K123787 · Jingzhou Haixin Green Cross Medical Products Co.,L · FXX · Jul 17, 2013 · General, Plastic Surgery
Device Facts
Record ID
K123787
Device Name
GREENCROSS SURGICAL MASK
Applicant
Jingzhou Haixin Green Cross Medical Products Co.,L
Product Code
FXX · General, Plastic Surgery
Decision Date
Jul 17, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The GreenCross Surgical Masks (for single use only) are indicated as a protective nose and mouth covering for health care workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situations where there is a risk of microorganism. body fluid, and particulate aerosol transfer.
Device Story
GreenCross Surgical Mask is a 3-ply, flat-pleated, single-use mask. Construction includes inner and outer layers of 100% spun-bond polypropylene and a middle layer of 100% meltblown polypropylene filter media. Features include latex-free elastic ear-loops or tie-on straps and a plastic wire nose piece. Used in clinical settings by healthcare workers and patients to provide a physical barrier against microorganisms, body fluids, and particulate aerosols. Device is non-sterile and intended for single use.
Clinical Evidence
Bench testing only. Performance evaluated via ASTM F2101 (BFE: 99.8%), ASTM F2299 (PFFE: 99.1%), ASTM F1862 (Fluid Resistance: 29/32 pass at 120 mm Hg), MIL-M-36945C (Delta-P: 2.6-2.9 mm H2O/cm²), 16 CFR 1610 (Flammability: Class 1), and ISO 10993 (Biocompatibility).
Technological Characteristics
3-ply flat-pleated mask. Materials: 100% spun-bond polypropylene (inner/outer), 100% meltblown polypropylene (filter). Plastic wire nose piece. Latex-free elastic ear-loops or tie-on straps. Dimensions: 175±5 mm x 90±5 mm. Biocompatible per ISO 10993.
Indications for Use
Indicated for health care workers and patients as a protective nose and mouth covering during medical and surgical procedures where there is a risk of exposure to microorganisms, body fluids, and particulate aerosols.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Surgical Mask (For Single Use Only) by Wellmien (Suzhou) Imp. & Exp. Trading Co., Ltd (K101000)
Submission Summary (Full Text)
{0}------------------------------------------------
K123787
# Section 2: 510(k) Summary:
# JUL 1 7 2013
This summary of 510k safety and effectiveness information is being submitted In accordance with the requirements of 21CFR 807.92
### Submitter & Foreign Manufacture Identification
Jingzhou Haixin Green Cross Medical Products Co., Ltd Century Building 19F A. Jianghan RD 206 Wuhan City, Hubei 430010, P. R, China Contact person: Bill Liao Tel: +86-278-281-5070
### Submitter's FDA Registration Number: N/A
### US Agent and Contact Person 1.
Xiangyang Song 6 Vanderblet Road, Acton, MA, 01720 Tel: 312-479-6204 Email: xysong@gmail.com
Date of Summary: December 03, 2012, Date of Revision 1: December 26, 2012, Date of Revision 2: May 17, 2013
### 2. Regulatory Information
Name of Device: GreenCross Surgical Mask Model No.: GFM12 Type: Tie-on, Earloop Common Name: Surgical Mask Panel: General & Plastic Surgery Device Classification: Class II Regulation Number: 878.4040 Product Code: FXX Classification Name: Mask, Surgical
### Predicate Device Information: 3.
K101000: Surgical Mask (For Single Use Only) by Wellmien (Suzhou) Imp. & Exp. Trading Co., Ltd
#### Device description: 4.
{1}------------------------------------------------
GreenCross surgical masks, Type: Earloop or Tie-on, are pleated 3-ply mask. Inner and outer layers are made of 100% spun-bond polypropylene. Middle player is made of 100% meltblown polypropylene filter media. Ear-loops are made of soft latex free elastic loops. All of the materials used in the construction of the GreenCross surgical masks are being used in currently marketed devices, except the nose piece used plastic wire instead of aluminum wire. (see predicate information)
### Intended Use: ડ.
The GreenCross Surgical Masks (for single use only) are indicated as a protective nose and mouth covering for health care workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situations where there is a risk of microorganism. body fluid, and particulate aerosol transfer.
### 6. Comparison to Predicate Devices
GreenCross Surgical Mask is compared with its Predicate devices: K101000: Surgical Mask (For Single Use Only) by Wellmien (Suzhou) Imp. & Exp. Trading Co., Ltd. The design, material, and other technical characteristics of these two devices are very similar, except that the nose piece for GreeCross Surgical Mask used plastic wire instead of aluminum wire, also the size, color & design type (Ear Loop or Tie-on) are different to predicate device. These features are not expected to affect the safety and performance of the device.
