Retrospective clinical samples were used to establish reference intervals and to validate the clinical sensitivity and specificity of the QUANTA Flash Sm and RNP assays against clinical diagnoses.
Clinical Validation Study (Sm); Retrospective clinical cohort analysis
Patients with SLE, systemic sclerosis, rheumatoid arthritis, Sjögren's syndrome, autoimmune liver disease, viral hepatitis, and other systemic rheumatic/infectious diseases.; Sample Size: 379; Number of Sites: 3 (Neuss Center for Rheumatology, Dr. Carlos von Mühlen, and other sources)
Predicate ELISA
Clinical sensitivity and specificity for SLE
Clinical Validation Study (RNP); Retrospective clinical cohort analysis
Patients with SLE, MCTD, systemic sclerosis, rheumatoid arthritis, Sjögren's syndrome, autoimmune liver disease, viral hepatitis, and other systemic rheumatic/infectious diseases.; Sample Size: 424; Number of Sites: 3 (Neuss Center for Rheumatology, Dr. Carlos von Mühlen, and other sources)
Predicate ELISA
Clinical sensitivity and specificity for SLE and MCTD
Indications for Use
The QUANTA Flash Sm is a chemiluminescent immunoassay for the semi-quantitative determination of lgG anti-Sm antibodies in human serum. The presence of antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of Systemic Lupus Erythematosus (SLE). The QUANTA Flash RNP is a chemiluminescent immunoassay for the semi-quantitative determination of lgG anti-ribonucleoprotein (RNP) antibodies in human serum. The presence of anti-RNP antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of Systemic Lupus Erythematosus (SLE) and Mixed Connective Tissue Disease (MCTD). QUANTA Flash Sm Calibrators are intended for use with the QUANTA Flash Sm chemiluminescent immunoassay for the determination of IgG anti-Sm antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values. QUANTA Flash RNP Calibrators are intended for use with the QUANTA Flash RNP chemiluminescent immunoassay for the determination of IgG anti-RNP antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values. QUANTA Flash Sm Controls are intended for use with the QUANTA Flash Sm chemiluminescent immunoassay for quality control in the determination of IgG anti-Sm antibodies in human serum. QUANTA Flash RNP Controls are intended for use with the QUANTA Flash RNP chemiluminescent immunoassay for quality control in the determination of IgG anti-RNP antibodies in human serum.
Device Story
Device is a chemiluminescent immunoassay (CIA) for semi-quantitative detection of IgG anti-Sm and anti-RNP antibodies in human serum; utilizes BIO-FLASH automated analyzer. Input: patient serum sample; process: sample diluted, incubated with native Sm/RNP antigen-coated paramagnetic beads, washed, reacted with isoluminol-labeled anti-human IgG, and triggered with Fe(III) coproporphyrin/urea-hydrogen peroxide; output: relative light units (RLU) converted to chemiluminescent units (CU) via lot-specific master curve. Used in clinical laboratories by trained personnel. Output aids clinicians in diagnosing SLE and MCTD in conjunction with other clinical findings.
Clinical Evidence
Clinical validation performed on 379 samples for Sm and 424 samples for RNP. Sm assay: 14.4% sensitivity, 97.9% specificity for SLE. RNP assay: 37.1% sensitivity, 95.1% specificity for SLE/MCTD combined; 75.0% sensitivity for MCTD. ROC analysis showed diagnostic efficiency (AUC 0.62 for Sm; 0.61-0.87 for RNP). Precision studies (CLSI EP5-A2) showed total CV <15%. Interference and hook effect studies confirmed no significant impact.
Technological Characteristics
Solid phase: paramagnetic microparticles (beads) coated with native Sm or RNP antigen (calf thymus). Detection: chemiluminescent immunoassay (isoluminol-conjugated anti-human IgG). Energy: luminometer (optical system). Connectivity: BIO-FLASH instrument (automated, closed system). Calibration: lot-specific Master Curve + two calibrators. Software: automated processing, auto-rerun capability for samples above AMR.
Indications for Use
Indicated for semi-quantitative determination of IgG anti-Sm or anti-RNP antibodies in human serum to aid in diagnosis of Systemic Lupus Erythematosus (SLE) and Mixed Connective Tissue Disease (MCTD).
