Post-marketing surveillance data was submitted to demonstrate the safety and efficacy of the device's SCOPE waveform and to support substantial equivalence.
Post-marketing surveillance; Safety and efficacy
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Post Marketing Surveillance; Post-marketing surveillance
Patients treated with the device in a real-world setting
Not applicable for this study
Safety and efficacy
Indications for Use
The HeartSine samaritan® PAD 350P (also known as PAD 350P and SAM 350P) is indicated for use on victims of cardiac arrest who are exhibiting the following signs: . Unconscious . Not breathing . Without circulation The samaritan® PAD 350P is intended for use by personnel who have been trained in its operation. Users should have received training in basic life support / AED, advanced life support or a physician-authorized emergency medical response training program. The samaritan® PAD 350P is indicated for use on patients greater than 8 years old or over 55 lbs (25 kg) when used with the adult samaritan® Pad-Pak (Pad-Pak-01 or Pad-Pak-07). The samaritan® PAD 350P is indicated for use on children between 1 and 8 years of age or up to 55 lbs (25 kg) when used with the samaritan® Pediatric-Pak (Pad-Pak-02).
Device Story
Portable, battery-operated AED for cardiac arrest treatment. Inputs: patient ECG and ICG via disposable electrodes. Operation: proprietary ECG analysis algorithm detects shockable/non-shockable rhythms; delivers escalating, truncated exponential biphasic waveform (SCOPE®) at 150J-200J (adult) or 50J (pediatric). User interface: voice/text/icon prompts; audible metronome guides CPR compression rate. Used by trained personnel in emergency settings. Output: therapeutic shock, CPR guidance, and event data recording. Benefits: automated rhythm analysis and real-time CPR coaching improve resuscitation outcomes. Data viewable via Saver EVO® software.
Clinical Evidence
Bench testing included biocompatibility (ISO 10993), usability (IEC 62366), EMC, and environmental testing (MIL-STD 810F). Algorithm validated using standardized AHA, MIT, and CU databases. Clinical evidence includes GLP animal data and post-market surveillance data, supplemented by clinical data from predicate device (K023854). No adverse events reported. Testing confirms safety and efficacy of SCOPE® waveform.
Technological Characteristics
Portable AED; battery-operated. SCOPE® biphasic waveform. Disposable Pad-Pak (battery/electrode unit). Connectivity: event data recording, viewable via Saver EVO® software. Standards: IEC 60601 (Edition 3), ISO 10993, IEC 62366, MIL-STD 810F. Software: major level of concern.
Indications for Use
Indicated for unconscious, non-breathing, pulseless cardiac arrest victims. Adult use (>8 years or >55 lbs) with adult Pad-Pak; pediatric use (1-8 years or ≤55 lbs) with pediatric Pad-Pak. Requires trained operators (BLS/AED, ALS, or emergency response training).
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
{0}------------------------------------------------
K123881 pg 1 of 6
Image /page/0/Picture/3 description: The image shows the title of a section. The section is titled "510(k) Summary". The section number is 5.
# JUL 1 1 2013
# Date Summary Prepared:
November 26, 2012
#### Submitter's Name and Address:
HeartSine Technologies Ltd. 203 Airport Road West Belfast, Northern Ireland, United Kingdom BT3 9ED
# Contact Person:
James McGuinness, Quality Manager
Telephone: +44 28 9093 9419 Fax Number: +44 28 9093 9401
#### Device Name:
Proprietary Name: HeartSine samaritan® PAD 350P (also known as SAM 350P) Common Name: Automated External Defibrillator Classification Name: DC-Defibrillator, Low Energy Product Code: MKJ
5-2 | P a g e
{1}------------------------------------------------
K123881 pg 2 of 6
# Predicate Device:
The features and functions of the HeartSine samaritan® PAD 350P are substantially equivalent to those of the HeartSine samaritan® PAD 300 (K041067) also known as SAM 300 and PAD300P (K052465) also known as SAM 300P.
# Device Description:
The HeartSine samaritan® PAD (Public Access Defibrillator) 350P (also called the SAM 350P) is a small, lightweight portable battery operated Automated External Defibrillator (AED) designed to treat victims of a cardiac arrest. The samaritan® PAD 350P incorporates a simple user interface of voice and text/icon prompts to guide the user in the use of the device. The samaritan® PAD 350P also incorporates an audible metronome to guide the user as to the correct rate at which chest compressions should be administered in accordance with current AHA resuscitation guidelines.
