ILLUMIGENE MYCOPLASAMA DNA AMPLIFICATION ASSAY, AND ILLUMIGENE MYCOPLASMA EXTERNAL CONTROLS KIT
K123423 · Meridian Bioscience, Inc. · OZX · Jun 5, 2013 · Microbiology
Device Facts
Record ID
K123423
Device Name
ILLUMIGENE MYCOPLASAMA DNA AMPLIFICATION ASSAY, AND ILLUMIGENE MYCOPLASMA EXTERNAL CONTROLS KIT
Applicant
Meridian Bioscience, Inc.
Product Code
OZX · Microbiology
Decision Date
Jun 5, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3980
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K123423 · Jun 5, 2013
ILLUMIGENE MYCOPLASAMA DNA AMPLIFICATION ASSAY, AND ILLUMIGENE MYCOPLASMA EXTERNAL CONTROLS KIT
Meridian Bioscience, Inc.
Leftover de-identified clinical specimens from routine laboratory testing
Retrospective clinical specimens were used to evaluate the clinical sensitivity and specificity of the illumigene Mycoplasma DNA Amplification Assay compared to a composite reference method.
2012 Clinical Evaluation; Retrospective and prospective clinical performance evaluation; Study Period: 2012
Patients suspected of Mycoplasma pneumoniae infection; Sample Size: 219 retrospective specimens (out of 334 total); Number of Sites: 5
Composite reference method (culture and validated real-time PCR with bi-directional sequencing)
Clinical sensitivity and specificity
Indications for Use
The illumigene Mycoplasma DNA amplification assay, performed on the illumipro-10™, is a qualitative in vitro diagnostic test for the direct detection of DNA from Mycoplasma pneumoniae in human throat and nasopharyngeal swabs obtained from patients suspected of having Mycoplasma pneumoniae infection. The illumigene Mycoplasma assay utilizes loop-mediated isothermal DNA amplification (LAMP) technology to detect Mycoplasma pneumoniae by targeting a segment of the Mycoplasma pneumoniae genome. Results from the illumigene Mycoplasma DNA amplification assay should be used in conjunction with clinical presentation, other laboratory findings, and epidemiological risk factors as an aid in the diagnosis of Mycoplasma infection and should not be used as the sole basis for treatment or other patient management. Positive results do not rule out co-infection with other organisms and negative results in persons with respiratory tract infections may be due to pathogens not detected by this assay. Lower respiratory tract infections due to M. pneumoniae may not be detected by this assay. If lower respiratory tract infection due to M. pneumoniae is suspected, additional laboratory testing using methods other than the illumigene Mycoplasma DNA Amplification Assay may be necessary. illumigene Mycoplasma is intended for use in hospital, reference or state laboratory settings. The device is not intended for point-of-care use.
Device Story
System comprises illumigene Mycoplasma DNA Amplification Assay kit and illumipro-10 Automated Isothermal Amplification and Detection System. Input: human throat or nasopharyngeal swab specimens. Process: manual DNA extraction/purification using Qiagen QIAamp DSP DNA Mini Kit; heat treatment; isothermal amplification (LAMP) in illumipro-10. Detection: illumipro-10 monitors light transmission changes caused by magnesium pyrophosphate precipitate formation during amplification. Output: qualitative results (POSITIVE, NEGATIVE, INVALID) reported by instrument. Used in hospital/reference labs by trained personnel. Results aid clinical diagnosis alongside patient presentation and other findings; not for sole management decisions.
Clinical Evidence
Clinical study (Feb-Sept 2012) evaluated 334 throat and nasopharyngeal swabs (prospective and retrospective) against a composite reference method (culture and real-time PCR with sequencing). Prospective nasopharyngeal sensitivity 100% (95% CI: 51.0-100%), specificity 100% (95% CI: 92.6-100%). Prospective throat sensitivity 100% (95% CI: 67.6-100%), specificity 100% (95% CI: 91.8-100%). Retrospective nasopharyngeal PPA 94.4%, NPA 95.6%. Retrospective throat PPA 84.6%, NPA 98.5%.
Technological Characteristics
Uses loop-mediated isothermal amplification (LAMP) technology. Instrumentation: illumipro-10 Automated Isothermal Amplification and Detection System. Detection principle: visible light transmission monitoring of magnesium pyrophosphate precipitate. Reagents: illumigene Assay Control Reagent, Reaction Buffer, Test Device. Connectivity: standalone instrument. Software: embedded firmware for signal processing and result interpretation.
Indications for Use
Indicated for qualitative detection of Mycoplasma pneumoniae DNA in throat and nasopharyngeal swabs from patients suspected of infection. For professional use in hospital, reference, or state laboratories. Not for point-of-care use.
