K123546 · 3M Company · FRC · Mar 18, 2013 · General Hospital
Device Facts
Record ID
K123546
Device Name
3M ATTEST(TM) AUTO-READER
Applicant
3M Company
Product Code
FRC · General Hospital
Decision Date
Mar 18, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The 3M Attest™ 390 Auto-reader is designed to incubate and automatically read the 3M Attest™ Rapid Readout Biological Indicators for Steam, 1291 and 1292, at 60°C for a final fluorescent result at 1 hour for 1291 and 3 hours for 1292. The 3M Attest™ 390G Auto-reader is designed to incubate and automatically read the 3M Attest™ Rapid Readout Biological Indicator for Ethylene Oxide, 1294, at 37°C for a final fluorescent result at 4 hours.
Device Story
Device functions as incubator and reader for biological indicators (BIs) in central sterilization departments. User activates BI in integrated crusher well, then places in incubation well. Device maintains constant temperature (60°C for 390; 37°C for 390G) to facilitate enzymatic reaction associated with spore growth. Fluorescence detection performed via UV LED and photodiode sensor. Output displayed on LCD panel (time remaining) and via audible alarm/visual '+' symbol upon detection of sterilization failure (fluorescence increase). Negative results indicated by '-' symbol. Optional PC connectivity via ethernet allows status monitoring and PDF documentation of last 100 results. Provides rapid verification of sterilization cycle adequacy, enabling timely release of sterilized items.
Clinical Evidence
Bench testing only. Qualification testing evaluated performance of 3M Attest™ Rapid Readout BIs (1291, 1292, 1294) in 390/390G readers. Endpoints included alignment with 7-day conventional incubation (>97% agreement) and survival/kill time verification. Temperature maintenance verified at 60°C and 37°C respectively. Electrical safety and EMC testing performed per IEC 61010-1, IEC 61010-2-010, and IEC 61326.
Technological Characteristics
Incubator/reader with 10 color-coded wells. Sensing: UV LED and photodiode for fluorescence detection. Temperature control: 60°C (390) or 37°C (37°C). Power: 100-240V AC input, 12V DC internal. Connectivity: Ethernet for PC interface. Standards: IEC 61010-1, IEC 61010-2-010, IEC 61326, FCC Part 15 Subpart B.
Indications for Use
Indicated for use by trained personnel in central sterilization departments to monitor sterilization processes using 3M Attest™ Rapid Readout Biological Indicators (1291, 1292 for steam; 1294 for ethylene oxide).
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
# Premarket Notification (510(k)) Summary
# MAR 1 8 2013
3M
#### 510(k) Number: K123546
# Sponsor Information:
3M Health Care 3M Center, Bldg. 275-5 W-06 St. Paul, MN 55144-1000
| Contact Person: | Suzanne Leung, Ph.D., RAC |
|-----------------|---------------------------|
| | Regulatory Affairs |
| Phone Number: | (651) 575-8052 |
| FAX Number: | (651) 737-5320 |
March 14, 2013 Date of Summary:
# Device Name and Classification:
| Common or Usual Name: | Reader/Incubator |
|-----------------------|-----------------------------------------------------------------------------------|
| Proprietary Name: | 3M Attest™ 390 Auto-reader<br>3M Attest™ 390G Auto-reader |
| Classification Name: | Accessory to Biological Sterilization Process Indicator<br>(21 CFR § 880.2800(a)) |
| Product Class: | Class II |
| Product Code: | Accessory to FRC |
# Predicate Devices:
| Product Name | 'Primary Predicate' | 'Design Predicate' |
|---------------------------|-----------------------------------------------------------------------------------------|---------------------------------------------------------------------|
| AttestTM 390 Auto-reader | Same Intended Use and Indications<br>for Use as AttestTM 290 Auto-reader<br>(K004009) | Similar in design to AttestTM 490<br>Auto-reader (K103277, K121484) |
| AttestTM 390G Auto-reader | Same Intended Use and Indications<br>for Use as AttestTM 290G Auto-<br>reader (K031012) | Similar in design to AttestTM 490<br>Auto-reader (K103277, K121484) |
{1}------------------------------------------------
#### Description of Device:
The Attest™ 390 and 390G Auto-readers are used in conjunction with Attest™ Rapid Readout Biological Indicators by trained personnel in central sterilization departments to monitor sterilization processes. The 390 and 390G Auto-readers, like its predecessors the Attest™ 290, 290G and 490 Auto-readers, have two functions. First, the devices serve to incubate the biological indicator (BI) at the temperature appropriate to the organism. Second, the devices determine whether a potential sterilization failure may have occurred by measuring the intensity of fluorescence produced by an enzymatic reaction associated with spore growth inside the selfcontained biological indicator.
