OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM

K123421 · Olympus Medical Systems Corp. · NEZ · Apr 29, 2013 · Gastroenterology, Urology

Device Facts

Record IDK123421
Device NameOLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM
ApplicantOlympus Medical Systems Corp.
Product CodeNEZ · Gastroenterology, Urology
Decision DateApr 29, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1300
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K123421 · Apr 29, 2013OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEMOlympus Medical Systems Corp.Retrospective analysis of recorded image data from a clinical studyRetrospective evaluation of antenna reception performance using recorded image data from a clinical study of healthy volunteers.Retrospective analysis; Antenna performance; Clinical study data

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective analysis of clinical study dataHealthy volunteers; Sample Size: 29 subjectsPredicate deviceVideo gap occurrence as an indicator of reception performance

Indications for Use

The OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM is intended for visualization of the small intestine mucosa. The Red Color Detection Function is intended to mark frames of the video suspected of containing blood or red areas.

Device Story

System captures small intestine images via ingestible capsule (EC-Y0006); capsule contains objective lens, image sensor, 4 white LEDs, radio transmitter (315MHz), and silver oxide batteries. Patient wears antenna unit (MAJ-Y0163) and recorder (RE-Y0002) to receive/store image data. Physician uses workstation (PC with ENDOCAPSULE software) to download, view, and analyze images. Software features include 3D tracking (capsule position estimation via antenna signal strength), red color detection (highlighting suspected blood/lesions), and various playback modes (Express viewing, Overview). System aids clinicians in identifying small intestine mucosal abnormalities; patient benefits from non-invasive visualization of small bowel.

Clinical Evidence

Clinical study with 30 healthy volunteers evaluated 3D Track function accuracy and image quality. 3D Track accuracy verified against radiographic ground truth. Image quality assessed via investigator questionnaire, showing superiority to predicate. Antenna performance evaluated retrospectively via video gap occurrence, confirming equivalence to predicate.

Technological Characteristics

Capsule (11mm x 26mm) made of biocompatible polymer composites. Sensing: CMOS image sensor with 4 white LEDs. Transmission: 315MHz radio. Power: Silver oxide batteries. Connectivity: Antenna unit to wearable recorder; data transfer to PC workstation via cradle. Software: Server-client architecture on commercial PC. Sterilization: Not specified.

Indications for Use

Indicated for visualization of small intestine mucosa in patients requiring endoscopic examination. No specific age or gender contraindications provided.

Regulatory Classification

Identification

An ingestible telemetric gastrointestinal capsule imaging system is used for visualization of the small bowel mucosa as an adjunctive tool in the detection of abnormalities of the small bowel. The device captures images of the small bowel with a wireless camera contained in a capsule. This device includes an ingestible capsule (containing a light source, camera, transmitter, and battery), an antenna array, a receiving/recording unit, a data storage device, computer software to process the images, and accessories.

