K123792 · Mennen Medical , Ltd. · DQX · Aug 30, 2013 · Cardiovascular
Device Facts
Record ID
K123792
Device Name
HORIZON XVU (FFR)
Applicant
Mennen Medical , Ltd.
Product Code
DQX · Cardiovascular
Decision Date
Aug 30, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
The Horizon XVu is a state-of-the-art computerized laboratory, capable of acquiring and displaying essential patient data such as ECG/Heart Rate, Intra Cardiac ECG (ICECG), invasive blood pressures, pulse oximetry, respiration, cardiac output, body temperatures, and EtCO2. BP waveforms from different heart and vascular sites, are continuously presented on the Physiological Waveform Display. The hemodynamic data, waveform, gradients and FFR, can be stored, recorded, analyzed and presented in a variety of report formats.
Device Story
Horizon XVu is a computerized catheterization laboratory system used in hospitals by clinicians to monitor patient vital signs during endovascular procedures. It acquires physiological signals (ECG, ICECG, invasive/non-invasive BP, SpO2, respiration, temperature, EtCO2) via a Front End unit (CFE). The system processes these signals in real-time on a UNIX-based workstation. The modification adds Fractional Flow Reserve (FFR) measurement, which calculates the ratio of distal coronary pressure (Pd) to proximal aortic pressure (Pa) using two invasive pressure channels. The system displays waveforms and numerical data, stores case history, and generates reports. Clinicians use the output to evaluate lesion severity and guide stent implantation decisions. The device benefits patients by providing integrated hemodynamic assessment without requiring additional standalone diagnostic equipment.
Clinical Evidence
Bench testing only. System-level software validation was performed comparing the modified Horizon XVu to the predicate Horizon XVu system. Performance testing ensured functional requirements and specifications were met. Electrical safety and EMC testing were conducted by an independent laboratory (Standard Institute of Israel) per IEC 60601-1, 60601-1-1, 60601-1-2, 60601-2-27, 60601-2-30, 60601-2-34, and 60601-2-49.
Technological Characteristics
System consists of a Front End unit (CFE) and a central UNIX-based workstation. Inputs: ECG, ICECG, invasive/non-invasive BP, SpO2, temperature, EtCO2. FFR uses two invasive pressure channels. Connectivity: Ethernet network between CFE and central system. Standards: IEC 60601-1, 60601-1-2, 60601-2-27, 60601-2-30, 60601-2-34, 60601-2-49. Software: Real-time control and processing. Defibrillator-protected ECG amplifier. Isolated inputs for vital signs.
Indications for Use
Indicated for patients undergoing coronary and peripheral endovascular procedures and angioplasty requiring hemodynamic monitoring, including ECG, heart rate, intra-cardiac ECG, invasive blood pressure, pulse oximetry, respiration, cardiac output, body temperature, EtCO2, and Fractional Flow Reserve (FFR) measurement. Prescription use only.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains the logo for Mennen Medical. The logo consists of a stylized letter "m" enclosed in a circle on the left, followed by the words "mennen" stacked on top of "medical" on the right. The text is in a simple, sans-serif font.
#### Topic: Traditional 510K for Horizon XVu-addition of FFR Programmable diagnostic computer
#### Establishment Name, Registration Number and Address:
| 510K number: | K123792 |
|----------------------|-----------------------------------------|
| Name: | Mennen Medical Ltd. |
| Registration Number: | 9611022 |
| Operator Number: | 9069173 |
| Address: | 4 Hayarden Street, Yavne, 81228, Israel |
| Postal Address: | PO Box 102,<br>Rehovot, 76100, Israel |
| Tel: | +972-8-9323333 |
Fax: +972-8-9328510 Contact person: Ifat Oren, Regulatory Affairs
To: Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville MD, 20850
**AUG 30 2013**
Attn.: Document Control Clerk From: Ifat Shwarts, Regulatory Affairs
The following information is being submitted in conformance with 21 CFR 807.87:
| 1. Classification Name | Programmable diagnostic computer |
|------------------------------------------------|------------------------------------------|
| 2. Classification Number: | 21 CFR 870.1425 |
| 3. Common/Usual Name | Programmable diagnostic computer |
| 4. Trade/Proprietary Name | Horizon XVu |
| 5. Part Number of Horizon SE | 960-100-310 (115V)<br>960-100-320 (230V) |
| 6. Establishment Registration Number | 9611022 |
| 7. FDA Classification | Class II |
| 8. Product Code | DQK |
| 9. Reviewing Panel | Cardiovascular |
| 10. Performance Standards | page 12 |
| 11. 510(k) Marketing clearance for Horizon XVu | K091165 – August 7, 2009 |
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Mennen Medical. The logo consists of a stylized letter "m" inside of a circle on the left, with the words "mennen medical" to the right of the circle. The word "mennen" is on top of the word "medical".
