K123576 · Philips Healthcare (Cleveland) · JAK · Jun 7, 2013 · Radiology
Device Facts
Record ID
K123576
Device Name
IMR SOFTWARE APPLICATION
Applicant
Philips Healthcare (Cleveland)
Product Code
JAK · Radiology
Decision Date
Jun 7, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K123576 · Jun 7, 2013
IMR SOFTWARE APPLICATION
Philips Healthcare (Cleveland)
Retrospective clinical image raw data sets
A panel of eight physicians evaluated 110 clinical image raw data sets reconstructed with both filtered backprojection and the IMR software application to compare image quality and confirm diagnostic utility.
Clinical patients undergoing CT scans; Sample Size: 110 clinical image raw data sets
Filtered backprojection (FBP)
Image quality comparison and physician preference
Indications for Use
The IMR reconstruction feature is intended as an alternative to standard reconstruction methods (filtered back projection) for the reconstruction of CT scanner data to produce diagnostic images. The IMR reconstruction feature is designed to reduce image noise, increase high-contrast spatial resolution, and improve low contrast detectability. IMR is designed to reduce dose required for diagnostic CT imaging. Image quality improvements and dose reduction depend on the clinical task, patient size, anatomical location, and clinical practice. IMR images will be used by a trained medical professional for diagnosis of clinical conditions in patients, including pediatrics and adults, who have been prescribed a CT scan as part of their clinical care.
Device Story
IMR Software Application is an iterative reconstruction feature for Philips CT systems. It processes raw CT data to generate diagnostic images, serving as an alternative to standard filtered backprojection (FBP). The software reduces image noise, enhances high-contrast spatial resolution, and improves low-contrast detectability, potentially enabling 60-80% radiation dose reduction. It is operated by trained medical professionals (radiologists/technicians) in clinical settings. The software resides on the CT system hardware. Physicians use the resulting images for clinical diagnosis; the improved image quality and dose reduction capabilities assist in clinical decision-making by providing clearer diagnostic information at lower radiation levels.
Clinical Evidence
Bench testing included phantom-based objective image quality testing (IEC 61223-3-5) and a 4-AFC human observer study for low-contrast detectability. Results showed 60-80% dose reduction, 43-80% LCD improvement, and 70-90% noise reduction. Clinical evaluation involved 110 raw data sets reconstructed with both FBP and IMR, evaluated by eight physicians; results confirmed diagnostic quality, with a majority preference for IMR images over FBP.
Technological Characteristics
Software-based iterative reconstruction for CT systems. Operates on raw CT data. Compliant with ISO 13485:2003 and QSR 21 CFR 820. Designed for integration with Philips CT hardware platforms. No specific hardware materials or energy sources listed as the device is a software application.
Indications for Use
Indicated for pediatric and adult patients prescribed a CT scan for clinical diagnosis. Used by trained medical professionals to reconstruct CT scanner data as an alternative to filtered backprojection to produce diagnostic images with reduced noise, improved high-contrast spatial resolution, and improved low-contrast detectability.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "PHILIPS" in large, bold, black letters. The font is a sans-serif typeface, and the letters are slightly distressed, giving them a textured appearance. The word is centered and takes up most of the frame.
## 510(k) Summarv
## Philips IMR Software Application
The summary of this 510(k) provides safety and effectiveness information submitted in accordance with the requirements of 21 CFR 807.92.
- 1. Submitter:
Philips Medical Systems (Cleveland), Inc. 595 Miner Road Cleveland, OH 44143
Contact:
Christine Anderson Regulatory Affairs Specialist Tel.: (440) 483-7732 Fax: (440) 483-4918
Date of Summary: November 16, 2012
- 2. Device Name and Classification
Philips IMR Software Application
Device Classification:
Device Name:
Computed Tomography X-Ray System. The FDA has classified the Computed Tomography X-Ray System and its accessories as Class II in 21 CFR 892.1750 (Product Code: 90 JAK)
## 3. Predicate Device Information
The Philips IMR Software Application is comparable in type and substantial equivalence to the legally marketed devices currently in commercial distribution, namely:
a. Philips IRT Software Application - K113483
- b. Philips Brilliance Volume CT System K060937
### 4. Device Description
The IMR reconstruction feature is intended as an alternative to standard reconstruction methods (filtered back projection) for the reconstruction of CT scanner data to produce diagnostic images. The IMR reconstruction feature is designed to reduce image noise, increase high-contrast spatial resolution, and improve low contrast detectability. IMR is designed to reduce dose required for diagnostic CT imaging. Image quality improvements and dose reduction depend on the clinical task, patient size, anatomical location, and clinical practice. The
JUN 7 2013
{1}------------------------------------------------
IMR Software Application will reside on any Philips CT System that meets minimum software platform and hardware requirements. IMR enables the user to apply iterative reconstruction techniques to reconstruct raw CT data to generate diagnostic CT images. The use of IMR to reconstruct images may be done prospectively or retrospectively.
