Classic sintering - Framework and reduced crowns in the anterior and posterior tooth region . - Bridge frameworks in the anterior and posterior tooth region with max. 2 pontics . - Crown caps in the anterior and posterior region . - Cone and telescoping crowns . Speed sintering - Framework and reduced crowns in the anterior and posterior tooth region . - Bridge frameworks in the anterior and posterior tooth region with max. 2 pontics and up to 9 units - Crown caps in the anterior and posterior region . - Cone and telescoping crowns Super speed sintering - Framework and reduced crowns with a maximum wall-thickness of 2 mm .
Device Story
inCoris ZI consists of zirconia ceramic (ZrO2) blocks used for custom dental restorations. Blocks are mounted to metal carriers and processed via Sirona CAD/CAM grinding machines. Following grinding, restorations undergo sintering (classic, speed, or super speed) to achieve final material properties. Used in dental clinics/labs by trained professionals. Output is a custom-fit dental restoration (crowns, bridges, caps). Benefits include biocompatible, high-strength ceramic restorations tailored to patient anatomy.
Clinical Evidence
No clinical data. Bench testing performed, including sintering tests, mechanical property evaluation, crack damage inspection, restoration fit, color suitability, and usability testing.
Technological Characteristics
Material: Yttria-stabilized zirconia ceramics (ZrO2). Standards: ISO 6872:2008, ISO 13356:2008, ISO 10993-1:2009. Physical properties: Density ≥ 6.05 g/cm3, Bending strength > 1100 MPa. Form factor: Block with metal carrier for CAD/CAM spindle mounting. Processing: CAD/CAM grinding followed by sintering.
Indications for Use
Indicated for dental patients requiring framework and reduced crowns, bridge frameworks (max 2 pontics, up to 9 units), crown caps, and cone/telescoping crowns in anterior and posterior regions.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
K.123.669
# APR 0 8 2013
## 510(k) Summary for Sirona Dental Systems inCoris ZI
### 1. Sponsor
Sirona Dental Systems GmbH
Fabrikstrasse 31.
D-64625 Bensheim
Germany
Contact Person: Fritz Kolle
Telephone: +49 6251 16 3294
Date Prepared: March 05, 2013
#### Device Name 2.
Proprietary Name: inCoris ZI
Common/Usual Name: Powder, Porcelain
Classification Name: Porcelain powder for clinical use
#### 3. Predicate Devices
inCoris ZI (K062509)
#### Intended Use 4.
Classic sintering
- · Framework and reduced crowns in the anterior and posterior tooth region
- . Bridge frameworks in the anterior and posterior tooth region with max. 2 pontics
- . Crown caps in the anterior and posterior region
- Cone and telescoping crowns .
### Speed sintering
- . Framework and reduced crowns in the anterior and posterior tooth region
- · Bridge frameworks in the anterior and posterior tooth region with max. 2 pontics and up to 9 . units
- · Crown caps in the anterior and posterior region
- · Cone and telescoping crowns
Sirona Dental Systems 510(k) inCoris ZI
Page 1
APPENDIX F
{1}------------------------------------------------
Super speed sintering
- Framework and reduced crowns with a maximum wall-thickness of 2 mm
#### న. Device Description and Function
The inCoris ZI are blocks of various sizes from which custom made dental restorations are grinded using Sirona CAD/CAM system. inCoris ZI ceramics constitute blocks comprised of zirconia ceramics (ZrO2). The blocks are initially manufactured in a partially sintered state; then, they are o individually processed to specification, and finally, densely sintered. One end plane of a block is mounted to a metal carrier that is inserted in the spindle's clamping chuck of the grinding machine. The blocks are available in different colors.
#### 6. Scientific Concept
The underlying scientific concept is
- . Processing dental restorations by Sirona Dental CAD/CAM System
- . Restorations are grinded from an inCoris ZI block by a Sirona CAM machine
- . Different sintering time to gain appropriate material properties
### 7. Physical and Performance Characteristics
### 7.1. Design
The design of the inCoris ZI is described in section 5, Device Description and Function.
