← Product Code [OCY](/productcode/OCY) · K123439

# NEO ERCP GUIDEWIRE (K123439)

_Neometrics, Inc. · OCY · Jan 3, 2013 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K123439

## Device Facts

- **Applicant:** Neometrics, Inc.
- **Product Code:** [OCY](/productcode/OCY.md)
- **Decision Date:** Jan 3, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The Neo ERCP Guidewire is intended for use in selective cannulation of the biliary ducts including the common bile, pancreatic, cystic, right and left hepatic ducts and to aid in the placement of diagnostic and therapeutic devices during endoscopic procedures.

## Device Story

Neo ERCP Guidewire is a medical device used during endoscopic procedures to facilitate selective cannulation of biliary and pancreatic ducts; aids in placement of diagnostic/therapeutic instruments. Device consists of nickel-titanium (Nitinol) alloy core wire, proximal polymer jacket, and 10cm radiopaque distal tip; available in .025" (Phantom) and .035" (Palomino) diameters and 260cm/480cm lengths; features hydrophilic coating. Operated by physicians in clinical settings. Radiopaque tip allows visualization under fluoroscopy to guide placement. Device provides mechanical support for navigating ducts; benefits patient by enabling access for diagnostic/therapeutic interventions.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (cytotoxicity, sensitization, irritation, systemic injection), sterility, packaging integrity, and mechanical performance (reverse bend, tensile strength, fracture resistance, corrosion resistance, dielectric insulation, and simulated use).

## Technological Characteristics

Nitinol alloy core wire; proximal polymer jacket; radiopaque distal polymer tip (10cm); hydrophilic coating. Diameters: 0.025" and 0.035". Lengths: 260cm and 480cm. Standards: ISO 11070 (reverse bend, tensile, fracture, corrosion, radiodetectability), ASTM F 2096-04 and F 88-09 (packaging).

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- JagWire ([K922302](/device/K922302.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

.

# Traditional 510(k)

K1234

## Section 5.0: 510(k) Summary

## 5.1 Administrative Information

# JAN 0 3 2013

| Name:           | NeoMetrics, Inc.                                       |
|-----------------|--------------------------------------------------------|
| Address:        | 2605 Fernbrook Lane North, Suite<br>Plymouth, MN 55447 |
| Contact Person: | Gene Champeau                                          |
| Title:          | President                                              |
| Phone:          | (763)-559-4440                                         |
| Fax:            | (763) 559-7676                                         |

. 2135342 FDA Registration Number: December 17, 2012 Date:

#### 5.2 Device Information

| Name of Device:      | Neo ERCP Guidewire                                        |
|----------------------|-----------------------------------------------------------|
| Common Name:         | Endoscopic Guidewire                                      |
| Classification Name: | Endoscopic Guidewire, Gastroenterology – Urology (876.15) |
| Product Code:        | OCY                                                       |

#### 5.3 Predicate Device Information

The following commercially available guidewire is a predicate device for Neo ERCP Guidewire.

| 510(k) Number | Trade or Model Name | Manufacturer            |
|---------------|---------------------|-------------------------|
| K922302       | JagWire             | Boston Scientific Corp. |

### 5.4 Device Description

The Neo ERCP Guidewire is constructed of a core wire of nickel titanium based alloy, a proximal polymer jacket and a radiopaque distal tip. It also features a hydrophilic coating and either straight or angled distal tip configurations. The distal end of the guidewire has a radiopaque distal tip (10cm). The wire family has two models, the Phantom refers to .025" OD models and the Palomino refers to .035" OD models. Both the Phantom and Palomino come in 260cm and 480cm lengths.

#### 5.5 Intended Use

The Neo ERCP Guidewire is intended for use in selective cannulation of the biliary ducts including the common bile, pancreatic, cystic, right and left hepatic ducts and to aid in the placement of diagnostic and therapeutic devices during endoscopic procedures.

