K123897 · Curative Medical, Inc. · BZD · Mar 13, 2013 · Anesthesiology
Device Facts
Record ID
K123897
Device Name
K SERIES CPAP SYSTEMS
Applicant
Curative Medical, Inc.
Product Code
BZD · Anesthesiology
Decision Date
Mar 13, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The K Series CPAP System, is designed for the treatment of Obstructive Sleep Apnea only in spontaneously breathing patients weighing >30 kg. It is intended to be used in the home or hospital/institutional environment.
Device Story
K Series CPAP System (Curasa CPAP SD) provides continuous positive airway pressure (4-20 cm H2O) to treat obstructive sleep apnea in patients >30 kg. Device includes a humidifier for warm, humidified air. Operated in home or hospital settings. New model adds an SD card interface to store and read patient compliance data for review on an external PC. System consists of CPAP unit, power supply, patient air circuit, and U-tube connection. Healthcare providers use compliance data to monitor therapy adherence. Benefits include reduced airway dryness and effective OSA treatment.
Clinical Evidence
Bench testing only. No clinical data provided. Compliance with EN 60601-1, EN 60601-1-2, EN ISO 8185:2007, and EN ISO 17510:2007 was demonstrated. Design verification tests (pressure, user interface, software, safety, ESD/EMC/EMI) passed.
Technological Characteristics
Continuous positive airway pressure device; pressure range 4-20 cm H2O; max flow 35 LPM. Class II equipment, Type B applied part, IPX1 ingress protection. Includes heated humidifier. Connectivity via SD card interface for data export. Software-controlled pressure regulation. Power: 90-264 VAC or 24 VDC.
Indications for Use
Indicated for treatment of Obstructive Sleep Apnea in spontaneously breathing patients weighing >30 kg. Contraindicated for patients with bullous lung disease, pathologically low blood pressure, bypassed upper airway, pneumothorax, or pneumocephalus. Caution advised for patients with CSF leaks, cribriform plate abnormalities, or history of head trauma.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Curative Medical Inc.
# MAR 1 3 2013
# Section 5. Special 510(k) Summary
## Special 510(k) SUMMARY
A summary of Special 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
| Submitter: | Curative Medical Inc. |
|--------------------------------|----------------------------------------------------|
| | 3227 Kifer Road |
| | Santa Clara, CA 95051 |
| | Establishment Number: 3008361782 |
| Company Contact<br>Person: | Cuong Tran, Director Research and Development |
| Phone: | (408) 414-2188 |
| Fax: | (408) 413-3000 |
| Email: | ctran@curativemedical.com |
| Submission<br>Correspondent: | Amy McKinney, Regulatory Affairs Consultant |
| Address: | 6518 Tamarind Sky Ln, Fulshear, TX 77441 |
| Phone: | (979) 236-1622 |
| Email: | amckinney29@att.net |
| Device Name: | K Series CPAP System - Curasa CPAP SD |
| Device Classification<br>Name: | Non-continuous ventilator (BZD)<br>21 CFR 868.5905 |
| Predicate Devices: | K Series CPAP System (K120285) |
| Date of Preparation: | February 6, 2013 |
## Device Description:
The K Series CPAP systems are used on adult patients for treatment of obstructive sleep apnea (OSA). The K Series CPAP Systems has 2 commercially available models, Curasa CPAP and Floton CPAP (K120285). This special 510(k) adds the Curasa CPAP SD model to the currently commercially available selection of models. The K series CPAP system provides a stable continuous positive airway pressure (CPAP). The humidifier, which works with all the K Series CPAP Systems, provides warm, humidified air for comfort to the patient, reducing nose and airway dryness. Each K Series CPAP system also includes the following accessories: a power supply, a Patient Air Circuit, and a U-tube connection between CPAP and humidifier.
The Curasa CPAP SD has similar electronic design and software as the Curasa CPAP, except for the addition of a SD card interface and associated software driver. The Curasa CPAP SD can store/read CPAP patient compliance data onto a SD card where it can be viewed by an external PC.