Special Controls

*Classification.* Class II (special controls). The special control is FDA's guidance, “Class II Special Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging Systems; Final Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 123421 117 ### 510(k) SUMMARY # APR 2 9 2013 #### OLYMPUS INTESTINAL ENDOSCOPE SMALL CAPSULE SYSTEM November 5, 2012 #### 1 General Information 를 Applicant: B Official Correspondent: 1 Manufacturer: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, 192-8507, Japan Establishment Registration No .: 8010047 Daphney Germain-Kolawole Regulatory Affairs Olympus America Inc. 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610 Phone: 484-896-5691 FAX: 484-896-7128 Email: daphney.germain-kolawole@olympus.com SHIRAKAWA OLYMPUS CO., LTD. 3-1, Aza-Ookamiyama, Ooaza-Odakura, Nishigo-mura, Nishishirakawa-gun, Fukushima, 961-8061, Japan Establishment Registration No .: 3002808148 OLYMPUS MEDICAL SYSTEMS CORP. HINODE PLANT 34-3 Hirai, Hinode-machi, Nishitama-Gun, Tokyo, 190-0182, Japan Establishment Registration Number: 3003637092 #### 2 Device Identification Device Trade Name: 트 Common Name: l Regulation Number: . Regulation Name: OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM Capsule Imaging System 21 CFR 876.1300 Ingestible telemetric gastrointestinal capsule imaging system {1}------------------------------------------------ | | Regulatory Class: | II (Special Controls) | |--|-----------------------|--------------------------| | | Classification Panel: | Gastroenterology/Urology | | | Product Code: | NEZ | #### 3 Legally Marketed Device to which Substantial Equivalence is Claimed The following table shows the device to which we claim substantial equivalence (predicate device). ## Table 18-1 Predicate Device of the OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM | Predicate Device | 510(k)<br>Number | |----------------------------------|------------------| | OLYMPUS CAPSULE ENDOSCOPE SYSTEM | K090210 | ## 4. Device Description OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM is a capsule imaging system used for visualization of the small intestine mucosa. The OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM consists of a CAPSULE ENDOSCOPE (EC-Y0006), an ACTIVATOR (MAJ-1478), a RECORDER (RE-Y0002), a CRADLE (MAJ-Y0160), a BATTERY PACK (MAJ-Y016), a RECORDER HOLDER (MAJ-Y0162), an ANTENNA UNIT (MAJ-Y0163), an ANTENNA HOLDER (MAJ-Y0164), an ENDOCAPSULE SOFTWARE SERVER-CLIENT (MAJ-Y0165) and an ENDOCAPSULE SOFTWARE CLIENT (MAJ-Y0166). ENDOCAPSULE SMALL INTESTINAL CAPSULE ENDOSCOPE SET (MAJ-Y0176) is a package containing five CAPSULE ENDOSCOPEs (EC-Y0006). ENDOCAPSULE RECORDER SET (MAJ-Y0177) is a package containing the ACTIVATOR (MAJ-1478), the RECORDER (RE-Y0002), the CRADLE (MAJ-Y0160), the BATTERY PACK (MAJ-Y0161), the RECORDER HOLDER (MAJ-Y0162), the ANTENNA UNIT (MAJ-Y0163), and the ANTENNA HOLDER (MAJ-Y0164). The CAPSULE ENDOSCOPE (EC-Y0006) has an outer diameter of 11mm, a total length of 26mm, and is designed and manufactured with polymer composites that are biocompatible and provide sufficient tolerance to exposure to body fluids (digestive tract juice) and external compression forces. The clear top cover contains a compact objective lens in front of the image sensor. Four white light emitting diodes (LEDs) are allocated around the objective lens. The exterior casing contains operational circuits for the image sensor, LEDs, radio transmitter, and antenna for radio transmission in 315MHz. The CAPSULE ENDOSCOPE (EC-Y0006) has two silver oxide batteries to power the circuit within the capsule. {2}------------------------------------------------ The RECORDER (RE-Y0002) is used in combination with the ANTENNA UNIT (MAJ-Y0163) for receiving signals from the CAPSULE ENDOSCOPE (EC-Y0006) and for recording image data captured by the CAPSULE ENDOSCOPE (EC-Y0006). The RECORDER (RE-Y0002) is powered by the BATTERY PACK (MAJ-Y0161). It is capable of continuously receiving and recording image data from the CAPSULE ENDOSCOPE (EC-Y0006). The RECORDER has a color LCD display panel for displaying received images, recorded images, patient ID, battery pack level, etc. The RECORDER (RE-Y0002) is inserted into a pouch on the RECORDER HOLDER (MAJ-Y0162) enabling the patient to move freely during the examination. The BATTERY PACK (MAJ-Y0161) is charged when the RECORDER (RE-Y0002) is attached to the CRADLE (MAJ-Y0160). Unlike the predicate device that includes Battery charger (MAJ-1478), no independent battery charger is required. The ANTENNA UNIT (MAJ-Y0163) is an integrated antenna sheet containing eight antenna elements and is stored in the ANTENNA HOLDER (MAJ-Y0164). The ANTENNA HOLDER is attached covering the patient's waist. A workstation for observation of the image data captured by the CAPSULE ENDOSCOPE (EC-Y0006) is constructed by installing the ENDOCAPSULE SOFTWARE SERVER-CLIENT (MAJ-Y0165) or the ENDOCAPSULE SOFTWARE CLIENT (MAJ-Y0166) to a commercially available personal computer. Functions of each software are shown in the table below. | Model | Functions | |--------------------------------------------|-------------------------------| | ENDOCAPSULE SOFTWARE | Observation of the image data | | SERVER-CLIENT (MAJ-Y0165) | Storage of the image data | | ENDOCAPSULE SOFTWARE<br>CLIENT (MAJ-Y0166) | Observation of the image data | When the user uses more than or equal to two workstations in the network environment, the user can select the ENDOCAPSULE SOFTWARE CLIENT (MAJ-Y0166) as the workstation from the 2nd workstation. The user can share data between two workstations or more. The ENDOCAPSULE SOFTWARE SERVER-CLIENT (MAJ-Y0165) includes the ENDOCAPSULE SOFTWARE LIGHT which can be installed on commercially available personal computer for viewing the image data captured by the CAPSULE ENDOSCOPE (EC-Y0006). The ENDOCAPSULE SOFTWARE LIGHT is provided for user convenience. The workstation facilitates the downloading of endoscopic image data recorded in the RECORDER (RE-Y0002) to the workstation via the CRADLE (MAJ-Y0160) for observation of patient image data. The workstation function sends patient information data to the RECORDER (RE-Y0002) from the workstation via the CRADLE (MAJ-Y0160). The workstation also provides the following functions: - Image display whose maximum rate of the image renewal is based upon the user setting: - Display / store images selected by the user as thumbnail data; - Development and printing of diagnostic reports with bookmarked images; - Image / video clip export; - Data storage to a network server. {3}------------------------------------------------ In addition, the image processing function offers the following user selectable functions: - . 3D Track function (new function) The signal reception strength of multiple antennas is calculated to estimate the capsule position, and the capsule track is displayed. - . Red Color Detection Function A software feature that highlights frames suspicious for blood or red lesions based upon analysis of red pixels within the image. - . Average Color Bar Calculates the average RGB colors of an image and displays the averaged color composite as a color line. - . Adjust mode Automatically adjusts the speed of frame display based upon both image movement and user setting. - . Express viewing function A software feature that allows for image playback and that has two playback modes. The Express-Sclected mode plays back images whose mucosal area is dissimilar from that of previous image. The Express-Skipped mode plays back the images skipped in the Express-Selected mode. - 사 Overview function A software feature that displays images whose mucosal area is dissimilar from that of previous images as an outline of the complete case. ### 5 Indications for Use The OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM is intended for visualization of the small intestine mucosa. The Red Color Detection Function is intended to mark frames of the video suspected of containing blood or red areas. #### Comparison of Technological Characteristics б The technological characteristics of the subject OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM are identical or similar to the predicate device identified in "3 Legally Marketed Device to which Substantial Equivalence is Claimed" above. A detailed comparison of the subject OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM with the predicate device is provided in the premarket notification. {4}------------------------------------------------ #### 7 Substantially Equivalent Discussion The subject OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM is a modification, to the previously cleared OLYMPUS CAPSULE ENDOSCOPE SYSTEM (K090210). The indications for use, principles of operation and fundamental technology of the subject OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM are similar or identical to the predicate device. The indicated patient population and anatomical site are also identical to the predicate The subject device also meets the guidance entitled "Class II Special device. Controls Guidance Document: Ingestible Telemetric Gastrointestinal Capsule Imaging System; Final Guidance for Industry and FDA, Document issued on: November 28, 2001." The proposed changes in this submission do not raise new safety and effectiveness issues related to device design. Based upon the results of bench testing and clinical testing performed, we believe that the subject OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM is substantially equivalent to the identified predicate device as demonstrated by this submission. #### Summary of non-clinical testing કિ The following non-clinical testing was performed. Risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2007. The design verifications tests and their acceptance criteria were identified and performed as a result of this risk analysis assessment. The following bench testing were performed. 1) Specification testing including "PH resistance testing", "Mechanical and structural integrity testing (including bite test)", "Battery life testing", "Field of view, depth of focus, distortion, color/illumination uniform, sensitivity of imaging sensor, color reproduction, resolution, and image noise testing", "Shock, vibration, temperature, and after the shelf life)". 2) Side-by-side bench testing between the subject device and the predicate device including "Image quality (including displayed lines and villus, field of view, optimum working distance)" and "Antenna reception performance". 