The Dental Barrier Membrane is intended for use during the process of guided bone regeneration (GBR) and guided tissue regeneration (GTR) as a biodegradable barrier membrane for: - Ridge augmentation for later implant insertion - Augmentation around implants placed in immediate/delayed extraction sockets - Localized ridge augmentation - Alveolar ridge preservation/reconstruction - Osseous fill around implants in peri-implantitis bone defects - Over the window in lateral window sinus elevation procedures - Intra-bony defects around teeth - Treatment of recession defects, together with coronally positioned flap - In furcation defects in multi-rooted teeth.
Device Story
Intraoral porcine collagen membrane; used in dental GBR/GTR procedures. Acts as physical barrier to contain bone graft material; excludes soft tissue in-growth to facilitate bone regeneration. Available in rough/smooth surface textures; supplied sterile in double blister packs. Used by dental clinicians in clinical settings. Provides stable barrier during healing; resorbable over time. Benefits patient by supporting new tissue/bone growth at defect sites.
Clinical Evidence
No clinical data submitted. Evidence based on bench testing (tensile strength, denaturation temperature, expansion, enzymatic resistance) and animal studies comparing resorption and inflammatory response to predicate Bio-Gide. Biocompatibility testing performed per ISO 10993-1.
Technological Characteristics
Porcine-derived Type I and III collagen and elastin fibers. Resorbable. Available in 15x20mm, 25x30mm, 30x40mm sizes. Surface weight >10 mg/cm². Sterilized by ethylene oxide. Characterized per ASTM F2212-11.
Indications for Use
Indicated for patients requiring guided bone or tissue regeneration in dental procedures, including ridge augmentation, implant site preparation, alveolar ridge preservation, peri-implantitis bone defects, sinus elevation, intra-bony defects, recession defects, and furcation defects.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
# Kr2369
FEB 2 4 2014
### Revised 510(k) Summary for the Matricel GmbH Matricel Dental Barrier Membrane (per 21CFR 807.92)
#### 1. SUBMITTER/510(K) HOLDER
Matricel GmbH
Kaiserstrasse 100
D-52134 Herzogenrath
Germany
| Contact Person: | Dr. Ingo Heschel |
|-----------------|---------------------|
| Telephone: | +49 2407 5644-11 |
| Fax: | +49 2407 5644-10 |
| Email: | heschel@matricel.de |
| Date Prepared: | July 24, 2013 |
#### 2. DEVICE NAME
| Proprietary Name: | Matricel Dental Barrier Membrane |
|----------------------|------------------------------------------|
| Common/Usual Name: | Barrier, animal source, intraoral |
| Classification Name: | Bone Grafting Material (21 CFR 872.3930) |
| Product Code: | NPL |
| Device Class: | 2 |
#### 3. PREDICATE DEVICES
- · Bio-Gide (Ed. Geistlich Soehne AG fuer Chemische Industrie, K042197 and K050446)
- · OssixTM Plus (ColBar LifeSciences Ltd., K053260)
#### 4. DEVICE DESCRIPTION
The Matricel Dental Barrier Membrane is an intraoral porcine membrane intended to be used in dental applications as a barrier membrane for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR) to support the growth of new tissue and bone. The Dental Barrier Membrane is composed of a network of highly purified porcine collagen (predominantly Type I with some Type III) and elastin fibers.
The Dental Barrier Membrane is offered in three sizes (15x20mm, 25x30mm and 30x40mm) with Rough and Smooth surfaces to accommodate physician preferences. The Smooth version is further processed to achieve a flattened surface, whereas the Rough version retains the natural fibrillar structure of the membrane. The Rough and
{1}------------------------------------------------
Smooth membrane have identical technological, mechanical and chemical characteristics.
The product is packaged in double blister packs and sterilized by ethylene oxide. The Dental Barrier Membrane is supplied sterile and is indicated for single use.
#### 5. INDICATION FOR USE/INTENDED USE
The Dental Barrier Membrane is intended for use during the process of guided bone regeneration (GBR) and guided tissue regeneration (GTR) as a biodegradable barrier membrane for:
- · Ridge augmentation for later implant insertion
- · Augmentation around implants placed in immediate/delayed extraction sockets
- · Localized ridge augmentation
- · Alveolar ridge preservation/reconstruction
- · Osseous fill around implants in peri-implantitis bone defects
- · Over the window in lateral window sinus elevation procedures
- · Intra-bony defects around teeth
- · Treatment of recession defects, together with coronally positioned flap
- · In furcation defects in multi-rooted teeth.