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
Predicate Devices
QUANTA Lite™ Sm ELISA (k922831)
QUANTA Lite™ RNP ELISA (k922833)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
k123593
B. Purpose for Submission:
New Device
C. Measurand:
Anti-Sm antibodies (IgG)
Anti-ribonucleoprotein (RNP) antibodies (IgG)
D. Type of Test:
Semi-quantitative, Chemiluminescent Immunoassay (CIA)
E. Applicant:
INOVA Diagnostics, Inc.
F. Proprietary and Established Names:
QUANTA Flash® Sm
QUANTA Flash® Sm Calibrators
QUANTA Flash® Sm Controls
QUANTA Flash® RNP
QUANTA Flash® RNP Calibrators
QUANTA Flash® RNP Controls
G. Regulatory Information:
1. Regulation section:
21 CFR § 866.5100 Antinuclear Antibody Immunological Test System
21 CFR § 862.1150 Calibrator
21 CFR § 862.1660 Quality Control Material (assayed and unassayed)
2. Classification:
Class II – Assay and Calibrator
Class I – Quality Control Material
3. Product code:
LKP - Anti-Sm Antibody, Antigen and Control
LKO - Anti-RNP Antibody, Antigen and Control
JIT - Calibrator, Secondary
JJX - Single (Specified) Analyte Controls (Assayed and Unassayed)
{1}
4. Panel:
Immunology (82)
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
QUANTA Flash Sm
The QUANTA Flash Sm is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-Sm antibodies in human serum. The presence of anti-Sm antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of Systemic Lupus Erythematosus (SLE).
QUANTA Flash Sm Calibrators are intended for use with the QUANTA Flash Sm chemiluminescent immunoassay for the determination of IgG anti-Sm antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.
QUANTA Flash Sm Controls are intended for use with the QUANTA Flash Sm chemiluminescent immunoassay for quality control in the determination of IgG anti-Sm antibodies in human serum.
QUANTA Flash RNP
The QUANTA Flash RNP is a chemiluminescent immunoassay for the semi-quantitative determination of IgG anti-ribonucleoprotein (RNP) antibodies in human serum. The presence of anti-RNP antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of Systemic Lupus Erythematosus (SLE) and Mixed Connective Tissue Disease (MCTD).
QUANTA Flash RNP Calibrators are intended for use with the QUANTA Flash RNP chemiluminescent immunoassay for the determination of IgG anti-RNP antibodies in human serum. Each calibrator establishes a point of reference for the working curve that is used to calculate unit values.
QUANTA Flash RNP Controls are intended for use with the QUANTA Flash RNP chemiluminescent immunoassay for quality control in the determination of IgG anti-RNP antibodies in human serum.
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
For Prescription Use only
4. Special instrument requirements:
BIO-FLASH® chemiluminescent analyzer (k083518)
I. Device Description:
Materials provided:
{2}
1. QUANTA Flash® Sm/RNP Reagent Cartridge contains reagents for 50 determinations. Each cartridge has a barcode that contains the assay name, the assay ID number, the lot number and expiration date, the four parameters of the lot specific master curve, and the reagent cartridge specific serial number. The BIO-FLASH software monitors the expiration dates of the onboard cartridges, as well as the reagent cartridge lots.
a. Sm/RNP coated paramagnetic beads, lyophilized.
b. Assay buffer -colored pink, containing Tris-buffered saline, Teen 20, protein stabilizers and preservatives.
c. Tracer IgG - Isoluminol labeled anti-human IgG antibody, containing buffer, protein stabilizers and preservatives.
