K123822 · Hk Surgical, Inc. · FRN · Sep 27, 2013 · General Hospital
Device Facts
Record ID
K123822
Device Name
KLEIN TOUCH PUMP AND DISPOSABLE TUBING SET
Applicant
Hk Surgical, Inc.
Product Code
FRN · General Hospital
Decision Date
Sep 27, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The KleinTouch Pump is an infiltration pump used to cause a flow of fluid from an IV bag into a patient in a manner controlled manually by a health care professional. The KleinTouch pump is not intended to be used as an IV infusion pump.
Device Story
KleinTouch Pump (KTP) is an electrically or battery-operated, manually controlled peristaltic pump; utilizes three rollers to create pumping action in sterile, single-use Klein Touch Tubing (KT20). Device features color graphic LCD and dual air-activated foot switches. Operated by trained healthcare professionals in clinical settings. Input is fluid from an IV bag; output is controlled fluid flow to patient. Healthcare provider monitors flow manually; device assists in infiltration procedures. Benefits include precise manual control of fluid delivery during infiltration. Device is not an IV infusion pump.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including electrical safety (UL 60601-1, CSA C22.2, EN/IEC 60601-1), electromagnetic compatibility (IEC/EN 60601-1-2, JIS T0601-1-2, AS/NZ 3200.1.2), biocompatibility of fluid path materials (ISO 10993-1), sterilization validation (ISO 11135), and human factors evaluation confirming safe use with no critical task errors.
Technological Characteristics
Peristaltic pump with 3-roller mechanism. Materials: Silicone (fluid path), plastic base. Dimensions: 11" x 10.75" x 5"; Weight: 8.4 lbs. Power: Electrical/Battery. Interface: Color graphic LCD, dual air-activated foot switches. Connectivity: None. Sterilization: ISO 11135 compliant. Safety: Class I, Type B protection against electric shock.
Indications for Use
Indicated for use by trained healthcare professionals to control the flow of fluid from an IV bag into a patient. Not for use as an IV infusion pump.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a logo for HK Surgical. The logo features the letters "HK" in a stylized font, with the word "SURGICAL" underneath. Below that is the website "HKSurgical.com". The address "1271 Puerta del Sol San Clemente, CA 92673" is printed below the website.
HK Surgical KleinTouch Pump and Tubing Set 510(k) Notification
## 5. 510(k) Summary
| Submitter: | HK Surgical, Inc. |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 1271 Puerta del Sol San Clemente, CA 92673 |
| Phone number: | 949-369-0101 |
| Fax number: | 949-369-9797, email peter@hksurgical.com |
| Contact person: | Pejman Farivar |
| Date prepared: | 12/10/2012 |
| 510(k) Number: | To be Determined |
| Trade name: | KleinTouch Pump (KTP) |
| Common name: | Infiltration Pump |
| Classification name: | Pump, Infusion |
| Classification: | Product Code: FRN,<br>Regulation Number: 880.5725<br>Class II |
| Substantial equivalence<br>claimed to: | Klein Infiltration Pump - K031432 |
| Description: | Device Functionality:<br><br>The KleinTouch Pump is an electrically or battery operated, manually<br>controlled, peristaltic pump using three rollers to create pumping<br>action in a section of sterile tubing.<br>Scientific Concept(s)<br><br>The KTP is designed to specifically function with sterile single use only<br>Klein Touch Tubing (KT20). Tubing must be primed before use.<br>Please refer to the KT20 Instructions for Use for further details.<br>Physical And Performance Characteristics |
| | |
| | Device Design Display: Color Graphic LCD Dual Foot Switches With Hose - Air Activated Dimensions: 11" x 10 ¾" x 5" Weight: 8.4 lbs. |
| | Components Used Motor Fluid Pathway Materials Silicone Inlet Hospital Power Cord Hospital Grade Plugs Barcode Reader Base, Plastic Battery Instrument Foam PCBA - Power Board PCBA, Main Circuit Board Pump, Peristaltic, 3-Roller, Fixed Clamp Switch, Rocker, Lighted |
| | Physical Properties The Pump is intended for use in combination with the KT20 tubing. The peristaltic pump using three rollers will create pumping action in KTP20 tubing set |
| | specifically designed for the KleinTouch Pump |