#### ComboMap
#### Traditional 510(k): Device Modification - Horizon XVu K123792
#### Terminology
#### Subject of this 510(k) = Horizon XVu with FFR
The Horizon XVu is a modified device, a system identical to the Horizon XVu Cathlab with the addition of Fractional Flow Ratio (FFR) measurement.
This submission is for a modification of the Horizon XVu that Mennen Medical has received 510K clearance for marketing by the FDA (K091165) August 7, 2009 (enclosed in part 21)
We intend to add to the Horizon XVu Computerized Catheterization Laboratory, FFR measurement option using as a predicated device the:
Volcano ComboMap = The predicated device for FFR measurement. The COMMBOMAP was cleared for marketing by FDA (K041134) June 2, 2004 (see attached FDA letter - part 21)
#### Intended Use of the Horizon XVu with FFR
The Horizon XVu is a state-of-the-art computerized laboratory, capable of acquiring and displaying essential patient data such as ECG/Heart Rate, Intra Cardiac ECG (ICECG), invasive blood pressures, pulse oximetry, respiration, cardiac output, body temperatures and EtCO2.
Heart rate, multi-lead ECG and BP waveforms from different heart sites are continuously presented on the Physiological Waveform Display. The hemodynamic data, and FFR waveform and numerical, can be stored, recorded, analyzed and presented in a variety of report formats.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for Mennen Medical. The logo consists of a circle with the letter "m" inside of it, followed by the words "mennen" in bold black letters and "medical" in light gray letters. The logo is simple and modern, and it is likely used to represent the company's brand.
### 1. Device Description: Horizon XVu
The prime function of the Horizon XVu (Cathlab) is to acquire and display vital-sign data and waveforms in real time during the catheterization process, creating a fully documented case history.
Image /page/2/Picture/4 description: This image shows a Horizon XVu Cathlab System, including the console option and a general view of the system. The image also shows the physician's remote, the CFE (presumably an electronic component), and the CFE mounting location. The Horizon XVu Cathlab System is labeled at the bottom of the image.
#### Page 3 of 12
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Mennen Medical. The logo consists of a circular graphic on the left and the words "mennen medical" on the right. The word "mennen" is on top of the word "medical". The circular graphic contains a stylized letter "m".
## 2. Functional Description of the Horizon XVu
The Horizon XVu is capable of acquiring and displaying essential patient data such as ECG/Heart Rate, ICECG, invasive blood pressures, pulse oximetry, respiration, cardiac output, and body temperature. Heart rate, multi-lead ECG, EtCO2 and BP waveforms from different heart and vascular sites are continuously presented on the Physiological Waveform Display. The hemodynamic data, waveform and numerical, can be stored, analyzed and presented in a variety of report formats.
The Horizon XVu software includes graphic presentation of the Heart and abdominal, cranial and peripheral vascular system, to support reporting of heart and vascular catheterization.
#### Catheterization procedure
During a typical catheterization procedure, the physician guides a catheter to the site of interest by observing an X-ray display and the Horizon XVu Patient Data Display, and then instructs the catheterization technician at the on-line workstation to acquire pressure waveform(s) data for that site. With a single keystroke, the technician labels the site and initiates waveform/data sample acquisition.
The waveform of interest changes to a pre-defined color at the Patient Data Display during acquisition for ease of identification. Another keystroke signals the Horizon XVu to stop the acquisition and begin analysis of the newly acquired pressure information via the dedicated Workstation Horizon XVu Computer, at which time the waveform for the corresponding pressure channel will change back to one of the default colors for non-defined pressures.
The Patient Data Procedure Display provides a view of the acquired ECG and heartbeats sampled for review, editing, and acceptance via the Horizon XVu Interface. The numerical results of calculations are based on the average of the accepted beats. These results will include such items as (for example) Systolic, Diastolic, and Mean pressure values.
The technician may at this point use the Interface to accept the displayed pressure information; doing so will store the values to the patient's on-line Procedure File.
