K123925 · Dentsply International, Inc. · NXC · May 24, 2013 · Dental
Device Facts
Record ID
K123925
Device Name
MTM CLEAR ALIGNER IN-OFFICE
Applicant
Dentsply International, Inc.
Product Code
NXC · Dental
Decision Date
May 24, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MTM® Clear Aligner In-Office is indicated for the treatment of anterior tooth malocclusions in patients with permanent dentition (i.e. all second molars). Utilizing a series of incremental minor tooth movements (torque, tipping, rotation and bodily movement), MTM® Clear Aligner sequentially positions teeth by way of continuous gentle force.
Device Story
System of thermoformed plastic aligners fabricated in-office for minor tooth movement; uses thermoplier accessories to create force points and spaces; optional auxiliaries (hooks, buttons, orthodontic elastics) attach to aligner to facilitate movement. Clinician-operated; aligners exert continuous gentle force on teeth; force points load polymeric shell; energy dissipates as bone responds. Aligners modified or replaced after incremental movement. Used in dental clinic; patient removes aligners as needed. Output is physical tooth repositioning; clinical decision-making based on clinician-developed treatment plan. Benefits patient via sequential, incremental correction of malocclusions.
Clinical Evidence
Prospective open-label three-phase study (pretreatment, in-treatment, post-treatment) evaluated safety and effectiveness for incremental minor tooth movement. Sample included 37 cases (5 torque, 7 tipping, 14 rotation, 11 bodily movement). Primary endpoints: successful incremental movement and safety. Results: successful movement achieved in all categories; no adverse events (e.g., root resorption) reported.
Technological Characteristics
Thermoformed plastic aligners; materials cleared under K062828. Complies with ISO 20795-2, ASTM D790, ASTM D570, and ASTM D256. Mechanical force application via thermoformed force points and optional orthodontic auxiliaries (hooks, elastics).
Indications for Use
Indicated for treatment of anterior tooth malocclusions in patients with permanent dentition (all second molars present).
Regulatory Classification
Identification
An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.
{0}------------------------------------------------
K123925
### SECTION 5: 510(k) SUMMARY
## MAY 2 4 2013
## 510(k) SUMMARY for MTM® Clear Aligner In-Office
#### 1. Submitter Information:
DENTSPLY International Susquehanna Commerce Center 221 West Philadelphia Street York, PA 17405
| Contact Person: | Helen Lewis |
|-------------------|--------------|
| Telephone Number: | 717-849-4229 |
| Fax Number: | 717-849-4343 |
Date Prepared: May 7, 2103
### 2. Device Name:
- Proprietary Name: MTM® Clear Aligner In-Office ●
- Classification Name:
Aligner, Sequential
- CFR Number: 872.5470
- Device Class:
- II Product Code: NXC
#### 3. Predicate Device:
| Product Name | 510(K) | Company Name |
|--------------------|---------|------------------------|
| Invisalign® System | K081960 | Align Technology, Inc. |
#### 4. Description of Device:
MTM® Clear Aligner In-Office is a system of plastic aligners fabricated in the dental professional's office that are modified with thermoplier accessories to create force points and spaces necessary for tooth movement by way of continuous gentle force. Commercially available auxiliaries such as hooks and orthodontic elastics may also be used in order to create the desired tooth movement. As the aligner is positioned on any particular tooth, the presence of the force points loads the polymeric shell material. The stored energy thus imparted into the elastomeric material of the aligner slowly dissipates over time as the bone underlying the tooth physiologically responds to the forces. After each desired incremental tooth movement has occurred, the aligner can be modified to increase the size of the force points or a new aligner may be fabricated. Each aligner moves the patient's teeth in small increments from their original state.
{1}------------------------------------------------
#### 5. Indications for Use:
MTM® Clear Aligner In-Office is indicated for the treatment of anterior tooth malocclusions in patients with permanent dentition (i.e. all second molars). Utilizing a series of incremental minor tooth movements (torque, tipping, rotation and bodily movement), MTM® Clear Aligner sequentially positions teeth by way of continuous gentle force.
#### 6. Description of Safety and Substantial Equivalence:
#### Technological Characteristics:
The MTM® Clear Aligners that are created in-office have the same technological characteristics as the predicate device, in that all the devices are made from commercially available plastic that is thermoformed to create a customized, patient-specific aligner. The aligners are then used for minor tooth movement by way of continuous gentle force.
In both systems, the aligners are created by thermoforming commercially available plastic material. The MTM® Clear Aligners are thermoformed by the healthcare practitioner in-office and in the case of the predicate device; the aligners are thermoformed by an outside laboratory. The plastic used for MTM Clear Aligners is cleared under K062828, Mouthguard and Aligner Material.
The MTM® Clear Aligner In-Office system incorporates the use of thermobiler accessories to create spaces and force points in order to cause minor tooth movement, These force points are located in specific areas and positioned in such a way that they provide a continuous force which slowly dissipates over time on the tooth to be moved for as long as the aligner is worn. The predicate device system also has the option of creating force points using pliers ("plier detailing") as a supplemental treatment to facilitate minor tooth movement.
