K123633 · Tissue Regeneration Systems, Inc. · GXR · Aug 16, 2013 · Neurology
Device Facts
Record ID
K123633
Device Name
TRS CRANIAL BONE VOID FILLER(TRS C-BVF)
Applicant
Tissue Regeneration Systems, Inc.
Product Code
GXR · Neurology
Decision Date
Aug 16, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TRS Cranial Bone Void Filler is intended for use in the repair of 13mm neurosurgical cranial burr holes. It should be gently packed into bony voids or gaps of the skeletal system that are not intrinsic to the stability of the bony structure.
Device Story
TRS C-BVF is a synthetic, porous, osteoconductive bone void filler; composed of polycaprolactone (PCL) matrix coated with hydroxylapatite and octacalcium phosphate. Manufactured via laser sintering; provided sterile. Designed as a 13mm diameter plug with a 20mm flange for burr hole repair. Used by neurosurgeons in clinical settings to fill cranial defects. Device acts as an osteoconductive scaffold; facilitates bone ingrowth; degrades in vivo via hydrolysis. Benefits patient by providing a resorbable matrix for natural bone regeneration. Healthcare providers place the device manually into the defect; output is the physical restoration of the bony void.
Clinical Evidence
No clinical studies performed. Evidence based on non-clinical bench testing (compressive mechanical properties, push-out force, FTIR material analysis) and animal testing (rabbit calvarial defect model). Animal study showed comparable new bone formation to predicate (chronOS) at 26 and 78 weeks. Biocompatibility testing (ISO 10993) confirmed non-cytotoxic, non-sensitizing, non-irritant, non-pyrogenic, and non-mutagenic status. Degradation testing confirmed PCL hydrolysis mechanism consistent with predicate devices.
Technological Characteristics
Synthetic, porous, osteoconductive scaffold. Materials: Polycaprolactone (PCL) matrix, hydroxylapatite, and octacalcium phosphate coating. Form factor: 13mm diameter plug with 20mm flange. Manufacturing: Laser sintering. Sterilization: Ethylene Oxide. Biocompatibility: Meets ISO 10993. Degradation: Hydrolysis.
Indications for Use
Indicated for repair of 13mm neurosurgical cranial burr holes and filling bony voids or gaps in the skeletal system not intrinsic to structural stability.
Regulatory Classification
Identification
A burr hole cover is a plastic or metal device used to cover or plug holes drilled into the skull during surgery and to reattach cranial bone removed during surgery.
CranioClamp - Bioabsorbable Cranial Bone Flap Fixation System (K071138)
Submission Summary (Full Text)
{0}------------------------------------------------
K123,633
Tissue Regeneration Systems, Inc. 510(k) Premarket Notification TRS CRANIAL BONE VOID FILLER (TRS C-BVF) August 14, 2013
# 510(k) Summary Tissue Regeneration Systems, Inc TRS Cranial Bone Void Filler Traditional 510(k)
### 1.0 Manufacturer Name
Tissue Regeneration Systems, Inc. 5400 Carillon Point Kirkland, Washington 98033
**AUG 16 2013**
### 2.0 Official Contact
Jim Fitzsimmons Chief Executive Officer
Phone: 425-576-4032 Fax: 425-576-4040 E-mail: Jim@tissuesys.com
# Alternate Contact
John Garinger Director, Regulatory Affairs and Quality Assurance Phone: 206-730-2815 Fax: 425-576-4040 E-mail: John@tissuesys.com ·
### 3.0 Date Prepared: November 16, 2012
### 4.0 Device Name and Classification
Proprietary Name: Common/Usual Name: Classification Name: Requlation Number: Device Class: Classification Name: Classification Panel:
TRS PCL Cranial Bone Void Filler Bone void filler Burr Hole Cover 8882.5250 Class II GXR
### 5.0 Indications for Use
TRS Cranial Bone Void Filler is intended for use in the repair of 13mm neurosurgical cranial burr holes. It should be gently packed into bony voids or gaps of the skeletal system that are not intrinsic to the stability of the bony structure.