The Horizon XVu is used for activities such as coronary and peripheral endovascular procedures and Angioplasty. The basic steps described above (catheter positioning, site definition/acquisition, analysis and acceptance) are performed as an integral part of these procedures.
The system has a computer that utilizes powerful, real-time, software to control the system operation and to process the vital patient sign data measurements acquired from the CFE or entered manually at the keyboard.
A Laser Printer is provided in the system. This provides printouts of textual and graphical summaries of all patient data and catheterization procedures.
・
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Mennen Medical. The logo consists of a circular graphic on the left and the words "mennen medical" on the right. The circular graphic appears to have the letter "m" inside of it. The words "mennen medical" are stacked on top of each other.
#### Base Configuration: Cathlab parameters
- ECG/Heart Rate/Respiration .
- 4 Invasive Blood Pressure channels .
- Non-invasive Blood Pressure .
- Pulse Oximetry (SpO2) .
- . EtCO2 (optional)
- . respiration
- temperature .
- ICECG Intra-Cardiac ECG .
#### Horizon XVu Options:
- . Full Disclosure
- Off-line workstations .
- Remote Interactive terminal .
- Angiography Analysis Package .
- . Fractional Flow Reserve - FFR
- CDR, DVD drive .
- Choice of Console Table regular, enhanced, compact or without consol .
### Main components of the Horizon XVu:
The Horizon XVu system consists of:
### (A) a Front End unit and
#### (B) a Central system
The "Cathlab Patient Front End" (CFE) acquires, processes, and converts vital signs (A) from the patient into digital signals. The CFE then sends the digitized signals and data, via a network connection, to the central system of the Horizon XVu for process and display.
The CFE can acquire the following physiological signals of the patient:
- · ECG the CFE acquires an ECG waveform and measures Heart Rate
- · ICECG the CFE measures 6 channels of differential intra-cardiac ECG
- · Blood Pressure the CFE acquires a BP waveform and measures Systole, Diastole and Mean Pressure
- · Temperature the CFE measures Temperature by means of a numeric value in Co or Fo
- · SpO2 the CFE acquires and measures oxygen saturation and creates a photoplethysmographic waveform and numeric value of the oxygen saturation
- · EtCO2 the CFE measures CO2 during the respiration cycle and present the end tidal (end expiratory) CO2 and the inspired CO2 - inCO2 and the respiration rate - RR
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Mennen Medical. The logo consists of a stylized letter "m" inside of a circle on the left, with the words "mennen" and "medical" stacked on top of each other to the right of the circle. The text and the letter "m" are in a simple, sans-serif font.
- The Central System contains the following main devices: (B)
- · Computer- see details below
- · Two local LCD displays
- Video line driver
- · AC Power Unit
- · Laser printer
- Hub
- Modem
#### FFR - Fractional Flow Reserve
FFR is a mode that uses two invasive pressure inputs.
Pa - Proximal pressure measured with external pressure transducer at the artery or aorta. and Pd - Distal pressure measured with a pressure wire sensor in the coronary arteries across lesion. Mean pressures are calculated and the ratio between the means Pd/Pa is calculated and displayed. The value of Pd/Pa is calculated continuously during intracoronary or venous pharmacological Hyperemic stimulus. The pressure gradient across lesion is measured and the pressure ration is calculated. Providing the FFR parameter that is used to evaluate the importance of the lesion in limiting blood flow to the distal coronary.
#### Computer Workstation
The computer workstation is a uni-processor system that runs the Horizon XVu program on a UNIX operating system. The workstation receives the digitized signals from the CFE via the Ethernet hub, displays real-time vital signs, analyzes, processes, and calculates the vital sign data and waveforms, cardiac status in real time during the catheterization process, creating a fully documented case history. The workstation continuously displays the vital signs waveforms and data on the local LCD displays.
The hemodynamic data. ECG and FFR waveform and numerical, can be stored, recorded, analyzed and presented in a variety of report formats.
#### 3. Horizon XVu 960-OPT-580/585 - similarity and Differences in Design:
Horizon XVu with FFR 960-OPT-580/585 vs. Horizon XVu
The following technological and other characteristics/features apply to both the Horizon XVu and to the Horizon XVu with FFR 960-OPT-580/585
- . Intended for use in hospitals
- . Do not change the functionality of the Cathlab system
- . Isolated inputs for vital signs sensors
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Mennen Medical. The logo consists of a stylized letter "m" inside of a circle on the left, followed by the words "mennen" and "medical" stacked on top of each other on the right. The text is in a simple, sans-serif font.
