SYNERGY is intended for use in: Augmentation or reconstructive treatment of alveolar ridge Filling periodontal defects Filling defects after root resection, apicoectomy, and cystectomy Filling extraction sockets to enhance preservation of the alveolar ridge Elevation of maxillary sinus floor Filling periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR). Filling peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)
Device Story
Synergy is an osteoconductive xenograft derived from bovine bone; processed into macro/micro-porous hydroxyapatite granules. Used by dental clinicians to restore bony defects in the mouth; supports new bone ingrowth; partially remodeled by osteoclasts/osteoblasts over time. Supplied as sterilized granules in glass vials; implanted during oral surgical procedures. Provides structural scaffold for bone regeneration; benefits patients by facilitating bone repair in periodontal and implant-related applications.
Clinical Evidence
No clinical data provided; substantial equivalence supported by bench testing (physical/chemical analysis), biocompatibility testing per ISO 10993, and performance evaluation in an anatomically relevant animal model.
Technological Characteristics
Anorganic bovine-derived hydroxyapatite bone mineral; porous granule form (0.30-0.84 mm or 0.84-2.0 mm). Sterilized via gamma irradiation. Complies with ASTM F1088-04a, ASTM F1185-03, ASTM F1581-08, and USP 31 standards.
Indications for Use
Indicated for patients requiring dental bone grafting for alveolar ridge augmentation, periodontal defect repair, root resection/apicoectomy/cystectomy defect filling, extraction socket preservation, maxillary sinus floor elevation, and peri-implant defect filling in conjunction with GTR/GBR procedures.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
# K123876
# 510(k) Summary
## 1. 510(k) Owner
2. Company Contact
3. 510(k) Preparer
- Odontit SA Azcuenaga 1077 4° D Buenos Aires 1115 Argentina Ph: 5411-48250221 5411-49039330 Fax: Email: info@odontit.com
## Mario Gersberg President
Blix Winston ACMD Consulting, LLC. Mullinix Mill Road Mt Airy, MD 21771 USA 301-607-9185 Ph: Email: fblixwinston@aol.com
4. Date of Preparation
May 19, 2014
- 5. Device Name and Classification:
Trade Name: Common Name: Classification Name: Synergy Anorganic Bovine Bone Grafting Material Bone Grafting Material, Animal Source
- 6. Predicate Devices:
Device: 510 (k) Numbers: Manufactured by: Bio-Oss® natural bone grafting material K871773, K952617, K970321, K033815 Geistlich Pharma AG Bahnhofstrasse 40 CH-61 10 Wolhusen Switzerland
{1}------------------------------------------------
Device: 510 (k) Number: Manufactured by: Equimatrix ™ K11816 Luitpold Pharmaceuticals, Inc. One Luitpold Drive, P0 Box 9001SE29Q1 Shirley, NY 11967SE29 Ph: 610-650-4200
## 7. Classification Names and Citations:
21CFR872.3930, NPM-bone grafting material, animal source, Class II
## 8. Compliance with Performance Standards:
The Device complies with the requirements listed in Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices
## 9. Device Description:
### a. General:
Synergy is a Xenograft of bovine origin. It contains macro and micro porous structures similar to Bio-Oss. Synergy is a purified osteoconductive mineral structure produced from bone in a multi-stage purification process. Synergy is chemically as well as structurally comparable to mineralized Bio Oss and Equimatrix. Synergy is used as an adjunctive therapy in restoring bony defects in the mouth
## b. Intended Use:
SYNERGY is intended for use in:
- . Augmentation or reconstructive treatment of alveolar ridge
- . Filling periodontal defects
- . Filling defects after root resection, apicoectomy, and cystectorny
- . Filling extraction sockets to enhance preservation of the alveolar ridge
- Elevation of maxillary sinus floor .
- . Filling periodontal defects in conjunction with products intended for Guided Tissue
- . Regeneration (GTR) and Guided Bone Regeneration (GBR).