21

{1}------------------------------------------------

K123 439 Page 2 of 3

## Traditional 510(k)

## 5.6 Differences in the Intended Use Statements

Differences in the indications for use statements between Neo ERCP Guidewire and the JagWire IFU are not significant. The JagWire IFU refers to bronchoscopic procedures. An endoscope is the tool used in bronchoscopic and endoscopic procedures. The Neo ERCP Guidewire is not indicated for use in bronchoscopic procedures.

### 5.7 Technological Characteristics

·

A comparison of the characteristics of the proposed device to the predicate device shows the proposed device to have the following same or similar technological characteristics to the device which has received 510(k) clearance:

- Same intended use; .
- Same operating principle; ●
- Same shelf life and sterilization process; .
- Similarities in Design, Material Types, and Technology include:
	- Nominal diameters: 0.025" and 0.035" 0
	- Similar Lengths: 260 & 480 cm 0
	- Nitinol alloy core wires 0
	- Distal radiopaque polymer tip O
	- Lubricious coatings O

To ascertain similarity, the following performance testing was conducted on both the proposed device and the predicate devices:

- FDA Coronary and Cerebrovascular Guidewire Guidance January 1995 sections 1, 2, . 3a, 3e, 3f and 3g
- Sterile Package Integrity per ASTM F 2096-04 .
- Sterile Package Integrity per ASTM F 88-09 ●
- Reverse Bend (Fatigue Failure) per ISO 11070 .
- Tensile testing per ISO 11070 �
- Fracture Resistance per ISO 11070 ●
- Corrosion Resistance per ISO 11070 .
- Radiodetectability per ISO 11070 ●
- Rail Support per NeoMetrics' test method TM-3002 .

Due to the destructive nature of some of the tests and the availability of predicate devices not all testing was conducted on each predicate device.

{2}------------------------------------------------

K123 439 Page 3 of 3

Traditional 510(k)

## 5.8 Summary of Non-Clinical Testing

Non-clinical testing of the Neo ERCP Guidewire includes bench testing, biocompatibility testing, shelf-life testing, packaging testing and sterilization evaluation. The following tests were conducted:

- . Sterility
- Biocompatibility . .
	- Cytotoxicity 0
	- Sensitization ಂ
	- Intracutaneous Irritation O
	- Acute Systemic Injection 0
- Packaging Integrity ●
- Device Compatibility ●
- Reverse Bend ●
- Tensile Strength
- Fracture Resistance .
- Corrosion Resistance .
- Dielectric Insulation ●
- Simulated Use .
- Shelf Life

Results of this testing demonstrate that the guidewire design meets the product specifications and intended uses.

### 5.9 Substantial Equivalence Conclusion

The Neo ERCP Guidewire described in this 510(k) is substantially equivalent to the device listed in section 5.3. The intended use, design, material types, technology, and performance of the Neo ERCP Guidewire are equivalent to the predicate device. There are no differences between devices which would raise issues of safety or effectiveness.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD. 20993-0002

January 3, 2013

Neometrics, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K123439

Trade/Device Name: Neo ERCP Guidewire Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: OCY Dated: December 7, 2012 Received: December 28, 2012

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{4}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Herbert P. Lerner

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Section 4: Indications for Use Statement

# Indications for Use

510(k) Number (if known): K (23439

Device Name: Neo ERCP Guidewire

Indications for Use:

The Neo ERCP Guidewire is intended for use in selective cannulation of the biliary ducts including the common bile, pancreatic, cystic, right and left hepatic ducts and to aid in the placement of diagnostic and therapeutic devices during endoscopic procedures.

Prescription Use: X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use: (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert P. Lerner

(Division Sign-Off) Division of Reproductive, Gastro-Renal, and Urological Devices

510(k) Number K123439

---

**Source:** [https://fda-staging.innolitics.com/device/K123439](https://fda-staging.innolitics.com/device/K123439)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).