← Product Code [NKB](/productcode/NKB) · K123552

# INSTINCT JAVA SYSTEM (K123552)

_Zimmer Spine · NKB · Apr 19, 2013 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K123552

## Device Facts

- **Applicant:** Zimmer Spine
- **Product Code:** [NKB](/productcode/NKB.md)
- **Decision Date:** Apr 19, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Instinct Java spinal fixation system is designed for spinal fixation procedures in skeletally mature patients performed through a posterior approach. The Instinct Java spinal fixation system is indicated for the temporary realignment and stabilization of one or more intervertebral segments from the thoracic spine to the sacrum until bony fusion is obtained. The Instinct Java spinal fixation system is indicated to achieve bony fusion via osteosynthesis at thoracic, lumbar and/or lumbosacral levels of the spine in documented cases of degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, fracture, spinal stenosis, kyphotic or lordotic spinal deformities, scoliosis, tumor and pseudoarthrosis, or for revision of a failed previous fusion.

## Device Story

Instinct Java System is a temporary spinal fixation implant system; used in posterior approach surgeries for skeletally mature patients. Implants include monoaxial/polyaxial pedicle screws, titanium alloy rods, hooks, autostable hooks, and connectors. System provides stabilization until bony fusion occurs; implants may be removed post-fusion at surgeon discretion. Used in OR by surgeons; instruments facilitate implant placement. System is compatible with specific previously cleared Zimmer Spine components (Sequoia, NexLink, Optima, Universal Clamp). Benefits include spinal realignment and stabilization for various degenerative and structural spinal pathologies.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing only, specifically mechanical performance testing per ASTM F1717-12 and ASTM F1798-97 (2008) standards.

## Technological Characteristics

Implants made of titanium alloy and commercially pure titanium. System includes monoaxial/polyaxial pedicle screws, rods, hooks, and connectors. Mechanical testing performed per ASTM F1717-12 (static/dynamic axial compression) and ASTM F1798-97 (2008) (static tightening torque, axial gripping). Non-sterile; requires user-validated cleaning and sterilization.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Predicate Devices

- Zimmer Spine Instinct™ Java® System ([K111301](/device/K111301.md))
- Zimmer Spine Instinct™ Java® System ([K113270](/device/K113270.md))
- Zimmer Spine Sequoia Spinal System ([K082032](/device/K082032.md))
- Medicrea PASS LP Spinal System ([K100297](/device/K100297.md))

## Reference Devices

- NexLink Band & In-Line Rod Connector ([K062505](/device/K062505.md), [K060634](/device/K060634.md), [K052566](/device/K052566.md), [K052247](/device/K052247.md), [K031985](/device/K031985.md))
- Zimmer Spine Universal Clamp 5.5mm Ti implants
- Nexlink 4.0 rods
- Optima 6.0 rods

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## APR 1 9 2013

Image /page/0/Picture/2 description: The image shows the logo for Zimmer Spine. The logo consists of a circle with a stylized letter "Z" inside it. Below the circle, the word "zimmer" is written in a sans-serif font, with a line underneath it. Below "zimmer", the word "spine" is written in a smaller, sans-serif font.

510(k) SUMMARY Zimmer Spine Instinct™ Java® System

K123552 510(k) Number ________________________________________________________________________________________________________________________________________________________________

| Date of Summary Preparation: | November 16, 2012                                                                                                                                 |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                   | Zimmer Spine, Inc.<br>7375 Bush Lake Road<br>Minneapolis, MN 55439                                                                                |
| Company Contact:             | Elsa A. Linke<br>Regulatory Affairs                                                                                                               |
| Manufacturer:                | Zimmer Spine<br>Cité Mondiale<br>23, parvis des Chartrons<br>33080 Bordeaux<br>France                                                             |
| Device Name:                 | Instinct Java System                                                                                                                              |
| Common Name:                 | Spinal Fixation System                                                                                                                            |
| Classification Name:         | Pedicle Screw Spinal System                                                                                                                       |
| Product Code:                | MNI, MNH, NKB                                                                                                                                     |
| Regulation Number:           | 888.3070                                                                                                                                          |
| Device Classification:       | Class III                                                                                                                                         |
| Predicate Devices:           | Zimmer Spine Instinct™ Java® System, K111301<br>K113270<br>Zimmer Spine Sequoia Spinal System, K082032<br>Medicrea PASS LP Spinal System, K100297 |

#### Description of Device:

The Instinct Java System is a temporary implant system used to correct spinal deformity in skeletally mature patients and facilitate the biological process of spinal fusion. This system is intended for non-cervical posterior use in the thoracic, lumbar and sacral areas of the spine. The Instinct Java spinal fixation system is indicated to achieve bony fusion via osteosynthesis at thoracic, lumbar and/or lumbosacral levels of the spine in documented cases of degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, fracture, spinal stenosis,

{1}------------------------------------------------

kyphotic or lordotic spinal deformities, scoliosis, tumor and pseudoarthrosis or for revision of a failed previous fusion.