The 390 and 390G Auto-readers include a Crusher Well in which the Attest™ Rapid Readout biological indicators are activated. After activation, the user places the biological indicator into one of the 10 color-coded Incubation Wells. The 390 and 390G Auto-readers have an incubation block controlled to the appropriate temperature for their respective biological indicators. The Auto-readers contain a LCD panel that shows the time remaining for incubation. Fluorescence detection at each incubation well is accomplished by the use of a UV LED and a photo diode sensor. As soon as a positive result is detected, a '+' sign is shown as an indication of sterilization cycle failure, accompanied by an audible alarm. At the end of the specified incubation time, if a negative result is detected, a '-' sign is shown on the LCD panel as indication of an adequate sterilization cycle. When the test is complete, as indicated by the (+) or (-) symbol on the Auto-reader LCD panel under the incubating BI, the customer records the result and disposes of the biological indicator.
#### 390 Auto-reader
The 3M Attest™ 390 Auto-reader is an accessory to the 3M Attest™ Rapid Readout selfcontained biological indicators 1291 and 1292 that are designed to incubate at 60°C. The 390 Auto-reader incubates and automatically reads the Attest™ 1291 and 1292 biological indicators for a fluorescent result within 1 hour for 1291 and 3 hours for 1292.
The 390 Auto-reader can be configured for three different configurations of biological indicators. The first configuration allows the 390 Auto-reader to read 1291 biological indicators in all 10 wells. The second configuration allows the 390 Auto-reader to read 1292 biological indicators in all 10 ten wells. The third configuration allows the 390 Autoreader to read 1291 biological indicators in the 5 wells to the left and 1292 biological indicators in the 5 wells to the right. The user inserts the biological indicator into the well with the matching color to the biological indicator's cap (blue cap 1291 into blue well, brown cap 1292 goes into brown well).
{2}------------------------------------------------
### 390G Auto-reader
The 3M Attest™ 390G Auto-reader is an accessory to the 3M Attest™ Rapid Readout selfcontained biological indicator 1294 that is designed to incubate at 37°C. The 390G Autoreader incubates and automatically reads the Attest™ 1294 biological indicator for a fluorescent result within 4 hours. The incubation wells are color coded green to match the cap of the 1294 biological indicator.
### Indications for Use:
The 3M Attest™ 390 Auto-reader is designed to incubate and automatically read the 3M Attest™ Rapid Readout Biological Indicators for Steam, 1291 and 1292, at 60°C for a final fluorescent result at 1 hour for 1291 and 3 hours for 1292.
The 3M Attest™ 390G Auto-reader is designed to incubate and automatically read the 3M Attest™ Rapid Readout Biological Indicator for Ethylene Oxide, 1294, at 37°C for a final fluorescent result at 4 hours.
# Comparison of the Attest™ 390 and 390G Auto-readers to their Predicate Devices:
#### Intended Use and Indications for Use
All of the Auto-readers share the same Intended Use. These devices incubate the biological indicators (BIs) to determine a positive or negative result based on a fluorescence signal. An increase in fluorescence signal, created by spore germination, indicates a failed sterilization cycle.
The Indications for Use for the 390 Auto-reader is the same as the predicate 290 Autoreader, to incubate and automatically read Attest™ 1291 and 1292 biological indicators. The Indications for Use for the 390G Auto-reader is the same as the predicate 290G Autoreader, to incubate and automatically read the Attest™ 1294 biological indicator.
