The RapidCross™ PTA Rapid Exchange Balloon Dilatation Catheter is intended to dilate stenosis in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
Device Story
RapidCross is a rapid exchange (RX) coaxial PTA balloon dilatation catheter compatible with 0.014" guidewires. It features a distally mounted semi-compliant inflatable balloon with an atraumatic, tapered tip and a lubricious coating. The device is used by physicians in clinical settings to treat peripheral artery stenoses and arteriovenous dialysis fistulae. The physician inserts the catheter over a guidewire, positioning the balloon across the lesion using radiopaque markers. The balloon is inflated/deflated via a manifold using a contrast/saline mixture. Proper positioning is facilitated by proximal depth marks on the shaft. By dilating the vessel, the device restores blood flow, potentially improving patient outcomes in peripheral vascular disease.
Clinical Evidence
Bench testing only. Testing included crossing profile, balloon burst strength, compliance, fatigue, inflation/deflation time, bond tensile strength, kink resistance, tracking, pushability, and radiopacity. Biocompatibility testing was performed per ISO 10993-1 for short-duration blood contact (<24 hours), including cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, hemolysis, pyrogenicity, complement activation, and thromboresistance. All tests met acceptance criteria.
Technological Characteristics
Rapid exchange (RX) coaxial catheter; 0.014" guidewire compatibility; 4F sheath compatibility. Features semi-compliant inflatable balloon, atraumatic tapered tip, and lubricious coating. Radiopaque marker bands for positioning. Available in 2-4 mm diameters and 20-210 mm lengths. Sterilization method is consistent with predicate devices.
Indications for Use
Indicated for patients requiring dilation of stenoses in iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
K1235244
P. 1 of 3
Image /page/0/Picture/1 description: The image shows the logo for ev3, a medical device company. The logo consists of the letters "ev3" in a stylized font, with three black circles above the letters. To the right of the logo is the tagline "Your endovascular company."
## 510(k) Summary
バ
FEB 1 2 2013
Page 1 of 3
| | FEB 1 2 20<br>RapidCross™ PTA Rapid Exchange Balloon Dilatation Catheter |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Summary | This 510(k) summary information is submitted in accordance<br>with the requirements of 21 C.F.R §807.92. |
| Applicant | ev3 Inc. |
| Submitter | ev3 Inc.<br>3033 Campus Drive<br>Plymouth, MN 55441-2651 |
| Contact Person | Laura J. Lind |
| Date Prepared | February 6, 2013 |
| Device Trade Name | RapidCross™ PTA Rapid Exchange Balloon Dilatation Catheter |
| Device Common Name | PTA Dilatation Catheter |
| Classification Name | Catheter, Angioplasty, Peripheral, Transluminal<br>(21 CFR 870.1250, Product Codes LIT, DQY) |
| Classification Panel | Cardiovascular |
| Predicate Devices | NanoCross™ .014 OTW PTA Dilatation Catheter (K082854,<br>K090849), PowerCross™ .018 OTW PTA Dilatation Catheter<br>(K093286), Sterling™ Monorail™ PTA Balloon Dilatation<br>Catheter (K053118), and Advance® 14LP Low Profile PTA<br>Balloon Dilatation Catheters (K090822). |
| Intended use | The RapidCross™ PTA Rapid Exchange Balloon Dilatation<br>Catheter is intended to dilate stenosis in the iliac, femoral, ilio-<br>femoral, popliteal, infra-popliteal, and renal arteries, and for the<br>treatment of obstructive lesions of native or synthetic<br>arteriovenous dialysis fistulae. |
| Device Description | The RapidCross PTA Rapid Exchange Balloon Dilatation<br>Catheter (RapidCross catheter) is a rapid exchange (RX) coaxial<br>catheter compatible with 0.014" guidewires, with a distally<br>mounted semi-compliant inflatable balloon and an atraumatic,<br>tapered tip. The distal portion of the catheter has a lubricious<br>coating. The manifold includes an inflation lumen, which is used<br>to inflate and deflate the dilatation balloon with a mixture of |
| | Tel: 763-398-7000<br>Fax: 763-591-3248 |
ev3 Corporate World Headquarters / Peripheral Vascular - 3053 Campus Dive, Flynnesh, MN 55441, USA - Ph - 1763 398 7050
ev3 Neurovascular • 9775 Tolodo V/ay, Irvine, CA 92618 • Ph • 1 949 837 3700
.
ev3 Europe / International Headquarters • 108/108 ₪e La Reètie. 75008 Paris, Firaco • Ph +33 156 89 59 1 3
{1}------------------------------------------------
contrast medium and saline solution. The balloon has two radiopaque markers for positioning the balloon relative to the stenosis. The radiopaque marker bands indicate the dilating or working section of the balloon. On the 150 mm and 210 mm devices, two additional marker bands denote the middle of the balloon body. A guidewire lumen starts at a guidewire port located 35 cm from the catheter tip and extends to the distal tip. The 90 cm useable length devices have proximal depth marks printed on the proximal shaft at lengths of 55 cm and 65 cm from the distal tip while the 170 cm useable length devices have depth marks at 90 cm and 100 cm to serve as a reference during catheter insertion.