### 6. PRINCIPLES OF OPERATION
The process of GTR/GBR is facilitated by the use of resorbable implants such as the Dental Barrier Membrane and legally marketed barrier products, including the predicate devices Bio-Gide (subject of K042197 and K050446) and Ossix"14 Plus (subject of K053260). The Dental Barrier Membrane, like the predicate devices, achieves its principal intended action solely by providing a physical barrier to contain the bone graft material at the defect site and to exclude in-growth of surrounding tissue for a period of time long enough to allow bone regeneration to take place.
## 7. TECHNOLOGICAL CHARACTERISTICS
The proposed Dental Barrier Membrane and the predicate devices Bio-Gide (subject of K042197 and K050446) and Ossix" Plus (subject of K053260) have substantially equivalent indications for use. These porcine membranes are used to provide a barrier to support GBR and GTR. Both the proposed and predicate devices are derived from porcine tissue and share the similar technological and biochemical characteristics. Specifically, the Dental Barrier Membrane and Bio-Gide membranes contain Type I with some Type III collagen and elastin fibers. Furthermore, the available sizes and weight per unit surface area (surface weight) of the Dental Barrier
{2}------------------------------------------------
Membrane is within the specifications of Bio-Gide. Both the proposed and predicate membranes are resorbable over time.
All membranes are sterile, packaged in double blister packages, and have a shelf life of 3 years. The Dental Barrier Membrane and Ossix "M Plus are sterilized by ethylene oxide and Bio-Gide is sterilized by gamma irradiation.
#### 8. NON-CLINICAL TESTING
Bench testing, consisting of tensile strength, peak denaturation temperature, expansion after wetting were completed to demonstrate that the performance of the Dental Barrier Membrane met specifications. The Dental Barrier Membrane was characterized in accordance with ASTM F2212-11. Biocompatibility testing to ISO 10993-1 was performed to demonstrate the biological safety of the device.
Biostability (enzymatic resistance) testing was performed to assess the potential for degradation due to colonization with dental bacteria. Results demonstrated that the Dental Barrier Membrane behaves as stable as the predicate Bio-Gide when subjected to enzymatic degradation.
Animal testing was performed to compare the resorption characteristics of the proposed Dental Barrier Membrane and the predicate Bio-Gide. Results demonstrate that the two products do not elicit an inflammatory response at the implantation site and exhibit comparable resorption characteristics.
### 9. CLINICAL TESTING
No clinical testing was submitted in support of this 510(k) premarket notification.
### 10. CONCLUSIONS
The indications for use, principles of operation, and technological characteristics of the proposed Dental Barrier Membrane are substantially equivalent to the predicate devices Bio-Gide (subject of K042197 and K050446) and Ossix™ Plus (subject of K053260). Differences between the proposed and predicate devices are limited to minor differences in visual appearance; texture of membrane, and sterilization methods. These differences are minor and do not impact the safety and effectiveness of the device.
The safety, performance and effectiveness of the Dental Barrier Membrane for its intended use are demonstrated by bench and animal studies. Based on the evidence provided, Matricel believes that the proposed Dental Barrier Membrane is
{3}------------------------------------------------
substantially equivalent to the predicates.
| Table 5-1. Side-by-Side Comparison of Matricel Dental Barrier Membrane with | | | | |
|------------------------------------------------------------------------------|--|--|--|--|
| Predicate Devices | | | | |
| | Matricel<br>Dental Barrier Membrane* | Bio-Gide®<br>(Ed. Geistlich Soehne Ag fuer<br>Chemische Industrie) | Ossix™ Plus<br>(ColBar LifeScience Ltd.) |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory Status | Proposed | K042197, K050446 | K053260 |
| Product Code | NPL | NPL | NPL |
| Intended Use | The Dental Barrier Membrane is<br>intended for use during the process<br>of guided bone regeneration (GBR)<br>and guided tissue regeneration<br>(GTR) as a biodegradable barrier<br>membrane for:<br>• Ridge augmentation for later<br>implant insertion<br>• Augmentation around implants<br>placed in immediate/delayed<br>extraction sockets<br>• Localized ridge augmentation<br>• Alveolar ridge<br>preservation/reconstruction<br>• Osseous fill around implants in<br>peri-implantitis bone defects<br>• Over the window in lateral<br>window sinus elevation<br>procedures<br>• Intra-bony defects around teeth<br>• Treatment of recession defects.