KOBOLD FLETCHER TYPE APPLICATOR SET, KOBOLD HENSCHKE TYPE APPLICATOR SET
Applicant
Spencer Fillmore
Product Code
JAQ · Radiology
Decision Date
May 14, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5700
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CT/MRI Compatible HDR Fletcher Type Applicator is indicated for use in any case where high dose rate (HDR) radiation treatment of the cervix and uterus is accepted by up-to-date clinical guidelines. The CT/MRI Compatible HDR Henschke Type Applicator is indicated for use in any case where high dose rate (HDR) radiation treatment of the cervix and uterus is accepted by up-to-date clinical guidelines.
Device Story
Kobold Fletcher and Henschke Type GYN T&O Applicator Sets are medical devices used for high dose rate (HDR) brachytherapy. The sets consist of intrauterine tandems and interlocking colpostats (Fletcher or Henschke style) with adjustable ovoids. These devices are used in clinical settings by physicians to deliver radiation boosts to the cervix and parametrical tissues in conjunction with external beam radiation therapy. The applicators are CT/MRI compatible, allowing for visualization and localization within the treatment volume. They feature a click-fit connector interface compatible with GammaMed and VariSource afterloaders. By providing a stable delivery system for radioactive sources, the applicators enable precise dosimetry, potentially improving treatment outcomes for patients with cervical or uterine conditions.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical testing and biocompatibility documentation were provided to support substantial equivalence.
Technological Characteristics
Materials: Titanium, Ultem®, Stainless Steel, Silicone, High Temperature Vinyl. Design: Intrauterine tandems (straight, 15°, 30°, 45° curvature) and interlocking colpostats with adjustable ovoids (1cm-3cm). Compatibility: CT/MRI, GammaMed/VariSource afterloaders. Sterilization: Steam (134°C, 44psi, 18 min).
Indications for Use
Indicated for patients requiring high dose rate (HDR) brachytherapy for the cervix and uterus. Contraindications: As per clinical guidelines and standard clinical practice.
Regulatory Classification
Identification
A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.
{0}------------------------------------------------
K123912
Page 1 of 5
б
Kobold Applicator Kits
# 510(k) Summary
MAY 1 4 2013 This summary of 510(k) safety and effectiveness information is furnished in accordance with requirements detailed in 21 CFR 807.92.
1 .
The assigned 510(k) number is K-TBD
Submitter's Identification: Spencer Fillmore 2410 S Sumner Lane Greenacres, WA, 99016
Correspondence:
Christina Bernstein BB Medical Surgical, Inc. 2670 Leavenworth Street San Francisco, CA 94133 Tel: 415-450-0515 Email tina@bbmedicalsurgical.com
Date of submission 07 December 2012
2.
Device name:
1. Fletcher type Applicator Set Proprietary name: Kobold Fletcher Type GYN T&O Applicator Set™
2. Henschke type Applicator Set Proprietary name: Henschke Tvpe GYN T&O Applicator Set™
- Regulation Section 892.5700 A.
- B. Classification: Class II
- C. Product Code: JAQ
- D. Panel: Radiology
3.
#### Intended Use:
1. The CT/MRI Compatible HDR Fletcher Type Applicator is indicated for use in any case where high dose rate (HDR) radiation treatment of the cervix and uterus is accepted by up-to-date clinical guidelines.
{1}------------------------------------------------
2. The CT/MRI Compatible HDR Henschke Type Applicator is indicated for use in any case where high dose rate (HDR) radiation treatment of the cervix and uterus is accepted by up-to-date clinical guidelines.
### 4.
## Device Description:
1. The CT/MRI Compatible HDR Fletcher Type Applicator Set™ design is based on the conventional modified Fletcher GYN applicator with a click fit connector interface. This applicator is particularly useful in providing an HDR radiation boost to the cervix and parametrical tissues in conjunction with external beam radiation therapy. The set consists of precision manufactured intrauterine tandems and a pair of interlocking colpostats with Fletcher type unshielded ovoids ranging in size from 1cm to 3cm. All components are designed to allow for customization of the treated geometry. The distance between the ovoids is adjustable with a lock knob. Four different tandems are provided: one straight, one with a slight curvature of 15 degrees, one with medium curvature of 30 degrees, and one with a maximal curvature of 45 degrees.
2. The CT/MRI Compatible HDR Henschke Type Applicator Set™ design is based on the conventional modified Henschke GYN applicator with a click fit connector interface. This applicator is particularly useful in providing an HDR radiation boost to the cervix and parametrical tissues in conjunction with external beam radiation therapy. The set consists of precision manufactured intrauterine tandems and a pair of interlocking colpostats with Henschke type unshielded ovoids ranging in size from 2cm to 3cm. All components are designed to allow for customization of the treated geometry. The distance between the ovoids is adjustable with a lock knob. Four different tandems are provided: one straight, one with a slight curvature of 15 degrees, one with medium curvature of 30 degrees, and one with a maximal curvature of 45 degrees.
#### 5.
### Substantial Equivalence Information:
Information presented supports substantial equivalence of the Kobold Fletcher and Henschke Type GYN T&O Applicator Set™ to the predicate device. Each of the two proposed devices has the same indications for use, has similar nature of body contact, is similar in shape and design, has the same fundamental technology and uses the same sterilization procedures. Please see predicate device brochure in Appendix C.
7
- Predicate device names: Mick® Ring Type GYN Applicator Set A.
- Predicate device K numbers: 011657 B.
- C. Comparison with predicate:
{2}------------------------------------------------
.
