The Philips Ingenuity Digital PET/CT System is a diagnostic imaging device that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem images anatomical cross-sections by computer reconstruction of x-ray transmission data. The PET subsystem images the distribution of PET anatomy-specific radiopharmaceuticals in the patient. The PET/CT system is used for the purpose of detecting, localizing, diagnosing, staging, re-staging and follow-up for monitoring therapy response of various diseases in oncology, cardiology and neurology. The system is intended to image the whole body, heart, brain, lung, gastrointestinal, bone, lymphatic, and other major organs for a wide range of patient types, sizes, and extent of diseases. Both subsystems can also be operated as fully functional, independent diagnostic tools including application of the CT scanner for diagnosis and for use in radiation therapy planning.
Device Story
Hybrid diagnostic imaging system combining PET and CT scanners in a single gantry; utilizes PET radiopharmaceuticals and X-ray transmission data. PET subsystem images radiopharmaceutical distribution; CT subsystem performs anatomical cross-section reconstruction. Operated by clinicians in clinical settings for oncology, cardiology, and neurology applications. Output consists of diagnostic images used by physicians to detect, localize, stage, and monitor disease or plan radiation therapy. Benefits include integrated anatomical and functional imaging for improved diagnostic accuracy and therapy management.
Clinical Evidence
Clinical images were provided in accordance with FDA guidance for Emission Computed Tomography Devices. No specific quantitative clinical study results (e.g., sensitivity, specificity) are reported in the summary.
Technological Characteristics
Hybrid PET/CT system; integrated gantry design; PET detector technology; X-ray transmission CT. Complies with NEMA NU-2 for performance and IEC 60601 series for electrical safety.
Indications for Use
Indicated for patients of all ages and sizes requiring diagnostic imaging for oncology, cardiology, and neurology. Used for detection, localization, diagnosis, staging, re-staging, and therapy monitoring of diseases in the whole body, heart, brain, lung, gastrointestinal, bone, and lymphatic systems. Can be used for radiation therapy planning.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The letters are black and appear to have a slightly rough or textured edge. The word is horizontally oriented and centered in the image.
K123599
page 1 of 3
Philips Medical Systems (Cleveland), Inc. 510(k) Premarket Notification Submission
.
. .
JUL 1 2 2013
## Section 005 510(k) Summary
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The letters are black and appear to be slightly pixelated. The word is centered and takes up most of the frame.
Philips Medical Systems (Cleveland), Inc. 510(k) Premarket Notification Submission
510(k) Summary in accordance with 21 CFR 807.92
| Date: | 20 November 2012 |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Philips Medical Systems (Cleveland), Inc.<br>595 Miner Road<br>Cleveland, OH 44143 USA<br>Establishment Registration: 1525965 |
| Primary Contact Person: | Michael S. Preto<br>Regulatory Affairs Specialist, CT/NM Regulatory Affairs<br>Philips Medical Systems (Cleveland), Inc.<br>Tel: +1 440 483 2295<br>Fax: +1 440 483 7355<br>E-mail: michael.preto@philips.com |
| Secondary Contact<br>Person: | Melinda Novatny<br>Senior Manager, CT/NM Regulatory Affairs<br>Philips Medical Systems (Cleveland), Inc.<br>Tel: +1 440 483 4255<br>Fax: +1 440 483 7355<br>E-mail: melinda.novatny@philips.com |
| Device Trade Name: | Ingenuity Digital PET/CT System |
| Common/Usual Name: | Combination PET/CT System |
| Classification Names: | Emission Computed Tomography System, 21 CFR 892.1200<br>X-ray Computed Tomography, 21 CFR 892.1750 |
| Product Code: | 90KPS<br>90JAK |
| Predicate Device(s): | K052640, Gemini Raptor |
| Device Description: | The device is a hybrid diagnostic imaging system that combines<br>Positron Emission tomography and X-ray computed tomography<br>scanners that can be utilized in fixed installations. The device is<br>comprised of the following system components/subsystems:<br>● Positron Emission Tomography (PET) scanner<br>● X-ray Computed Tomography (CT) scanner<br>● Patient Table |
{2}------------------------------------------------
K123599
Page 3 of 3
Image /page/2/Picture/1 description: The image shows the word "PHILIPS" in a bold, sans-serif font. The letters are black and have a slightly textured appearance. The word is centered and takes up most of the frame.
