RE-TRACE URETERAL ACCESS SHEATH, 10/12 FRENCH, LENGTH 35 AND 45 CM, URETERAL ACCESS SHEATH, 12/14 CH-FR, LENGTH 35CM, UR

K123675 · Coloplast A/S · FED · Mar 1, 2013 · Gastroenterology, Urology

Device Facts

Record IDK123675
Device NameRE-TRACE URETERAL ACCESS SHEATH, 10/12 FRENCH, LENGTH 35 AND 45 CM, URETERAL ACCESS SHEATH, 12/14 CH-FR, LENGTH 35CM, UR
ApplicantColoplast A/S
Product CodeFED · Gastroenterology, Urology
Decision DateMar 1, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

To establish a continuous conduit during urological endoscopic procedures facilitating the in and out passage of endoscopes and other instruments into the urinary tract.

Device Story

The Re-Trace Ureteral Access Sheath and Ureteral Access Sheath are medical devices used to create a continuous conduit in the urinary tract during endoscopic procedures. The device consists of a reinforced tube/sheath, an introducer/dilator, a connector, and a clip. The 10/12 Ch/Fr model includes a stainless steel ring at the distal tip for reinforcement. The introducer/dilator features a guidewire entry eye. The device is used by urologists in a clinical setting to facilitate the insertion and withdrawal of endoscopes and other surgical instruments. By providing a stable pathway, the sheath protects the ureter and simplifies instrument exchange, potentially reducing procedural time and trauma to the urinary tract.

Clinical Evidence

Bench testing only. Performance testing included sheath/introducer/component break force, friction, kink resistance, injection, and guidewire pullout force testing. Biocompatibility testing was conducted in accordance with ISO 10993.

Technological Characteristics

Reinforced tube/sheath, introducer/dilator, connector, and clip. 10/12 Ch/Fr model features a stainless steel ring at the distal tip. Biocompatibility per ISO 10993. Non-powered, mechanical device.

Indications for Use

Indicated for patients undergoing urological endoscopic procedures requiring a conduit for the passage of endoscopes and instruments into the urinary tract.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K123675 #### 1. 510(k) SUMMARY #### MAR 1 2013 510(K) Owner's Name: Address: Phone/Fax/Email: 3050 Humlebaek, Denmark (612) 302-4987 Office: Mobile: (612) 968-9567 (612) 287-4138 Fax: Email: usbes@coloplast.com Coloplast A/S Brian E. Schmidt January 3, 2013 Regulatory Affairs Manager 1601 West River Road N Re-Trace Ureteral Access Sheath Minneapolis, MN 55411 Ureteral Access Sheath Ureteral Access Sheath Endoscope and Accessories 21CFR section 876.1500 Gastroenterology-Urology Devices . Holtedam 1 Name of Contact Person: Address/Contact: Date Prepared: Trade Name: Common Name: ﻢ Classification Name: Product code: FED Class II ## Legally Marketed Device To Which Your Firm Is Claiming Equivalence: The Re-Trace Ureteral Access Sheath 10/12 Ch/Fr & Ureteral Access Sheath are substantially equivalent in performance, indication, design and materials to Re-Trace Ureteral Access Sheath 12/14 Ch/Fr from Coloplast A/S, cleared under Premarket notification # K 102485. {1}------------------------------------------------ ## Description of the Device: The Re-Trace Ureteral Access Sheath 10/12 Ch/Fr & Ureteral Access Sheath are line extensions of the Re-Trace Ureteral Access Sheath 12/14 Ch/Fr. All these devices comprise the following components: - Reinforced tube/sheath . - Introducer/dilator . - Connector . - Clip . The only design addition to the 10/12 Ch/Fr reinforced tube/sheath compared to the 12/14 Cly/Fr sheath is the presence of a reinforcing Stainless Steel ring at the distal tip. For the Ureteral Access Sheath, the introducer/dilator also only has a guidewire entry eye at the distal tip compared with guidewire entry and exit eyes and three exit holes for fluid delivery on the Re-Trace Ureteral Access Sheath introducer/dilator. Apart from these modifications, all the Re-Trace Ureteral Access Sheath & Ureteral Access Sheath range is very similar in design. ### Intended Use of the Device: To establish a continuous conduit during urological endoscopic procedures facilitating the in and out passage of endoscopes and other instruments into the urinary tract. ### Predicate Device: The Re-Trace Ureteral Access Sheath 10/12 Ch/Fr & Ureteral Access Sheath are substantially equivalent in performance, indication, design and materials to Re-Trace Ureteral Access Sheath 12/14 Ch/Fr from Coloplast A/S, cleared under Premarket notification # K102485. ### Reference Device: Regarding the proposed size range for Re-Trace Access Sheath & Ureteral Access Sheath, the Cook Flexor® Access Sheath from Cook Urological, Inc., cleared under Premarket notification #K043418, will be used as a reference device. {2}------------------------------------------------ ## Summary and Conclusions from the Nonclinical Tests Submitted: Product Performance testing comparing the subject device to the predicate device included the following tests/analysis: Sheath/ introducer/component break force testing, friction testing, kink resistance testing, injection testing, guidewire pullout force testing, packaging testing. Biocompatibility testing was performed according to ISO 10993 on the Re-Trace Ureteral Access Sheath. #### Conclusion Substantial equivalence is supported by successful completion of the performance testing comparing Re-Trace Ureteral Access Sheath 10/12 Ch/Fr & Ureteral Access Sheath to the predicate device and the biocompatibility testing conducted on the Re-Trace Ureteral Access Sheath. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in a bold, sans-serif font. The text is arranged on a single line and is centrally aligned. To the left of the text is a graphic element that appears to be a stylized representation of a human figure or symbol, possibly related to health or human services. The overall impression is that of an official heading or title, likely from a document or publication associated with the Department of Health & Human Services. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 01. 2013 Coloplast A/S Coloplast Corp. % Mr. Brian E. Schmidt Regulatory Affairs Manager 1601 West River Road North MINNEAPOLIS MN 55411 Re: K123675 Trade/Device Name: Re-Trace Ureteral Access Sheath (Models ACXL10 and AXXL10) and Ureteral Access Sheath (Models ACXS12, AXXS12, ACXS10 AXXS10) Regulation Number: 21 CFR& 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FED Dated: November 27, 2012 Received: December 4, 2012 Dear Mr. Schmidt: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misb randing and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Brian E. Schmidt Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # 2. STATEMENT OF INDICATIONS FOR USE # Indications for Use 510(k) Number (if known): K 123675 Device Name: Re-Trace Ureteral Access Sheath 10/12 Ch/Fr & Ureteral Access Sheath Indications for Use: To establish a continuous conduit during urological endoscopic procedures facilitating the in and out passage of endoscopes and other instruments into the urinary tract. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Benjami 2013.03 oductive, Gastro-P no 20 of 265
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...