Regulatory Classification
Identification
A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.
Special Controls
*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.
Predicate Devices
FilmArray® Respiratory Panel (RP) System (K120267)
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k123423
B. Purpose for Submission:
To obtain substantial equivalence determination for the illumigene® Mycoplasma DNA Amplification Assay
C. Measurand:
Specific DNA sequence from Mycoplasma pneumoniae
D. Type of Test:
Molecular assay based on loop-mediated isothermal amplification technology (LAMP)
E. Applicant:
Meridian Bioscience, Inc.
F. Proprietary and Established Names:
illumigene® Mycoplasma DNA Amplification Assay
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3980 – Respiratory Viral Panel Multiplex Nucleic Acid Assay
21 CFR 862.2570 – Instrumentation for Clinical Multiplex Test Systems
2. Classification:
Class II
3. Product code:
OZX – Mycoplasma pneumoniae DNA Assay System
1
{1}
OOI – Real Time Nucleic Acid Amplification System
4. Panel:
83 Microbiology
H. Intended Use:
1. Intended use(s):
The illumigene Mycoplasma DNA amplification assay, performed on the illumipro-10™, is a qualitative in vitro diagnostic test for the direct detection of DNA from Mycoplasma pneumoniae in human throat and nasopharyngeal swabs obtained from patients suspected of having Mycoplasma pneumoniae infection.
The illumigene Mycoplasma assay utilizes loop-mediated isothermal DNA amplification (LAMP) technology to detect Mycoplasma pneumoniae by targeting a segment of the Mycoplasma pneumoniae genome.
Results from the illumigene Mycoplasma DNA amplification assay should be used in conjunction with clinical presentation, other laboratory findings, and epidemiological risk factors as an aid in the diagnosis of Mycoplasma infection and should not be used as the sole basis for treatment or other patient management. Positive results do not rule out co-infection with other organisms and negative results in persons with respiratory tract infections may be due to pathogens not detected by this assay. Lower respiratory tract infection due to M. pneumoniae may not be detected by this assay. If lower respiratory tract infection due to M. pneumoniae is suspected, additional laboratory testing using methods other than the illumigene Mycoplasma DNA Amplification assay may be necessary.
illumigene Mycoplasma is intended for use in hospital, reference or state laboratory settings. The device is not intended for point-of-care use.
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
For Prescription Use Only
4. Special instrument requirements:
Testing is performed on the illumipro-10™ Automated Isothermal Amplification and Detection System from Meridian Bioscience, Inc.
I. Device Description:
The illumigene Molecular Diagnostic Test System is comprised of the illumigene®
2
{2}
Mycoplasma DNA Amplification Assay Test Kit, the illumigene® Mycoplasma External Control Kit and the illumipro-10™ Automated Isothermal Amplification and Detection System.
The illumigene Mycoplasma assay utilizes loop-mediated isothermal amplification (LAMP) technology to detect the presence of Mycoplasma pneumoniae in human respiratory specimens (throat and nasopharyngeal swab specimens). Each illumigene Mycoplasma assay is completed using an illumigene Assay Control Reagent containing Control material, an illumigene Reaction Buffer, an illumigene Mycoplasma Test Device and microcentrifuge tubes. Respiratory specimens are combined with the illumigene Assay Control Reagent. The Specimen/Control sample is manually extracted and purified using a commercially available extraction kit (Qiagen, QIAamp® DSP DNA Mini Kit). Extracted DNA is heat-treated. Target and Control DNA are made available for isothermal amplification via heat-treatment. The heat-treated Specimen/Control sample is added to the illumigene Reaction Buffer. DNA amplification occurs in the illumigene Test Device.
The illumipro-10™ heats each illumigene Mycoplasma Test Device containing prepared sample and control material, facilitating amplification of target DNA. When M. pneumoniae is present in the specimen, a 208 base pair sequence of the M. pneumoniae genome is amplified and magnesium pyrophosphate is generated. Magnesium pyrophosphate forms a precipitate in the reaction mixture.
The illumipro-10™ monitors the absorbance characteristics of the reaction solutions at the assay Run Start (Signalinitial, S_i) and at the assay Run End (Signalfinal, S_f). The illumipro-10™ calculates the ratio of the Run End (Signalfinal or S_i) reads with the Run Start (SignalInitial or S_i) reads and compares the ratio to an established cut-off value. The illumipro-10™ performs this ratio calculation for both the TEST chamber and the CONTROL chamber.
The illumigene Mycoplasma External Controls Kit contains a Positive and Negative Control Reagent. External Control reagents are provided to aid the user in detection of reagent deterioration, adverse environmental or test conditions, or variance in operator performance that may lead to test errors. External Control reagents are provided for use in routine Quality Control testing.