The RapidCross catheter is available in balloon sizes ranging from 2 mm to 4 mm in diameter and 20 mm to 210 mm in length; and, all sizes are compatible with 4 F sheaths.
Bench testing and biocompatibility testing were performed to support a determination of substantial equivalence. Results from this testing provide assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use.
Using internal Risk Analysis procedures, the following tests were performed:
| • Crossing Profile | • Radiopacity |
|--------------------------------|------------------------------|
| • Balloon Burst Strength | • Presence of Coating |
| • Balloon Compliance | • Coating Durability |
| • Balloon OD | • Particle Generation |
| • Inflation/Deflation Time | • Pushability |
| • Balloon Fatigue | • Support Wire Securement |
| • Bond Tensile Strength | • Tip ID / RX Port ID |
| • Kink | • Tip / Lesion Entry Profile |
| • Device Tracking | • Re-Insertion Force |
| • Insertion Force | • Catheter Working Length |
| • Balloon Pull-back Force | • RX Port Length |
| • Repeat Inflations (In Stent) | • RX Port OD |
| • Torque Strength | |
The device was tested for biocompatibility per ISO 10993-1 for <24 hours). The testing 1, intracutaneous reactivity, , pyrogen, complement tial thromboplastin time, and platelet/leukocyte count.
The RapidCross catheter met all acceptance criteria for the
| short duration contact with blood |
|--------------------------------------|
| included cytotoxicity, sensitization |
| acute systemic toxicity, hemolysis |
| activation, thromboresistance, part |
| platelet/leukocyte count |
## Performance data
{2}------------------------------------------------
bench testing with results similar to the predicates. Based on the bench test results, comparison to legally marketed predicates, and non-clinical test results, the RapidCross catheter is determined to perform as safely and effectively as the predicates for its intended use.
The RapidCross PTA Rapid Exchange Balloon Dilatation Catheter has the following similarities to the predicate devices:
- Same intended use (all predicates) ●
- Same indications for use (NanoCross PTA Dilatation . Catheter)
- Similar fundamental scientific technology (all predicates) .
- Similar operating principle (all predicates) .
- Similar technological characteristics .
- Same sterility assurance level and sterilization method . (NanoCross)
All devices are compatible with 0.014" wires and 4F sheaths. All devices have similar construction and principles of operation. All devices are used by the physician in a similar manner typical of PTA balloon catheters.
The RapidCross catheter and the predicates have the same intended use - all devices are intended to treat peripheral arteries. All devices are intended to treat the same target population. The manner in accessing and treating lesions is similar for the devices.
The RapidCross catheter indications for use are the same as the ev3 NanoCross PTA Dilation Catheter indications for use. The minimal differences between RapidCross and indications for use of the other predicates do not raise new safety and effectiveness questions.
Based on the intended use, technological characteristics, and results from safety and performance testing, the RapidCross PTA Rapid Exchange Balloon Dilatation Catheter is considered substantially equivalent to the legally marketed predicate devices NanoCross .014 OTW PTA Balloon Dilatation Cathetat (K082854, K090849), PowerCross PTA Dilatation Catheter (K093286) Sterling Monorail PTA Balloon Dilatation Catheter (K053118), and Advanced LP Dilatation Catheter (K090822).
## Summary of Substantial Equivalence
Conclusion
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a stylized emblem. The emblem is a symbol of an eagle with outstretched wings, representing protection and care, which is a common symbol used by the HHS.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -- WO66-G609 Silver Spring, MD 20993-0002
FEB 1 2 2013
ev3 Inc. c/o Mr. Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 55313
Re: K123544
Trade/Device Name: RapidCross PTA Rapid Exchange Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT, DQY Dated: January 23, 2013 Received: January 24, 2013
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Matthew G. Hillebrenner
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
## Indications for Use Statement
510(k) Number (if known): K123544
Device Name: RapidCross™ PTA Rapid Exchange Balloon Dilatation Catheter
Indications for Use:
The RapidCross™ PTA Rapid Exchange Balloon Dilatation Catheter is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M.A. Hillel
(Division Sign-Off) Division of Cardiovascular Devices
K123544 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.