- . Filling peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)
{2}------------------------------------------------
## c. Features:
Synergy is produced from selected bovine bone. Physical and chemical analysis has demonstrated that the composition of Synergy is similar to the predicate devices which have been shown to be anorganic, osteoconductive hydroxyapatite bone mineral. The product is milled to generate granules of two sizes (0.30 - 0. 84 mm or 0.84 - 2.0 mm) which are put into glass vials and sterilized by gamma irradiation.
## d. Working Principle:
Synergy has porous structure and composition, encouraging the formation and ingrowth of new bone at the implantation. In the course of time Synergy is partially remodelled by osteoclasts and osteoblasts (physiological remodelling).
{3}------------------------------------------------
# e. Device Comparison Table:
| Descriptive<br>Information | Synergy | Bio-OSS | EQUIMATRIX™ |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | N/A | K871773, K952617,<br>K970321, K033815 | K111816 |
| Intended Use | SYNERGY is<br>intended for use in: | Bio-Oss® is<br>recommended for: | EQUIMATRIX™<br>cancellous and<br>cortical granules are<br>recommended for: |
| | • Augmentation or<br>reconstructive<br>treatment of<br>alveolar ridge | • Augmentation or<br>reconstructive<br>treatment of the<br>alveolar ridge. | • Augmentation or<br>reconstructive<br>treatment of the<br>alveolar ridge. |
| | • Filling periodontal<br>defects | • Filling of intra<br>bony periodontal<br>defects. | • Filling of infrabony<br>periodontal defects. |
| | • Filling defects<br>after root<br>resection,<br>apicoectomy, and<br>cystectomy | • Filling of defects<br>after root<br>resection,<br>apicoectomy, and<br>cystectomy. | • Filling of defects<br>after root resection,<br>apicoectomy, and<br>cystectomy. |
| | • Filling extraction<br>sockets to<br>enhance<br>preservation of<br>the alveolar ridge | • Filling of<br>extraction sockets<br>to enhance<br>preservation of<br>the alveolar ridge. | • Filling of extraction<br>sockets to enhance<br>preservation of the<br>alveolar ridge. |
| | • Elevation of<br>maxillary sinus<br>floor | • Elevation of the<br>maxillary sinus<br>floor. | • Elevation of the<br>maxillary sinus floor. |
| | • Filling periodontal<br>defects in<br>conjunction with<br>products intended<br>for Guided Tissue<br>Regeneration<br>(GTR) and<br>Guided Bone<br>Regeneration<br>(GBR). | • Filling of<br>periodontal<br>defects in<br>conjunction with<br>products intended<br>for Guided Tissue<br>Regeneration<br>(GTR) and<br>Guided Bone<br>Regeneration<br>(GBR). | • Filling of periodontal<br>defects in<br>conjunction with<br>products intended<br>for Guided Tissue<br>Regeneration (GTR)<br>and Guided Bone<br>Regeneration(GBR). |
| | • Filling peri-implant<br>defects in | | • Filling of peri-implant<br>defects in<br>conjunction with<br>products intended<br>for Guided Bone<br>Regeneration<br>(GBR). |
| | conjunction with<br>products intended<br>for Guided Bone<br>Regeneration<br>(GBR) | • Filling of peri-<br>implant defects<br>inconjunction with<br>products intended<br>for Guided Bone<br>Regeneration<br>(GBR).<br>Bio-Oss® blocks<br>are recommended<br>for:<br>• Filling of large oral<br>and maxillofacial<br>osseous cavities. | |
| Descriptive<br>Information | Synergy | Bio-Oss | EQUIMATRIX™ |
| Device Design<br>Form | $0.30 - 0. 84$ mm or<br>$0.84 - 2.0$ mm<br>granules | 0.2 mm to 1.0 mm<br>or 1.00mm to<br>2.00mm granules<br>Bio-Oss<br>Cancellous- Bone<br>Block<br>Block Size1 x 1 x 2<br>c m | 0.2 mm to 1.0 mm or<br>1.00mm to 2.00mm<br>granules<br>N/A |
| Source Bone<br>Supplied as | Bovine<br>Granules supplied<br>in single use<br>sterilized vial. | Bovine<br>Granules supplied<br>single use sterilized<br>vial | Equine<br>Granules supplied as<br>single use sterilized<br>screw cap container<br>or prefilled syringe |
