SEAL BLOCK PRO (NOT FINALIZED), SEAL BLOCK OTC (NOT FINALIZED)
K123653 · Bisco, Inc. · LBH · Jun 3, 2013 · Dental
Device Facts
Record ID
K123653
Device Name
SEAL BLOCK PRO (NOT FINALIZED), SEAL BLOCK OTC (NOT FINALIZED)
Applicant
Bisco, Inc.
Product Code
LBH · Dental
Decision Date
Jun 3, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3260
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The indication for use of Seal Block PRO is to relieve sensitive teeth. Seal Block PRO is used to relieve sensitive teeth by blocking the pain caused by: - Hot . - . Cold - Sweet . - . Acidic - . Dental whitening agents Seal Block PRO can also be used to relieve sensitivity: - Prior to temporization (placement of provisional restorations). . - Prior to permanent cementation of indirect restorations. ◆ - Prior to placement of direct restorations. . - When root surfaces are exposed. .
Device Story
Seal Block is a dental desensitizer applied via applicator to sensitive teeth. It functions by forming calcium oxalate crystals that occlude dentinal tubules, thereby reducing dentin permeability and desensitizing natural dentition. Used in both professional (Pro) and over-the-counter (OTC) contexts. The device provides relief from sensitivity associated with thermal, chemical, or mechanical stimuli, including dental procedures like restorations or whitening. It benefits patients by mitigating pain from exposed dentin or pulp sensitivity.
Clinical Evidence
Bench testing only. Evidence includes laboratory R&D protocols evaluating dentin permeability reduction, shear bond strength, and pH. Biocompatibility was assessed via an ADA Oral Toxicity Study (10 rats, 14 days), concluding the device is safe for its intended use.
Technological Characteristics
Water-based oxalate solution. Low viscosity liquid. Applied via applicator. Industry-standard chemistry consistent with predicate devices. No electronic components or software.
Indications for Use
Indicated for patients experiencing tooth sensitivity caused by heat, cold, sweets, acidic foods/drinks, or dental whitening agents; also indicated for use prior to temporization, permanent cementation of indirect restorations, direct restorations, or when root surfaces are exposed.
Regulatory Classification
Identification
Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.
Special Controls
*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
Super Seal Tooth Desensitizer by Phoenix Dental, Inc. (K120109)
Submission Summary (Full Text)
{0}------------------------------------------------
K123653
Image /page/0/Picture/1 description: The image shows the word "BISCO" in a bold, sans-serif font. Above the word, there are four geometric shapes that appear to be connected. The shapes are arranged in a row, and they have a three-dimensional appearance. The image is black and white, and the text and shapes are clearly visible.
JUN 0 3 2013
## 510 (k) SUMMARY
Applicant:
Contact Person:
Date Prepared: Trade Name: Common Name: Product Code: Classification/Name:
Bisco, Inc. 1100 W. Irving Park Road Schaumburg IL, 60193 Michelle Schiltz-Taing Tel: 847-534-6146 Fax: 847-534-6146 24 April 2013 Seal Block Tooth Desensitizer LBH Varnish, Cavity Class Il per 21 CFR 872.3260
#### Predicate Devices:
Seal Block is substantially equivalent to:
BisBlock by Bisco, Inc. Schaumburg IL K033521 Super Seal by Phoenix Dental, Inc. Fenton, MI K983477 Super Seal Tooth Desensitizer by Phoenix Dental, Inc. Fenton, MI K120109
#### Indications for Use Seal Block PRO Version:
The indication for use of Seal Block PRO is to relieve sensitive teeth.
Seal Block PRO is used to relieve sensitive teeth by blocking the pain caused by:
- . Hot
- . Cold
- Sweet .
- . Acidic
- . Dental whitening agents
Seal Block PRO can also be used to relieve sensitivity:
- Prior to temporization (placement of provisional restorations). .
- Prior to permanent cementation of indirect restorations. ◆
- Prior to placement of direct restorations. .
- When root surfaces are exposed. .
BISCO, Inc. 1100 W. Irving Park Road Schaumburg, IL 60193 U.S.A. 800-247-3368 or 847-534-60C Fax: 847-891-5049 www.bisco.com
.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "BISCO" in a stylized font, with the top half of the letters outlined and the bottom half filled in with black. Above the word, there is a row of four geometric shapes that resemble cubes or hexagons, arranged in a slightly staggered pattern. The shapes are also black and white, with some having vertical lines.
## 510 (k) SUMMARY (continued)
Indications for Use Seal Block OTC Version:
Seal Block relieves sensitive teeth by blocking the pain caused by: heat, cold, sweets, acidic foods and drinks, dental whitening agents.
Description of Applicant Device:
Seal Block is applied using an applicator to the sensitive tooth/teeth. It forms calcium oxalate crystals that occlude dentinal tubules resulting in desensitization of natural dentition.