### 7.2. Material Used
inCoris ZI ceramics constitute blocks comprised of zirconia ceramics (ZrO2). One end plane of a block is mounted to a metal carrier that is inserted in the spindle's clamping chuck of the grinding machine. The material is biocompatible according to ISO 10993-1: 2009, "Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process".
### 7.3. Physical Properties
Final technical data of densely sintered inCoris ZI.
| Density: | $\geq$ 6.05 g cm-3 |
|-------------------------------------------------|--------------------|
| Thermal expansion<br>coefficient (20 - 500 °C): | 11.0 10-6 K-1 |
| Bending strength: | > 1100 MPa |
Sirona Dental Systems 510(k) inCoris ZI
Page 2
APPENDIX F
7.4. Chemical Properties
{2}------------------------------------------------
| Component | inCoris ZI |
|----------------|------------|
| ZrO2+HfO2+Y2O3 | ≥ 99.0% |
| Y2O3 | 5.2% |
| HfO2 | 2% |
| Al2O3 | < 0.35% |
| Fe2O3 | < 0.3% |
#### 8. Summary of the technological characteristics
Both, Proposed and Predicate Sirona inCoris ZI are made of zirconia ceramics(ZrO2) and block shaped. Both devices meet ISO 6872: 2008, "Dentistry -- Ceramic materials" and ISO 13356: 2008, "Implants for surgery, Ceramic materials based on yttria-stabilized tetragonal zirconia (Y-TZP)". Physical and chemical properties are similar.
### Nonclinical Testing 9.
Sirona performed a series of tests to assess whether the device is appropriate for the indications for use. Sintering tests coupled with bench mechanical testing highlight that the mechanical properties are appropriate. Furthermore, crack damage inspection, restoration fit, color suitability, and overall usability tests were conducted.
#### 10. Clinical Testing
Clinical tests have not been performed.
### 11. Conclusion
Based on the comparison of intended use, indications, contra-indications, material properties and processing/fabrication, Sirona Dental Systems believes that the Proposed and Predicate (K062509) Sirona inCoris ZI blocks are substantially equivalent.
Sirona Dental Systems 510(k) inCoris ZI
APPENDIX F
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three abstract human figures in a row, representing health and human services. The figures are depicted in a dark color, contrasting with the background.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
### April 8, 2013
Mr. Fritz Kolle Sirona Dental Systems GmbH Fabrikstrasse 31 Bensheim. Germany D-64625
Re: K123664
Trade/Device Name: inCoris ZI Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: II Product Code: EIH Dated: March 5, 2013 Received: March 7, 2013
Dear Mr. Kolle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Kolle
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kwame O. Ulmer -S for
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and
Dental Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K 12 3664
Device Name: ____ inCoris ZI
Indications for Use:
Classic sintering
- Framework and reduced crowns in the anterior and posterior tooth region .
- Bridge frameworks in the anterior and posterior tooth region with max. 2 pontics .
- Crown caps in the anterior and posterior region �
- Cone and telescoping crowns ●
Speed sintering
- Framework and reduced crowns in the anterior and posterior tooth region ●
- . Bridge frameworks in the anterior and posterior tooth region with max. 2 pontics and up to 9 units
- Crown caps in the anterior and posterior region .
- . Cone and telescoping crowns
Super speed sintering
- Framework and reduced crowns with a maximum wall-thickness of 2 mm .
Prescription Use X (Part 21 CFR 801 Subpart D) OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Sirona Dental Systems 510(k) | March 05, 2013 |
|------------------------------|----------------|
| inCoris ZI | |
| | Mary S. Runner -S |
|--|----------------------------------------------|
| | Susan Runner, DDS, MA |
| | 2013.04.04 10:34:20 |
| | (Division Sign-Off) |
| | Division of Anesthesiology, General Hospital |
| | Infection Control, Dental Devices |
| 510(k) Number: | K123664 |
|----------------|---------|
|----------------|---------|
510(k) Number:_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.