← Product Code [JJY](/productcode/JJY) · K123725

# FASTPACK CONTROL KIT (K123725)

_Qualigen, Inc. · JJY · Mar 8, 2013 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K123725

## Device Facts

- **Applicant:** Qualigen, Inc.
- **Product Code:** [JJY](/productcode/JJY.md)
- **Decision Date:** Mar 8, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1660
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The FastPack® Control Kit is an assayed quality control for the verification of the accuracy and precision of the FastPack® and FastPack® IP Systems when used for the quantitative determination of the analytes listed in the package insert. The following analytes are included in the package insert: - Free Thyroxine (FT4) - Human Chorionic Gonadotropin (hCG) - Testosterone - Total Prostate Specific Antigen (tPSA) - Thyroid Stimulating Hormone (TSH)

## Device Story

FastPack® Control Kit is a liquid, multi-analyte assayed quality control product; contains synthetic matrix with chemicals, preservatives, stabilizers, and human/synthesized analyte constituents (FT4, hCG, testosterone, tPSA, TSH). Used in clinical laboratories to monitor performance of FastPack® and FastPack® IP Systems. Healthcare providers compare observed control values against established ranges to verify assay accuracy and precision. Benefits include ensuring consistent reagent and instrument performance, thereby supporting reliable clinical decision-making for patient diagnostic testing.

## Clinical Evidence

No clinical data. Bench testing only. Stability protocols (shelf-life and open-vial) were reviewed and accepted. Value assignment performed via replicate analysis using six FastPack analyzers, three reagent lots, and two calibrations.

## Technological Characteristics

Matrix: Bovine serum albumin (BSA). Form: Ready-to-use liquid. Fill volume: 5 mL. Storage: 2–8°C. Stability: 18 months shelf-life, 120 days open-vial. Analyte traceability: Commercially available standards.

## Regulatory Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

## Predicate Devices

- Immunology Control (k960824)
- FastPack® Controls for PSA (k003095)
- FastPack® TSH Controls (k052301)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k123725

B. Purpose for Submission:
New device

C. Measurand:
Multi-analyte control material for Free Thyroxine (FT4), Human Chorionic Gonadotropin (hCG), Testosterone, Total Prostate Specific Antigen (tPSA), and Thyroid Stimulating Hormone (TSH)

D. Type of Test:
Not applicable

E. Applicant:
Qualigen, Inc.

F. Proprietary and Established Names:
FastPack Control Kit

G. Regulatory Information:

|  Regulation Section | Classification | Product Code | Panel  |
| --- | --- | --- | --- |
|  21 CFR § 862.1660 | Class I, reserved | JJY, multi-analyte controls, all kinds (assayed) | Clinical Chemistry (75)  |

H. Intended Use:

1. Intended use(s):
Refer to Indications for Use below.

{1}

2. Indication(s) for use:

The FastPack Control Kit is an assayed quality control for the verification of the accuracy and precision of the FastPack System and FastPack IP Systems when used for the quantitative determination of the analytes listed in the package insert. The following analytes are included in the package insert:

- Free Thyroxine (FT4)
- Human Chorionic Gonadotropin (hCG)
- Testosterone
- Total Prostate Specific Antigen (tPSA)
- Thyroid Stimulating Hormone (TSH)

3. Special conditions for use statement(s):

For in vitro diagnostic use.

For prescription use only.

4. Special instrument requirements:

FastPack System

I. Device Description:

The FastPack Control Kit consists of a low and high control. Each control is prepared from bovine serum albumin (BSA) as the primary protein source and to provide a viscosity similar to human serum. Other components added include constituents of human origin, stabilizers, inhibitors, and preservatives. Added analyte materials (hCG, TSH, FT4, Testosterone, and PSA) are purchased commercially. The controls are supplied in ready to use liquid form and packaged in a 5 mL bottle each.

All human source materials were tested and found to be negative for, HIV ½, HBV, and HCV using FDA approved methods.

J. Substantial Equivalence Information:

1. Predicate device name(s):

Immunology Control (containing FT4, testosterone, and hCG)
FastPack® Controls (containing PSA)
FastPack® TSH Controls

{2}

2. Predicate 510(k) number(s):

k960824

k003095

k052301

3. Comparison with predicate:

|  Similarities |   |   |   |
| --- | --- | --- | --- |
|  Item | Device FastPack® Control Kit | Predicate 1 Immunology Control (containing FT4, testosterone, and hCG) (k960824) | Predicate 2 and 3 FastPack® Controls for PSA (k003095) and FastPack® TSH Controls (k052301)  |
|  Intended Use | Same | Assayed quality control for the verification of the accuracy and precision of the analytes listed in the package insert. | Assayed quality control for the verification of the accuracy and precision of the analytes listed in the package insert  |
|  Form | Same | Liquid | Liquid  |
|  Fill Volume | Same | 5 mL | 5 mL  |
|  Differences |   |   |   |
| --- | --- | --- | --- |
|  Item | Device FastPack® Control Kit | Predicate 1 Immunology Control (containing FT4, testosterone, and hCG) (k960824) | Predicate 2 and 3 FastPack® Controls for PSA (k003095) and FastPack® TSH Controls (k052301)  |
|  Matrix | Synthetic | Human Serum | Bovine Serum Albumin  |
|  Opened Vial Stability at 2 – 8°C | 120 days | 30 days | 9 months  |
|  Shelf Life Stability | 18 months at 2 – 8°C | 3 years at -20°C | 12 months at 2 – 8°C  |

{3}

K. Standard/Guidance Document Referenced (if applicable):

None referenced

L. Test Principle:

Not applicable

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Not applicable

b. Linearity/assay reportable range:

Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Traceability

The analytes contained in the FastPack Control Kit were obtained from commercially available sources and are traceable to commercially available standards.

Stability

Shelf-life and open-vial stability protocols and acceptance criteria were reviewed and found to be acceptable. The closed vial (shelf-life) stability claim at the recommended storage temperature of 2 to 8 °C is 18 months based on ongoing real-time studies. The opened vial stability claim at the recommended storage temperature of 2 to 8 °C is 120 days based on real time data.

Value Assignment

Values are assigned to the calibrators based on replicate analysis using six FastPack analyzers, three reagent lots, and two calibrations. The sponsor calculates a range based on ± 3 standard deviations from the mean for each level for each analyte, but states that each laboratory should establish their own acceptable ranges.

d. Detection limit:

{4}

Not applicable

e. Analytical specificity:
Not applicable

f. Assay cut-off:
Not applicable

2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
Expected values are provided in the value assignment sheets provided with the package insert.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

{5}

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

6

---

**Source:** [https://fda-staging.innolitics.com/device/K123725](https://fda-staging.innolitics.com/device/K123725)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).