2. Resuspension buffer, 1 vial, one time use - colored pink, containing buffer, protein stabilizers and preservatives.
Materials required but not provided:
1. BIO-FLASH® instrument with operating computer
2. BIO-FLASH® System Rinse (part number: 3000-8205)
3. BIO-FLASH® Triggers (part number: 3000-8204)
4. BIO-FLASH® Cuvettes (part number: 3000-8206)
5. QUANTA-FLASH® Sm/RNP Calibrators (part number 701121 / 701116);
6. QUANTA-FLASH® Sm/RNP Controls (part number: 701122 / 701117
J. Substantial Equivalence Information:
1. Predicate device name(s) and 510(k) number(s):
| Predicate device | Predicate (510k) number(s) |
| --- | --- |
| QUANTA Lite™ Sm ELISA | k922831 |
| QUANTA Lite™ RNP ELISA | k922833 |
2. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | QUANTA Flash Sm/RNP | QUANTA Lite Sm/RNP |
| Intended Use
QUANTA Flash Sm | For the semi-quantitative determination of IgG anti-Sm antibodies in human serum. The presence of anti-Sm antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of Systemic Lupus | For the semi-quantitative detection of Sm/RNP antibodies in human serum. The presence of Sm/RNP antibodies can be used in conjunction with clinical findings and other laboratory tests |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | QUANTA Flash Sm/RNP | QUANTA Lite Sm/RNP |
| QUANTA Flash RNP | Erythematosus (SLE).For the semi-quantitative determination of IgG anti-ribonucleoprotein (RNP) antibodies in human serum.The presence of anti-RNP antibodies, in conjunction with clinical findings and other laboratory tests, can aid in the diagnosis of Systemic Lupus Erythematosus (SLE) and Mixed Connective Tissue Disease (MCTD). | to aid in the diagnosis of System Lupus Erythematosus (SLE) and related connective tissue diseases. |
| Assay Type | Semi-quantitative | Same |
| Traceability | In-house standards | Same |
| Analyte detected | Human IgG, antibodies to Sm/RNP | Same |
| Antigen | Native Sm/RNP antigen, purified from calf thymus | Same |
| Sample matrix | Serum | Same |
| Shelf life (reagent kit, calibrators, controls) | One year at 2-8°C | Same |
| Controls | Negative and Positive | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | QUANTA Flash Sm/RNP | QUANTA Lite Sm/RNP |
| Assay methodology | Chemiluminescent immunoassay | Enzyme-linked immunosorbent assay (colorimetric) |
| Instrument platform | BIO-FLASH® chemiluminescent analyzer | Manual |
| Solid phase | Paramagnetic micro particles (beads) | 96-well plate |
| Conjugate and detection antibody | Isoluminol conjugated anti-human IgG (monoclonal) | HRP conjugated anti-human IgG (polyclonal) |
| Calibration and | Lot specific Master Curve + | Sm/RNP single point |
| spectroscopy | Semi-quantitative | Semi-quantitative |
{4}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Calibrators | two Calibrators (Sold separately) | calibrator, low positive (Included in the kit) |
| Units | CU (chemiluminescent units), arbitrary | Units, arbitrary |
| Cutoff | 20 CU | 20 Units |
| Assay Range | | |
| Sm | 3.3 to 693.5 CU | N/A |
| RNP | 3.5 to 643.8 CU | |
# K. Standard/Guidance Document Referenced (if applicable):
CLSI EP7-02, Interference Testing in Clinical Chemistry; Approved Guideline
CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement. Procedures:
A Statistical Approach; Approved Guideline
CLSI EO17-A, Protocols for Determination of Limits of Detection and Limits of
Quantitation; Approved Guideline
CLSI EP5-A2, Evaluation of Precision Performance of Quantitative Measurement
Procedures - Approved Guideline
CLSIEP09-A2-IR Method Comparison and Bias Estimation Using Patient Samples;
Approved Guideline - Second Edition
CLSI C28-A# Defining, Establishing, and Verifying Reference Intervals in the
Clinical Laboratory; Approved Guideline
# L. Test Principle:
The QUANTA Flash Sm and RNP assays are designed to run on the BIO-FLASH instrument. This platform is a fully automated closed system with continuous load and random access capabilities that automatically processes the samples, runs the assay and reports the results. It includes liquid handling hardware, luminometer and computer with software-user interface. The QUANTA Flash Sm and RNP assays utilize a reagent cartridge format, which is compatible with the BIO-FLASH instrument.