3) Evaluation of 3D track function including "Accuracy evaluation" and "Display evaluation" 4) Stability testing The testing results showed the subject device has sufficient performance or equivalent performance to the predicate device. {5}------------------------------------------------ The following biocompatibility testing were performed for the patient contacting parts of the EC-Y0006 and testing results met the biocompatibility requirement. - Cytotoxicity - Sensitization - Acute Intracutaneous Reactivity - Acute Oral Toxicity - Acute Systemic Toxicity - Implantation instead of sub-chronic toxicity The software validation activities were performed in accordance with the FDA Guidance, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The device software is considered a "Moderate Level of Concern." The following standards were used during the design and validation of the subject devices: -ISO 14971: 2007 -IEC 60601-1: 1988, Amendment 1: 1991, Amendment 2: 1995 -IEC 60601-1-1: 2000 -IEC 60601-1-2: 2001, Amendment 1: 2004 -IEC 60601-2-18: 1996, Amendment 1: 2000 -IEC 60950-I : 2005 -180 11135-1: 2007 -ISO 10993-1: 2009 -ISO 10993-5: 2009 -ISO 10993-6: 2007 -ISO 10993-10: 2010 -1SO 10993-11: 2006 #### ਹੈ Summary of clinical testing A clinical study was performed and thirty healthy volunteers were involved in this study. The objective of the study was to evaluate the accuracy of the 3D Track function and image quality. Additionally, the performance of the new antenna was retrospectively evaluated utilizing the recorded image data in the clinical study above. The summary of the clinical study results is described below. ### 3D Track The accuracy of the 3D Track function was evaluated. Healthy volunteers swallowed the capsule endoscope and underwent x-ray photography. Based upon the radiographic image data, positions of the capsule endoscope were derived. Then the positions of the capsule endoscope derived from the radiographic image and the estimated positions of the capsule endoscope at the time of the radiographic image Three subjects were excluded, and therefore twenty-seven shooting were compared. subjects were eligible. The test results met the predefined acceptance criteria. Therefore, it was verified that the accuracy of the 3D Track function was maintained as we presumed. {6}------------------------------------------------ ### Image quality Image quality of the subject device was assessed by the principal investigator questionnarie regarding resolution, brightness including evenness of illumination, color tone, and field of view. The principal investigator's comments showed that the subject device is greatly superior to the predicate device in regard to image quality. ### Antenna Performance The performance of the new antenna was evaluated since the shape and type of the antennas of the subject device is different from those of the predicate device. The evaluation was made using video gap occurrence as the indicator of the reception performance. One subject was excluded, and therefore twenty-nine subjects were eligible. The evaluation results met the predefined acceptance criteria. Thus, the reception performance of the subject device is equivalent to that of the predicate device. #### 10 Conclusion When compared to the predicate device, the OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM does not incorporate any significant changes in intended use, method of operation, material, or design that could affect the safety or effectiveness of the device. {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the symbol. The text is in all caps and is evenly spaced around the circle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 29, 2013 OLYMPUS MEDICAL SYSTEMS CORP. % Ms. Daphney Germain-Kolawole Regulatory Affairs Project Manager Olympus America, Inc. 3500 Corporate Parkway, P.O. Box 610 CENTER VALLEY PA 18034-0610 Re: K123421 Trade/Device Name: OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM Regulation Number: 21 CFR& 876.1300 Regulation Name: Ingestible telemetric gastrointestinal capsule imaging system Regulatory Class: II Product Code: NEZ Dated: March 15, 2013 Received: March 18, 2013 Dear Ms. Germain-Kolawole: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {8}------------------------------------------------ # Page 2 - Ms. Daphney Germain-Kolawolė Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to - http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours. Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {9}------------------------------------------------ 1/1 # Indications for Use 510(k) Number (if known): K123421 Device Name: OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM Indications For Use: The OLYMPUS SMALL INTESTINAL CAPSULE ENDOSCOPE SYSTEM is intended for visualization of the small intestine mucosa. The Red Color Detection Function is intended to mark frames of the video suspected of containing blood or red areas. Prescription Use______________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Benjamin R. Fisher S 2013.04.29 15:44:16 -04'00' Page 1 of __ l________________________________________________________________________________________________________________________________________________________________ K123421 12
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...