K123710 · Bio-Med USA, Inc. · FMF · May 9, 2013 · General Hospital
Device Facts
Record ID
K123710
Device Name
UNITOX BOTOX SYRINGE
Applicant
Bio-Med USA, Inc.
Product Code
FMF · General Hospital
Decision Date
May 9, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
UniTox Syringe is single use, sterile, intended use for the subcutaneous injection of Botox®Cosmetic into parts of body below the surface of skin.
Device Story
UniTox® Syringe is a sterile, single-use, disposable hypodermic syringe with a permanently affixed lumen needle. Device consists of a barrel, plunger rod with synthetic rubber gasket, and yellow end-cap. Operates as a manual piston syringe for subcutaneous delivery of Botox®Cosmetic. Used by clinicians in a medical setting. Output is the manual delivery of the drug; healthcare provider uses graduated markings on the barrel to measure dosage. Benefits include precise administration of Botox®Cosmetic.
Clinical Evidence
No clinical data. Substantial equivalence is based on design, material, and performance characteristics identical to the predicate device.
Technological Characteristics
Piston syringe; sterile, single-use, disposable. Includes barrel, plunger rod, synthetic rubber gasket, and permanently affixed lumen needle. Conforms to ISO 7864 (hypodermic needles) and ISO 8537 (sterile syringes). No electronic components or software.
Indications for Use
Indicated for subcutaneous injection of Botox®Cosmetic into body areas below the skin surface. Prescription use.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
# KIZ3710
MAY 9 2013
Bio-Med USA Inc
UniTox®Syringe
# Summary of safety and effectiveness
In accordance with Section 513(1) of the SMDA as defined in 21CFR part 807.3, This summary is submitted to obtain Premarket 510(K) notification
- 1. Submitter Mr. Young Chi. Register Number. 2246683 Bio-Med USA Inc. 111 Ellison Street Paterson .. NJ 07505 Tel: 973 870 2361, 973 278 5222 Fax: 201 934 6030 E mail: biomedusa@msn.com
- 2. Name of Medical Device Trade name UniTox® Syringe . Piston Syringe Classification name : Regulation : 880.5860 Class : II Product code : FMF/FMI
- Substantial Equivalence (Identification of legally Marketed Device) 3. The UniTox @ Syringe are substantially equivalent in Design, Function, Performing and all used material to already cleared DMJECT Uni-body Insulin Syringe by K993017, and other several needle permanently attached Insulin Syringe, but different graduate to use Botox®Cosmetic ..
## 4. Device Description
UniTox® Syringe are sterile, single use, disposable hypodermic syringes with a permanently affixed lumen needle to the tip of syringe. The syringe consists of a barrel, a plunger rod with Synthetic Rubber Gasket, and yellow colored end-cap over the needle to prescrve sterility of the fluid path.
In addition the UniTox ® Syringe are pyrogen free, and available , 0.5cc (20units), 30G X 5/16"needle length and conformed following standard (graduate drawing attached)
ISO 7864 Sterile, Hypodermic Needles for single use
ISO 8537: second edition,: 2007 Sterile, Single use Syringe with/without needle
{1}------------------------------------------------
### 5. Device Intended use
UniTox Syringe is single use, sterile, intended use for the subcutaneous injection of Botox®Cosmetic into parts of body below the surface of skin.
- 6 Summary of Technological characteristics.
UniTox® Syringe are the same function, performing as those currently on market in permanently needle attached uni-body insulin syringe, and there are no difference in technological characteristic between the UniTox® Syringe and cited predicate device, only has different graduate (mark ) to use Botox®Cosmetic.
accordingly, no any new issues of safety or effectiveness raised. This device operates on principles of piston syringe.
## 7 Packaging and Labeling
The UniTox® Syringe is sterilized, and packed in blister individually, and one hundred I live on roles are packed in a Chipboard box, and also provided 10 sterilize syringe in a poly bag, 10 poly bags in a Chipboard box. Each blister pack, poly bag and Chipboard box are labeled with the contents and appropriate information per the FDA's quality systems regulation and labeling requirement (21 CFR part 801)
- Inner box drawing sample attached 0
- 10pcs poly bag. Individual blister, Graduate mark drawing attached t
## 8. Biocompatibility certify
UniTox® Syringe using exact same raw material, which has been previously tested and accepted Biocompatibility, and also manufactured with same process as alteady cleared Predicate Device DM JECT, SUREJECT uni-body insulin syringe K993017, so additional Biocompatibility testing to ISO10993 standard is not required
## Conclusion
The UniTox® Syringe submitted in this Pre-market notification is substantially equivalent to the DMFCT, SUREJECT insulin syringe cleared K993017 in respect to design, used material, producing process, Technology / Principle of Operation and Performance. material, producing process, res do not raise any new issues of safety or effectiveness.
End of Summary
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of a stylized human figure embracing a globe, representing the department's mission to protect the health of all Americans and provide essential human services.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 9, 2013
Mr. Young Chi President BioMed USA, Incorporated 111 Ellison Street PATTERSON, NJ 07505
Re: K123710
Trade/Device Name: UniTox® Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF, FMF, FMI Dated: March 25, 2013 Received: April 8, 2013
## Dear Mr. Chi
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Chi
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Image /page/3/Picture/8 description: The image shows a stylized logo or signature. The text "for" is visible at the top left, followed by a more elaborate script that includes the words "with" and "u-owits". The overall design is bold and somewhat abstract, with thick lines and looping elements. The right side of the image features a geometric shape resembling a stylized letter "A".
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{4}------------------------------------------------
## Bio-Med USA Inc.
# 510(K) submission
## UniTox® Syringe
# Indication for use statement
K 123710 510(K) Nr
: UniTox® Syringe Device Name
: Single use, Sterilized, Intended use for the subcutaneous injection Indication for use of Botox®Cosmetic into parts of body below the surface of skin.
or/and Over the counter Prescription use XX 21CFR 801 Sub part C Part 21CFR 801 Sub part D
PLEASE DO NOT USE BELOW THE LINE- CONTINUED ANOTHER PAGE IF NEEDED
Concurrence of CDRH, office of Device Evaluation (OED)
Image /page/4/Picture/10 description: The image shows the FDA logo. The logo is made up of the letters F, D, and A. The letters are all capitalized and are in a bold, sans-serif font. The letters are arranged in a horizontal line, with the F on the left, the D in the middle, and the A on the right. The logo is black and white.
Richard C. Chapman 2013.05.08 11:04:38 -04'00'
(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices
K123710 510(k) Number:
Add: 111 Ellison St, Paterson, NJ 07505. U.S.A. 1 973 278 5222
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.