| Description | Our Device | Predicate Device (K101000) |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for<br>Use | Nose and mouth covering for health<br>care workers and patients to prevent<br>microorganism, body fluid, and<br>particulate aerosol transfer. | Same |
| Layers | Three | Three |
| Materials | Outer layer is made of 100% spun-<br>bond polypropylene.<br>Middle player is made of 100%<br>meltblown polypropylene filter<br>media.<br>Inner layer is made of 100% spun-<br>bond polypropylene.<br>Ear-loops are made of soft latex free<br>elastic loops.<br>The nose piece is plastic wire | Outer layer is made of 100% spun-<br>bond polypropylene.<br>Middle player is made of 100%<br>meltblown polypropylene filter<br>media.<br>Inner layer is made of 100% spun-<br>bond polypropylene.<br>Ear-loops are made of soft elastic<br>loops.<br>The nose piece is malleable<br>aluminum wire |
### Comparison of Intended Use, Design, Material and Specifications
{2}------------------------------------------------
| Dimensions | 175±5 mm x90±5 mm | Large: 180x90mm, Medium:<br>175 x95mm, Small: 145x90 mm |
|------------------------|-------------------|---------------------------------------------------------|
| Mask style | Flat Pleated | Flat Pleated |
| Design | EarLoop, Tie-on | EarLoop |
| Anti Fog<br>Shield | N/A | N/A |
| Color | white | Blue |
| NIOSH<br>certification | N/A | N/A |
## Comparison of Device Performance
| Test | Our Device | Predicate Device (K101000) |
|----------------------------------------------------|-----------------------------------|----------------------------|
| Fluid Resistance<br>Performance<br>ASTM F1862 | 29 out of 32 pass<br>at 120 mm Hg | 31 out of 32 pass |
| Particle Filtration<br>Efficiency<br>ASTM F2299 | 99.1% | 99.8% |
| Bacteria Filtration<br>Efficiency<br>ASTM F2101 | 99.8% | 99.9% |
| Differential Pressure<br>(Delta-P)<br>MIL-M-36945C | 2.6-2.9<br>(mm Water/cm²) | 3.7-4.0<br>(mm Water/cm²) |
| Flammability<br>16CFR 1610 | Class 1 None Flammable | Class 1 None Flammable |
| Biocompatibility<br>ISO 10993-5, -10 | Biocompatible | Biocompatible |
#### 7. Discussion of Non-Clinical Tests Performed to Determine Substantial Equivalence
The following non-clinical tests were performed to determine substantial equivalence. Tests were conducted following the recommended procedures outlined in the following standards. Test results met all relevant requirements in the test standards, and are comparable to the predicate device.
- (1) ASTM F2101 Bacterial Filtration Efficiency (BFE)
- Pressure Differential (Delta P) MIL-M-36945C Cl. MIL-M-36945C (2) 4.4.1.1.1
- ASTM F2299 Latex Particle Challenge (PFFE) (3)
- 16 CFR Part 1610 Flammability (4)
- Biocompatibility per ISO 10993 (5)
{3}------------------------------------------------
- (6) ASTM F1862 Fluid Resistant - Synthetic Blood Penetration Resistant Test
It is our conclusion that performance testing met all relevant requirements of the above said test standards. More detail comparison of the design, technical, and performance characteristics to the predicted device are summarized in "510(k) Premarket Notification for Jingzhou Haixin Surgical Mask" chapter 4 summary report, section 4.1 to 4.4. More details of non-clinical tests are summarized in Chapter 4, section 4.5 and 4.6
### Discussion of Clinical Tests Performed 8. Not applicable
#### 9. Conclusions
.
The GreenCross surgical mask has the same intended use and technological characteristics as the predicate devices. Moreover, bench testing contained in this submission demonstrates that the technological characteristics of GreenCross Surgical masks do not raise any new questions of safety or effectiveness. Therefore, GreenCross surgical masks By Jingzhou Haixin Green Cross Medical Products Co., Ltd are substantially equivalent to the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings outstretched, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circle around the eagle. The logo is black and white and has a simple, clean design. The eagle is a symbol of strength and freedom, and the logo represents the department's mission to protect the health of all Americans.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### July 17, 2013
Jingzhou Haixin Green Cross Medical Products Company, Limited C/O Mr. Xiangyang Song 6 Vanderbelt Road ACTON MA 01720
Re: K123787
Trade/Device Name: GreenCross Surgical Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FXX Dated: May 16, 2013 Received: June 14, 2013
Dear Mr. Song:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Song
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mary S. Runner -S
Kwame Ulmer, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K123787
## Section 1: Indication for Use Statement
## Device name: GreenCross Surgical Mask
The GreenCross Surgical Masks are indicated as a protective nose and mouth covering for health care workers and patients involved in medical and surgical procedures. The masks are indicated in any procedure or situations where there is a risk of microorganism, body fluid, and particulate aerosol transfer.
Prescription Use
Over the Counter Use _________________________________________________________________________________________________________________________________________________________
Elizabeth F. Claverie
2013.07.16 18:50:22 -04'00'
(Division Sign-Off) (Division of Anesthestology, General Hospital n Control, Dental I
**510(k) Number:** K123787
5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.