# Incubation Temperature
The 390 Auto-reader reaches an incubation temperature of 60+/-2°C within 30 minutes of power up at room temperature and maintains its temperature throughout the incubation period at labeled environmental conditions (16-40°C, 20-80% RH, non-condensing). The predicate device, the Attest 290 Auto-reader, has the same incubation temperature specifications.
{3}------------------------------------------------
The 390G Auto-reader reaches an incubation temperature of 37+/-2°C within 30 minutes of power up at room temperature and maintains its temperature throughout the incubation period at labeled environmental conditions (16-29°C, 20-80% RH, non-condensing). The predicate device, the Attest 290G Auto-reader, has the same incubation temperature specifications.
## Basis of Rapid Readout
The Attest 390/390G Auto-readers and their predicate devices, the Attest 290/290G and 490 Auto-readers, all have Rapid Readout capabilities due to photodiodes that detect fluorescence produced by enzymatic activity that results from growing biological indicator organisms.
### Voltage Range
All Auto-readers share the same voltage range, operating at 100-240 Volts AC with a 12 Volt DC conversion for internal circuitry.
### Visual Display and Audible Alert Tone
The Attest 390/390G Auto-readers contain an update to the visual display from the 290/290G Auto-readers. Like the 'design predicate' the 490 Auto-reader, the 390/390G Auto-readers contain a LCD panel, which provides feedback to the user for each BI inserted into an incubator/reader well and displays a constant readout of time remaining for each BI. In contrast, the predicate 290/290G Auto-readers contain LEDs located at each well to indicate passed, failed, or in-process status to the user for each BI. All Auto-readers contain an audible alert tone that beeps to notify the user when a positive result has been detected.
### PC Connectivity
The 390/390G Auto-readers can be connected to a PC via an ethernet connection to allow the user to view current status of BIs, view the last 100 BI results, and save documentation as a PDF file if desired. This is an option that also exists on the 490 Auto-reader but that the 290/290G Auto-readers do not have.
# Summary of Predicate Comparisons
The 390 and 390G Auto-readers share the same Intended Use. Indications for Use, and fundamental scientific technology such as Rapid Readout capability and energy source as the predicate devices the 290 and 290G Auto-readers. Additionally, the 390 and 390G Auto-readers share updated design features with the 'design predicate' the 490 Auto-reader by providing a LCD panel and PC connectivity.
{4}------------------------------------------------
# Evaluation Data for Determining Substantial Equivalence:
### Biological Indicator Qualification
The validation of the 390 and 390G Auto-readers for use with their respective biological indicators was conducted following the FDA Guidance for Industry and FDA Staff, Biological Indicator (BI) Premarket Notification [510(k)] Submissions; October 4, 2007. Multiple lots of 3M Attest™ Rapid Readout biological indicators were evaluated for performance when used with the 3M Attest™ 390 and 390G Auto-readers. A Summary of biological indicator testing is shown below.
| Auto-<br>reader | Testing | Acceptance Criteria | Result |
|-----------------|------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| 390 | Qualification testing<br>with Attest™<br>1291 BI | Meets FDA's requirements for Reduced<br>Incubation Time with > 97% alignment with<br>the conventional incubation time of 7 days<br>for the following readout time<br>• Fluorescent result in 1 hour<br>Survival/Kill: All BIs survive after exposure<br>at the survival time and all BIs are killed<br>after exposure at the kill time | Pass |
| 390 | Qualification testing<br>with Attest™<br>1292 BI | Meets FDA's requirements for Reduced<br>Incubation Time with > 97% alignment with<br>the conventional incubation time of 7 days<br>for the following readout time<br>• Fluorescent result in 3 hours<br>Survival/Kill: All BIs survive after exposure<br>at the survival time and all BIs are killed<br>after exposure at the kill time | Pass |
| 390 | Auto-reader<br>Maintenance of<br>Incubation<br>Temperature | Maintain 60+/- 2ºC over the specified<br>incubation period | Pass |
| 390G | Qualification testing<br>with Attest™<br>1294 BI | Meets FDA's requirements for Reduced<br>Incubation Time with > 97% alignment with<br>the conventional incubation time of 7 days<br>for the following readout time<br>• Fluorescent result in 4 hours<br>Survival/Kill: All BIs survive after exposure<br>at the survival time and all BIs are killed<br>after exposure at the kill time | Pass |
| 390G | Auto-reader<br>Maintenance of<br>Incubation<br>Temperature | Maintain 37+/- 2ºC over the specified<br>incubation period | Pass |
Summary of Biological Indicator Qualification
The results of these evaluations showed that the new Attest™ 390 and 390G Auto-readers are qualified for use with the Attest™ Rapid Readout biological indicators.