<br>together with coronally<br>positioned flap<br>• In furcation defects in multi-<br>rooted teeth | Bio-Gide® is recommended for:<br>• Simultaneous use of GBR-<br>membrane (Bio-Gide®) and<br>implants:<br>• Augmentation around implants<br>placed in immediate extraction<br>sockets;<br>• Augmentation around implants<br>placed in delayed extraction<br>sockets;<br>• Localized ridge augmentation for<br>later implantation;<br>• Alveolar ridge reconstruction for<br>prosthetic treatment:<br>• Filling of bone defects after root<br>resection, cystectomy, removal of<br>retained teeth:<br>• Guided bone regeneration in<br>dehiscence defects; and<br>• Guided tissue regeneration<br>procedures in periodontal defects. | Ossix™-Plus biodegradable<br>collagen membrane is intended for<br>use during the process of guided<br>bone regeneration (GBR) and<br>guided tissue regeneration (GTR) as<br>biodegradable barrier for:<br>• Ridge augmentation for later<br>implant insertions.<br>• Simultaneous ridge augmentation<br>and implant insertions.<br>• Ridge augmentation around<br>implants inserted in delayed<br>extraction sites.<br>• Ridge augmentation around<br>implants inserted in immediate<br>extraction sites.<br>• Alveolar ridge preservation<br>consequent to tooth (teeth)<br>extraction(s).<br>• Over the window in lateral<br>window sinus elevation<br>procedures.<br>• In implants with vertical bone<br>loss due to infection, only in cases<br>where satisfactory debridement<br>and implant surface disinfection<br>can be achieved.<br>• In intra bony defects around teeth<br>• For treatment of recession<br>defects, together with coronally<br>positioned flap<br>• In furcation defects in multi<br>rooted teeth |
| Material Source | Porcine | Porcine | Porcine |
| Composition | Collagen Types I and III, and elastin | Collagen Types I and III, and elastin | Collagen Type I*** |
| Size | 15x20mm, 25 x 30mm, 30 x 40 mm | 13 x 25 mm, 25 x 25 mm, 40 x 50<br>mm | 15 mm x 25 mm, 25 mm x 30 mm<br>and 30 mm x 40 mm |
| Surface weight ** | >10 mg/cm² | >10 mg/cm2 | Data not available |
| Form | Membrane | Membrane | Membrane |
| Cross-linking | No | No | Yes |
| Resorbable | Yes | Yes | Yes |
| Conformability | The membrane is flexible and<br>conforms to the defect site | The membrane is flexible and<br>conforms to the defect site | The membrane is flexible and<br>conforms to the defect site |
| Non-pyrogenic | Yes | Data not available | Data not available |
| Biocompatibility | Yes | Yes | Yes |
| Sterilization | Ethylene oxide | Gamma radiation | Ethylene oxide |
| Shelf Life | 3 years | 3 years | Data not available |
| Packaging | Double blister package | Double blister package | Double blister package |
* The Matricel Dental Barrier Membrane is offered in rough and smooth versions
** Surface weight [weight of the product in mg divided by the surface area in cm³]
*** Source: http://iadr.confex.com/iadr/2010dc/webprogramed/Paper129086.html
*** Source: <http://iadrc.confex.com/iadrc/2010dc/webprogrammod/Paper129086.htm>
Matricel GmbH Additional Information - K123697 July 24, 2013
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or other bird in flight. The logo is printed in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Comrol Center - WO66-G609 Silver Spring, MD 20993-0002
February 24, 2014
Matricel GmbH Dr. Ingo Heschel CEO Matricel GmbH Kaiserstrasse 100 D-52134 Herzogenrath Germany
Re: KI23697
Trade/Device Name: Matricel Dental Barrier Membrane Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: NPL Dated: February 4, 2014 Received: February 4, 2014
Dear Dr. Heschel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA bas made a determination that your device complies with other requirements of the Act
{5}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Image /page/5/Picture/8 description: The image shows the name Kwame Ulmer in bold black font on the left side of the image. On the right side of the image is a logo that is made up of the letters EDA and a triangle. The letters and triangle are made up of lines and dots.
for
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Revised Indications for Use
510(k) Number (if known): K123697
Device Name: Matricel Dental Barrier Membrane
#### Indications for Use:
... ...
The Dental Barrier Membrane is intended for use during the process of guided bone regeneration (GBR) and guided tissue regeneration (GTR) as a biodegradable barrier membrane for:
- · Ridge augmentation for later implant insertion
- · Augmentation around implants placed in immediate/delayed extraction sockets
- · Localized ridge augmentation
- · Alveolar ridge preservation/reconstruction
- · Osseous fill around implants in peri-implantitis bone defects
- · Over the window in lateral window sinus elevation procedures
- · Intra-bony defects around teeth
- · Treatment of recession defects, together with coronally positioned flap
- · In furcation defects in multi-rooted teeth.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mary S. Runner -S
2014.02.24
14:44:55 -05'00'
Susan Runner
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.