:
| SUBSTANTIAL<br>EQUIVALENCE<br>TABLE | KOBOLD LLC | KOBOLD LLC | MICK RADIO-NUCLEAR<br>INSTRUMENTS, INC.® |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K-Number | TBD | TBD | K-011657 |
| Device Description | Applicator Set.<br>Fletcher Type | Applicator Set.<br>Henschke Type | Applicator Set, Ring Type |
| Indications for Use | The CT/MRI<br>Compatible HDR<br>Fletcher Type<br>Applicator is<br>indicated for use in<br>any case where high<br>dose rate (HDR)<br>radiation treatment<br>of the cervix and<br>uterus is accepted by<br>up-to-date clinical<br>guidelines. | The CT/MRI<br>Compatible HDR<br>Henschke Type<br>Applicator is<br>indicated for use in<br>any case where high<br>dose rate (HDR)<br>radiation treatment<br>of the cervix and<br>uterus is accepted by<br>up-to-date clinical<br>guidelines. | The Mick Radio-Nuclear<br>Instruments, Inc. HDR<br>Tandem/Ring Applicator with<br>Rectal Retractor is intended for<br>use in Brachytherapy. The<br>delivery of intra-cavitary<br>radiation therapy requires not<br>only proper visualization and<br>localization of the applicator<br>within the treatment volume,<br>but precise dosimetry and a<br>stable delivery system from<br>which treatment can be<br>administered. The Mick Radio-<br>Nuclear HDR Tandem/Ring<br>Applicator with Rectal<br>Retractor meets these<br>requirements. |
| Afterloader<br>Compatibility | GammaMed®<br>VariSource® with<br>click-fit connector. | GammaMed®.<br>VariSource® with<br>click-fit connector. | GammaMed®.<br>VariSource® with<br>click-fit connector. |
| Design Geometry | Flexible | Flexible | Flexible |
| Materials | Titanium. Ultem®<br>Stainless Steel,<br>Silicone, High<br>Temperature Vinyl | Titanium. Ultem®.<br>Stainless Steel,<br>Silicone, High<br>Temperature Vinyl | Titanium, Stainless Steel,<br>Polysulphone, Silicone |
| Packing | Tray | Tray | Cassette |
| Sterility | Provided Non-Sterile | Provided Non-Sterile | Provided Non-Sterile |
| Sterilization<br>Methods,<br>Sterilization<br>Conditions. | Steam 18 min at 273°F<br>(134°C), 44psi (3 bar)<br>Dry 15 min; no<br>Sterrad® compatibility<br>documented | Steam 18 min at 273°F<br>(134°C), 44psi (3 bar)<br>Dry 15 min; no<br>Sterrad® compatibility<br>documented | Steam 20 min at 132°C<br>Dry 15 min; no Sterrad®<br>compatibility documented |
| Biocompatibility | Documented | Documented | Not Documented |
| Anatomical Sites | Uterus, Cervix | Uterus, Cervix | Cervix |
| Environmental<br>Compatibility | CT/MRI Compatible | CT/MRI Compatible | CT/MRI Compatible |
・ :
8
·
{3}------------------------------------------------
Kobold Applicator Kits
6.
# Test Principle, Performance Characteristics:
FDA has not established special controls or performance standards for this device.
# 7.
.
# Bench Top Testing
Extensive testing in accordance with known standards is documented by the manufacturer. The standard is a specification for the materials used in surgical implant applications, predicate brachytherapy devices and manufacturer's acceptance procedures are detailed in Appendix A. Mechanical testing of the finished devices is additionally described in Appendix A.
### 8.
### Conclusions
Both versions of Kobold`s Applicator Set are similar in intended use and technological characteristics to predicate devices reviewed. Each of the two versions of the device is similar with respect to indications for use and physical characteristics to predicate devices in terms of section 510(k) substantial equivalency.
Contraindications: As per clinical guidelines and standard clinical practice.
Warnings and Precautions: The precautions are provided in the device labeling for the Kobold Fletcher Style GYN T&O Applicator Set™ and Kobold Henschke Style GYN T&O Applicator Set™. There is no warning associated with this type of device.
### ੇ.
### Summary
| Description | Comparison with Predicate Device |
|------------------|----------------------------------|
| Biocompatibility | Safe as Predicate Device |
9
{4}------------------------------------------------
K123912
Page 5 of 5
1 0
Kobold Applicator Kits
:
.
| Performance Characteristics | Substantially equivalent |
|-----------------------------|--------------------------|
| Intended Use | Substantially equivalent |
| Performance Tests | Not Required |
Each version of the device 1. Kobold Fletcher Type GYN T&O Applicator Set™ and 2. Kobold Henschke Style GYN T&O Applicator Set™, based on the information submitted in this 510(k) application has been demonstrated to be substantially equivalent to the predicate devices.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the United States Department of Health and Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 14, 2013
Kobold, LLC % Ms. Christina Bernstein Regulatory Director 2670 Leavenworth Street SAN FRANCISCO CA 94133
Re: K123912
Trade/Device Name: Fletcher Type GYN T&O Applicator Set™ Regulation Number: 21 CFR 892.5700 Regulation Name: Remote Controlled Radionuclide Applicator System Regulatory Class: Class II Product Code: JAQ Dated: April 19, 2013 Received: May 2, 2013
Dear Ms. Bernstein:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device. subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{6}------------------------------------------------
Page 2 - Ms. Christina Bernstein
comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of nedical device-related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesTorYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Michael D. DiHara
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K123912
Device Name: Fletcher Type GYN T&O Applicator Set™
Indications for Use:
The CT/MRI Compatible HDR Fletcher Type Applicator is indicated for use in any case where high dose rate (HDR) radiation treatment of the cervix and uterus is accepted by up-to-date clinical guidelines.
Prescription Use X
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) Michal D. O-Ling
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostic and Radiological Health K123912 510(k)
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.