| Intended use: | The KleinTouch Pump is an infiltration pump used to cause a flow of<br>fluid from an IV bag into a patient in a manner controlled manually by a<br>health care professional. The KleinTouch pump is not intended to be<br>used as an IV infusion pump. |
| Safety and Labeling | WARNINGS AND CAUTIONS<br>The device labeling contains cautions and warnings which are<br>considered essential to the safety of personnel, patients, equipment,<br>and property. |
| | This device must be used by a trained professional with current<br>knowledge regarding safety of total dosage. Orders for dosage must<br>be by a physician and legible.<br>Protection against electric shock: Class I, Type B<br>Refer to the Instructions for Use for the Pump and Infusion Set for<br>Warnings and cautions. |
| Performance Testing | MET LABS tested per UL Standards 2601-1; 2nd ed. Medical<br>Electrical Equipment- Part 1 General Requirements for Safety<br>• CSA C22.2<br>• UL 60601-1<br>• EN/IEC 60601-1 |
| | EMC:<br>• IEC/EN 60601-1-2:2007-ED30<br>• JSA-JIS T0601-1-2:2002<br>• AS/NZ 3200.1.2:2005<br>Biocompatibility Testing demonstrates all fluid path materials are<br>biocompatible in accordance with ISO10993-1.<br>Sterilization Validation to SAL 10-6 in accordance with ISO11135.<br>Human Factors evaluation demonstrates the KleinTouch Pump can be<br>used safely with no critical task errors.<br>Performance testing demonstrates the KleinTouch Pump meets the specifications. |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo and contact information for HK Surgical. The logo features the letters 'HK' in a stylized font, with the word 'SURGICAL' written below. The website 'HKSurgical.com' is also listed. The address '1271 Puerta del Sol, San Clemente, CA 92673' is provided, indicating the company's location.
## Substantial Equivalence Conclusion:
HK has provided test data to assess the technological differences and demonstrate Subject Device is substantially equivalent to the predicate device (K031432) and at least is safe and effective.
| Page 15 of 92 | HK Surgical KleinTouch Pump and Tubing Set 510(k) Notification | Confidential |
|---------------|----------------------------------------------------------------|--------------|
|---------------|----------------------------------------------------------------|--------------|
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three legs, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 27, 2013
HK Surgical, Inc. Ms. Clare Bennett General Manager 1271 Puerta del Sol SAN CLEMENTE, CA 92673
Re: K123822
Trade/Device Name: KleinTouch Pump (KTP) Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: February 1, 2013 Received: February 1, 2013
Dear Ms. Bennett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - Ms. Clare Bennett
comply with all the Act 's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events} (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
· Sincerely yours.
Mary S. Runner -S
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation -Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo for HK Surgical. The logo features the text "HK SURGICAL" in a bold, sans-serif font, with "HKSurgical.com" written below it. The address "1271 Puerta del Sol, San Clemente, CA 92673" is printed below the website address. The logo is simple and professional, and it is likely used to identify the company on its products and marketing materials.
K123822
HK Surgical KleinTouch Pump and Tubing Set 510(k) Notification
## 4. Indications for Use Statement
The KleinTouch Pump is an infiltration pump used to cause a flow of fluid from an IV bag into a patient in a manner controlled manually by a health care professional. The KleinTouch pump is not intended to be used as an IV infusion pump.
. . Prescription Use
(21 CFR 807 Subpart C)
AND/OR Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(Please DO NOT WRITE BELOW THIS UNE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Cancurrence of CDRH, Office of Dovice Evaluation (ODE)
Image /page/4/Picture/10 description: The image shows the logo of the Food and Drug Administration (FDA). The logo is composed of the letters "FDA" in a stylized font. The letters are made up of small, geometric shapes, giving the logo a textured appearance. The logo is black and white.
Richard C. Chapman 2013.09.27 15:31:41 -04'00'
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.