Native Sm or RNP antigen that is purified from calf thymus is coated onto paramagnetic beads. The bead suspension is lyophilized and stored in the bead tube. Prior to use in the BIO-FLASH system, the sealed reagent tubes are pierced with the reagent cartridge lid and the beads are rehydrated and resuspended using resuspension buffer by pipetting up and down with a transfer pipette. The reagent cartridge is then loaded onto the BIO-FLASH® instrument. Samples are also loaded onto the instrument in sample racks. A patient serum sample is prediluted by the BIO-FLASH with system rinse in a small disposable plastic cuvette. Small amounts of the diluted patient serum, the beads, and assay buffer are all combined into a second cuvette, and mixed. This cuvette is then incubated at $37^{\circ}\mathrm{C}$ . The beads are magnetized and washed several times. Isoluminol conjugated anti-human IgG antibodies are then added to the cuvette, and again incubated at $37^{\circ}\mathrm{C}$ . The beads are magnetized and washed repeatedly. The Isoluminol conjugate is oxidized when Trigger 1 (Fe (III) coproporphyrin in sodium hydroxide solution) and Trigger 2
{5}
(urea-hydrogen peroxide in sodium chloride solution) are added to the cuvette, and the flash of light produced from this reaction is measured as Relative Light Units (RLU) by the BIO-FLASH optical system. The RLU are proportional to the amount of isoluminol conjugate that is bound to the human IgG, which is in turn proportional to the amount of anti-RNP antibodies bound to the Sm or RNP on the beads.
For quantitation, the QUANTA Flash Sm and RNP assays utilize a predefined lot specific Master Curve that is uploaded onto the instrument through the reagent cartridge barcode. Every new lot of reagent cartridge must be calibrated before first use with the QUANTA Flash Sm and RNP Calibrators. Based on the results obtained with the two Calibrators included in the Calibrator Set (sold separately), an instrument specific Working Curve is created, which is used to calculate chemiluminescent units (CU) from the instrument signal (RLU) obtained for each sample.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
The precision of the QUANTA Flash Sm/RNP assay was evaluated on samples containing various concentrations of Sm antibodies (5 serum samples) or RNP antibodies (7 serum samples). Samples were run in duplicates, twice a day, for at least 20 days. The %CV values were within the acceptance limit. The within run, between run, between day and total precisions are summarized in the table below:
QUANTA Flash Sm
| N | Mean Value (CU) | Within Run %CV | Between Run %CV | Between Day %CV | Total Imprecision %CV |
| --- | --- | --- | --- | --- | --- |
| 80 | 13.1 | 6.5 | 1.5 | 5.6 | 8.7 |
| 80 | 22.0 | 9.7 | 4.6 | 4.2 | 11.5 |
| 84 | 93.0 | 6.4 | 0.5 | 5.2 | 8.3 |
| 88 | 237.7 | 8.7 | 7.1 | 2.7 | 11.5 |
| 84 | 338.6 | 5.3 | 2.9 | 5.8 | 8.4 |
QUANTA Flash RNP
| N | Mean Value %CV | Within Run %CV | Between Run %CV | Between Day %CV | Total Imprecision %CV |
| --- | --- | --- | --- | --- | --- |
| 92 | 6.7 | 3.5 | 0.0 | 5.3 | 6.4 |
| 84 | 24.8 | 3.7 | 0.0 | 3.0 | 4.8 |
| 88 | 31.6 | 3.4 | 2.2 | 8.5 | 9.4 |
| 88 | 218.6 | 4.2 | 3.6 | 9.3 | 10.8 |
| 92 | 120.7 | 4.1 | 3.7 | 3.7 | 6.6 |
| 88 | 319.5 | 3.7 | 4.4 | 3.6 | 6.8 |
| 96 | 409.6 | 4.8 | 4.5 | 5.7 | 8.7 |
{6}
# Reproducibility:
Three samples for Sm and RNP each were tested on two different regent lots, using two different lots of calibrators, by two operators. Samples were run in quadruplicates, two times a day, for 10 days, to generate 80 data points. Data were analyzed and within run, between reagent lots, between calibrator lots, between operators and total precision were calculated and the results are summarized in the tables below. The $\% \mathrm{CV}$ values ranged between 1.7 to $10.7\%$ .