J. Substantial Equivalence Information:
1. Predicate device name(s):
FilmArray® Respiratory Panel (RP) System
2. Predicate 510(k) number(s):
K120267
3. Comparison with predicate:
3
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Qualitative test for the direct detection and identification of Mycoplasma pneumoniae in nasopharyngeal swabs and throat swabs | Qualitative test for the direct detection and identification of Mycoplasma pneumoniae (and other organisms) in nasopharyngeal swabs |
| Organism Detection | Mycoplasma pneumoniae | Same |
| Detection | Self contained and automated | Self contained and automated |
| Assay Target | DNA | DNA |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Specimen Type | Nasopharyngeal swabs, throat swabs | Nasopharyngeal swabs |
| Assay Platform | Illumipro-10TM Automated isothermal Amplification and Detection System | FilmArray Instrument |
| Amplification Technology and Target Sequence Detected | Singleplex DNA Amplification Assay; Loop-Mediated Isothermal Amplification (LAMP) for the detection of a 208 base pair sequence of the Mycoplasma genome. The illumipro-10TM detects changes in reaction solution absorbance by visible light transmission | Multiplex Nucleic Acid Amplification Assay; Assay utilizes freeze-dried reagents to perform nucleic acid purification, reverse transcription, and nested multiplex PCR with DNA melt analysis |
| Extraction | Separate manual extraction method using Qiagen reagents | Automated extraction included in FilmArray RP pouch |
| External Assay Controls | External Positive and Negative Controls provided in illumigene Mycoplasma External Control Kit | External controls not provided |
| Reading Method | Visible Light Transmission | Fluorescence Emissions |
| Interpretation of Results | Results of the illumigene Mycoplasma Assay are interpreted by the illumipro-10TM and reported as INVALID, POSITIVE and NEGATIVE based on change in light transmission of the reaction mixtures. EMPTY WELL is reported when an illumigene Test | Results of the FilmArray Respiratory Panel (RP) Assay report are interpreted by the FilmArray Instrument for M. pneumoniae and reported as Detected, Not Detected or Invalid. |
{4}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | Device is not detected by the *illumipro-10*^{TM} or when questionable Signal Initial (S_{i}) transmission is detected. | |
K. Standard/Guidance Document Referenced:
- Clinical and Laboratory Standards Institute. 2008. User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline Second Edition (EP12A)
- Clinical and Laboratory Standards Institute. 2005. User Verification of Performance for Precision and Trueness; Approved Guideline Second Edition (EP15A2)
- Clinical and Laboratory Standards Institute. 2005. Interference Testing in Clinical Chemistry; Approved Guideline Second Edition (EP7A2)
L. Test Principle:
The <i>illumigene</i> Mycoplasma assay is based on loop mediated isothermal amplification technology (LAMP). Loop mediated amplification of DNA is accomplished by the use of specially designed primers that provide specific and continuous isothermal amplification. Magnesium pyrophosphate is produced as a by product of LAMP amplification. The magnesium pyrophosphate forms a white precipitate in the reaction solution, giving the reaction solution a turbid appearance. Change in sample absorbance created by precipitation of magnesium pyrophosphate indicates the presence of target DNA and is considered a positive reaction. The absence of target DNA results in no detectable change in sample absorbance and is considered a negative reaction.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Reproducibility Study:
Blind coded panels of 10 samples were supplied to three independent laboratories. Samples were randomly sorted within each panel to mask sample identities. Each panel consisted of true negative (1 replicate), high negative (three replicates), low positive (three replicates), and moderate positive (three replicates) concentrations. Panel members were prepared in natural negative matrix in M4 non-nutritive transport medium (polyester swab inoculated with normal nasal flora, screened negative for M. pneumoniae) and were spiked with M. pneumoniae strain M129 to the final concentrations indicated in the table below. The low positive sample concentration was set near the Limit of Detection (LoD) which should generate
{5}
positive results $\geq 95\%$ of the time and the high negative sample was set at a concentration that should generate negative results $\geq 95\%$ of the time.
