K123622 · Siemens Medical Solutions USA, Inc. · IYN · Dec 5, 2012 · Radiology
Device Facts
Record ID
K123622
Device Name
ACUSON
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
IYN · Radiology
Decision Date
Dec 5, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal, Abdominal Intraoperative, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
Device Story
The Acuson SC2000 is a mobile, software-controlled diagnostic ultrasound system. It acquires primary/secondary harmonic ultrasound echo data via various transducers (transthoracic/transesophageal/intracardiac catheters). Inputs include ultrasound signals and optional ECG for image synchronization. The system processes these to display B-Mode, M-Mode, Pulsed/Continuous/Color/Power Doppler, 3D, and 4D imaging. Operated by clinicians in clinical settings, the device provides anatomical measurements and calculation packages. Output is displayed on a flat-panel monitor to assist physicians in evaluating pathology and guiding intraoperative procedures. It benefits patients by providing non-invasive visualization of anatomy and physiology, supporting clinical diagnosis and procedural guidance.
Clinical Evidence
No clinical data was required for this submission. Substantial equivalence was established through bench testing, including safety and performance verification against recognized standards (e.g., IEC 60601-1, IEC 60601-2-37, ISO 10993-1, AIUM/NEMA UD-2/UD-3).
Technological Characteristics
Mobile ultrasound system; supports multiple transducers (phased array, TEE, intracardiac catheters). Imaging modes: B, M, PW/CW/Color/Power Doppler, 3D/4D, Harmonic. Connectivity: Wireless, DICOM (Print, Storage, Structured Reporting). Safety: EN60601-1, EN60601-2-37, ISO 10993-1. Software-controlled with workflow acceleration tools (eSie Measure, SpaceTime, TEQ).
Indications for Use
Indicated for adult, pediatric, neonatal, and fetal patients requiring ultrasound imaging or fluid flow analysis of cardiac, vascular, abdominal, neurological, and musculoskeletal structures. Contraindications: None stated.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Acuson SC2000 Diagnostic Ultrasound System (K113179)
{0}------------------------------------------------
# 510(k) Summary Prepared October 18, 2012
DEC 0 5 2012
- 1. Sponsor: Siemens Medical Solutions, Inc., Ultrasound Division 685 East Middlefield Road Mountain View, California 94043
- Contact Person: Patrick J. Lynch (650) 694-5658 Telephone: (650) 694-5580 Fax:
- Acuson SC2000™ Diagnostic Ultrasound System 2. Device Name:
- Common Name: Diagnostic Ultrasound System
# Classification:
Requlatory Class: ll Review Category: Tier II Classification Panel: Radiology
| Ultrasonic Pulsed Doppler Imaging System | FR # 892.1550 | Product Code 90-IYN |
|------------------------------------------|---------------|---------------------|
| Ultrasonic Pulsed Echo Imaging System | FR # 892.1560 | Product Code 90-IYO |
| Diagnostic Ultrasound Transducer | FR # 892.1570 | Product Code 90-ITX |
| Diagnostic Intravascular Catheter | FR # 870.1200 | Product Code OBJ |
# 3. Legally Marketed Predicate Devices
The Acuson SC2000™ Ultrasound System in this 510k is a modification to SC2000™ Diagnostic Ultrasound System previously cleared in K113179, the Acuson S2000 (K111674) and the Acuson X300 (K121699).
# 4. Device Description:
The SC2000™ Diagnostic Ultrasound System is a multi-purpose mobile, software controlled diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bio-effect mechanisms. Its function is to acquire primary or secondary harmonic ultrasound echo data and display it in B-Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, Color Doppler Mode, Amplitude Doppler Mode, a combination of modes, 3D Imaging, or Harmonic Imaging and 4D imaging on a Flat Panel Display.
The SC2000™ Ultrasound System has been optimized for user ergonomics with adjustable keyboard height and rotation and independently adjustable Flat Panel Display. There is an available off-line workstation (SC2000WP)
{1}------------------------------------------------
# 5. Intended Use
The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac. Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal, Intraoperative Abdominal, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are:
## Cardiac Imaging Applications and Analysis
The system transmits ultrasound energy into adult, pediatric, neonatal, and fetal cardiac creating 2D (B), 3D, M-Mode (M), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, creat vessels, and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.
The system also supports catheters which are intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients.
The system has Cardiac Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
### Vascular Imaging Applications and Analysis
The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the carotid arteries or juggler veins in the neck; superficial and deep veins and arteries in the arms and legs; and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.
