K123619 · Merete Medical GmbH · LZO · May 24, 2013 · Orthopedic
Device Facts
Record ID
K123619
Device Name
INTRABLOCK BIOBALL HIP SYSTEM (IBS)
Applicant
Merete Medical GmbH
Product Code
LZO · Orthopedic
Decision Date
May 24, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IntraBlock™ BioBall™ Hip System (IBS) is intended for use in total hip arthroplasty. The IBS is indicated for: Patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, Avascular necrosis of the femoral head, Correction of functional deformity, Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques, Revision procedures where other treatment or devices have failed. The IBS is indicated for cementless use only.
Device Story
IntraBlock™ BioBall™ Hip System (IBS) is a total hip replacement prosthesis for cementless primary or revision arthroplasty. System components include femoral stems, heads, taper adapter sleeves, and acetabular cups. Stems are straight, collarless, titanium alloy (Ti-6Al-4V ELI) with proximal titanium plasma spray (TPS) coating; available in multiple sizes with 135° neck-shaft angle and 12/14 morse taper. Ceramic heads are Biolox® delta (Alumina Matrix Composite). Acetabular cups are titanium alloy with TPS coating and UHMWPE inlays. Surgeons implant the device in an OR setting. The system restores hip joint function, alleviates pain, and corrects deformities. Clinical benefit is achieved through mechanical reconstruction of the hip joint.
Clinical Evidence
Bench testing only. Testing included distal and proximal fatigue testing per FDA guidance for femoral stem prostheses, range of motion analysis, and characterization of the titanium plasma spray (TPS) coating. Modular connection analysis (fretting and corrosion) and static compression testing were performed for the ceramic ball hip system and liner locking mechanism.
Technological Characteristics
Materials: Ti-6Al-4V ELI (ASTM F-136) with titanium plasma spray (TPS) coating; Biolox® delta (ISO 6474-2) ceramic; UHMWPE. Design: Modular hip prosthesis with 12/14 morse taper trunnion. Stems: 6.25-15.00 mm sizes, 135° neck-shaft angle. Cups: 48-62 mm diameter. Connectivity: None (mechanical implant). Sterilization: Not specified.
Indications for Use
Indicated for patients with severe hip pain/disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, avascular necrosis of femoral head, functional deformity, or proximal femur fractures (non-union, neck, trochanteric) unmanageable by other techniques. Also indicated for revision procedures. For cementless use only.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
# MAY 2 4 2013
## 2. 510(k) Summary of Safety and Effectiveness Information
| Date Prepared: | November 2012 |
|----------------|-----------------------|
| Submitted by: | Merete Medical GmbH |
| | Alt Lankwitz 102 |
| | 12247 Berlin, Germany |
#### FDA Registration Number: 3002949614
Contact Person:
:: : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : : :
Rüdiger Hilken Merete Medical, Inc. 4 Crotty Lane - Suite 118 New York International Plaza New Windsor, NY 12553 Phone: 914 967 1532
Device Name: IntraBlock™ BioBall™ Hip System (IBS)
Common Name: Total hip replacement device
Classification Names: Hip Joint metal/ceramic/polymer semi-constrained. cemented or nonporous uncemented prosthesis (21 CFR 888.3353)
Product Code: LZO
Proposed Regulatory Class: Class II
K073567 - Biolox® delta Option Ceramic Heads Predicate Devices: K043537 - Taperloc® 12/14 Taper Femoral Components K121563 - Corin Trift TS Hip K001534 - Pinnacle™ Acetabular Cup K000306 - Pinnacle™ Acetabular Cup K093472 - Trinity Acetabular System
#### Device Description:
The IntraBlock™ BioBall™ Hip System (IBS) is composed of femoral stems, heads. taper adapter sleeves and acetabulum cups. The System is designed for uncemented use for either primary or revision hip arthroplasty.
The IBS stem is straight, collarless, and flat with a tapered design. The stem is manufactured from titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F-136 with a proximal titanium plasma spray (TPS) coating. The stems are proportionally sized and shaped in sizes 6.25 mm to 15.00 mm to match the geometry of the femur. A lateralized version of the device is available. Both versions have a neck shaft angle of 135° and a 12/14 morse taper trunion. The Ceramic Head is made of Biolox® delta (Alumina Matrix Composite ISO 6474-2) and available in diameter 28 mm.
The taper adapter sleeves are made from titanium alloy (Ti-6AI-4V ELI) with a 12/14 morse taper trunion in the four neck lengths S, M, L and XL.