Technological Characteristics:
All components of Seal Block are based upon industry standard chemistry and are found in the legally marketed predicate devices Super Seal (K983477 and K120109) and BisBlock (K033521). Comparisons of the chemical composition of Seal Block to the predicates are provided in the following table:
| Chemical<br>Composition | Super Seal<br>K983477 | Super Seal Tooth<br>Desensitizer<br>K120109 | BisBlock<br>K033521 | Seal Block<br>Pro and OTC |
|-------------------------|-----------------------|---------------------------------------------|---------------------|---------------------------|
| Oxalate Solution | X | X | X | X |
| Water Based | X | X | X | X |
Performance Data:
The physical/mechanical properties of Seal Block were tested in the lab using R&D testing protocols to determine dentin permeability reduction, shear bond strength, and pH. The information provided in this 510(k) for Seal Block compared to the predicates demonstrates that it is effective for its indications of use. A comparison of the physical/mechanical properties are included below:
| Physical / Mechanical<br>Property Comparison | Super Seal<br>K983477 | Super Seal<br>Tooth<br>Desensitizer<br>K120109 | BisBlock<br>K033521 | Seal Block |
|----------------------------------------------|-----------------------|------------------------------------------------|---------------------|------------|
| Low viscosity | X | X | X | X |
| SEM Comparison | X | | X | X |
BISCO, Inc. 1100 W. Irving Park Road - -Schaumburg, IL 60193 U.S.A. 800-247-3368 or 847-534-600 Fax: 847-891-5049 www.bisco.com
An ISO 13485 Certified Company
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "BISCO" in a stylized font, with a geometric pattern above it. The letters of the word are partially filled in with black, creating a contrast with the white outlines. The geometric pattern consists of interconnected shapes, some of which are filled with black while others are white, creating a visually interesting design element above the word.
## 510 (k) SUMMARY (continued)
Biocompatibility:
An evaluation of biocompatibility was conducted to determine the safety of Seal Block. It is concluded from the safety evaluation and the results of the ADA Oral Toxicity Study (10 rats, 14 days) that Seal Block is safe for its intended uses.
Conclusion:
Side by side comparisons clearly demonstrate that the applicant device is substantially equivalent to the other legally marketed devices. It is concluded that the information supplied in this submission has proven the safety and efficacy of these products.
> BISCO, Inc. 1100 W. Irving Park Road Schaumburg, IL 60193 U.S.A.
800-247-3368 or 847-534-600 Fax: 847-891-5049 www.bisco.com
An ISO 13485 Certified Company
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the bird symbol. The logo is black and white.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 3, 2013
Ms. Michelle Schiltz-Taing Bisco, Incorporated 1100 West Irving Park Road SCHAUMBURG IL 60193
Re: K123653
Trade/Device Name: Seal Block OTC, Seal Block Pro Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity Varnish Regulatory Class: II Product Code: LBH Dated: April 24, 2013 Received: May 9, 2013
Dear Ms. Schiltz-Taing:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices.that.have.been.reclassified.in.accordance.with.the.provisions.of.the.Federal. F.ood., Drug. and Cosmetic Act (Act (Act) that do not require approval of a premarket approval application (PMA): You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Suzan Ruaner, DDS, MA
Digitally signed by Mary S. Runner -5
DN: c=US, o=U.S. Government; ou=HHS,
ou=FDA, ou=People, cn=Mary S. Runner
0.9.2342.19200300.100.1.1=1300087950
Date: 2013.06.03:14:24:10 -04'00'
Kwame Ulmer, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510 (k) Number (if known): _ K123653
Device Name: _ Seal Block Pro
Indications for Use:
The indication for use of Seal Block PRO is to relieve sensitive teeth.
Seal Block PRO is used to relieve sensitive teeth by blocking the pain caused by:
- Hot .
- . Cold
- Sweet .
- Acidic .
- Dental whitening agents .
Seal Block PRO can also be used to relieve sensitivity:
- Prior to temporization (placement of provisional restorations). .
- Prior to permanent cementation of indirect restorations. .
- Prior to placement of direct restorations. ●
- When root surfaces are exposed. ●
AND/OR Prescription Use P (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Digitally signed by Mary S. Runner -S
DN: C=US, o=US. Government,
ou-HHS, ou FDA, ou=People,
Susan Runner, DDS,MA cn=Mary S. Runner -S,
0.9.2342.19200300.100.1.1=1300087
950
Date: 2013.06.03 14:24:48-04'00'
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
**510(k) Number:** K123653
{6}------------------------------------------------
510 (k) Number (if known): _ Kl23653
K125653
Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications for Use:
Seal Block relieves sensitive teeth by blocking the pain caused by: heat, cold, sweets, acidic foods and drinks, dental whitening agents
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
510(k) Number:
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Runner, DDSBIA
(Division Sign-Off)
Digitally signed by Mary S. Runner -S
DN: c=US, o=U.S. Government,
ou=HHS, ou=FDA, ou=People, cn=Mary
S. Runner -S
0.9.2342.19200300.100.1.1=130008795
Date: 2013.06.03 14:25:11 -04'00'
Division of Anestheslology, General Hospital Infection Control, Dental Devices
KI23653
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.