The system has Vascular Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
#### Superficial Imaging Applications
The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of conventional or superficial musculoskeletal structures and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.
## Intraoperative Imaging Applications
The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), and Pulsed Wave Doppler (PWD) to obtain images and blood flow velocity that provide guidance during neurological intraoperative procedures.
#### Transcranial Imaging Applications
The system transmits ultrasound energy into the cranium of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the brain and surrounding anatomical structures to evaluate the presence or absence of pathology.
The system provides Measurement Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
{2}------------------------------------------------
# 6. Summary of Technological Characteristics -- New Device Compared to Predicate
The submission device is a modification to SC2000™ Diagnostic Ultrasound System previously cleared in K072365, K102017 and K113179 with regard to both intended use and technological characteristics.
| Description | Acuson<br>SC2000TM<br>K113179 | Acuson<br>S2000<br>K111674 | Acuson<br>X300<br>K121699 | Acuson<br>SC2000TM<br>This<br>submission |
|----------------------------------------------------------------------------------------------------------------------------------|-------------------------------|----------------------------|---------------------------|------------------------------------------|
| System: | | | | |
| Hardware Safety - EN60601-1 Certified | X | | | X |
| Software Safety - EN60601-4 Certified | X | | | X |
| Acoustic Thermal Safety - EN60601-2-37 Certified | X | | | X |
| Transducers: | | | | |
| See Table in Section 4.4.1.4 - All previously cleared | X | | | X |
| Biocompatibility - Same transducers no new materials | X | | | X |
| V7M Transducer | | X | X | X |
| 10V4 Transducer | | X | X | X |
| AcuNav 8F/10F | X | | | X |
| AcuNav V | X | | | X |
| Soundstar 10F | X | | | X |
| Biocompatibility - Same transducers no new materials | X | | | X |
| Accessories: | | | | |
| Respirometer | X | | | X |
| Biocompatibility | X | | | X |
| AcuNav Joystick | X | | | X |
| Imaging: | | | | |
| SpaceTimeTM resolution control | X | | | X |
| NativeTM Tissue Harmonic Imaging | X | | | X |
| MultiHertz multiple frequency imaging | X | | | X |
| RESTM enhanced resolution imaging format | X | | | X |
| Output display standard compliance | X | | | X |
| Native TEQTM dynamic ultrasound technology | X | | | X |
| TEQTM ultrasound technology | X | | | X |
| TEQTM technology for Spectral PW and CW Doppler | X | | | X |
| Dual screen and live dual imaging | X | | | X |
| Acoustic clip capture | X | | | X |
| eSie Measure Workflow Acceleration | X | | | X |
| Preset functionality with preset ordering and grouping<br>capabilities and linking to MultiHertzTM multiple frequency<br>imaging | X | | | X |
| Integrated Stress echo capability | X | | | X |
| Cardiac imaging and quantification package | | | X | |
| DTI™ Doppler tissue imaging includes the following color<br>Doppler capabilities: | X | | X | |
| DTI Velocity (DTV) | X | | X | |
| Color Doppler Harmonic capability in DTI | X | | X | |
| High Frame Rate Tissue Doppler (HTD) Capability | X | | X | |
| DTI Energy (DTE) | X | | X | |
| Color Doppler Velocity (CDV) | X | | X | |
| Color Harmonic imaging | X | | X | |
| DTI Pulsed Wave (DTI PW) capability | X | | X | |
| Color Doppler M-Mode for: | X | | X | |
| CDV | X | | X | |
| DTV | X | | X | |
| DTE | X | | X | |
| Contrast Imaging | | | | |
| Cadence™ contrast pulse sequencing technology (CPS) | X | | X | |
| PrecisonBurst programmable triggering for contrast agent<br>destruction | X | | X | |
| TEQ and NTEQ ultrasound technology for Cadence™ CPS | X | | X | |
| Cardiac Calculations: | | | | |
| Cardiac calculations package | X | | X | |
| Slope Line calculations | X | | X | |
| Generic Calculations/Ratio function | X | | X | |
| Vascular imaging package | | | | |
| B-mode, M-mode, NTHI, CPS, PW Doppler, Triplex imaging<br>and High Resolution Color Flow (HRCF) | X | | X | |
| Cadence™ CPS Capture | X | | X | |
| Advanced SieClear™ spatial compounding with Dynamic TCE<br>TM (tissue contrast enhancement) technology | X | | X | |
| Advanced Vascular Analysis report package | X | | X | |
| Clarify™ Vascular Enhancement Technology | X | | X | |
| Linear and Vector imaging formats | X | | X | |
| Auto Doppler option | X | | X | |
| Time averaged velocity (TAV) calculation | X | | X | |
| Vascular Calculation package | X | | X | |
| Transcranial imaging | X | | X | |
| Application packages: | | | | |
| syngo® Mitral Valve Assessment | X | | X | |
| syngo® auto Ejection Fraction technology | X | | X | |
| D1ART with syngo® fourSight™ TEE view | X | | X | |
| syngo® Quantitative Synch Tools™ technology (QST) | X | | X | |