- ECG amplifier front end with defibrillator protection .
- Invasive BP input circuit .
- Non Invasive BP measurement .
- . SpO2 measurement
- . Selectable filters for ECG and BP
- . Analog output for ECG and BP
- . Doctor's display of vital signs and physiological waveforms
- . Show doctor feature for diagrams from the technician screen
The main difference between the Horizon XVu and the Horizon XVu with FFR 960-OPT-580/585 .:
On the Horizon XVu with FFR 960-OPT-580/585 it is possible to display the FFR waveform and value.
The SW changes of the Horizon XVu with FFR 960-OPT-580/585 vs. the Horizon XVu are not related to the type, size, method of integration, and assembly of the HW components.
The following table compares the major software element and/or changes done in the Horizon XVu vs. the Horizon XVu with FFR 960-OPT-580/585, Cathlab:
| SW Component | Horizon XVu<br>960-800-XXX | Horizon XVu with<br>9600OPT580/585<br>FFR |
|----------------------------------------------------------------|------------------------------------------|-------------------------------------------|
| Doctors display | All waveforms and numeric vital<br>sings | Same |
| Display of leads and<br>waveforms on Cathlab<br>display screen | Yes | Yes |
| Solaris Operating<br>System for Cathlab | Yes | Yes |
| Technician Screen | Controls only | Same |
| GUI | Same | Same |
| Menus | Full set | Same |
| Vital Signs | Full set | Full set + FFR |
| | Horizon XVu | Horizon XVu- FFR |
|--------------------------|-------------|---------------------------------------------------------------------|
| Part/Option Number | 960-800-XXX | 960-OPT-580 for use of RADI<br>wire<br>960-OPT-585 for Volcano wire |
| Input Circuit Parameters | | |
| Surface ECG | Yes | Yes |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for Mennen Medical. The logo consists of a circular graphic on the left and the words "mennen medical" on the right. The word "mennen" is on top of the word "medical". The circular graphic contains a stylized letter "m".
| ICECG | | |
|-------------|------------|-----------------------------|
| NIBP | Yes | Yes |
| Invasive BP | 4 channels | 4 channels – 2 used for FFR |
| Respiration | Yes | Yes |
| SpO2 | Yes | Yes |
| EtCO2 | Yes | Yes |
#### Menu Details
#### Technician Display
The technician display of the Horizon XVu FFR 960-OPT-580/585 is identical to the Horizon XVu with the addition of control of the FFR procedure.
The GUI control keys of the Horizon XVu 960-OPT-580/585 are same as to the Horizon XVu, with the addition of FFR procedure
We submit that the addition of the FFR 960-OPT-580/585 option to the Horizon XVu is limited to the addition of a new parameter - FFR. This change does not amount to a change in the "fundamental scientific technology" of the Cathlab and does not disqualify the Horizon XV u from being the subject of a 510(k).
#### Comparison to the predicate device
The Horizon XVu with FFR 960-OPT-580/585 option is comparable to the FFR mode of the predicated device Volcano - ComboMap (see also part 12- Substantial equivalence) The functional menus of Horizon XVu were modified by adding the option to measure and calculate the FFR vital sign and waveform, without any change to the other vital sign monitoring capabilities
| Subject | Volcano - ComboMap<br>Predicated Device | FFR mode on Horizon<br>XVu |
|--------------------|-----------------------------------------|------------------------------------------------------|
| FDA 510 K | K 041134 | K 091165 prior to change |
| Features | FFR + Flow velocity | FFR |
| Pressure wire | Volcano | Volcano or Radi |
| Transducer input | 5μV/V/mmHg | 5μV/V/mmHg |
| Pressure range | -30 to +330 mmHg | -50 to +300 mmHg |
| Pressure bandwidth | DC to 25 Hz | DC to 6 Hz, DC to 12 Hz,<br>DC to 24 Hz, DC to 40 Hz |
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Mennen Medical. The logo consists of a stylized letter "m" inside of a circle on the left, followed by the words "mennen" on the top line and "medical" on the bottom line. The text is in a simple, sans-serif font.