| Sample ID | Final Concentration (CFU/Test) | Final Concentration (CFU/mL) | Expected Result |
| --- | --- | --- | --- |
| Moderate Positive | 30 | 800 | Positive |
| Low Positive | 15 | 400 | Positive |
| High Negative | 0.03 | 0.8 | Negative |
| Negative | N/A | N//A | Negative |
Three lots of illumigene Mycoplasma reagents and five illumipro-10™ instruments were used in reproducibility studies. Each clinical site tested two panels each day for five days with testing performed by two operators at each site on the same day (intra-assay variability). Positive and Negative Controls were tested each day of testing. Testing results are provided in the table below:
| | Site 1 | | Site 2 | | Site 3 | | Total | | Average S_{r}:S_{t} | SD | %CV |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample Type | Percent agreement | | Percent agreement | | Percent agreement | | Percent agreement | | | | |
| Negative | 10/10 | 100% | 10/10 | 100% | 10/10 | 100% | 30/30 | 100% | 100.4 | 2.85 | 2.83% |
| High Negative | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 100% | 101.1 | 3.32 | 3.29% |
| Low Positive | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 90/90 | 100% | 61.7 | 6.18 | 10.02% |
| Moderate Positive | 29/30 | 96.7% | 30/30 | 100% | 29/30 | 96.7% | 88/90 | 97.8% | 62.3 | 6.25 | 10.04% |
| Negative Control | 10/10 | 100% | 10/10 | 100% | 10/10 | 100% | 30/30 | 100% | 101.5 | 4.03 | 3.97% |
| Positive Control | 10/10 | 100% | 10/10 | 100% | 10/10 | 100% | 30/30 | 100% | 60.5 | 4.29 | 7.10% |
b. Linearity/assay reportable range:
Not applicable as the illumigene Mycoplasma DNA Amplification Assay is a qualitative assay.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Specimen Stability:
Specimen storage and hold time parameters were evaluated for the illumigene Mycoplasma assay. Validation studies performed at Meridian were completed using Rayon and polyester swabs with M4, M4RT, M5, $0.85\%$ Saline, UTMRT, and Liquid Stuart Transport Medium. Samples were prepared in both simulated nasal and throat matrices spiked with the FH strain of M. pneumoniae at high negative and low
{6}
positive (2x LoD) concentrations.
Testing was performed in triplicate for each storage condition; (1) upper limit of room temperature (32.5°C) and (2) refrigerated (2-8°C). All samples tested yielded the correct results for sample storage of up to 49 hours for elevated room temperature storage and up to 6 days at 2-8°C. Study results demonstrated that throat and nasopharyngeal swabs (both rayon and polyester) in the above described transport media can be held at the claimed storage conditions: 19-29°C for up to 48 hours or at 2-8°C for up to 5 days prior to testing.
## Extracted Specimen Stability:
Studies were performed to evaluate storage parameters for storage of specimen extracts prior to heat treatment. Samples consisting of rayon and polyester swabs with M5 media in a simulated negative matrix were spiked with *M. pneumoniae* at high negative, LoD and 5x LoD concentrations. Contrived negative samples were also tested. Extracted samples were held at the upper limit of room temperature (29°C) prior to heat treatment and testing on the illumipro-10™. All extracted samples yielded the expected results for storage up to 75 minutes. Study results validated the claimed stability of 19-29°C for up to 1 hour prior to heat treatment for specimen extracts.
## Controls
**External Controls:** The illumigene Mycoplasma External Control Kit includes external positive and negative controls and is sold separately from the illumigene Mycoplasma DNA Amplification Assay Kit. The positive external control contains non-infectious plasmid DNA with *M. pneumoniae* and *S. aureus* inserts. The negative external control consists of non-infectious plasmid DNA with a *S. aureus* insert. Testing of external controls was performed for all analytical and clinical studies with no discordant results. It is recommended that external controls are run with each new lot and new shipment of illumigene Mycoplasma kits. External control testing should be performed thereafter in accordance with appropriate federal, state and local guidelines. Results from the illumigene Mycoplasma test kit should not be reported if external controls do not produce the expected results.
**Internal Control:** Primers for amplification of the internal control are incorporated into each device in the CONTROL chamber. Internal control DNA (Assay Control II Reagent- phosphate-buffered solution containing formalin-treated *E. coli* harboring plasmid containing a segment of the *Staphylococcus aureus* genome) is combined with each specimen and processed through all steps of the procedure. The internal control is present to monitor for sample processing, amplification inhibition, and assay reagent integrity.