Philips Medical Systems (Cleveland), Inc. 510(k) Premarket Notification Submission
| Intended Use: | The Philips Ingenuity Digital PET/CT System is a diagnostic imaging device that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem images anatomical cross-sections by computer reconstruction of x-ray transmission data. The PET subsystem images the distribution of PET anatomy-specific radiopharmaceuticals in the patient. The PET/CT system is used for the purpose of detecting, localizing, diagnosing, staging, re-staging and follow-up for monitoring therapy response of various diseases in oncology, cardiology and neurology. The system is intended to image the whole body, heart, brain, lung, gastrointestinal, bone, lymphatic, and other major organs for a wide range of patient types, sizes, and extent of diseases. Both subsystems can also be operated as fully functional, independent diagnostic tools including application of the CT scanner for diagnosis and for use in radiation therapy planning. |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technology: | The Ingenuity Digital PET/CT System has replaced the PET detector design and has integrated both PET and CT subsystems into a single gantry. The system operates on the same principles as the predicate device. |
| Determination of Substantial Equivalence: | The Ingenuity Digital PET/CT System has comparable indications for use as its predicate device.<br><br>Summary of Nonclinical Tests:<br>In accordance with "Guidance for the Submission of Premarket Notifications for Emission Computed Tomography Devices and Accessories (SPECT and PET) and Nuclear Tomography Systems ", system performance is provided using NEMA NU-2. Safety evidence is provided using IEC 60601 series of standards.<br><br>Summary of Clinical Tests:<br>In accordance with "Guidance for the Submission of Premarket Notifications for Emission Computed Tomography Devices and Accessories (SPECT and PET) and Nuclear Tomography Systems ", clinical images are included. |
| Conclusion: | Philips Medical Systems (Cleveland), Inc. considers the Ingenuity Digital PET/CT System to be as safe, as effective, and performance is substantially equivalent to the predicate device. |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services (HHS). The logo features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The caduceus is encircled by the words "U.S. Department of Health & Human Services" in a circular arrangement.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 12, 2013
Philips Medical Systems (Cleveland), Inc. % Mr. Michael S. Preto Regulatory Affairs Specialist 595 Miner Road CLEVELAND OH 44143
Re: K123599
Trade/Device Name: Ingenuity Digital PET/CT System Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS & JAK Dated: June 6, 2013 Reccived: June 11, 2013
Dear Mr. Preto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Insting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Preto
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm.
Sincerely yours,
Michael D. O'Hara
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
for
Enclosure
{5}------------------------------------------------
## Indications for Use Form
K123599
510(k) Number (if known): Device Name: Ingenuity Digital PET/CT System Indications for Use:
The Philips Ingenuity Digital PET/CT System is a diagnostic imaging device that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem images anatomical crosssections by computer reconstruction of x-ray transmission data. The PET subsystem images the distribution of PET anatomy-specific radiopharmaceuticals in the patient. The PET/CT system is used for the purpose of detecting, localizing, diagnosing, staging, re-staging and follow-up for monitoring therapy response of various diseases in oncology, cardiology and neurology. The system is intended to image the whole body, heart, brain, lung, gastrointestinal, bone, lymphatic, and other major organs for a wide range of patient types, sizes, and extent of diseases. Both subsystems can also be operated as fully functional, independent diagnostic tools including application of the CT scanner for diagnosis and for use in radiation therapy planning.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
AND/OR
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Michael D. O'Hara
Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety
510(k)_ K123599
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Page 4-2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.