Neurology
### 6.0 Device Description
{1}------------------------------------------------
## Tissue Regeneration Systems. Inc. 510(k) Premarket Notification TRS CRANIAL BONE VOID FILLER (TRS C-BVF) August 14, 2013
TRS Cranial Bone Void Filler (TRS C-BVF) is a synthetic, porous, osteoconductive, bone void filler made from PCL polycaprolactone (CaH100-)X which will degrade and resorb fully in vivo by hydrolysis and is subsequently metabolized by the body, and hydroxylapatite (Ca10(PO4)6.(OH)2) with a calcium phosphate bone mineral coating (Hydroxylapatite and Octacalcium phosphate). TRS BVF has an interconnected porous structure that acts as an osteoconductive matrix for the ingrowth of bone.
TRS C-BVF is available in single size, which is a 13mm diameter x 5mm "plug" with a 20mm diameter x .90mm thick flange.
TRS C-BVF is manufactured using a laser sintering process and is then coated with the calcium phosphate bone mineral coating. The product is shipped to a contract manufacturer who packages, labels and sterilizes the C-BVF devices. They are then returned to TRS and inventoried as Finished Goods.
### 7.0 Predicate Devices
Osteopore PCL Scaffold Bone Void Filler (BVF), (K051093, product code GXP cleared on March 17, 2006) Note: This device would be cleared under product code GXR if it were reviewed using today's more product specific product codes.
Synthes chronOS Composite resorbable bone void filler (K071046, product code MQV, cleared on October 23, 2007).
Synthes chronOS, porous, osteoconductive, resorbable bone void filler (K041350, product code GXP, cleared on July 8, 2004).
Synthes Rapid Resorbable Cranial Clamp burr hole cover (K041611, product code GXR, cleared on September 8, 2004.
### 8.0 Comparison to Marketed Devices
Tissue Regeneration Systems, Inc. purports the information contained in this 510(k) Submission demonstrates that the TRS Cranial Bone Void Filler is substantially equivalent to Osteopore PCL Scaffold Bone Void Filler cleared under K051093. Synthes chronOS Composite cleared under K071046, Synthes chronOS, cleared under K041350 and Synthes Rapid Resorbable Cranial Clamp cleared under K041611.
6 - 3
{2}------------------------------------------------
510(k) Premarket Notification
CRANIAL BONE VOID FILLER (TRS C-BV
August 14, 20
# Table 6-1 Predicate Device Regulatory Comparisor
| Information | Subject Device | Predicate Device | Predicate Device | Predicate Device | Predicate Device |
|----------------------|-----------------------------------|------------------------------------------|-------------------------------------|-----------------------------------------|---------------------------------------------------------------|
| Manufacturer | Tissue Regeneration Systems, Inc. | Synthes | Synthes chronOS | Osteopore | Synthes |
| Trade Name | TRS C-BVF | Synthes chronOS Composite | Synthes chronOS | Osteopore PCL Scaffold Bone Void Filler | CranioClamp - Bioabsorbable Cranial Bone Flap Fixation System |
| 510(k) # | To Be Determined | K071046 | K041350 | K051093 | K071138 |
| Product Code | GXR | MQV | GXP | GXP | GXR |
| Regulation Number | 882.55250 | 888.3045 | 882.5300 | 882.5300 | 882.5250 |
| Classification Name | Burr Hole Cover | Resorbable Calcium Salt Bone Void Filler | Methylmethacrylate for Cranioplasty | Methylmethacrylate for Cranioplasty | Burr Hole Cover |
| Device Class | Class II | Class II | Class II | Class II | Class II |
| Classification Panel | Neurology | Orthopedic and Rehabilitation Devices | Neurology | Neurology | Neurology |
{3}------------------------------------------------
| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>1 | |
| | |
| and in the warm and we were was new<br>0<br>1 | |
| 1<br>1<br>1 | |
| Table 6-2: Device Characteristic Comparison | | | | | | | |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|---------|-----------|
| Characteristic | TRS C-BVF<br>(subject device) | Synthes chronOS<br>Composite<br>(predicate/reference) | Synthes chronOS<br>(predicate device) | Osteopore PCL<br>Scaffold<br>(predicate device) | Synthes<br>CranioClamp -<br>Bioabsorbable<br>Cranial Bone<br>Flap Fixation<br>System (predicate<br>device) | Similar | Different |
| Materials | PCL (poly-ε-<br>caprolactone),<br>Hydroylapatite,<br>Octacalcium<br>Phosphate<br>Bone contacting<br>surface:<br>Hydroxyapatite and<br>Calcium Phosphate<br>PCL Matrix<br>Ceramic Particles | PCL (polylactide-co-ε-<br>caprolactone) matrix<br>with imbedded β-<br>Tricalcuim Phosphate<br>granules.<br>Bone contacting<br>surface:<br>Calcium Phosphate<br>PCL Matrix<br>Ceramic Particles | β-Tricalcuim<br>Phosphate<br>granules.<br>Bone contacting<br>surface:<br>Calcium Phosphate | PCL (poly-ε-<br>caprolactone)<br>Bone contacting<br>surface: PCL (poly-<br>ε-caprolactone)<br>PCL Matrix<br>No Ceramic<br>Particles | 85:15 Poly (L-lactide-<br>co-glycolide)<br>Bone contacting<br>surface:85:15 Poly<br>(L-lactide-co-<br>glycolide)<br>No PCL<br>No Ceramic Particles | X | |