QUANTA Flash Sm:
| N | Mean Value (CU) | Within Run %CV | Between reagent Lots %CV | Between Calibrator Lots %CV | Between operators %CV | Total %CV |
| --- | --- | --- | --- | --- | --- | --- |
| 80 | 15.2 | 2.8 | 9.4 | 6.9 | 6.9 | 7.0 |
| 80 | 23.3 | 2.4 | 10.7 | 7.6 | 7.3 | 7.7 |
| 80 | 157.3 | 3.5 | 10.1 | 9.6 | 8.4 | 8.3 |
QUANTA Flash RNP:
| N | Mean Value (CU) | Within Run %CV | Between reagent Lots %CV | Between Calibrator Lots %CV | Between operators %CV | Total %CV |
| --- | --- | --- | --- | --- | --- | --- |
| 80 | 15.3 | 2.9 | 3.1 | 3.3 | 1.8 | 2.8 |
| 80 | 23.1 | 2.3 | 2.8 | 3.1 | 1.7 | 2.6 |
| 80 | 173.1 | 3.1 | 3.6 | 2.9 | 1.8 | 2.9 |
# b. Linearity/assay reportable range:
# Assay linear range:
Five serum samples with various anti-Sm or anti-RNP concentrations were diluted with a low negative serum sample to obtain values that cover the whole analytical measuring range. The percent recovery was more than $90\%$ . The results are summarized for all samples:
QUANTA Flash Sm
| Dilution range (CU) | Slope (95%CI) | Y-Intercept (95%CI) | R2 |
| --- | --- | --- | --- |
| 3.3 – 8.2 | 1.07 (0.98 to1.16) | -0.51 (-1.05 to 0.04) | 0.98 |
| 4.3-42.7 | 1.03 (0.98-1.07) | -0.34 (-1.57 to 0.90) | 0.99 |
| 6.1-61.2 | 1.02 (1.00 to 1.04) | -0.62 (-1.41 to 0.17) | 1.00 |
| 43.5-435.4 | 0.98 0.94 to 1.03 | -4.32 (-16.52 to 7.89) | 0.99 |
| 123.8-866.4 | 0.93 (0.86-1.01) | 38.2 (-3.8 to 80.3) | 0.99 |
{7}
The claimed linear range for Sm IgG is 3.3 to 693.5 CU
QUANTA Flash RNP
| Dilution range (CU) | Slope (95%CI) | Y-Intercept (95%CI) | R² |
| --- | --- | --- | --- |
| 3.8 to 37.7 | 1.02
(0.97 to 1.08) | 0.41
(-0.88 to 1.69) | 0.99 |
| 13.1 to 130.5 | 1.02
(0.98 to 1.06) | -5.66
(-8.84 to -2.48) | 0.99 |
| 12.9 to 128.6 | 1.02
(0.98 to 1.06) | -2.56
(-5.37 to 0.25) | 1.00 |
| 63.2 to 631.8 | 1.01
(0.98 to 1.05) | -15.99
(-29.12 to -2.86) | 1.00 |
| 81.1 to 730.2 | 1.07
(1.04 to 1.11) | -33.9
(-47.7 to -88.2) | 0.99 |
| 3.8 to 730.2
Combined | 1.01
(1.00 to 1.02) | -5.17
(-8.27 to -2.07) | 1.00 |
The claimed linear range for RNP IgG is 3.5 to 643.8 CU
Over the range detection:
The BIO-FLASH software has an auto-rerun option available. If this option is selected, the instrument will automatically rerun any sample that has a result >693.5 CU for Sm and >643.8 CU for RNP by further diluting it by a factor of 10, and calculate the actual CU using this additional dilution factor. Samples as high as 6935 CU for Sm, and 6438 CU for RNP were tested.
High hook effect:
Five Sm and six RNP high positive samples were tested. The assay does not appear to demonstrate a hook effect at high concentration: up to 2429 CU in the Sm assay and up to 3140 CU in the RNP assay.
d. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
There are no recognized international standards for the measurement of anti-Sm and anti-RNP antibodies. Calibrators and controls are traceable to in-house standards that are used to create the master curve for the QUANTA Flash Sm and QUANTA Flash RNP assays.