## d. Detection limit:
Analytical Sensitivity studies were designed to determine the analytical limit of detection (LoD) of the illumigene Mycoplasma DNA Amplification Assay. The
{7}
LoD is the lowest number of colony forming units (CFUs) per test aliquot that can be distinguished from negative samples with a high degree of probability (95%). Two M. pneumoniae strains, FH (ATCC 15531) and M129, were evaluated. Aliquots from the two strains were sent to an independent testing facility for concentration determination by colony count. Samples were prepared in a natural negative matrix (rayon swabs in M4 non-nutritive Transport Medium inoculated with normal nasal flora previously screened negative for M. pneumoniae). Replicates for each dilution were tested with the illumigene Mycoplasma DNA Amplification Assay; however, testing for a specific concentration was discontinued when more than one negative result was obtained. The lowest dilution producing positive results in a minimum of 19 of 20 replicates was identified as the LoD. Testing was performed using three lots of reagents and six illumipro-10™ instruments. External Positive and Negative Controls were tested each day throughout the study. The LoD for the assay was confirmed as 88 CFU/Test (2,350 CFU/mL) for the FH strain (ATCC 15531) and 7.5 CFU/Test (200 CFU/mL) for the M129 strain.
e. Analytical Reactivity/Inclusivity:
Nine additional strains (see table below) of M. pneumoniae were tested with the illumigene Mycoplasma DNA Amplification Assay and produced the expected positive results at or below stated assay LoD of 88 CFU/Test (2,350 CFU/mL). Samples were diluted in sterile saline and spiked into negative matrix which consisted of rayon swabs inoculated with normal nasal flora and placed into M4 Transport Media. The nasal matrix was previously screened as negative for M. pneumoniae. Three replicates of each strain were tested resulting in 100% detection. Both M. pneumoniae Type 1 and Type 2 strains were represented.
| Strain Identification | Strain Identification |
| --- | --- |
| PI1428 | Mutant 22 (ATCC 39505) |
| MAC (ATCC 15492) | UAB 55612 |
| M52(ATCC 15293) | UAB 56317 |
| Bru (ATCC 15377) | UMTB-10G (ATCC 49899)
M. pneumoniae and M. genitalium) |
| M129-B170 (ATCC 29343) | |
f. Interference:
An interference study was carried out to evaluate the effect of potentially interfering substances on the accuracy of test results obtained with the illumigene Mycoplasma DNA Amplification Assay. Potentially interfering substances were diluted in sterile saline and added to M4 medium with rayon swabs (negative sample) and to M4 medium with polyester swabs (positive samples). Two sets of positive samples were spiked separately with M. pneumoniae strain M129 and with M. pneumoniae strain FH (ATCC 15531) at
{8}
concentrations near the LoD. Testing included three replicates for each interferent alone (negative sample) and three replicates for each interferent mixed with each of the two M. pneumoniae strains (six positive samples per interferent).
Testing of the following biological substances at the concentrations indicated demonstrated no interference (no false positive or false negative results): Mucus (5.0 mg/mL), White blood cells (0.5% v/v), and whole blood (5% V/V).
Testing of the following chemical substances at the concentrations indicated demonstrated no interference (no false positive or false negative results): Acetaminophen (18.1 mg/mL), Albuterol Sulfate (20 mg/mL), Aspirin (9.1 mg/mL), Azithromycin dehydrate (2.0 mg/mL), Cepacol® Mouthwash [Ethanol, denatured (1.4% v/v), Cetylpyridinium chloride (.005% v/v)], Contac® Cold + Flu Tablets [Acetaminophen (14.8 mg/mL), Chlorpheniramine maleate (0.06 mg/mL), Phenylephrine HCl (0.15 mg/mL)], Diphenhydramine HCl (2.6 mg/mL), Erythromycin (20.0 mg/mL), HALLS® Cough [Menthol (0.06 mg/mL)], Ibuprofen (12.7 mg/mL), Phenylephrine HCl (0.595 mg/mL), Prednisone (20.0 mg/mL), Robitussin® Cough+Chest Congestion Cough Syrup [Dextromethorphan HBr (0.20 mg/mL), Guaifenesin (2.0 mg/mL)], Saline Nasal Spray [Sodium chloride (0.65 mg/mL)].
The presence of phenylephrine HCl (found in nasal decongestants and flu and cold tablets) at concentrations above 0.595 mg/mL yielded false negative results with M. pneumoniae strain M129. This information is included as a limitation in the package insert.
g. Analytical specificity/Cross-reactivity:
Potentially cross-reacting microorganisms expected to be present in throat or nasopharyngeal swab specimens were added to samples both negative and positive for M. pneumoniae. Negative samples were prepared with M4 transport medium inoculated with nasal flora on rayon swabs and did not contain M. pneumoniae. Positive samples were prepared by spiking confirmed negative sample matrix (M4 transport medium inoculated with nasal flora on polyester swabs) with Mycoplasma pneumoniae, FH strain (ATCC 15531) at concentrations near the LoD. Three replicates were tested for both positive and negative samples and for each potential cross-reacting organism.
Samples of each potentially cross-reactive (or interfering) microorganism were prepared at minimum concentrations of ≥1.0 × 10⁶ CFU/mL for bacteria/fungi and ≥1.0 × 10⁵ TCID₅₀/mL for viruses. Human DNA was tested at 2.0 ng/test.