| | Not intended to be<br>trimmed | Can be trimmed with a<br>scalpel or molded | Can be trimmed<br>with a scalpel or<br>molded | Can be trimmed<br>with a scalpel | Not intended to be<br>trimmed | X | |
| Biocompatible | Resorbable<br>materials<br>Meets ISO 10993<br>Interconnected<br>macroporous<br>structure<br>PCL Matrix<br>Ceramic Particles | Resorbable materials<br>Meets ISO 10993<br>Interconnected<br>macroporous structure<br>PCL Matrix<br>Ceramic Particles | Resorbable<br>materials<br>Meets ISO 10993<br>Interconnected<br>macroporous<br>structure<br>No PCL<br>Ceramic Particles | Resorbable<br>materials<br>Meets ISO 10993<br>Interconnected<br>macroporous<br>structure<br>PCL Matrix<br>No Ceramic<br>Particles | Resorbable materials<br>Meets ISO 10993<br>Smooth surface<br>No PCL<br>No Ceramic Particles | X | |
| Design | Not intended to be<br>trimmed<br>Cylindrical plug with<br>flange | Can be trimmed with a<br>scalpel or molded<br>Various forms/sizes | Can be trimmed<br>with a scalpel or<br>molded<br>Various forms/sizes | Can be trimmed<br>with a scalpel<br>Various forms/sizes | Not intended to be<br>trimmed<br>18 mm diameter disc | X | |
| Characteristic | TRS CBVF<br>(subject device) | Synthes chronOS Composite<br>(predicate/reference) | chronOS<br>(predicate device) | Osteopore PCL Scaffold<br>(predicate device) | CranioClamp - Bioabsorbable<br>Cranial Bone<br>Flap Fixation<br>System | Similar | Different |
| Intended Use | Is intended to be gently packed or placed into bony voids or gaps of the skeletal system. | Is intended to be gently packed or placed into site. | Is intended to be gently packed or placed into site. | Is intended to be gently packed or placed into bony voids or gaps of the skeletal system. | Is intended for covering burr holes and for fixation of cranial bone flaps, in pediatric and adult patients. | X | |
| Indications for Use | Is indicated for use in the repair of neurosurgical cranial burr holes. | Is indicated for bony voids or gaps that are not intrinsic to the stability of the bony structure. Synthes chronOS Composite is indicated for use in the treatment of bony defects created surgically or through traumatic injury. | Is intended for the repair or filling of craniofacial defects and craniotomy cuts with a surface area no larger than 25 cm². It is also indicated for the restoration or augmentation of bony contours of the craniofacial skeleton; including the fronto-orbital, malar and mental areas. | Is indicated for use in the repair of neurosurgical burr holes, craniotomy cuts and other cranial defects. It is also for use in the augmentation or restoration of bony contour in the craniofacial skeleton. | Is intended for covering burr holes and for fixation of cranial bone flaps, in pediatric and adult patients. | | X |
| Function | Fill bony voids or gaps of the skeletal system until the device is replaced by bone during the healing process. | Fill bony voids or gaps of the skeletal system until the device is replaced by bone during the healing process. | Fill bony voids or gaps of the skeletal system until the device is replaced by bone during the healing process. | Fill bony voids or gaps of the skeletal system until the device is replaced by bone during the healing process. | Covering burr holes and for fixation of cranial bone flaps | X | |
| Sterilization | Sterile Ethylene Oxide | Sterile Irradiation | Sterile Irradiation | Sterile Irradiation | Sterile | | X |
{4}------------------------------------------------
6 - 6
{5}------------------------------------------------
# Tissue Regeneration Systems, Inc. 510(k) Premarket Notification TRS CRANIAL BONE VOID FILLER (TRS C-BVF) August 14, 2013
Comparison to Marketed Devices continued: The subject device and the predicate devices have the same intended use, are of similar design, perform the same function and are composed of similar resorbable materials. The subject and predicate devices are osteoconductive and both provide an interconnected, porous scaffold and an environment for new bone ingrowth. Both devices are available in similar forms and are provided sterile. Both the subject and predicate devices are shown to be biocompatible, perform similarly in in-vitro and animal testing and are composed of component materials with a history of use in implantable medical devices.