Value assignment:
Calibrators and controls are manufactured by diluting human serum that contains high titer of anti-Sm or anti-RNP antibodies with a buffer stabilizers and preservative. The human serum is obtained from commercial sources and it is tested for markers of infectious substances. The target CU value is achieved through trial dilutions on a small scale. Once a dilution is selected,
{8}
the calibrators and controls are bulked, tested, and adjusted. Upon completion of the manufacturing process, the calibrators and controls are tested on at least two instruments, on at least two lots, of reagent cartridge in replicates of 10 to determine final value assignment.
The QANTA Flash® Sm/RNP Calibrators are designed to adjust the predefined Master Curve into an instrument specific working curve.
Calibrators and controls are specified in the labeling but are not supplied with the assay. The target values and target ranges are the same for Sm and RNP. The table below summarizes the controls and calibrators target values and target ranges:
| | Sm/RNP |
| --- | --- |
| Calibrators | |
| Calibrator 1 | 12.0 CU |
| Calibrator 2 | 300.0 CU |
| Controls | |
| Negative Control | 10.0 CU |
| Positive Control | 50.0 CU |
**Kit stability:**
The claimed stability is based on accelerated studies and is summarized in the table below. Real time stability is ongoing. The BIO-FLASH software monitors the expiration dates of the onboard cartridges, as well as the reagent cartridge lots. Real time stability study is ongoing.
| | Sm | RNP |
| --- | --- | --- |
| Shelf Life | | |
| Reagent Cartridge, Calibrators and Controls | 1 year at 2-8C (until expiration date on label) | 1 year at 2-8C (until expiration date on label) |
| After being opened | | |
| Reagent Cartridge | 33 days | 28 days |
| Calibrators | 8 hours (onboard only); up to 4 uses. | 8 hours (onboard only); up to 4 uses. |
| Controls | 10 min per use, up to 2.5 hours total onboard or up to 15 uses (whichever is fulfilled first | 10 min per use, up to 2.5 hours total onboard or up to 15 uses (whichever is fulfilled first |
**e. Detection limit:**
**Limit of Blank (LoB):**
LoB was determined with 60 measurements on two blank (immunoglobulin stripped) serum samples. The LoB RLU values are lower than the bottom
{9}
limit of the four parameter logistic curve that the instrument uses to calculate CU, and therefore cannot be converted into CU.
## Limit of detection (LoD):
Five (5) serum samples were serially diluted until the RLU values no longer show a decreasing trend, to low-level samples that are in between the LoB to approximately 4 x LoB. The samples were run in replicates of 4 for 4 days to obtain 16 measurements per sample for a total of 80 measurements. The LoD of both assays is below lower limit of the reportable range which is 3.3 CU for Sm and 3.5 CU for RNP.
## e. Analytical specificity:
### Interfering substances:
Three samples for each analyte were tested: Sm -16.3 CU, 28.6 CU and 190.7 CU; RNP - 8.8 CU, 55.5 CU, and 280.3 CU. Interfering substances were spiked into the specimen in 10% of total specimen volume. The samples were tested in triplicates. Recovery of the unit values was calculated compared to control samples spiked with the same volume of diluent. No interference was observed up to the concentrations listed in the table below:
| Potential Interfering Compound | Concentration |
| --- | --- |
| Bilirubin | 10 mg/dL |
| Hemoglobin | 200 mg/dL |
| Triglycerides | 1000 mg/dL |