- 5. Indications for Use
The IMR reconstruction feature is intended as an alternative to standard reconstruction methods (filtered backprojection) for the reconstruction of CT scanner data to produce diagnostic images. The IMR reconstruction feature is designed to reduce image noise, increase high-contrast spatial resolution, and improve low contrast detectability. IMR is designed to reduce dose required for diagnostic CT imaging. Image quality improvements and dose reduction depend on the clinical task, patient size, anatomical location, and clinical practice. IMR images will be used by a trained medical professional for diagnosis of clinical conditions in patients, including pediatrics and adults, who have been prescribed a CT scan as part of their clinical care.
- 6. Comparison to Predicate
In the opinion of Philips, the IMR Software Application is of a comparable type and substantially equivalent to the Philips Brilliance Volume CT System and the IRT Software Application, the legally marketed devices described in section 3 above. IRT reconstructs images from raw data based on iterative processing using the raw and image data. IMR is identical to IRT except for additional use cases and additional claims for image quality, as measured on a phantom, and dose reduction claims, as evaluated via a human observer study. Therefore, IMR expands on the capabilities of IRT.
Predicate 510(k) Number Brilliance Volume K060937 Philips IRT Software K113483 Application
Cleared June 5, 2006 September 26, 2012
- 7. Safetv
The Philips IMR software application is manufactured in accordance with the Quality System Requlation (QSR) 21 CFR 820 and to International Standards ISO 13485:2003. Potential hazards are identified in a hazard analysis and controlled in the following manner:
Software: Safety is assured by the company procedures that conform to accepted practices, including the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
Instructions for Use are provided with the software application for the safe and effective operation of the application by the user.
{2}------------------------------------------------
## 8. Performance Testing Summary
## Non-clinical Testing
Obiective image quality testing was conducted using phantoms and following the methodology in IEC 61223-3-5, Evaluation and routine testing in medical departments - Parts 3-5: Acceptance test - Imaging performance of computed tomography x-ray equipment for determining noise, CT number uniformity and high contrast spatial resolution. The CT scan data was reconstructed both with filtered backproiection and the IMR software application for comparison. Additionally, a Four Alternative Force Choice (4-AFC) Detection human observer study was conducted with a four Low Contrast pin phantom for evaluating low contrast detectability (LCD). In this study, a cohort of human subjects was required to identify a low contrast pin in a panel of four images. In order to characterize the statistical nature of the detection of a low contrast object in an image with noise, the test was repeated multiple times for each test subject, and multiple test subjects were used. The resulting data confirmed the improved LCD using the IMR application.
IMR may simultaneously enable 60% - 80% lower radiation dose; and 43% -80% low-contrast detectability improvement; and 70% - 83% less image noise, relative to filtered backprojection, as demonstrated through phantom-based tests. Low-contrast detectability was assessed using a reference abdomen protocol, based on a 4-AFC phantom test.
IMR may alternatively enable 1.2x - 1.7x high-contrast spatial-resolution improvement; or 2.5x - 3.6x low-contrast detectability improvement; or up to.90% image noise reduction, relative to filtered backprojection, as demonstrated through phantom-based tests. Lower image noise assessed using a Reference Chest Protocol; Improved high-contrast spatial resolution using Reference Abdomen and Thorax Protocols: Improved low-contrast detectability using a Reference Abdomen Protocol. All metrics tested on phantoms.