None of the following organisms demonstrated interference or cross-reactivity with the illumigene Mycoplasma DNA Amplification Assay: Acinetobacter baumannii, Acinetobacter calcoaceticus, Actinomyces odontolyticus, Bacillus subtilis, Bacteroides fragilis, Bordetella parapertussis, Bordetella pertussis, Burkholderia cepacia, Candida albicans, Candida glabrata, Candida parapsilosis, Chlamydia pneumoniae, Citrobacter freundii, Clostridium difficile, Corynebacterium diphtheriae, Enterobacter cloacae, Enterococcus faecalis, Escherichia coli,
9
{9}
Escherichia coli (ESBL), Fusobacterium nucleatum, Haemophilus ducreyi, Haemophilus influenzae, Haemophilus parainfluenzae, Helicobacter pylori, Klebsiella pneumoniae, Klebsiella pneumoniae (KPC), Legionella pneumophila, Listeria monocytogenes, Mycoplasma genitalium, Mycoplasma hominis, Neisseria cinerea, Neisseria gonorrhoeae, Neisseria meningitidis, Peptostreptococcus anaerobius, Proteus mirabilis, Proteus vulgaris, Pseudomonas aeruginosa, Salmonella paratyphi (Group A), Salmonella typhimurium (Group B), Serratia liquefaciens, Staphylococcus aureus, Staphylococcus epidermidis, Streptococcus agalactiae (Group B), Streptococcus anginosus (Group F), Streptococcus bovis (Group D), Streptococcus canis (Group G), Streptococcus equisimilis, Streptococcus mitis, Streptococcus pneumoniae, Streptococcus pyogenes, Streptococcus salivarius, Ureaplasma urealyticum, Adenovirus, Coxsackievirus, Cytomegalovirus, Epstein Barr virus, Herpes simplex virus 1, Herpes simplex virus 2, Influenza A, Influenza B, Human metapneumovirus, Parainfluenza virus 1, Parainfluenza virus 2, Parainfluenza virus 3, Respiratory syncytial virus A, Respiratory syncytial virus B, Rhinovirus, Human DNA.
Moraxella catarrhalis, Nocardia asteroides, and Coronavirus each produced false negative results in one of three M. pneumoniae-positive replicates tested. Testing of an additional four replicates for each of the three potentially interfering organisms was performed with all additional replicates yielding the expected positive results. Therefore for each of the three organisms, false-negative results occurred in one of seven replicates.
h. Assay cut-off:
The illumigene Mycoplasma is manufactured with fixed cut-off values. The product is designed around a pre-selected cut-off value and amplification reagent concentrations are optimized to ensure appropriate reactions are obtained. Development optimization includes evaluation of characterized positive and negative clinical specimens. Amplification reagent concentrations are adjusted during design as needed to ensure illumigene results are aligned with clinical specimen reported results.
Cut-off values are applied in the following manner:
The illumipro-10™ calculates the ratio of the Run End (Signal final or Sf) reads with the Run Start (Signal Initial or Si) reads and compares the ratio to an established cut-off value. The illumipro-10™ performs this ratio calculation to both the TEST chamber and the CONTROL chamber.
Fixed cut-off values for the CONTROL chamber are used to determine validity. CONTROL chamber Sf:Si ratios less than 90% are considered valid and allow for reporting of TEST chamber results (POSITIVE, NEGATIVE). CONTROL chamber Sf:Si ratios greater than or equal to 90% are considered invalid. Test chamber results are not reported for samples with an "INVALID" control result. More stringent cut-off criteria are applied to the Control chamber reaction in order to ensure that amplification is not inhibited, reagents are performing as intended, and sample processing was performed appropriately.
10
{10}
Fixed cut-off values for the TEST chamber are used to report sample results. TEST chamber Sf:Si ratios less than 82% are reported as ‘POSITIVE’ and TEST chamber Sf:Si ratios greater than or equal to 82% are reported as ‘NEGATIVE’. Numerical values for the CONTROL and The TEST chamber results are not displayed for the user.
i. Fresh/Frozen Studies:
Meridian performed a fresh versus frozen study with 60 samples contrived from 30 fresh nasal wash specimens determined to be negative for *M. pneumoniae* using a validated PCR assay. Each nasal wash specimen was tested both with and without *M. pneumoniae* spiked into the sample. Spiked specimens were prepared at concentrations near the LoD (7.5 CFU/test or 200 CFU/mL, initially tested with the illumigene assay, and tested again after each of three freeze/thaw cycles. Study results demonstrated 100% agreement for all negative samples for each freeze/thaw cycle. For positive specimens, 29/30 specimens demonstrated the expected positive result. One positive sample gave a false negative result after the first freeze/thaw cycle but this same sample was positive after the second and third freeze/thaw cycles, indicating that the sample had not deteriorated. Results of the study sufficiently demonstrated that fresh and frozen specimens should provide equivalent results with the illumigene Mycoplasma DNA Amplification Assay.