### 9.0 Performance Testing
TRS Cranial Bone Void Filler has undergone a comprehensive battery of non-clinical testing, including chemical, physical, animal and biocompatibility. Testing has provided reasonable assurance of safety and effectiveness for its intended use and supports a determination of substantial equivalence.
### 9.1 Biocompatibility, including Degradation Testing:
Comprehensive biocompatibility testing of the TRS C-BVF material demonstrates that it is biocompatible; non-genotoxic, nonpyrogenic, non-toxic and a non-irritant. See Table 6-3 for the specific tests performed including the results and conclusions.
Results of the degradation testing, which included accelerated and real-time polymer degradation, as well as extreme simulation ceramic degradation, have shown comparable performance between the TRS and chronOS predicate device. Both devices demonstrated a similar PCL degradation mechanism by hydrolysis of the ester linkage to give a carboxylic acid and an alcohol as by products. In addition, similar trace elements were detected for both devices. Molecular ions detected from the extracts from both devices at 90-day real-time test demonstrated similar retention times, and indicate the degraded compounds are similar or related. These results provide further evidence that the TRS C-BVF is substantially equivalent to the chronOS predicate device.
While degradation testing was not performed on the OsteoPore PCL Scaffold predicate device, an infrared analysis of both the TRS C-BVF device and the OsteoPore PCL Scaffold device was undertaken using Fourier Transform Infrared Spectroscopy (FTIR) to qualitatively establish similarity of PLC polymeric material components between these devices. FTIR results for both TRS device and Osteopore PCL Scaffold showed strong peaks
{6}------------------------------------------------
# Tissue Regeneration Systems, Inc. 510(k) Premarket Notification TRS CRANIAL BONE VOID FILLER (TRS C-BVF) August 14, 2013
consistent with traditional peaks found in the infrared spectra of PCL indicated that PCL is a major component for these devices. (See Section 11.4 of this 510(k) submission and Appendix B, Tab B-5 for a summary of this testing.) Given that both the TRS C-BVF device and the OsteoPore PCL Scaffold predicate are composed of Poly-E-caprolactone material, degradation results for the OsteoPore material are presumed substantially similar to the TRS C-BVF device.
In conclusion, the TRS C-BVF device is biocompatible and demonstrates comparable degradation performance to the predicate devices, supporting a claim of substantial equivalence to the predicate chronOS and OsteoPore PCL Scaffold devices.
| ISO<br>10993<br>Standard | Test | Results | Conclusions |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 10993-5 | Cytotoxicity (ISO Elution<br>Method- 1X MEM<br>Extract) | No cytoxicity or cell lysis was noted in any of the test wells. No<br>pH shift was observed at 48 hours. Reactivity grade was 0<br>(none). | Non-cytotoxic |
| 10993-10 | Sensitization (Guinea<br>pig maximization<br>sensitization test) | All Animals were clinically normal throughout the study. Test<br>article extracts showed no evidence of causing delayed dermal<br>contact sensitization in guinea pig. | Non-sensitizer |
| 10993-10 | Intracutaneous<br>Reactivity | There was no erythema and no edema from the 0.9% sodium<br>chloride solution test extract. There was very slight erythema<br>and very slight edema from the sesame oil test extract.<br>However, the difference from the control was 1.0 or less. | Non-irritant |
| 10993-11 | Systemic Toxicity (Acute<br>systemic toxicity in mice) | There was no mortality or evidence of systemic toxicity from<br>the test extracts. Body weight data were acceptable. | No acute systemic<br>toxicity |
| 10993-3 | Genotoxicity (Gene<br>mutation): Bacterial<br>Reverse Mutation study | The DMSO and saline extracts from TRS device were<br>considered to be non-mutagenic to Salmonella typhimurium<br>tester strains TA98, TA100, TS1535, and TA1537, and to<br>Escherichia coli tester strain WP2uvrA. | Non-mutagenic |
| 10993-3 | Genotoxicity (in-vivo;<br>Mouse Peripheral Blood<br>Micronucleus Study) | The saline and sesame oil test extracts did not induce<br>micronuclei in mice. There were no statistically significant<br>differences between the test and negative control groups. | Non-clastogenic |