## Clinical Image Evaluation
A total of 110 clinical image raw data sets were reconstructed with filtered backprojection and then with the IMR software application to compare image quality. The FBP and IMR images were evaluated image quality by a panel of eight physicians. Resulting data confirmed that the IMR software application provides diagnostic quality images and in majority of the cases, the physicians preferred the IMR images to standard (filter back projection) reconstruction.
## Conclusion
IMR may simultaneously enable:
- · 60% 80% lower radiation dose; and
- · 43% 80% low-contrast detectability improvement; and
- · 70% 83% less image noise.
{3}------------------------------------------------
relative to filtered backprojection, as demonstrated through phantom-based tests (1).
IMR may alternatively enable:
- · 1.2x 1.7x high-contrast spatial-resolution improvement(1): or
- · 2.5x 3.6x low-contrast detectability improvement 13; or
- · Up to 90% image noise reduction(4),
relative to filtered backprojection, as demonstrated through phantom-based tests.
Note : In clinical practice, the use of IMR may reduce CT patient dose depending on the clinical task, patient size, anatomical location, and clinical practice. A consultation with a radiologist and a physicist should be made to determine the appropriate dose to obtain diagnostic image quality for the particular clinical task.
| IMR Image Quality<br>Parameter | Results | Test Conditions |
|------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Dose Reduction:<br>60%-80% lower<br>dose with<br>simultaneous<br>improved image<br>quality [1] | 60% lower radiation<br>dose with a 43 %<br>improvement in low<br>contrast detectability<br>and 83% less image<br>noise | Lower (60%) limit as specified by:<br>IMR settings:<br>• Reference Protocol1: Abdomen<br>• CTDIvol: 4 mGy<br>• Slice thickness: 0.8/0.4mm<br>• Clinical indication: Sharp<br>• Level: L3<br>As compared to:<br>FBP settings<br>• Reference Protocol: Abdomen<br>• CTDIvol: 10mGy,<br>• Slice thickness: 0.8/0.4 mm,<br>• Standard filter: C<br>As tested on the MITA CT IQ<br>Phantom manufactured by Phantom<br>Laboratories, model CCTI83 and low<br>contrast detectability testing using<br>human observers. |
The table below represents the test conditions used to achieve the above performance parameters
<sup>4</sup> Reference protocols refer to the standard factory reference protocols provided on the CT scanner
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the word "PHILIPS" in all capital letters. The text is in a bold, sans-serif font. The letters are black and the background is white.
.
.
| 80% lower radiation<br>dose with a 80%<br>improvement in low<br>contrast detectability<br>and 70% less image<br>noise. | Upper limit (80%) as specified by<br>IMR settings:<br>• Reference protocol: Abdomen<br>• CTDIvol: 2mGy<br>• Slice thickness: 0.8/0.4mm<br>• Clinical indication: Smooth<br>• Level: L3<br>As compared to:<br>FBP settings<br>• Reference Protocol: Abdomen<br>• CTDIvol: 10mGy<br>• Slice thickness: 0.8/0.4 mm,<br>• Standard filter: B |
|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | As tested on the MITA CT IQ<br>Phantom manufactured by Phantom<br>Laboratories, model CCT183 and low<br>contrast detectability testing using<br>human observers. |
:
