VASCULAR POSITIONING SYSTEM (VPS) G4 CONSOLE

K123813 · Vasonova, Inc. · OBJ · Jan 18, 2013 · Cardiovascular

Device Facts

Record IDK123813
Device NameVASCULAR POSITIONING SYSTEM (VPS) G4 CONSOLE
ApplicantVasonova, Inc.
Product CodeOBJ · Cardiovascular
Decision DateJan 18, 2013
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1200
Device ClassClass 2

Indications for Use

The intended use of the VPS Stylet and Console (VPS System) is to quickly and accurately guide market available central catheters to the desired location which is the lower third of the SVC or at the cavo-atrial junction.

Device Story

VPS G4 Console provides real-time guidance for central venous catheter tip placement. System inputs include electrical cardiac signals (ivECG) and blood flow data (Doppler ultrasound) via a polyimide stylet with a distal Doppler sensor and stainless steel ivECG wire. Console processes these signals to provide visual guidance via a touchscreen tablet display. Used in clinical settings by healthcare providers to navigate catheters to the lower third of the SVC or cavo-atrial junction. Output consists of colored icons and waveform data; a 'Blue Bullseye' indicates the catheter tip has reached the desired location. System assists clinicians in confirming tip placement without immediate reliance on fluoroscopy or chest X-ray, potentially reducing radiation exposure and procedure time.

Clinical Evidence

No clinical data provided. Substantial equivalence is supported by bench testing, including software verification and validation, and compliance with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and IEC 60601-2-37 (ultrasonic diagnostic equipment safety).

Technological Characteristics

System comprises a console with tablet display and a polyimide stylet. Sensing utilizes a distal Doppler sensor and stainless steel ivECG wire. Connectivity is via extension cable to the console. Compliant with IEC 60601-1, IEC 60601-1-4, and IEC 60601-2-37. Software-driven data acquisition and processing.

Indications for Use

Indicated for guidance and tip positioning of central venous catheters in adult patients as an alternative to fluoroscopy or chest x-ray. Contraindicated/limited in patients with altered cardiac rhythms (atrial fibrillation, atrial flutter, severe tachycardia, pacemaker-driven rhythm) or femoral/saphenous vein access, where P-wave presentation is altered; these patients require additional confirmation methods.