| Cholesterol | 224.3 mg/dL |
| Rheumatoid factor (RF) | 500 IU/mL |
## Cross reactivity:
To test potential cross-reactivity, patient samples with various antibodies to autoimmune or infectious disease markers were tested. The results are summarized in the tables below:
QUANTA Flash Sm
| Patient group/Disease | N= | No (% Positive) |
| --- | --- | --- |
| Autoimmune liver disease | 2 | 0 (0.0%) |
| Viral hepatitis | 19 | 0 (0.0%) |
| Scleroderma | 74 | 3 (4.1%) |
| Sjögren’s Syndrome | 5 | 0 (0.0%) |
| Rheumatoid arthritis | 70 | 0 (0.0%) |
| Systemic rheumatic disease, other | 53 | 1 (1.9%) |
| Infectious disease (HIV + syphilis) | 10 | 1 (10%) |
| Total | 233 | 5 (2.1%) |
{10}
Quanta Flash RNP
| Patient group/Disease | N | No (% Positive) |
| --- | --- | --- |
| Autoimmune liver disease | 2 | 0 (0.0%) |
| Scleroderma | 76 | 8 (10.5%) |
| Sjögren’s Syndrome | 6 | 0 (0.0%) |
| Rheumatoid arthritis | 70 | 2 (2.9%) |
| Polymyositis/Dermatomyositis | 14 | 1 (7.1%) |
| Other systemic rheumatic disease, other | 48 | 0 (0.0%) |
| Viral hepatitis | 20 | 0 (0.0%) |
| Infectious disease (HIV + syphilis) | 10 | 1 (10%) |
| Total | 246 | 12 (4.9%) |
f. Assay cut-off:
The cutoff was determined by testing samples from reference subjects, 232 samples for Sm and 255 samples for RNP. The cutoff was established as $>99^{\text{th}}$ percentile of the results obtained on the reference subjects. The cutoff for both Sm and RNP assay was set at:
Negative < 20 CU
Positive ≥ 20 CU
2. Comparison studies:
a. Method comparison with predicate device:
QUANTA Flash Sm
Samples for method comparison analysis include samples from the clinical validation studies (SLE patients, other disease controls). There were 9 discrepant results: 3 SLE, one HIV, one other autoimmune, 2 scleroderma, one ANA human reference sera, and one blood donor. Only samples within the assay measuring range were included. The results are summarized below:
QUANTA Flash Sm
| | Sm ELISA | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| QUANTA Flash Sm CIA | Positive | 35 | 6 | 41 |
| | Negative | 3 | 75 | 78 |
| | Total | 38 | 81 | 119 |
Positive % agreement: 92.1% (35/38) (95% CI: 78.6 – 98.3%)
Negative % agreement 92.6% (35/38) (95% CI: 84.6 – 97.2%)
Total % agreement 92.4% [(75+35)/119] (95% CI: 86.1 – 96.5%)
{11}
12
# QUANTA Flash RNP
Samples for method comparison analysis include samples from the clinical validation studies (SLE patients, other disease controls, and 5 blood donor samples). 14 samples were added to increase the number of samples around the cutoff, 13 of them were diluted to achieve the desired antibody level. There were 14 discrepant results: 9 SLE, one HIV and 3 diluted serum samples. Only samples within the assay measuring range were included. The results are summarized below:
| | RNP ELISA | | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| QUANTA Flash RNP CIA | Positive | 46 | 6 | 52 |
| | Negative | 11 | 104 | 115 |
| | Total | 57 | 110 | 167 |
Positive % agreement: 80.7% (46/57) (95% CI: 68.1 – 90.0%)
Negative % agreement 94.5% (104/110) (95% CI: 88.5 – 98.0%)
Total % agreement 89.8% [(46+104)/173] (95% CI: 84.2 – 94.0%)
b. Matrix comparison:
Not applicable; assay for human serum only
3. Clinical studies:
a. Clinical Sensitivity and Specificity:
A set of samples, none of which were used in establishing the reference range, was used to validate the clinical performance of the Sm and RNP CIA. None of the SLE samples are from drug-induced lupus patients.