The system consists of implants and instruments. The implants consist of monoaxial and polvaxial pedicle screws of varying diameters and lengths, blockers, pre-contoured and straight Titanium alloy rods, transverse connectors of varying lengths, hooks, autostable hooks and axial and side-by-side connectors. All implants are made of titanium alloy and one commercially pure titanium component within the transverse connectors.

Re-usable surgical instruments are provided to facilitate placement of the implants.

In addition, the Instinct Java System is compatible with the transverse connectors currently cleared for the market as part of the Sequoia Spinal System, identified in K082032. Furthermore, the Instinct Java titanium alloy rods may be connected to the NexLink Band & In-Line Rod Connector, identified in K062505, K060634, K052566, K052247, K031985. The system may also be used in combination with the Zimmer Spine Universal Clamp 5.5mm Ti implants. Axial and side-by-side connectors may be used with Nexlink 4.0 rods and Optima 6.0 rods.

The implants and instruments are provided non-sterile. Instructions for Use are provided that contain validated cleaning and sterilization instructions for the user.

This system is intended to provide stabilization until a solid spinal fusion develops. The system may then be removed, per the surgeon's discretion. This decision should be made based on the risk/benefit evaluation for each patient.

#### Intended Use:

The Instinct Java spinal fixation system is designed for spinal fixation procedures in skeletally mature patients performed through a posterior approach. The Instinct Java spinal fixation system is indicated for the temporary realignment and stabilization of one or more intervertebral segments from the thoracic spine to the sacrum until bony fusion is obtained.

The Instinct Java spinal fixation system is indicated to achieve bony fusion via osteosynthesis at thoracic. Iumbar and/or lumbosacral levels of the spine in documented cases of degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, fracture, spinal stenosis, kyphotic or lordotic spinal deformities, scoliosis, tumor and pseudoarthrosis, or for revision of a failed previous fusion.

#### Comparison of Technological Characteristics:

The purpose of this 510(k) is to seek clearance for the addition of hooks, autostable hooks, axial and side-by-side connectors, and associated instruments. The Instinct Java Spinal System shares the same technological characteristics as the predicate devices. These characteristics include similar design, materials, range of sizes, technical requirements, and intended use. Determination of substantially equivalent performance characteristics in regard to the predicate devices was confirmed through compliance with ASTM F1717-12 and ASTM F1798-97 (2008). including dynamic and static axial compression, axial gripping capacity, and static tightening torque. In addition, this 510(k) establishes compatibility of the Instinct Java axial and side-by-side connectors in combination with Nex-Link 4.0mm CpTi rods and Optima 6.0 Ti alloy

{2}------------------------------------------------

rods. Furthermore, validated cleaning and sterilization are provided for the nonsterile components of the system.

#### Substantial Equivalence:

The Instinct Java Spinal System is substantially equivalent to the predicate devices in design, materials, function and intended use.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract emblem resembling an eagle or bird-like figure with stylized wings and body.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Zimmer Spine, Incorporated % Mr. Ron Yarbrough Director of Regulatory Affairs 7375 Bush Lake Road Minneapolis, Minnesota 55439

Re: K123552

Trade/Device Name: Instinct™ Java® System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH Dated: March 14, 2013 Received: March 15, 2013

Dear Mr. Yarbrough:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

Letter dated: April 19, 2013

{4}------------------------------------------------

Page 2 - Mr. Ron Yarbrough

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

### Mark N.Mielkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### Indications for Use Statement

510(k) Number (if known): K123552

Device Name: Instinct™ Java® System

Indications for Use:

The Instinct Java spinal fixation system is designed fixation procedures in skeletally mature patients performed through a posterior approach.

The Instinct Java spinal fixation system is indicated for the temporary realignment and stabilization of one or more intervertebral segments from the thoracic spine to the sacrum until bony fusion is obtained.

The Instinct Java spinal fixation system is indicated to achieve bony fusion via osteosynthesis at thoracic. Iumbar and/or lumbosacral levels of the spine in documented cases of degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondvlolisthesis, fracture, spinal stenosis, kyphotic or lordotic spinal deformities, scoliosis, tumor and pseudoarthrosis, or for revision of a failed previous fusion.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH. Office of Device Evaluation (ODE)

# Ronald® → Jean - S

(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K123552

---

**Source:** [https://fda-staging.innolitics.com/device/K123552](https://fda-staging.innolitics.com/device/K123552)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