| syngo® Velocity Vector Imaging™ technology (VVI) rotation | X | | X | |
| syngo® ACQ auto-tracking contrast quantification | x | | | x |
| Connectivity: | | | | |
| Wireless Network Connectivity | | | x | x |
| DICOM Print Service | x | | | x |
| DICOM Media Storage Service | x | | | x |
| DICOM Structured Reporting | x | | | x |
{3}------------------------------------------------
{4}------------------------------------------------
## 7. A brief discussion of nonclinical tests submitted, referenced, or relied on in the 510(k) for a determination of substantial equivalence
The SC2000™is designed, verified, and validated according to the company's design control process and has been subjected to extensive safety and performance testing before release. Final testing of the SC2000 included various safety and performance testing designed to ensure the device meets all of its specifications. Safety tests have been performed to ensure the device complies with applicable industry and safety standards including:
The Acuson SC2000™ has been designed to meet the following product safety standards:
- UL 60601-1, Safety Requirements for Medical Equipment .
- 이 IEC 60601-2-37 Diagnostic Ultrasound Safety Standards
- CSA C22.2 No. 601-1, Safety Requirements for Medical Equipment
- AIUM/NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical ■ Acoustic Output Indices on Diagnostic Ultrasound Equipment
- . AIUM/NEMA UD-2, Acoustic Output Measurement Standard for Diagnostic Ultrasound
- . Safety and EMC Requirements for Medical Equipment
- · IEC 60601-1
- IEC 60601-1-1 .
- . IEC 60601-1-2
- . IEC 1157 Declaration of Acoustic Power
- . ISO 10993-1 Biocompatibility
## 8. A summary discussion of the clinical tests submitted, referenced, or relied on for a determination of substantial equivalent
Since the SC2000 uses the same technology and principles as existing devices, clinical data is not required.
## 9. Summary
Intended uses and other key features are consistent with traditional clinical practice and FDA quidelines. The design and development process of the manufacturer conforms with 21CFR820 Quality System Requlation and ISO 13485:2003 quality system standards. The product is designed to conform with applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound has accumulated a long history of safe and effective performance. Therefore it is the opinion of Siemens Medical that the SC2000 is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the bird symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
December 5, 2012
Siemens Medical Solutions USA, Inc. Ultrasound Group % Mr. Mark Job Responsible Third Party Official · Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313
Re: K123622
Trade/Device Name: SC2000TM Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX, and OBJ Dated: November 21, 2012 Received: November 23, 2012
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the SC2000™ Diagnostic Ultrasound System, as described in your premarket notification:
### Transducer Model Number
| 9L4 | AcuNav 8F and 10F Ultrasound Catheter |
|---------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| V5M TEE | ACUSON AcuNav™ V 10F Ultrasound |
| 4V1c | Catheter |
| 8V3c | SoundStar 10F Ultrasound Catheter |
| AUX CW2 | V7M TEE |
| 4Z1c | 10V4 |
| | ー・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ |
If your device is classified (see above) into either class II (Special Controls) or class III (PMA),
{6}------------------------------------------------
it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Jeffrey Ballyns at (301) 796-6105.
Sincerely Yours,
Michael D. O'Hare
Janine M. Morris Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{7}------------------------------------------------
## 510(k) Number (if known):
#### Device Name:
SC2000™ Diagnostic Ultrasound System
#### Indications for Use:
The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Abdominal, Abdominal Intraoperative, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are:
#### Cardiac Imaging Applications and Analysis
The system transmits ultrasound energy into adult, pediatric, neonatal, and fetal cardiac patients creating 2D (B), 3D, M-Mode (M), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.
The system also supports catheters which are intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients.
The system transmits ultrasound energy from either a transthoracic or transesophageal approach in adult and pediatric patients; and from a transthoracic approach in neonatal and fetal cardiac patients creating 2D (B), 3D, M-Mode (M), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.
The system has Cardiac Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
## Vascular Imaging Applications and Analysis
The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the carotid arteries or juggler veins in the neck; superficial and deep veins and arteries in the arms and legs; and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.