{1}------------------------------------------------
#### K123897-S01
#### Curative Medical Inc.
The K Series CPAP system, model Curasa CPAP SD, has the following similarities to the previous cleared predicate device:
- Same intended use .
- Same operating principle .
- Same technology .
- . Same manufacturing process
## Intended Use:
The K Series CPAP (Continuous Positive Airway Pressure) systems are intended for the treatment of Obstructive Sleep Apnea only in spontaneously breathing patients weighing >30 kg. It is intended to be used in the home or hospital/institutional environment.
Contraindications:
- . Bullous Lung Disease
- . Pathologically Low Blood Pressure
- . Bypassed Upper Airway
- Pneumothorax ●
- Pneumocephalus has been reported in a patient using nasal Continuous Positive Airway . Pressure. Caution should be used when prescribing CPAP for susceptible patients such as those with: cerebral spinal fluid (CSF) leaks, abnormalities of the cribriform plate, prior history of head trauma, and/or pneumocephalus.
- . The use of positive airway pressure therapy may be temporarily contraindicated if you exhibit signs of sinus or middle ear infection. Not for use with patients whose upper airways are by-passed.
## Summary of Performance Data and Substantial Equivalence:
The K Series CPAP system, model Curasa CPAP SD, was designed and verified in accordance with the risk analysis and product requirements. All tests were conducted on the new model to establish substantial equivalence to the predicate. The K Series CPAP system, model Curasa CPAP SD, was tested and shown to be compliant with the following standards:
- 1. EN 60601-1-1:1995 + A1:1993 + A2:1995
Medical Electrical equipment - Part 1: General requirement for Safety
- 2. EN 60601-1-2:2001 Medical Electrical equipment - Part 1-2: General requirement for Safety - Collateral Standard: Electromagnetic compatibility - Requirements and tests
- 3. EN ISO 8185:2007 Respiratory Tract humidifiers for medical use - Particular requirements for respiratory humidifier systems
- 4. EN ISO 17510:2007 Sleep Apnoea Breathing therapy - Part 1: Sleep apnoea breathing therapy equipment
The following testing was conducted to demonstrate the performance of K series CPAP system. model Curasa CPAP SD, is as safe and effective as its predicate device in its intended environment:
{2}------------------------------------------------
| Design Verification Test | Result |
|-----------------------------------|--------|
| Pressure Testing | Pass |
| System and User Interface Testing | Pass |
| IFU Verification Testing | Pass |
| ESD / EMC / EMI | Pass |
| Safety Testing (IPX1 / ESD) | Pass |
| Software Verification Testing | Pass |
Test data leveraged from the predicate device includes the following:
- Sound Test .
- Reliability Test .
- Packaging Test �
- . Humidity ISO 8185 Test
- VOC and PM2.5 Test .
- . Biocompatibility Tests
This Special 510(k) submission presents the results of the testing and detailed description to demonstrate that K series CPAP system, model Curasa CPAP SD, is substantially equivalent to the K Series CPAP with heated humidifier systems (K120285), models Curasa CPAP and Floton CPAP.
The table below provides a design comparison of the Curasa CPAP SD to the predicate K Series CPAP with heated humidifier, Curasa model.
| Physical Features | Predicate Device -<br>K Series Curasa CPAP<br>(K120285) | K Series Curasa CPAP SD |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| CPAP Device: | | |
| Device Size (cm) | Curasa: 17 x 11.7 x 9.3 | Similar,.17 x 11.8 x 9.7 |
| Weight (kg) | Curasa: 1.4 | Same |
| Humidifier | Yes | Same |
| Memory Card | No | Yes |
| Technology | Stable, continuous positive<br>airway pressure | Same |
| Indications | The K Series CPAP System, is<br>designed for the treatment of<br>Obstructive Sleep Apnea only<br>in spontaneously breathing<br>patients weighing >30 kg.<br>It is intended to be used in the<br>home or hospital/institutional<br>environment. | Same |
