested according to the Quidel Molecular Direct C. difficile Assay draft instructions for use. We elected to calculate clinical performance based on the initial test result obtained for each specimen. Therefore, the data below is for the remaining seven hundred and nimety-one (791) specimens.
| Combined Sites – Combined Ages | | | | | | | | |
|--------------------------------------------------------|-------|------|--------|-------|-------------|-------|-------|-------|
| Enhanced Toxigenic Culture | | | 95% CI | | | | | |
| | | POS | NEG | Total | Sensitivity | 87.3% | 81.1% | 91.6% |
| Quidel<br>Molecular<br>Direct C.<br>difficile<br>Assay | POS | 137 | 8* | 145 | Specificity | 98.7% | 97.5% | 99.4% |
| | NEG | 20** | 626 | 646 | | | | |
| | Total | 157 | 634 | 791 | | | | |
* Eight (8) discordant specimens (Quidel Molecular Positive/Enhanced Toxigenic Culture Negative) reported were tested with a FDA-cleared molecular device. Two (2) of these specimens were positive for C. difficile, and six (6) were negative.
** Seventeen (17) out of twenty (20) discordant specimens (Quidel Negative/ Enhanced Toxigenic Culture Positive) reported, were tested with a FDA-cleared molecular device. Three (3) specimens were unavailable for testing. Eleven (11) of these specimens were found negative for C. difficile, and six (6) were positive.
### Conclusion
This summary of safety and effectiveness information provides the necessary detail for a determination of substantial equivalence for the QuantStudio™ Dx Real-Time PCR Instrument.
{10}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 8, 2013
Life Technologies Deanna Vella 5791 Van Allen Way Carlsbad, California 92008
Re: K123955
Trade/Device Name: Life Technologies Regulation Number: 21 CFR §862.2570 Regulation Name: QuantStudio™ DX Real-Time PCR Instrument Regulatory Class: Class II Product Code: OOI Dated: December 20, 2012 Received: December 21, 2012
Dear Ms. Vella:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may; therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{11}------------------------------------------------
Page 2 - Ms. Vella
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807:97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.
Sincerely yours,
## Uwerschierf -S for
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{12}------------------------------------------------
### Indications for Use Statement
510(k) Number (if known): k123955
Device Name: QuantStudio™ Dx Real-Time PCR Instrument
### Indications for Use:
The QuantStudio™ Dx Real-Time PCR Instrument with QuantStudio™ Dx Software is intended to perform fluorescence-based PCR to provide detection of FDA cleared and approved nucleic acid sequences in human-derived specimens. The QuantStudio™ Dx Real-Time PCR Instrument with QuantStudio™ Dx Software is intended for in vitro diagnostic use by trained laboratory technologists in combination with nucleic acid reagent kits/tests manufactured and labeled for diagnostic purposes on this instrument.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Center for Devices and Radiological Health
Raquel Aृङ्घिeat -- 3
2013.03.03.07 -05.00 -05.00'
Division Sign-Off CDRH, Center for Devices and Radiological Health
510(k) K123955
Page _1_ of _1__
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.