| 10993-3 | Genotoxicity (Mouse<br>Lymphoma Assay) | The undiluted RPMI₀ and 1.0% DMSO TRS device extracts did<br>not cause any positive increase in the mean mutant frequency<br>in the L5178Y/TK⁺/⁻ cell line either in the presence or absence<br>of metabolic activation. | Non-mutagenic |
| 10993-6 | Local effects after<br>Implantation (Muscle<br>implantation study in<br>rabbits-2 weeks) | The macroscopic reaction of TRS device was not significant as<br>compared to the ChronOS and negative control (HDPE).<br>Microscopically, TRS device was classified as a non-irritant as<br>compared to ChronOS and HDPE. | Non-irritant as<br>compared to the<br>negative control<br>(HDPE) and<br>predicate device<br>(ChronOS) |
| ISO<br>10993 | Test Description | Results | August 14, 2013<br>Conclusions |
| 10993-6 | Local effects after<br>Implantation (Muscle<br>implantation study in<br>rabbits-6 weeks) | The macroscopic reaction of TRS device was not significant as<br>compared to the ChronOS and negative control (HDPE).<br>Microscopically, TRS device was classified as a non-irritant as<br>compared to ChronOS and a slight irritant when compared to<br>HDPE. | Non-irritant as<br>compared to the<br>predicate device<br>(ChronOS) and a<br>slight irritant when<br>compared to the<br>negative control<br>(HDPE) |
| 10993-11 | Systemic toxicity (USP<br>Pyrogen study) | One (out of 3) of the animals showed a rise of 0.5°Cabove its<br>baseline temperature. The result was inconclusive and the<br>protocol required 5 additional animals to be injected. | Inconclusive - retest<br>required |
| 10993-11 | Systemic toxicity (USP<br>Pyrogen study-retest) | A total of 3 out of 8 rabbits had a temperature rise of ≥ 0.5°C,<br>and the total temperature rise of the 8 animals did not exceed<br>3.3°C. The USP test requirement was met. TRS device was<br>considered non-pyrogenic. | Non-pyrogenic |
| 10993-11 | Systemic toxicity: Sub-<br>chronic toxicity (13 week<br>study in rats following<br>subcutaneous<br>implantation) | No evidence of systemic toxicity from TRS device following<br>subcutaneous implantation in the rat. Daily clinical<br>observations, body weights, necropsy findings, organ weights<br>and organ/body weight ratios were within acceptable limits and<br>were similar between TRS device and control (HDPE)<br>treatment group. There were no changes in histopathology,<br>hematology values or clinical chemistry values in either male<br>or female rats. Microscopic evaluation of the selected tissues<br>revealed no evidence of a treatment related response.<br>Microscopic evaluation of the implant sites revealed that the<br>test implant site scores were higher than control implant site<br>scores. | No evidence of<br>systemic toxicity,<br>local macroscopic<br>tissue reaction not<br>significant as<br>compared to HDPE<br>control,<br>microscopically<br>classified as slight<br>irritant as compared<br>to HDPE control |
| 10993-11 | Systemic toxicity:<br>Chronic toxicity (26<br>week study in rats<br>following subcutaneous<br>implantation) | No evidence of systemic toxicity from TRS device following<br>subcutaneous implantation in the rat. Daily clinical<br>observations, body weights, necropsy findings, organ weights<br>and organ/body weight ratios were within acceptable limits and<br>were similar between TRS device and control (HDPE)<br>treatment group. There were no changes in histopathology,<br>hematology values or clinical chemistry values in either male<br>or female rats that were considered to be biologically<br>significant or related to treatment with TRS device. Microscopic<br>evaluation of the selected tissues revealed no evidence of a<br>treatment related response. Microscopic evaluation of the<br>implant sites revealed that the test implant site scores were<br>higher than control implant site scores | No evidence of<br>systemic toxicity,<br>local macroscopic<br>tissue reaction not<br>significant as<br>compared to HDPE<br>control,<br>microscopically<br>classified as<br>moderate irritant as<br>compared to HDPE<br>control.<br>The degrading<br>sample elicited an<br>expected mild<br>macrophage/giant<br>cell response. |
| 10993-4 | ASTM F756 Hemolysis<br>Test - Direct contact<br>and extract test | The hemolytic index of the test article in direct contact with<br>blood was 0.4% and the hemolytic index for the test article<br>extract was 0.1%. The test article in direct contact with blood<br>and the test article extract were both non-hemolytic. | Non-hemolytic |