| Product Use, Transport,<br>Storage: | | |
| Operation<br>(degree Celsius.) | 5 to 35 | Same |
{3}------------------------------------------------
## Curative Medical Inc.
| . .<br>. . | . . |
|------------|-------|
| . . | ----- |
| Physical Features | Predicate Device K Series Curasa CPAP (K120285) | K Series Curasa CPAP SD |
|---------------------------------------------------------------|--------------------------------------------------|-------------------------|
| Transport & Storage<br>(degree Celsius.) | -20 to 60 | Same |
| Atmosphere Pressure<br>(Operation) | 70 to 106 kPa | Same |
| Standards Compliance | | |
| IEC-60601-1 | Yes | Same |
| IEC-60601-2 | Yes | Same |
| ISO 17510-1 | Yes | Same |
| Mode of Operation | Continuous | Same |
| AC Power Consumption | 90 to 264 VAC, 50-60 Hz, 2A@115VAC and 1A@230VAC | Same |
| DC Power Consumption | 24 VDC, 2.5 A | Same |
| Type of Protection Against<br>Electric Shock | Class II Equipment | Same |
| Degree of Protection Against<br>Electric Shock | Type B Applied Part | Same |
| Degree of Protection Against<br>Ingress of Water | IPX1 | Same |
| Pressure Range (cm H2O) | 4-20 | Same |
| Pressure Stability<br>(cm H2O), as measured by<br>ISO 17510-1 | 4-20 cm H20 +/- 2.0 cm H20<br>ISO17510 compliant | Same |
| Maximum Flow (LPM), as<br>measured by ISO 17510-1 | 35 | Same |
| Compliance | | |
| | IEC-60601-1<br>ISO 8185:2007 | Same |
| Power Consumption | | |
| | 24VDC derived from A/C<br>power supply | Same |
| Electrical shock protection: | Class II | Same |
| Drip Proof Equipment | IPX1 | Same |
| Heater Setting | continuous | Same |
The table below provides a comparison of the features of the Curasa CPAP SD to the predicate K Series CPAP with heated humidifier, Curasa model.
| Functional Features | Same/Different |
|---------------------|---------------------------------------------------------------------------------------------|
| Software | Same with the addition of SD card<br>reader driver to Curasa CPAP SD |
| Hardware Components | Same with the addition of SD card<br>reader circuit board to Curasa CPAP SD |
| Humidifier | Same |
| User Interface | Same |
| Labelling | Same with the addition of handling the<br>SD card to Curasa CPAP SD and new<br>product name |
| Packaging | Same |
-
{4}------------------------------------------------
## K123897-S01
## Curative Medical Inc.
## Conclusion:
Concilision:
The information and data provided in this Special 510(k) Notification establishes that the K i he information and data provided in this Openal of o(t) , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , , substantially equivalent to the legally marketed predicate device (K120285).
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### March 13, 2013
Curative Medical, Incorporated C/O Ms. Amy E. McKinney, MS, RAC Regulatory Affairs Consultant 6518 Tamarind Sky Lane FULSHEAR TX 77441
Re: K123897
Trade/Device Name: K Series CPAP System - Curasa CPAP SD Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: February 8, 2013 Received: February 11, 2013
Dear Ms. McKinney:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/6/Picture/7 description: The image shows the text "Kwame O. Ulmer for". The text is in a simple, sans-serif font. The words "Kwame" and "Ulmer" are in bold, while the "O." and "for" are not. The letters "O" and "U" are stylized with a pattern of concentric lines.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use Statement
1123897 510(k) Number (if known): 长什2XXXXX
Device Name: K Series CPAP System - Curasa CPAP SD Indications For Use:
The K Series CPAP System, is designed for the treatment of Obstructive Sleep Apnea only in spontaneously breathing patients weighing >30 kg.
It is intended to be used in the home or hospital/institutional environment.
Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| c=US, o=U.S. Government, |
|--------------------------------|
| ou=HHS, ou=FDA, ou=People, |
| 0.9.2342.19200300.100.1.1=130 |
| 0059331, cn=Albert E. Moyal -5 |
| 2013.03.13:13:35:19-04'00' |
Albert
Moyal-S for LS
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
**510(k) Number:**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.