{5}------------------------------------------------
# Electrical Safety and EMC Testing
The Attest™ 390 and 390G Auto-readers were also tested for safety by Underwriters Laboratory to verify compliance to:
- IEC 61010-1 (2001) Second Edition; Safety requirements for electrical equipment . for measurement, control, and laboratory use – Part 1: General requirements,
- IEC 61010-2-010 (2003) Second Edition; Safety requirements for electrical . equipment for measurement, control, and laboratory use - Part 2-010: Particular requirements for laboratory equipment for the heating of materials
In addition, the Attest™ 390 and 390G Auto-readers have been tested by a certified Testing Laboratory to verify electromagnetic compatibility per:
- USA Title 47, Code of Federal Regulations (2009) for: .
- O Radiated Emissions (FCC Part 15, Subpart B, Class A)
- Conducted Emissions (FCC Part 15, Subpart B, Class A), and
- IEC 61326: Electrical Equipment for Measurement, Control and Laboratory Use-● EMC Requirements.
# Substantial Equivalence Conclusions:
The Attest 390/390G Auto-readers are substantially equivalent to the Attest 290/290G Autoreaders, which share the same Intended Use and same Indications for Use. Both systems use the same fundamental scientific technology to incubate biological indicators and detect a failure in a sterilization cycle. They both provide the results to the user through a visual display and an audible alarm. Data has been generated that qualifies the 390/390G Autoreaders for use with the Attest Rapid Readout biological indicators and data has been generated that demonstrates compliance to electrical safety and EMC standards. The minor updates in design features provided with the 390/390G Auto-readers do not present any new questions of safety and effectiveness with the devices.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the Department of Health and Human Services (HHS) of the United States. The logo features the department's name arranged in a circular fashion around a stylized symbol. The symbol consists of three curved shapes that resemble a stylized human form or a wave-like design, and it is positioned to the right of the text.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### March 18, 2013
Suzanne Leung, Ph.D., RAC Regulatory Affairs 3M Health Care 3M Center, Building 275-5W-06 ST. PAUL MN 55144-1000
Re: K123546
Trade/Device Name: 3M Attest™ 390 Auto-reader and 3M Attest™ 390G Auto-reader Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: FRC Dated: February 6, 2013 Received: February 8. 2013
Dear Dr. Leung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
for
Ruth
urohitu.s.
++
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and
Radiological Health
Enclosure
{8}------------------------------------------------
### Indications for Use Statement
510(k) Number: K123546
Device Name: 3M Attest™ 390 Auto-reader and 3M Attest™ 390G Auto-reader
Indications for Use:
The 3M Attest™ 390 Auto-reader is designed to incubate and automatically read the 3M Attest™ Rapid Readout Biological Indicators for Steam, 1291 and 1292, at 60°C for a final fluorescent result at 1 hour for 1291 and 3 hours for 1292.
The 3M Attest™ 390G Auto-reader is designed to incubate and automatically read the 3M Attest™ Rapid Readout Biological Indicator for Ethylene Oxide, 1294, at 37°C for a final fluorescent result at 4 hours.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use ਮ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth-F.-Claverie
2013:03:14 17:18:55
-04'00"
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital Division of Anesu Dental Devices
**510(k) Number:**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.