| ASTM<br>F2382 | Partial Thromboplastin<br>Time (PTT) | Average clotting time of the test article was 77% of the<br>negative control | Passed as a minimal<br>activator of the<br>intrinsic coagulation<br>pathway<br>(ASTM F2382<br>defines the test<br>result of > 50% of the<br>negative control as a<br>passing result) |
| ISO<br>10993 | Test Description | Results | Conclusions |
| 10993-4 | C3a Complement<br>Activation Assay | C3a concentration of the test article was statistically<br>significantly higher than the activated NHS (normal human<br>serum) control and negative control (low density polyethylene). | Complement<br>activation by a<br>device is primarily a<br>surface related<br>phenomenon. As<br>TRS is not a direct<br>blood-contacting<br>device, there is no<br>direct interaction of<br>the blood with the<br>surface of the<br>device. |
| 10993-4 | SC5b-9 Complement<br>Activation Assay | SC5b-9 concentration of the test article was statistically<br>significantly higher than the activated NHS (normal human<br>serum) control and negative control (low density polyethylene). | Complement<br>activation by a<br>device is primarily a<br>surface related<br>phenomenon. As<br>TRS is not a direct<br>blood-contacting<br>device, there is no<br>direct interaction of<br>the blood with the<br>surface of the<br>device. |
| 10993-13 | Identification and<br>quantification of<br>degradation products<br>from polymeric devices<br>(Real time polymer<br>degradation<br>Polymer:<br>30,90,180,365,540, and<br>730 days) | After incubation in phosphate buffered saline (PBS) at 37 °C,<br>TRS C-BVF lost 10% and 23% of its molecular weight and<br>0.26% and 0.47% of its initial mass at 365 and 730 days,<br>respectively. Extracts of TRS device revealed one compound<br>(2-hexenoic acid, butyl ester, (E)-) at 30, 90, 180, and 365 day<br>at a concentration in the range of 2-6 ppm. ChronOS<br>composite resulted in one compound (di-n-octyl phthalate)<br>detected in the 30 day replicate 2 extract. The average<br>molecular weight (Mw) shows that TRS devices were<br>unchanged after 30 days, but did exhibit degradation by 90<br>days with approximately 5% reduction in average molecular<br>weight. Mw at 365 days showed a decrease of approximately<br>5%with a similar trend for the later time points. The ChronOS<br>composite, on the other hand, exhibited a 57% reduction in<br>average molecular weight by 30 days and a considerable 90%<br>reduction by 90 days, 95% at 180 days, 98% by 365 days, and<br>was completely degraded prior to the 540 day time point. The<br>decrease in pH of the solution of ChronOS composite, may<br>have affected the degradation profile. The number average<br>molecular weight (Mn) showed that the Mn of TRS device<br>decreased approximately 10% over 365 days and further<br>decreased to 23% after 730 days. No other compounds were<br>detected at any other time point above the quantitation limit. | TRS device real time<br>degradation results<br>demonstrated that<br>the device is in the<br>first stage of<br>degradation with the<br>decrease in<br>molecular weight<br>without mass loss<br>and deformation.<br>(Additional detail to<br>following in<br>Executive Summary) |
| 10993-14 | Identification and<br>quantification of<br>degradation products<br>from ceramics<br>(extreme/simulation) | The extreme solution test caused approximately 26-27% of the<br>ceramic and hydroxylapatite to dissolve from both the coated<br>and uncoated discs. After the testing had been performed, the<br>buffer solutions showed the presence of calcium, magnesium,<br>phosphorus, sulfur, and in some cases aluminum, barium, and<br>iron.<br>The simulation solution test caused a mass loss of less than<br>1% for both coated and uncoated discs. The solutions after<br>the simulation test contained calcium and phosphorus for both<br>sample types. In the case of the coated discs, magnesium<br>and sodium were also observed. The SEM images for the<br>coated samples after the simulation test still showed the flake-<br>like structure, but some erosion was apparent. After the<br>simulation test, no significant differences in the EDS data were<br>observed, as compared to the coated samples that had not<br>been subjected to the simulation test. | Both extreme and<br>simulation test<br>demonstrated that<br>the ceramic<br>component of TRS<br>device is composed<br>of calcium and<br>phosphate ceramic.<br>Quantitative analysis<br>of four individual<br>heavy metal<br>elements (As, Cd,<br>Pb, Hg) was within<br>acceptable limit as<br>stated in ISO 13779.<br>(Additional detail to<br>following in<br>Executive Summary) |