| Measurement accuracy | ± 3 mmHg (-30 to +100 mmHg)<br>± 3% (> 100 mmHg) | +/- 2 mmHg, or +/- 2%,<br>whichever is greater,<br>exclusive of transducer |
|------------------------|-------------------------------------------------------------|----------------------------------------------------------------------------|
| # pressure channels | 2 | 4 (2 used for FFR) |
| Analog Pressure output | 1V/100 mmHg. | same |
| Pa | Mean Arterial Pressure | Same |
| Pd | Mean Distal Pressure at<br>the tip of the pressure<br>wire | Same |
| FFR = Pd/Pa | Ratio | Same |
| Tracing | ECG,<br>Pa dynamic , Pa mean,<br>Pd dynamic, Pd mean<br>FFR | Same |
#### Explanation of differences
The basic amplifier specifications are the same for the Horizon XVu (FDA approved) and the Horizon XVu with FFR
The number of pressure channels of the Horizon XVu is four (4). Two (2) of these channels are used for the FFR calculation. Same as in the predicated Volcano ComboMap. The FFR option of the Horizon XVu is comparable only to the pressure measurement of the COMBOMAP, and not to the ComboMap capabilities to measure flow.
### 4. Reasons to the change
### FFR
Mennen Medical has added to the FFR capability, for the following reasons
- . Users request to add to the Horizon XVu the FFR software capability to enable use of the pressure measurements that are performed in any case of hemodynamic catheterization.
- . Use of FFR feature on the Horizon XVu provides an important measurement without the need for an addition device.
- . The Horizon XVu is used for coronary diagnosis and implantation of stents and FFR is a diagnostic tool in the decision if and where to implant a stent.
We submit that no changes were made:
- · to the control mechanism of the Horizon XVu.
- · to the operating principle of the Horizon XVu.
- · to the energy type of the Horizon XVu
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for Mennen Medical. The logo consists of a stylized letter "m" inside of a circle on the left, followed by the words "mennen medical" in a simple sans-serif font. The word "mennen" is on top of the word "medical."
The FFR option is a software option , thus sterilization is not an issue.
FFR capability uses single use pressure wire sensors not manufactured by Mennen Medical which are supplied sterile. Thus sterilization is not an issue.
#### 5. Conclusion of comparison of technological characteristics:
We consider the FFR option of the Horizon XVu to be substantially Equivalent to the FFR in the Volcano ComboMap
We submit that any difference between the FFR measurement of the Horizon XVu and the FFR measurement of the ComboMap.
- . fall within the scope of a 510(k) Device Modification and
- do not raise any new issues of safety and effectiveness .
### 6. Verification, Validation and Testing
The Horizon XVu has been subject to extensive performance testing to ensure that: 1. The acquisition and display of the patient data and waveforms by the Horizon XVu with 960-OPT-580/585 remain the same for the predicate device Horizon XVu.
At the system level, SW Validation of the performance of the Horizon XVu with 9600PT580/585 as compared to the Horizon XVu Cathlab system, was carried out in accordance with the test plan described in the Mennen Medical Validation Test Procedure for the Horizon XVu.
The SW Test Description for the Horizon XVu with 9600PT580/585 was derived from the SW Test Description for the Horizon XVu Cathlab system, with the necessary addition of the FFR measurements
Final testing for the Horizon XVu system included performance tests designed to ensure that the device meets all functional requirements and performance specifications, in accordance with the requirements of the Final Test Procedure for the Horizon XVu system.
Electrical Safety testing and EMC testing were performed by an independent testing laboratory (Standard Institute of Israel SII) to ensure that the device complicable industry and safety standards (attached in Part 17)
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image contains the logo for "mennen medical". The logo consists of a gray circle with a white lowercase "m" inside of it. To the right of the circle, the word "mennen" is written in black, and below that, the word "medical" is written in gray.
#### 7. Proposed Labeling
The system will be called Horizon XVu with FFR 960OPT580/585
Page IV of the introduction to the Horizon XVu User Guide contains the following
Prescription Notice: "Federal United States law restricts the sale and use of this instrument to qualified medical personnel only." The following symbols appear on page IX of the Horizon XVu User's Guides under the section entitled "Label Locations & Symbol Descriptions," and on the front panel of the CFE. See the image of the front panel of the CFE on page 11 below.