The system has Vascular Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
#### Superficial Imaging Applications
The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of conventional or superficial musculoskeletal structures and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.
{8}------------------------------------------------
### Intraoperative Imaging Applications
The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), and Pulsed Wave Doppler (PWD) to obtain images and blood flow velocity that provide guidance during neurological intraoperative procedures.
#### Transcranial Imaging Applications
The system transmits ultrasound energy into the cranium of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the brain and surrounding anatomical structures to evaluate the presence or absence of pathology.
The system provides Measurement Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In-Vitro Diagnostics and Radiological Health (OIR)
Page 1 of
Muhal D. O'kin
(Division Sign Off)
(Division Sign Off
Division of Radiological Health
Office of In Vitro Diagnostics and Radiological Health
510(k) K123622
{9}------------------------------------------------
510(k) Number (if known):
Device Name:
## SC2000 Diagnostic Ultrasound System
Ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>3D | Other:<br>Real<br>Time<br>3D |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|--------------|------------------------------|
| Ophthalmic | | | | | | | | | | | | |
| Fetal | P | P | P | P | P | P | P | | P* | P | | P |
| Abdominal | P | P | P | P | P | P | P | | P* | P | | |
| Intraoperative<br>Abdominal | P | P | P | P | P | P | P | | P* | P | | |
| Intraoperative<br>Neurological | P | P | P | P | | P | P | | P* | P | | |
| Pediatric | P | P | P | P | P | P | P | | P* | P | P | P |
| Small Organ<br>(specify) | | | | | | | | | | | | |
| Neonatal<br>Cephalic | | | | | | | | | | | | |
| Adult Cephalic | P | P | P | P | P | P | P | | P* | P | | |
| Cardiac | P | P | P | P | P | P | P | | P* | P | P | P |
| Trans-esophageal | P | P | P | P | P | P | | | P* | | P | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | |
| Intra-Luminal | P | P | P | P | P | P | P | | P* | | | P |
| Peripheral Vessel | P | P | P | P | P | P | P | P | P* | P | | |
| Laparoscopic | | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | P | P | P | P | | P | P | P | P* | P | | |
| Musculo-skeletal<br>Superficial | P | P | P | P | | P | P | P | P* | P | | |
| Other (Neonatal<br>Cardiac) | P | P | P | P | P | P | P | | P* | P | | |
| Other (Intra-<br>Cardiac) | P | P | P | P | P | P | P | | P* | | | P |
N=new indication. P = Previously Cleared in 510(k) K072365, K102017, K113179
#### Additional Comments:
*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics and Radiological Health (OIR)
Prescription Use (Per 21 CFR 801.109)
(Division Sign Off)
Division of Radiological Health
Page 2 of
Office of in Vitro Diagnostics a
510(k) K123622
{10}------------------------------------------------
510(k) Number (if known):
Device Name:
### 9L4
Indications for Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging |
|----------------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | P | P | P | | P | P | P | P* | P |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(specify) | | | | | | | | | | |
| Neonatal<br>Cephalic<br>Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intra-Luminal | | | | | | | | | | |
| Peripheral Vessel | | P | P | P | | P | P | P | P* | P |
| Laparoscopic | | P | P | P | | P | P | P | P* | P |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | P | P* | P |
| Musculo-skeletal<br>Superficial | | P | P | P | | P | P | P | P* | P |
| Other (specify) | | | | | | | | | | |
N=new indication. P = Previously Cleared in 510(k) K072365, K102017, K113179
#### Additional Comments:
*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics and Radiological Health (OIR) Prescription Use (Per 21 CFR 801.109)
(Division Sign Off)
Page 3 of
Page 3 of
Division of Radiological Health
Office of In Vitro Diagnostics and Radiologica
510(k) K123622
{11}------------------------------------------------
## 510(k) Number (if known):
Device Name:
## V5M TEE
Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>3D | |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|--------------|--|
| Ophthalmic | | | | | | | | | | | | |
| Fetal | | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | | |
| Intraoperative<br>Abdominal | | | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | | | |
| Pediatric | | P | P | P | P | P | | | P* | | P | |
| Small Organ<br>(specify) | | | | | | | | | | | | |
| Neonatal<br>Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac | | P | P | P | P | P | | | P* | | P | |
| Trans-esophageal | | P | P | P | P | P | | | P* | | P | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | | |
| Intra-Luminal | | | | | | | | | | | | |
| Peripheral Vessel | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | | |
N=new indication. P = Previously Cleared in 510(k) K072365, K102017, K113179
#### Additional Comments:
*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE .
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics and Radiological Health (OlR) Prescription Use (Per 21 CFR 801.109)
Prescription Use (Per 21 CFR 801.109)
(Division Sign Off)
Page 4 of
Division of Radiological Health
Office of In Vitro Diagnostics and Rac
510(k) K123622
{12}------------------------------------------------
510(k) Number (if known):
Device Name:
## 4V1c
Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging |
|----------------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | P | P | P | | P* | P |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | P | P | P | P | P | | P* | P |
| Small Organ<br>(specify) | | | | | | | | | | |
| Neonatal<br>Cephalic<br>Adult Cephalic | | P | P | P | P | P | P | | P* | P |
| Cardiac | | P | P | P | P | P | P | | P* | P |
| Trans-esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intra-Luminal | | | | | | | | | | |
| Peripheral Vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (Neonatal<br>Cardiac) | | P | P | P | P | P | P | | P* | P |
N=new indication. Previously Cleared in 510(k) K072365, K102017, K113179
510k
#### Additional Comments:
*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE
(PLESE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics and Radiological Health (OIR) Prescription Use (Per 21 CFR 801.109)
Michael D. O'Hare
(Division Sign Off)
Page 5 of
(Division Sign Only)
Division of Radiological Health
Office of In Vitro Di ical Health
{13}------------------------------------------------
510(k) Number (if known):
Device Name:
# 8V3c
Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging |
|----------------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | P | P | P | | P * | P |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | P | P | P | P | P | P | | P * | P |
| Small Organ<br>(specify) | | | | | | | | | | |
| Neonatal<br>Cephalic<br>Adult Cephalic | | | | | | | | | | |
| Cardiac | | P | P | P | P | P | P | | P * | P |
| Trans-esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intra-Luminal | | | | | | | | | | |
| Peripheral Vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (Neonatal<br>Cardiac) | | P | P | P | P | P | P | | P * | P |
N=new indication. Previously Cleared in 510(k) K102017, K113179
#### Additional Comments:
*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE
(PLESE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics and Radiological Health (OIR) Prescription Use (Per 21 CFR 801.109)
Minh D. O'Kee
(Division Sign Off)
Division of Radiological Healt
Jitice of In Vitro Diagnostics and Radiological Health
510(k) K123622
{14}------------------------------------------------
510(k) Number (if known):
Device Name:
# AUX CW2
Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>Abdominal | | | | | | | | | | |
| Intraoperative<br>Neurological | | | | | | | | | | |
| Pediatric | | | | | P | | | | | |
| Small Organ<br>(specify) | | | | | | | | | | |
| Neonatal<br>Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | P | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intra-Luminal | | | | | | | | | | |
| Peripheral Vessel | | | | | P | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N=new indication. Previously Cleared in 510(k) K072365, K102017, K113179
### Additional Comments:
(PLESE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics and Radiological Health (OIR) Prescription Use (Per 21 CFR 801.109)
Michael D'John
Page 7 of
(Division Sign Off)
Division of Radiological Health
Office of In Vitro Diagnostics and Radiological Health
510(k) K123622
{15}------------------------------------------------
510(k) Number (if known):
Device Name:
### 4Z1c
Intended Use:
Ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>Real<br>Time<br>3D |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|------------------------------|
| Ophthalmic | | | | | | | | | | | |
| Fetal | | P | P | P | P | P | | | P* | P | P |
| Abdominal | | | | | | | | | | | |
| Intraoperative<br>Abdominal | | | | | | | | | | | |
| lntraoperative<br>Neurological | | | | | | | | | | | |
| Pediatric | | P | P | P | P | P | | | P* | P | P |
| Small Organ -<br>(specify) ** | | | | | | | | | | | |
| Neonatal<br>Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | P | P | P | P | P | | | P* | P | P |
| Trans-esophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intra-Luminal | | | | | | | | | | | |
| Peripheral Vessel | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
N=new indication. P = Previously Cleared in 510(k) K072365, K102017, K113179
#### Additional Comments:
*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics and Radiological Health (OIR) Prescription Use (Per 21 CFR 801.109)
Mhhel D.O.Thm
(Division Sign Off)
Page 8 of
Division of Radiological Health
Office of In V
5100 K123622
{16}------------------------------------------------
510(k) Number (if known):
Device Name:
## AcuNav 8F and 10F Ultrasound Catheter
Intended Use:
Catheter is intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients as follows:
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>Real<br>Time<br>3D |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|------------------------------|
| Ophthalmic | | | | | | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.