| Descriptive<br>Information | Synergy | Bio-Oss | EQUIMATRIX™ |
| Device Design<br>Material | Anorganic<br>osteoconductive<br>hydroxyapatite<br>bone mineral | Anorganic naturally<br>derived<br>osteoconductive<br>hydroxyapatite<br>bone mineral | Anorganic naturally<br>derived<br>osteoconductive<br>hydroxyapatite bone<br>mineral |
| Sterilization | Gamma irradiation | Gamma irradiation | Gamma irradiation |
| Descriptive<br>Information | Synergy | Bio-Oss | EQUIMATRIX™ |
| Device Design | | | |
| Shelf Life | 1 year | Manufacturer | 3 Years |
| Calcium/Phos<br>horous (Ca/P<br>Ratio) | 2.1 / 1 | 1.66 / 1 | Similar to BioOss |
.
{4}------------------------------------------------
{5}------------------------------------------------
## f. Substantial Equivalence
Based upon comparison of the intended use, biocompatibility, sterility, physical and chemical testing, and the performance evaluation of the subject and predicate device in an anatomically relevant animal model, and the results of clinical cases, Odontit SA concludes that Synergy is substantially equivalent to the predicate devices Bio-Oss and Equimatrix™.
# g. Standards
Synergy complies with the standards listed below:
| FDA<br>Recognized<br>Standards | Synergy | Bio-Oss | EQUIMATRIX |
|---------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|
| | ISO 10993 | ISO 10993 | ISO 10993 |
| | ASTM F1088 - 04a | | |
| Additional<br>Standards<br>Used | Synergy | Bio-Oss | EQUIMATRIX |
| | USP 31 | | |
| | ASTM F1185 - 03 | | |
| | ASTM F1581 - 08 | | |
| | Argentine<br>Pharmacopeia | | |
| FDA Special<br>Controls | Synergy | Bio-Oss | Equimatrix™ |
| | Class II Special<br>Control Guidance<br>Bone Grafting<br>Material Devices | Class II Special<br>Control Guidance<br>Bone Grafting<br>Material Devices | Class II Special<br>Control Guidance<br>Bone Grafting<br>Material Devices |
{6}------------------------------------------------
## h. Conclusion:
Based on the information provided within this submission, Odontit concludes that Synergy is substantial equivalent to Bio-Oss and Equimatrix™. .
.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, rendered in a bold, black color.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### May 22, 2014
Odontit SA C/O Mr. Blix Winston, MPA, MS Submission Correspondent ACMD Consulting, Limited Liability Company 2600 Mullinix Mill Road Mt. Airy, MD 21771
Re: K123876
Trade/Device Name: Synergy Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material, Animal Source Regulatory Class: II Product Code: NPM Dated: May 19, 2014 Received: May 19, 2014
Dear Mr. Winston:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,
{8}------------------------------------------------
Page 2 - Mr. Winston
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
MarySRunner -S
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K123876
#### Device Name
Synergy, Anorganic Cancellous Bovine Bone Graft Material
Indications for Use (Describe)
SYNERGY is intended for use in: Augmentation or reconstructive treatment of alveolar ridge Filling periodontal defects Filling defects after root resection, apicoectomy, and cystectomy Filling extraction sockets to enhance preservation of the alveolar ridge Elevation of maxillary sinus floor Filling periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR). Filling peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR)
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
CALL SERVED FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FO
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Sheena A. Green 2014.05.21 14:36
FORM FDA 3881 (9/13)
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
{10}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
.
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.