and the same of the same
{5}------------------------------------------------
# PHILIPS
/
# Philips Medical Systems (Cleveland) Inc.
| High-contrast<br>spatial resolution<br>improvement [2] | 1.2x – 1.7x high-<br>contrast spatial-<br>resolution improvement | High-contrast spatial resolution (x-y)<br>and image noise.<br>Lower limit as specified by:<br>IMR settings Reference Protocol: Abdomen CTDIvol: 4mGy Slice thickness: 1.0/0.5 mm Clinical Indication: Sharp Level: L3. As compared to:<br>FBP Settings Reference Protocol: Abdomen CTDIvol: 4mGy Slice thickness: 1.0/0.5 mm, Standard filter: B Upper limit as specified by:<br>IMR settings Reference Protocol: Thorax/High<br>Res Helical Protocol CTDIvol: 20mGy Slice thickness: 1.0/0.5 mm Matrix: 512 matrix Clinical Indication: Sharp Level: L3 As compared to:<br>FBP Settings Reference Protocol: Abdomen CTDIvol: 20mGy Slice thickness: 1.0/0.5 mm, Standard filter: B Tested on the CatPhan 600 |
|--------------------------------------------------------|------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Low-contrast<br>detectability<br>improvement [3] | 2.5x – 3.6x low-contrast<br>detectability<br>improvement | Lower limit as specified by:<br>IMR Settings:<br>• Reference Protocol: Abdomen<br>• CTDIvol: 10 mGy<br>• Slice thickness: 0.8/0.4 mm,<br>• Clinical Indication: Sharp<br>• Level: L3<br>As compared to:<br>FBP Settings<br>• Reference Protocol: Abdomen<br>• CTDIvol: 10mGy<br>• Slice thickness: 0.8/0.4 mm,<br>• Standard filter: C<br>Upper limit as specified by:<br>IMR Settings:<br>• Reference Protocol: Abdomen<br>• CTDIvol: 4 mGy<br>• Slice thickness: 0.8/0.4 mm,<br>• Clinical Indication: Smooth<br>• Level: L3<br>As compared to:<br>FBP Settings<br>• Reference Protocol: Abdomen<br>• CTDIvol: 4mGy<br>• Slice thickness: 0.8/0.4 mm,<br>Standard filter: B<br>As tested on the MITA CT IQ<br>Phantom manufactured by Phantom<br>Laboratories, model CCT183 and low<br>contrast detectability testing using<br>human observers |
| Noise reduction [4] | 90% image noise<br>reduction | IMR Settings:<br>• Reference Protocol: Chest<br>• CTDIvol: 20 mGy<br>• Slice thickness: 0.8/0.4 mm,<br>• Clinical Indication: Smooth<br>• Level: L3<br>As compared to:<br>FBP Settings:<br>• Reference Protocol: Chest<br>• CTDIvol: 20mGy<br>• Slice thickness: 0.67/0.34 mm,<br>• Standard filter: B<br>Tested on the CatPhan 600 with<br>cylindrical body ring |
:
:
:
{6}------------------------------------------------
# PHILIPS
# Philips Medical Systems (Cleveland) Inc.
-
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "PHILIPS" in a bold, sans-serif font. To the right of the word, there is some handwritten text that reads "K123576 Page 8 of 8". The text is written in a cursive style.
Based on the above considerations, it is Philips' opinion that the results of the verification and validation testing and the results of the risk analysis demonstrates safety and effectiveness of the Philips IMR Software Application and that it is substantially equivalent to the predicate devices documented above.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring. MD 20993-0002
June 7, 2013
Philips Medical System (Cleveland) Inc. % Ms. Christine Anderson Regulatory Affairs Specialist 595 Miner Road CLEVELAND OH 44143
Re: K123576
Trade/Device Name: IMR Software Application Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: Class II Product Code: JAK Dated: May 31, 2013 Received: June 04, 2013
Dear Ms. Anderson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). -You-may=therefore=market-the-device-subject-to-the-general-controls-provisions-of-the-Act-The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting vour device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR.803); good manufacturing practice requirements as set
{9}------------------------------------------------
Page 2 - Ms. Anderson
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Michael D. O'Hara
for
Janine M. Morris Director, Division of Radiological Health Office of in vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
# Indications for Use
510(k) Number (if known):
K123576
Device Name:
IMR Software Application
Indications for Use:
The IMR reconstruction feature is intended as an alternative to standard reconstruction methods (filtered back projection) for the reconstruction of CT scanner data to produce diagnostic images. The IMR reconstruction feature is designed to reduce image noise, increase high-contrast spatial resolution, and improve low contrast detectability. IMR is designed to reduce dose required for diagnostic CT imaging, Image quality improvements and dose reduction depend on the clinical task, patient size, anatomical location, and clinical practice. IMR images will be used by a trained medical professional for diagnosis of clinical conditions in patients, including pediatrics and adults, who have been prescribed a CT scan as part of their clinical care
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE-DO-NOT-WRIFE-BELOW-THIS-LINE-EON-ANOTHER-PAGE-IF-NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
Michael D. O'Hara
(Division Sign Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
K123576 510(k)
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.