Image /page/10/Figure/6 description: This image shows a table with different medical symbols and their descriptions. The first row shows an exclamation point inside of a triangle, with the text "Attention - see Accompanying Instructions for Use". The second row shows a stick figure inside of a square, with the text "Type BF Applied part (next to NIBP, EtCO2 and SpO2 connectors)". The third row shows a heart inside of a square, with the text "Type CF Applied Part (next to IBP, Temperature and CO connectors)". The fourth row shows a heart inside of a square with two symbols on either side of the heart, with the text "Type CF Applied Part - Defibrillation Proof (next to ECG connector)".
Image /page/10/Picture/7 description: The image shows a medical device with several labeled sections. The sections include ECG/RESP, SpO2, CO/2 Temp, NIBP, Temp 1, and Temp 2/EtCO2. On the right side of the device, there are four IBP sections labeled 1 through 4, each with a 'Zero' button.
Symbols and labeling on the front panel of the CFE
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for Mennen Medical. The logo consists of a stylized letter 'm' enclosed in a circle on the left, followed by the word 'mennen' in a simple, sans-serif font on the top line. Below 'mennen', the word 'medical' is written in the same font, completing the logo.
#### 8. Voluntary Standards
Appropriate voluntary standards for this device, to which conformance have been demonstrated:
- * IEC 60601-1: (2005) Medical Electrical Equipment Part:1 General Requirements for Safety
- * IEC 60601-1-1 (2000) Medical Electrical Equipment Part 1-1: General Requirements for Safety Collateral Standard: Safety Requirements for Medical Electrical Systems
- * IEC 60601-1-2 (2007): Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests.
- ☆ IEC 60601-2-27 (2005): Medical electrical equipment, Part 2, Requirements for safety of electrocardiograph monitoring equipment.
- 今 IEC 60601-2-30 (1999): Medical electrical equipment, Part 2 - requirements for safety of automatic cycling indirect blood pressure monitoring equipment
- ☆ IEC 60601-2-34 (2005): Medical electrical equipment, Part 2 - Particular requirements for the safety of direct blood pressure monitoring equipment
- や IEC 60601-2-49 (2006): Particular Requirements for the safety of multifunction patient monitoring equipment
#### 9. Indications for Use
The Horizon XVu is a state-of-the-art computerized laboratory, capable of acquiring and displaying essential patient data such as ECG/Heart Rate, Intra Cardiac ECG (ICECG), invasive blood pressures, pulse oximetry, respiration, cardiac output, body temperatures and EtCO2.
Heart rate, multi-lead ECG and BP waveforms from different heart sites are continuously presented on the Physiological Waveform Display. The hemodynamic data, and FFR waveform and numerical, can be stored, recorded, analyzed and presented in a variety of report formats.
#### Attachment:
Horizon XVu User Manual - part 13 FTP for FFR option in Horizon XVu - Part 16 j Horizon XVu, Risk Analysis - part 16 d
{12}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hamoshire Avenue Document Control Center . WO66-G60 Silver Spring, MD 20993-002
Image /page/12/Picture/3 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
August 30, 2013
Mennen Medical Ltd. c/o: Ifat Shwarts Regulatory Affairs 4 Hayarden st., Yavne P.O. Box 102 Rehovot, IS 76100
K123792 Re:
Trade Name: Horizon XVu (FFR) Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable diagnostic computer Regulatory Class: Class II Product Code: DQX Dated: August 2, 2013 Received: August 5, 2013
Dear Ifat Shwarts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{13}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm 1 15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, Misbranding by reference to premarket notification (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDeyices/Safety/ReportaProblem/default.htmfor the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Owen P. Faris -S
for Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{14}------------------------------------------------
#### K123792
# Indications for Use
510(k) Number (if known):
Device Name: Horizon XVu
Indications For Use:
The Horizon XVu is a state-of-the-art computerized laboratory, capable of acquiring and displaying essential patient data such as ECG/Heart Rate, Intra Cardiac ECG (ICECG), invasive blood pressures, pulse oximetry, respiration, cardiac output, body temperatures, and EtCO2.
BP waveforms from different heart and vascular sites, are continuously presented on the Physiological Waveform Display. The hemodynamic data, waveform, gradients and FFR, can be stored, recorded, analyzed and presented in a variety of report formats.
*The Intended Use of the Horizon XVu as indicated above is same as the Indications For Use.
Prescription Use_YES (Part 21 CFR 801 Subpan D) AND/OR
Over-The-Counter Use NO (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Com
Digitally signed by
:Owen P. Faris -S
Date: 2013.08.30
15:27:18-04'00'
Page I of I
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.