A proprietary ECG analysis algorithm automatically determines whether a victim has a shockable or non-shockable rhythm and advises a shock when appropriate. If a shock is required, the samaritan will automatically charge to the appropriate energy level and prompt the user to press the illuminated shock button. This enables the delivery of therapeutic energy to the patient.
An escalating, truncated exponential biphasic waveform pulse is delivered to the patient via two disposable defibrillator electrodes. This waveform is known as SCOPE® (Self-compensating Output Pulse Envelope). A 150 Joule, 150 Joule, 200 Joule escalating energy sequence is used in accordance with current AHA resuscitation guidelines.
After initial analysis and shock delivery (if appropriate), the PAD 350P will advise that CPR (cardiopulmonary resuscitation) may be commenced via a number of voice prompts such as "It is safe to touch the patient" and "Begin CPR" in addition to emitting an audible metronome.
5-3 | P a g e
{2}------------------------------------------------
#### K123881 pg 3 of 6
The defibrillator records the patient's electrocardiogram (ECG) and the patient's ICG (Impedance Cardiogram). The ECG can be viewed using HeartSine's Saver EVO® software.
The Pad-Pak is a combined battery and electrode unit which is single use. The electrodes used with the samaritan® PAD 350P are two non-sterile, single-use, self-adhesive, conductive gelled defibrillation electrodes. The Pad-Pak is available in three versions: an adult version, a pediatric version, and an adult version meeting FAA temperature, shock and flammability requirements for use on commercial aircraft.
The samaritan® PAD 350P incorporates the following features:
- . Controls for Power ON/OFF and Shock
- Automated charging at escalating energies of 150J, 150J, 200J .
- Automated self-tests and LED status indicator .
- Combined, disposable battery and electrodes (Pad-Pak™) .
- Electrode placement guidance voice prompts and LED/icon indicators .
- CPR voice prompts and metronome .
- Pediatric function for victims between the ages of 1 and 8 years at non-. escalating energy of 50 J
- . Integral event data recording
# Indications for Use:
The HeartSine samaritan® PAD 350P is indicated for use on victims of cardiac arrest who are exhibiting the following signs:
- . Unconscious
- Not breathing .
- Without circulation .
{3}------------------------------------------------
#### K123881 pg 4 of 6
The samaritan® PAD 350P is intended for use by personnel who have been trained in its operation. Users should have received training in basic life support / AED, advanced life support or a physician-authorized emergency medical response training program.
The samaritan® PAD 350P is indicated for use on patients greater than 8 years old or over 55 lbs (25 kg) when used with the adult samaritan® Pad-Pak (Pad-Pak-01 or Pad-Pak-07). The samaritan® PAD 350P is indicated for use on children between 1 and 8 years of age or up to 55 lbs (25 kg) when used with the samaritan® Pediatric-Pak (Pad-Pak-02).
#### Substantial Equivalence:
The PAD 350P is substantially equivalent to the HeartSine PAD Models 300 (K041067) and 300P (K052465) in intended use, technological characteristics and performance:
Indications: The indications for use of the PAD 350P are the same as the indications for the PAD 300P. The PAD 350P indications are also very similar to the PAD 300 indications, except that the PAD 300 was not indicated for pediatric use.
Technological Characteristics: The PAD 350P has very similar technological characteristics to the predicate devices. All three devices are prescription use automated external defibrillators designed to treat victims of a cardiac arrest that incorporate a simple user interface of voice and textlicon prompts to guide the user in the use of the device. The key differences between the PAD 350P and the predicate devices are that the PAD 350P has the following additional features:
- 1. Addition of CPR coaching prompts
- 2. Complies with 2010 AHA guidelines, including a CPR metronome that signals rate of chest compressions during the 2-minute CPR period(s) at 110 compressions-per-minute
5-5 | P a g e
{4}------------------------------------------------
#### K123881 pg 5 of 6
- 3. Larger capacity battery which allows longer standby and operating times and the delivery of a greater number of shocks
- 4. More intuitive LED status indicator
- 5. A refined arrhythmia analysis algorithm
- 6. Longer event recording
- 7. Logging of additional types of event information
- 8. Addition of a new version of the Pad-Pak (electrode/battery pack) meeting US FAA requirements for use in commercial aircraft
- 9. Meets current IEC 60601 standards
The above do not raise any concerns of safety and effectiveness for PAD 350P when compared to its predicate devices, HeartSine PAD Models 300 (K041067) and 300P (K052465).