# QUANTA Flash Sm
A total of 379 samples were included in the Validation Set for the QUANTA Flash Sm. None of these samples were used in establishing the reference range. This validation set includes: 146 SLE patients, 202 samples from patients with other rheumatic diseases (including 74 with systemic sclerosis and 70 with RA), 2 samples from patients with autoimmune liver disease, 29 infectious disease serum samples (10 HBV, 9 HCV, 5 HIV, 5 syphilis). All samples were run on the QUANTA Flash Sm CIA. The results were analyzed to calculate sensitivity and specificity for SLE and are summarized below
QUANTA Flash Sm
| | Diagnostic Group | | |
| --- | --- | --- | --- |
| | + (SLE) | - (Not SLE) | Total |
| Positive test >20 CU | 21 | 5 | 26 |
| Negative test ≥20 CU | 125 | 228 | 353 |
| Total | 146 | 233 | 379 |
{12}
Sensitivity (95% CI): 14.4% (9.1 - 21.1%)
Specificity (95% CI); 97.9% (95.1- 99.3)
# Quanta Flash RNP
A total of 424 samples were included in the Validation Set for the QUANTA Flash RNP. None of these samples were used in establishing the reference range. This validation set includes: 146 samples from SLE patients, 32 samples from MCTD patients, 48 samples from patients with other rheumatic diseases (not SLE and not MCTD), 2 patients with autoimmune liver disease, 6 patients with Sjögren's syndrome, 76 samples from patients with systemic sclerosis, 70 samples from rheumatoid arthritis patients, 14 samples from patients with polymyositis/dermatomyositis, 30 infectious disease serum samples (10 HBV, 10 HCV, 5 HIV, 5 syphilis). All samples were run on the QUANTA Flash RNP CIA. The results were analyzed to calculate sensitivity and specificity for SLE and are summarized below:
QUANTA Flash RNP in SLE
| | Diagnostic Group | | |
| --- | --- | --- | --- |
| | + (SLE) | - (Not SLE) | Total |
| Positive test >20 CU | 42 | 12 | 54 |
| Negative test ≥20 CU | 104 | 234 | 338 |
| Total | 146 | 246 | 392 |
Sensitivity (95% CI): 28.8% (21.6 - 36.8%)
Specificity (95% CI); 95.1% (91.6-97.5%)
QUANTA Flash RNP in MCTD
| | Diagnostic Group | | |
| --- | --- | --- | --- |
| | + (MCTD) | - (Not SLE/MCTD) | Total |
| Positive test >20 CU | 24 | 12 | 36 |
| Negative test ≥20 CU | 8 | 234 | 242 |
| Total | 32 | 246 | 278 |
Sensitivity (95% CI): 75.0% (56.6 - 88.5%)
Specificity (95% CI); 95.1% (91.6 - 97.5%)
The tables below show the prevalence of antibodies for each clinical subgroup and each analyte, as determined by the QUANTA Flash tests:
Prevalence of Sm determined by QUANTA Flash Sm
| Patient group/Disease | N= | No (% Positive) |
| --- | --- | --- |
| SLE | 146 | 21(14.4%) |
| Autoimmune liver disease | 2 | 0 (0.0%) |
| Viral hepatitis | 19 | 0 (0.0%) |
{13}
| Patient group/Disease | N= | No (% Positive) |
| --- | --- | --- |
| Scleroderma | 74 | 3 (4.1%) |
| Sjögren’s Syndrome | 5 | 0 (0.0%) |
| Rheumatoid arthritis | 70 | 0 (0.0%) |
| Systemic rheumatic disease, other | 53 | 1 (1.9%) |
| Infectious disease (HIV + syphilis) | 10 | 1 (10%) |
| Total | 379 | |
Prevalence of RNP determined by QUANTA Flash RNP
| Patient group/Disease | N | No (% Positive) |
| --- | --- | --- |
| SLE | 146 | 42 (28.8%) |
| MCTD | 32 | 24 (75.0%) |
| Autoimmune liver disease | 2 | 0 (0.0%) |
| Scleroderma | 76 | 8 (10.5%) |
| Sjögren’s Syndrome | 6 | 0 (0.0%) |
| Rheumatoid arthritis | 70 | 2 (2.9%) |
| Polymyositis/Dermatomyositis | 14 | 1 (7.1%) |
| Other systemic rheumatic disease, other | 48 | 0 (0.0%) |
| Viral hepatitis | 20 | 0 (0.0%) |
| Infectious disease (HIV + syphilis) | 10 | 1 (10%) |
| Total | 424 | |
b. Other clinical supportive data (when a. is not applicable):
Not applicable
4. Clinical cut-off:
Same as assay cutoff
5. Expected values/Reference range:
The anti-Sm and anti-RNP antibody levels were analyzed on a panel of 101 apparently healthy blood donors (71 female and 30 males, ages 18 to 55, with an average age and median age of 34 years). With the cutoff of 20 CU, 1% of the samples were positive. The mean concentration was:
QUANTA Flash Sm: 3.6 CU (ranging from <3.3 to 20.5 CU)
QUANTA Flash RNP: 5.7 CU (ranging from <3.5 to 196.8 CU) for
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.