With the MTM® Clear Aligner In-Office system, auxiliaries such as hooks, buttons, and orthodontic elastics may be used by the healthcare practitioner together with the aligner for the purpose of tooth movement. The H.I.T.® - Hook Inserting Tack System, is designed to insert external hooks onto the MTM® Clear Aligners for the purposes of attaching elastics. The H.I.T.® system consists of a H.I.T.® punch plier (used to punch a hole into the aligner into which the hook is inserted), a H.I.T.® insertion plier (used to insert the hook into the MTM® Clear Aligners), and the H.I.T.® hooks (a hook used to attach the elastics). Commercially available orthodontic elastics can be used to attach to the hooks. These auxiliaries are attached to the aligner and not the teeth, and the usage of these auxiliaries is determined by the dental practitioner. With the predicate system, the healthcare practitioner also has the option of using these types of auxiliary components in-office with the aligner for the purpose of tooth movement.
In both systems, the aligners can be removed by the patient at any time, and treatment can be discontinued at any time.
{2}------------------------------------------------
#### Non-Clinical Performance Data:
The MTM® Clear Aligner In-Office System is composed of existing, commercially available DENTSPLY Mouthguard and Aligner Materials which are cleared in premarket notification K062828. Material properties data derived from the methods described in ISO 20795-2 (Dentistry - Base polymers - Part 2: Orthodontic base polymers), ASTM D790 (Standard Test Methods for Flexural Properties of Unreinforced and Reinforced Plastics and Electrical Insulating Materials), ASTM D570 (Standard Test Method for Water Absorption of Plastics), and ASTM D256 (Standard Test Method for Determining the Izod Pendulum Impact Resistance of Plastics) for the aligner materials have been included to support substantial equivalence. No additional in vitro testing of finished, customized aligners has been included to support substantial equivalence. Clinical data is included to validate that the MTM® Clear Aligner meets the requirements of its intended use.
#### Clinical Performance Data
A clinical study was performed in order to support the safety and efficacy of the MTM® Clear Aligner In-Office System. The objective of the study was to demonstrate the ability of the MTM® Clear Aligner In-Office to achieve incremental minor tooth movement (including torque, tipping, rotation and bodily movement), using a series of aligners consistent with and fully supportive of the product's indications for use.
The three phase prospective open-label safety and effectiveness study included a pretreatment period for establishing baseline root integrity and tooth position and to develop a treatment plan, an in-treatment monitoring period to document progression of the case, and a post-treatment period designed to establish final tooth position and root integrity. This data was also used to determine the magnitude of total tooth movement.
The study results included the following number of teeth successfully treated by sequential aligners:
| • Torque | 5 cases |
|-------------------|----------|
| • Tipping | 7 cases |
| • Rotation | 14 cases |
| • Bodily Movement | 11 cases |
No adverse events, including root resportion, were reported during the study.
{3}------------------------------------------------
#### Conclusion Regarding Substantial Equivalence:
MTM® Clear Aligner In-Office moves teeth by way of continuous gentle force through a sequence of clear aligners that follow the treatment plan developed by the clinician as does the predicate, Align System.
MTM® Clear Aligner In-Office is substantially equivalent to the predicate Invisalign System originally cleared under premarket notification K981095 and subsequently modified under premarket notification K081960. The MTM® Clear Aligner In-Office incorporates the same fundamental technology, is composed of similar materials, has the same intended use, and similar indications for use. Clinical data has been further provided to support the safety and efficacy of the MTM® Clear Aligner system when used within its proposed intended use. The conclusion of the design and intended use comparison of the MTM® Clear Aligner In-Office to the predicate device and the results of the clinical data included in this premarket notification support substantial equivalence.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular pattern around the caduceus symbol. The logo is black and white.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### May 24, 2013
Ms. Helen Lewis Director Corporate Regulatory Affairs DENTSPLY International, Incorporated Susquehanna Commerce Center 221 West Philadelphia Street YORK PA 17405
Re: K123925
Trade/Device Name: MTM® Clear Aligner In-Office Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: II Product Code: NXC Dated: April 30, 2013 Received: May 2, 2013
#### Dear Ms. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Lewis
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Susan Runner, DDS,BA
Digitally signed by Mary S. Runner -5
DN: c=US, o=U.S. Government,
ou=HHS, ou=FDA, ou=People,
cn=Mary S. Runner -S.
0.9.2342.19200300.100.1.1=130008795
Date: 2013.05.24 14:33:43-04'00'
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## SECTION 4: INDICATIONS FOR USE STATEMENT
1123925 510(k) Number (if known):
Device Name: MTM® Clear Aligner In-Office
### Indications for Use:
MTM® Clear Aligner In-Office is indicated for the treatment of anterior tooth malocclusions in patients with permanent dentition (i.e. all second molars). Utilizing a series of incremental minor tooth movements (torque, tipping, rotation and bodily movement), MTM® Clear Aligner sequentially positions teeth by way of continuous gentle force.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runner, DDS, MA
(Division Sign-Off)
Digitally signed by Mary S. Runner -S
DN: c=US, o=U.S. Government, ou=HHS,
ou=FDA, ou=People, cn=Mary S. Runner
0.9.2342.19200300.100.1.1=1300087950
Date: 2013.05.10 08:33-22 04:00
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
510(k) Number: K123925
MTM Clear Aligner In-Office
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.