Merete Medical GmbH
Submission Date: 2012-11-16 Rev. 2: 2013-04-18_Rev.3: 2013-05-17
{1}------------------------------------------------
#### 510(k) Submission - IntraBlock™ BioBall™ Hip System (IBS)
The cementless MultiCup Locking PressFit is manufactured from titanium alloy (Ti-6Al-4V ELI) conforming to ASTM F-136 with a titanium plasma spray (TPS) coating on the outer surface. It is used with a UHMWPE inlay sized for Ø28 mm heads. 5 holes with threaded plugs give the option for an additional fixation with locking screws. The MultiCup is available in 9 different sizes ranged from 48mm up to 62mm outer diameter.
#### Indications for Use
The IntraBlock™ BioBall™ Hip System (IBS) is intended for use in total hip arthroplasty.
The IBS is indicated for:
- Patients suffering from severe hip pain and disability due to rheumatoid l arthritis, osteoarthritis, traumatic arthritis
- Avascular necrosis of the femoral head
- Correction of functional deformity
- Treatment of non-union, femoral neck fracture, and trochanteric . fractures of the proximal femur with head involvement, unmanageable by other techniques
- Revision procedures where other treatment or devices have failed ﺳ
The IBS is indicated for cementless use only.
#### Substantial Equivalence:
The IntraBlock™ BioBall™ Stems, MultiCup™ Locking PressFit, Femoral Heads and Titanium Taper Adapter Sleeves are technologically similar to the predicate devices.
## Legally marketed Devices to which substantial Equivalence is claimed
K073567 - Biolox® delta Option Ceramic Heads
K043537 - Taperioc® 12/14 Taper Femoral Components
- K121563 Corin Trift TS Hip
K001534 - Pinnacle™ Acetabular Cup
K000306 - Pinnacle™ Acetabular Cup
K093472 - Trinity Acetabular System
#### Non-Clinical Performance Data:
Non-clinical testing and analysis were provided, including mechanical testing and coating characterization. Mechanical testing included distal and proximal fatique testing of the worst case consistent with the "Guidance for Industry and FDA staff Non-Clinical Information for femoral Stem Prostheses". Range of Motion analysis was also performed. The TPS coating underwent characterization per FDA´s .Guidance for Industry on Testing of Metallic Plasma Sprayed Coatings on Orthopaedic Implants to support reconsideration of Post Market Surveillance Requirements". The Modular connection analysis including fretting and corrosion were also performed as well as static compression testing consistent with the "Guidance document for the preparation of premarket notifications for ceramic ball hip systems". Furthermore the connections between the stem, the adapter and the head, as well as the liner locking mechanism were tested.
Merete Medical GmbH
Submission Date: 2012-11-16 Rev. 2: 2013-04-18 Rev.3: 2013-05-17
Page 15 of 75
{2}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a bird in flight, composed of three curved lines.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
## May 24, 2013
Merete Medical GmbH % Mr. Emanuel Anapliotis President Alt-Lankwitz 102 Berlin 12247 Germany
Re: K123619
Trade/Device Name: IntraBlock™ BioBall™ Hip System (IBS) Regulation Number: 21 CFR 888.3353 · Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis. Regulatory Class: Class II Product Code: LZO Dated: April 18, 2013 Received: April 24, 2013
Dear Mr. Anapliotis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
#### Page 2 - Mr. Emanuel Anapliotis
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Image /page/3/Picture/8 description: The image shows the name "Erin DKeith" in a bold, sans-serif font. The letters "E", "r", "i", "n", "e", "i", "t", and "h" are solid black, while the letters "D" and "K" have a patterned design. The name appears to be a logo or a stylized signature.
For
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## 1. Indications for Use Statement
## Indications for Use
510(k)·Number (if known): K123619
Device Name: IntraBlock™ BioBall™ Hip System (IBS)
Indications for Use:
The IntraBlock™ BioBall™ Hip System (IBS) is intended for use in total hip arthroplasty.
The IBS is indicated for:
- Patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis,
- -Avascular necrosis of the femoral head,
- Correction of functional deformity. ー
- -Treatment of non-union, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement, unmanageable by other techniques,
- Revision procedures where other treatment or devices have failed. -
AND/OR
The IBS is indicated for cementless use only.
Prescription Use
(Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth Frank -S
Division of Orthopedic Devices
Merete Medical GmbH
Submission Date: 2012-11-16 Rev. 2: 2013-04-18_Rev.3: 2013-05-17
Page 13 of 75
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.