Performance: As summarized in the section below, the 510(k) includes performance testing demonstrating that the PAD 350P and its accessories have successfully completed comprehensive bench, animal and clinical studies as well as biocompatibility evaluation for the patient contacting materials and software validation appropriate for a major level of concern device. This testing demonstrates the safety and effectiveness of the PAD 350P and that its performance is substantially equivalent to the predicate devices and other legally marketed AEDs.
# Summary of Performance Information:
#### Non-Clinical Testing:
Extensive biocompatibility, Usability, EMC and Environmental testing was conducted in accordance with ISO 10993, IEC 62366, IEC60601 (Edition 3) and MIL-STD 810F.
Software validation testing was carried out as appropriate for a major level of concern device. Validation also included testing on key device subassemblies and performance testing of the device as a whole in accordance with device
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
{5}------------------------------------------------
specifications. Standardised AHA, MIT and CU databases were used to extensively validate the HeartSine samaritari® PAD 350P algorithm.
#### Clinical and Usability Testing:
GLP Animal data and Post Marketing Surveillance data were submitted in addition to the referencing of clinical data used to support the 510(k) clearance of the predicate device for the Samaritan AED (K023854). No adverse events or complications were reported. This data demonstrate the safety and efficacy of the HeartSine samaritan® PAD 350P SCOPE® waveform and its substantial equivalence to other automated external defibrillators.
The performance testing demonstrates that the PAD 350P is as safe, as effective, and performs as well as the predicate devices (HeartSine samaritan® PAD 300 (K041067) and PAD300P (K052465)).
#### Conclusion:
The information in this 510(k) submission demonstrates the HeartSine samaritan® PAD 350P is substantially equivalent to the predicate devices (HeartSine samaritan® PAD 300 (K041067) and PAD300P (K052465)) with respect to intended use, technological characteristics and performance.
510(k) Summary
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002
July 11, 2013
Heartsine Technologies Inc. c/o Mr. James McGuinness Quality Manager Canberra House 203 Airport Road West Belfast. BT3 9ED Northern Ireland
Re: K123881
Trade/Device Name: Samaritan PAD 350P AED Regulation Number: 21 CFR 870.5310 Regulation Name: Automated External Defibrillator Regulatory Class: Class III (three) Product Code: MKJ Dated: June 18. 2013 Received: June 19, 2013
Dear Mr. McGuinness:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
ff your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{7}------------------------------------------------
Page 2 - Mr. James McGuinness
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Owen P. Faris -S
for
Bram D. Zuckerman Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
K123881
Image /page/8/Picture/3 description: The image shows a title card for Section 4, which is titled "Indications for Use". The text is white and bolded, and the background is black. The text is centered on the card.
510(k) Number (if known):
Device Name: HeartSine Technologies, Ltd. samaritan® PAD 350P
The HeartSine samaritan® PAD 350P (also known as PAD 350P and SAM 350P) is indicated for use on victims of cardiac arrest who are exhibiting the following signs:
- . Unconscious
- . Not breathing
- Without circulation
The samaritan® PAD 350P is intended for use by personnel who have been trained in its operation. Users should have received training in basic life support / AED, advanced life support or a physician-authorized emergency medical response training program. The samaritan® PAD 350P is indicated for use on patients greater than 8 years old or over 55 lbs (25 kg) when used with the adult samaritan® Pad-Pak (Pad-Pak-01 or Pad-Pak-07). The samaritan® PAD 350P is indicated for use on children between 1 and 8 years of age or up to 55 lbs (25 kg) when used with the samaritan® Pediatric-Pak (Pad-Pak-02).
#### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Owen P. Faris -S
2013.07.11 12:51:12
-04'00'
Prescription Use: ✓ or Over-the-Counter Use: **__**
4-2 | P a g e
Section 04
Indications for Use Statement
Page 16 of 2400
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.