Regulatory Classification

Identification

An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 12. 510(k) Summary . | Submitter | VasoNova, Inc., a wholly owned subsidiary of Teleflex, Inc.<br>155 Jefferson Drive, Suite 100<br>Menlo Park, CA 94025 | | | |--------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|--| | Name of contact person | Christine Ford, RAC | | | | | Office phone: 610-378-0131, ext. 3338<br>Mobile phone: 484-797-7110 | | | | | Fax: 610-478-3179<br>Email: christine.ford@teleflex.com | | | | Date summary prepared | December 7, 2012 | | | | Device trade or<br>proprietary name | Vascular Positioning System™ (VPSTM) G4 Console | | | | Common or usual name | Catheter, Ultrasound, Intravascular | | | | Classification name | Diagnostic intravascular catheter, 21 CFR §870.1200, Class II, product code OBJ | | | | Classification panel | Cardiovascular | | | | Legally marketed device(s)<br>to which equivalence is<br>claimed | K103260 VasoNova Vascular Positioning System™ (VPSTM) Console | | | | Device description | The VPS System consists of a VPS Console with tablet display and a VPS<br>Stylet.<br>The VPS G4 Console consists of two key integrated software-driven<br>components: a data acquisition module and an embedded computer with<br>processing capabilities. The data acquisition module within the console<br>transmits and receives ultrasound data while receiving electrical signals from<br>the heart through sensors mounted at the tip of the VPS Stylet. Using signals<br>gathered by the data acquisition module, the graphical user interface provides<br>the user with guidance for tip positioning of central venous access devices.<br>The separately provided VPS Stylet is a polyimide tube containing a Doppler<br>sensor on a coax cable and an intravascular electrocardiogram (ivECG) signal<br>sensing stainless steel wire. The Doppler sensor and the exposed portion of the<br>ivECG are located at the distal end of the stylet and are used to detect and<br>transmit physiological information to the VPS Console for analysis.<br>The<br>proximal end of the stylet connects to an extension cable that connects to the<br>VPS Console. | | | | Intended use of the<br>device | The intended use of the VPS Stylet and Console (VPS System) is to quickly and<br>accurately guide market available central catheters to the desired location which<br>is the lower third of the SVC or at the cavo-atrial junction. | | | | SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF THE SUBJECT DEVICE COMPARED TO THE<br>PREDICATE DEVICE | | | | | Characteristic | Predicate VPS Console | Subject VPS G4 Console | | | Guidance indicator | Colored icons, waveform data | Colored icons, waveform data | | | Guidance signal type(s) | ivECG, Doppler ultrasound | ivECG, Doppler ultrasound | | | Signal Conductor | Two conductor stylet wires<br>(coax cable and ivECG wire) | Two conductor stylet wires<br>(coax cable and ivECG wire) | | | Acoustic Output | One setting; not controlled by the user | One setting; not controlled by the user | | | User Interface | PC-based system with keyboard and touch screen | PC-based system connected to a touchscreen tablet display | | | Software Interface | Custom GUI | Custom GUI | | | PERFORMANCE TESTS | | | | | Characteristic | Standard/Test | Results Summary | | | Electromagnetic Compatibility | IEC 60601-1-2: 2007-03 - Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and essential performance:<br>Collateral standard: Electromagnetic compatibility - Requirements<br>and tests | Compliant | | | Electrical Safety and Acoustic<br>Safety | IEC 60601-1:1988 (A1:1991 + A2:1995) - Medical electrical<br>equipment - Part 1: General requirements for safety IEC 60601-1-4: 1996 (A1:1999)/EN 60601-1-4:2000- Medical<br>electrical equipment - Part 1-4: Collateral Standard: Programmable<br>electrical medical systems IEC 60601-2-37:2001 (A1:2004 + A2:2005) - Medical electrical<br>equipment - Part 2-37: Particular requirements for the safety of<br>ultrasonic medical diagnostic and monitoring equipment | Compliant | | | Software verification and<br>validation | VasoNova Software and Algorithm Verification and Validation testing | Compliant | | | Conclusions: | The intended use, technological characteristics, method of collecting and<br>interpreting physiological data, and performance of the predicate VasoNova<br>VPS Console and the VPS G4 Console along with testing to voluntary standards<br>and applicable FDA guidance documents provide evidence that the subject VPS<br>G4 console is substantially equivalent to the predicate VPS Console in terms of<br>safety, efficacy, and performance.<br>The described physical, software, and hardware modifications are enhancements<br>to various aspects of the predicate VPS Console. They do not alter the<br>fundamental scientific technology of the device; they do not alter the method in<br>which physiological data is collected from the patient; they do not alter the<br>method in which physiological data is interpreted by the software algorithm to<br>determine the symbols used to indicate catheter migration through the<br>vasculature and final tip placement. Any differences between the subject VPS<br>G4 Console and the predicate VPS Console do not raise new issues of safety or<br>effectiveness. | | | {1}------------------------------------------------ . : . . . . . 1 : 1 : 1 : 1 : {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the eagle. The logo is black and white and appears to be a scanned image. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 JAN 1 8 2013 Vasonova Inc. c/o: Christine Ford Regulatory Affairs Manager 155 Jefferson Drive, Suite 100 Menlo Park, CA 94025 Re: K123813 Trade Name: Vascular Positioning System (VPSTM) G4 Console Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic intravascular catheter Regulatory Class: Class II Product Code: OBJ Dated: December 7, 2012 Received: December 19, 2012 Dear Ms. Ford: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. {3}------------------------------------------------ Page 2 -- Ms. Christine Ford If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. **Owen P. Faris -S** for Bram D. Zuckerman, M.D. Director Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): __ K123813 ## Vascular Positioning System (VPS™) G4 Console Device Name: _________________________________________________________________________________________________________________________________________________________________ Indications For Use: The VPS Stylet and Console are indicated for guidance and tip positioning for central The Console is intended for use with the stylet supplied by venous catheters. VasoNova. The Vascular Positioning System consists of a Console and associated power supply and cabling and a stylet. The VPS Console, when used with the VPS Stylet, provides real-time catheter tip location information by using the patient's physiological (cardiac electrical activity and blood flow) information. When the Vascular Positioning System guidance indicator shows a Blue Bullseye, the catheter tip is in the desired location. The Vascular Positioning System is indicated for use as an alternative method to fluoroscopy or chest x-ray for central venous catheter tip placement confirmation in adult patients. NOTE: If a steady blue bull's eye is not obtained, standard hospital practice should be followed to confirm catheter tip location. Limiting but not contraindicated situations for this technique are in patients where alterations of cardiac rhythm change the presentation of the P-wave as in atrial fibrillation, atrial flutter, severe tachycardia and pacemaker-driven rhythm, and in central venous catheterization procedures performed through femoral or saphenous vein access which change the presentation of the P-wave. In such patients, who are easily identifiable prior to central venous catheter insertion, the use of an additional method is required to confirm catheter tip location. Prescription 'Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OIR) | |-----------------------------------------------------------------------------------| |-----------------------------------------------------------------------------------| (Division Sign-Off) Division of Cardiovascular Devices Page 1 of
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...