Table 6-3: Biocompatibility Tests, Results and Conclusions
{7}------------------------------------------------
# Tissue Regeneration Systems, Inc.
{8}------------------------------------------------
# Tissue Regeneration Systems, Inc. 10 - 10 - 10 - 2010 - 10 - 510 (K) Premarket Notification
TRS CRANIAL BONE VOID FILLER (TRS C-BVF)
TRS CRANIAL BONE VOID FILLER (TRS C-BVF)
{9}------------------------------------------------
### 9.2 Bench Testing:
Bench testing to evaluate the compressive mechanical properties and push out force of the TRS C-BVF, the chronOS and Ostepore devices showed comparable performance and demonstrated that the TRS C-BVF device to possess appropriate performance characteristics for its intended use. In addition, material testing using FTIR demonstrated the same PCL material composition in both the predicate Osteopore PCL Scaffold device and the TRS C-BVF device. These results provide further technical evidence supporting a claim of substantial equivalence between the subject and predicate devices.
### 9.3 Animal Testing:
Results of the Rabbit Calvarial Defect testing have demonstrated similar performance of the TRS C-BVF and predicate chronOS devices with respect to new bone formation. By 26 weeks, total bone formation in both the subject and predicate devices was approximately equal and by 78 weeks, CT showed bone volumes for both devices to be nearly identical. The data shows that TRS C-BVF facilitates a constant, sustained bony healing response over time. Results of this testing demonstrate that the TRS C-BVF performs in a similar manner to the predicate chronOS device in a cranial defect in an animal model. This testing provides further technical evidence that the TRS C-BVF device performs comparably to the predicate chronOS device and provides additional support to a claim of substantial equivalence to the predicate devices.
### 0.4 Clinical (Literature):
The clinical literature summarized utilizes PCL in a number of material forms and indications. Clinical studies reported good outcomes in applied applications, expected absorption rates and with no reported material-mediated complications.
No clinical studies have been performed in support of this 510(k) application due to the similarities of the TRS C-BVF to the predicate devices. Further, the constituent materials of the subject device have a long history of safe and successful clinical use in numerous implantable medical applications as identified above. The PCL containing materials and material combinations referenced in this literature are comparable to that utilized in the TRS C-BVF. As such, the reported successful clinical history provides evidence of PCL's safety profile in clinical use and is representative of the expected safety profile for TRS C-BVF.
{10}------------------------------------------------
# 10.0 Conclusion
Product characterization and testing on the TRS Cranial Bone Void Filler when compared to its predicate devices demonstrate that it is substantially equivalent to the Osteopore PCL Scaffold Bone Void Filler, Synthes chronOS, Synthes chronOS Composite and Synthes Rapid Resorbable Cranial Clamp devices, commercially available cranial burr hole covers.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
# August 16, 2013
Tissue Regeneration Systems, Inc. Mr. William J. Fitzsimmons President and Chief Executive Officer 5400 Carillon Point Kirkland, WA 98033
Re: K123633
Trade/Device Name: Tissue Regeneration Systems Cranial Bone Void Filler (TRS C-BVF) Regulation Number: 21 CFR 882.5250 Regulation Name: Burr Hole Cover Regulatory Class: Class II Product Code: GXR Dated: July 10, 2013 Received: July 18, 2013
Dear Mr. Fitzsimmons:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{12}------------------------------------------------
Page 2 – Mr. William J. Fitzsimmons
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Joyce M. Whang -S
for Victor Krauthamer, Ph.D. Acting Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{13}------------------------------------------------
# Indications for Use
510(k) Number (if known): K123633
Device Name: Tissue Regeneration Systems Cranial Bone Void Filler (TRS C-BVF))
Indications For Use:
TRS C-BVF is intended for use in the repair of 13 mm neurosurgical cranial burr holes. It should be gently packed into bony voids or gaps of the skeletal system that are not intrinsic to the stability of the bony structure.
V Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# lovce M. Whana -
Division Sign Off) Division of Neurological and Physical Medicine Devices (DNPMD)
510(k) Number K123633
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.