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable, performance of the assay was evaluated in comparison to a composite reference method of culture and PCR followed by bi-directional sequencing.
b. Matrix comparison:
Sample Type Equivalence Study:
Sample type equivalency studies were performed to confirm the suitability of several swab types as well as several non-nutritive transport media for use with the illumigene Mycoplasma DNA Amplification Assay. The following swab types were evaluated: Cotton, Flocked Nylon, Foam, Polyester, and Rayon. The following non-nutritive transport mediums were evaluated: M4, M4-RT, M5, Liquid Stuart, 0.85% Saline and UTM-RT (Universal Transport Medium).
Contrived positive samples were prepared by combining confirmed *M. pneumoniae*-negative matrices with the FH strain of *M. pneumoniae* at concentrations near the LoD (2670 CFU/mL for NP swabs, and 920 CFU/mL for throat swabs. Leftover nasal wash and throat specimens were used to prepare specimens for the study. Replicates of both positive and negative samples were tested for all possible combinations of specimen matrix, swab, and media type. All negative samples yielded correct results.
11
{11}
Three positive samples gave unexpected negative results: One replicate of a NP matrix/rayon swab/M5 media combination and two replicates of a throat matrix/polyester swab/M4 media combination.
Testing of four additional NP/rayon/M5 media samples yielded no discordant results. Additionally, each of three variables (NP, rayon, M5) was challenged in combination with other matrices, swab types and media types resulting in no additional false negative results. Therefore, it is likely that the one original discordant result was likely due to the use of *M. pneumoniae* concentrations near the LoD.
Twenty additional samples prepared in throat matrix/polyester swab/M4 media were tested in a simulated matrix equivalency study with all 20 samples yielding expected positive results. Therefore, it is likely that the two original discordant results for this swab/media combination were likely due to the use of *M. pneumoniae* concentrations near the LoD.
Swab and transport medium types with demonstrated performance are described in the illumigene Mycoplasma DNA Amplification Assay Package Insert.
## 3. Clinical Studies:
### a. Clinical Sensitivity and specificity
The illumigene Mycoplasma DNA Amplification Assay was evaluated in 2012 by five independent clinical test sites located in the Midwestern, Southern, and Central regions of the United States. A total of 334 throat and nasopharyngeal (NP) swab specimens, collected from patients suspected of *Mycoplasma pneumoniae* infection were evaluated with the test device to establish performance characteristics.
The study included testing of leftover de-identified specimens submitted to the testing laboratories for routine *M. pneumoniae* testing. Specimens were presumed to be from symptomatic patients. Testing included prospectively collected and tested fresh specimens, prospectively collected and retrospectively tested frozen specimens (all comers), and retrospectively tested frozen selected specimens.
The performance of illumigene Mycoplasma DNA Amplification Assay was compared to a composite reference method that included *M. pneumoniae* culture and identification and a validated real-time PCR assay followed by bi-directional sequencing for positive specimens. The culture portion of the composite reference method was performed at four of the testing sites with one testing site sending specimens to a reference laboratory for culture. All reference PCR assays were performed by Meridian BioScience, Inc with sequencing performed by a reference laboratory for positive specimens. For performance calculations, specimens positive by culture or positive by real-time PCR and bi-directional sequencing were considered positive. Specimens negative by both culture and PCR were considered negative.
A total of 12 specimens were excluded from the performance analysis due to inconclusive culture results and negative PCR results. A total of 103 (30.8%)
12
{12}
prospective and 219 retrospective specimens (65.6%) were tested with one initial invalid result (0.30%). The illumigene assay performance for both prospective and retrospective nasopharyngeal and throat swab specimens as compared to the composite reference method are shown in the tables below. Also shown is the illumigene assay performance as compared to the reference PCR and bi-directional sequence testing.
illumigene Mycoplasma Assay Performance: Nasopharyngeal Swabs
| Specimen Description | Positive Specimens | | | Negative Specimens | | | Invalid Results |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | illumigene vs. Comparator | % Sensitivity or PPA | 95% CI | illumigene vs. Comparator | % Specificity or NPA | 95% CI | |
| Composite Method Comparator | | | | | | | |
| Prospective | 4/4 | % Sensitivity 100% | 51.0 – 100.0% | 48/48 | % Specificity 100% | 92.6 – 100.0% | 0 |
| Retrospective | 34/36 | PPA 94.4% | 81.9 – 98.5% | 86/90 | NPA 95.6% | 89.1 – 98.3% | 0 |
| Validated PCR Assay with Bi-directional Sequencing Comparator | | | | | | | |
| Prospective | 4/4 | % Sensitivity 100% | 51.0 – 100.0% | 48/48 | % Specificity 100% | 92.6 – 100.0% | 0 |
| Retrospective | 34/36 | PPA 94.4% | 81.9 – 98.5% | 86/90 | NPA 95.6% | 89.1 – 98.3% | 0 |
illumigene Mycoplasma Assay Performance: Throat Swabs
| Specimen Description | Positive Specimens | | | Negative Specimens | | | Invalid Results |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | illumigene vs. Comparator | % Sensitivity or PPA | 95% CI | illumigene vs. Comparator | % Specificity or NPA | 95% CI | |
| Composite Method Comparator | | | | | | | |
| Prospective | 8/8 | % Sensitivity 100% | 67.6 – 100.0% | 43/43 | % Specificity 100% | 91.8 – 100.0% | 0 |
| Retrospective | 22/26^{a} | PPA 84.6% | 66.5 – 93.9% | 66/67 | NPA 98.5% | 92.0 – 99.7% | 1 |
{13}
| Validated PCR Assay with Bi-directional Sequencing Comparator | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Prospective | 8/8 | % Sensitivity 100% | 67.6 – 100.0% | 43/43 | % Specificity 100% | 91.8 – 100.0% | 0 |
| Retrospective | 21/21^{b} | PPA 100% | 84.5 – 100.0% | 70/72 | NPA 97.2% | 90.4 – 99.2% | 1 |
a. Four specimens originally identified by culture as positive were not confirmed by either the illumigene assay or the independent PCR method. Results suggest sample degradation during storage.
b. One specimen originally identified by the illumigene assay and culture as positive was negative by the independent PCR method. This specimen is classified as a false positive relative to PCR.
Age information was known for 83.5% (269/322) of patients included in the performance analysis. Seven (2.6%) patients tested were between 0-28 days of age; 38 (14.1%) patients were between 29 days and up to 2 years of age; 139 (51.7%) patients were between 2 and up to 12 years of age; 61 (22.7%) patients were between 12-18 years, and 9 (3.3%) patients were between 18-21 years of age. The remaining 15 (5.6%) study patients were 21 years or older. No performance differences were noted based on chronological age.
The study population included 90 (27.9%) female patients and 91 (28.3%) male patients. Gender was unknown for 141 (43.8%) of the study participants. No performance differences were noted based on gender.
Clinical performance of the illumigene Mycoplasma DNA Amplification Assay was assessed by testing de-identified nasopharyngeal and throat swab specimens lacking clinical information; accordingly, the number of patients with M. pneumoniae pneumonia included in the clinical studies was unknown and performance for this group cannot be described separately.
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
Overall incidence of Mycoplasma pneumoniae in prospectively collected and tested specimens during the 2012 clinical study was 11.7% (12/103).
N. Instrument Name:
illumipro-10™
O. System Descriptions:
1. Modes of Operation:
{14}
The illumipro-10™ instrument heats each illumigene Mycoplasma Test Device containing prepared samples and Control Reagent, facilitating amplification of target DNA. When M. pneumoniae is present in a nasopharyngeal or throat swab specimen, a conserved M. pneumoniae DNA sequence is amplified and magnesium pyrophosphate is generated. Magnesium pyrophosphate forms a precipitate in the reaction mixture. The illumipro-10™ instrument detects the change in light transmission through the reaction mixture created by precipitating magnesium pyrophosphate. Sample results are reported as Positive or Negative based on the detected change in light transmission.
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ x ☐ or No ☐
3. Specimen Identification:
Specimens are identified by position. Default Sample Identification is based on Block and Well position (e.g., Block A, Well 1). The user may input Sample Identification information using the keypad or the optional external keyboard.
4. Specimen Sampling and Handling:
Specimens are prepared manually. Incubation, amplification and detection are automated using the illumipro-10™.
5. Calibration:
Calibration of the illumipro-10™ is not required.
6. Quality Control:
The illumigene Mycoplasma External Control Kit contains a Positive and a Negative Control Reagent. The External Positive and Negative Controls are not built into the test system, but are tested in the same manner as a patient specimen. The illumigene Mycoplasma External Control Kit is provided to aid the user in detection of reagent deterioration, adverse environmental or test conditions, or variance in operator performance that may lead to test errors. The illumigene Mycoplasma External Controls are recommended for routine Quality Control.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
Not Applicable
{15}
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
1. The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
16
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.