K123411 · Bionix Development Corp. · PDH · Mar 7, 2013 · General Hospital
Device Facts
Record ID
K123411
Device Name
LITTLE ANGELS SWADDLING BLANKET
Applicant
Bionix Development Corp.
Product Code
PDH · General Hospital
Decision Date
Mar 7, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5700
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Little Angels Swaddling Blanket developed and manufactured by Bionix Development Corp., Toledo, Ohio, is intended to be used as an adjunct to neonatal phototherapy systems used for the treatment of neonatal hyperbilirubinemia in the clinical or home setting. It is intended to be used by or under the direction of a licensed physician.
Device Story
Little Angels Swaddling Blanket is a spun bond polypropylene fabric sheet designed to swaddle neonates undergoing phototherapy for hyperbilirubinemia. The device features a die-cut, heat-welded pattern allowing for tight swaddling with only a single layer of fabric between the infant and the light source. It includes a rear pocket to hold light-emitting phototherapy pads. Used in clinical or home settings, the blanket facilitates phototherapy by maintaining infant position while ensuring high light transmittance (>90%) from overhead or pad-based light sources. The device is operated by healthcare providers or caregivers under physician direction. By securing the infant, it enables consistent phototherapy delivery, potentially improving treatment efficacy for neonatal jaundice.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by bench testing comparing light transmittance properties of the subject device against the predicate device using GE Giraffe SPOT PT LITE and Drager PT 4000 light sources.
Technological Characteristics
Material: Spun bond polypropylene. Form factor: Die-cut, heat-welded swaddling blanket with integrated pocket for light-emitting pads. Light transmittance: ≥90%. Fabric weight: 15.29 g/m2. Connectivity: None (passive). Sterilization: Not specified.
Indications for Use
Indicated for neonates with hyperbilirubinemia undergoing phototherapy treatment in clinical or home settings. Intended for use by or under the direction of a licensed physician.
Regulatory Classification
Identification
A neonatal phototherapy unit is a device used to treat or prevent hyperbilirubinemia (elevated serum bilirubin level). The device consists of one or more lamps that emit a specific spectral band of light, under which an infant is placed for therapy. This generic type of device may include supports for the patient and equipment and component parts.
Predicate Devices
BiliSoft Phototherapy System Disposable fiber optic pad covers (K053568)
Submission Summary (Full Text)
{0}------------------------------------------------
K123411
# 510(K) Summary:
MAR 0 7 2013
Image /page/0/Picture/3 description: The image shows the logo for Bionix. The logo consists of a stylized letter "B" inside a circle on the left, followed by the word "BIONIX" in bold, sans-serif font. A registered trademark symbol is located to the upper right of the "X".
Corporate Office 5154 Enterprise Blvd., Toledo, Ohio 43612 Phone 419.727.8421 · Fax 419.727.8426
October 25, 2012
The following 510(k) summary has been prepared pursuant to requirements specified in 21CFR 807.92
# 1. Submitter Information:
a. Applicant: Bionix Development Corporation 5154 Enterprise Blvd. Toledo, Ohio 43612
James Huttner M.D., Ph.D. b. Contact: Vice President, New Product Development Phone: (419) 727-8421 Fax: (419) 727-4430 Email: jhuttner@bionix.com
#### 2. Device Name:
- a. Trade Name: Little Angels Swaddling Blanket
- Swaddling blanket for use with neonatal phototherapy systems b. Common Name:
- c. Classification Name: Blanket, Neonatal Phototherapy (Per CFR section 880.5700)
### 3. Intended Use:
The Little Angels Swaddling Blanket is intended to be used as an adjunct to neonatal phototherapy systems used for the treatment of neonatal hyperbilirubinemia in the clinical or home setting.
Page 11
{1}------------------------------------------------
#### 4. Substantial Equivalence Device(s):
The Little Angels Swaddling Blanket is similar in composition and function to the following device used as an adjunct to current neonatal phototherapy systems used to treat neonatal hyperbilirubinemia.
- a. Disposable fiber optic pad covers use with the BiliSoft Phototherapy System manufactured and legally marketed by Lumitex, Inc., Strongsville, OH. This device is listed under regulation number 880.5700, Unit, Neonatal Phototherapy, and is classified as a Class II device. This device has been cleared by the FDA under K053568.
#### 5. Device Description:
The Little Angels Swaddling Blanket is intended to be used as an adjunct to neonatal phototherapy systems used for the treatment of neonatal hyperbilirubinemia in the clinical or home setting. The Little Angels Swaddling Blanket is similar in composition and function to other devices used as adjuncts to current neonatal phototherapy systems used to treat neonatal hyperbilirubinemia.
The Little Angels Swaddling Blanket is comprised of a sheet of spun bond polypropylene material that is die-cut and heat welded in a pattern that allows an infant undergoing phototherapy for hyperbilirubinemia to be tightly swaddled in a fashion such that only a single layer of material is ever interposed between the infant and the phototherapy source.
Spun bond polypropylene is a commonly used fabric in medical products and garments. Spun bond polypropylene is non-irritating to the skin. soft, and has a good drape that allows it to easily conform to body contours. Spun bond polypropylene is available in different fabric weights, depending on the tightness of the weave and fabric density. Light-weight spun bond polypropylene fabric is used in the Little Angels Swaddling Blanket to ensure maximum light transmittance from the phototherapy source to the patient's skin. {See section on "Bench Testing".)
In use, an infant undergoing phototherapy for hyperbilirubinemia is first laid atop the Little Angels Swaddling Blanket, and the blanket is then folded around the infant in a prescribed fashion that ensures the infant is tightly swaddled and that only a single layer of fabric lies between the phototherapy source and the infant's skin. When an overhead phototherapy source is used, the swaddled infant is then placed under the light source in the usual fashion in order to receive the phototherapy treatment. The Little Angels Swaddling Blanket also has a pocket formed in its rear panel (under the infant). If a light emitting pad type phototherapy source is to be used, the light emitting pad is placed in this pocket and phototherapy is administered to the swaddled infant in the usual fashion. Also, double phototherapy can be administered to an infant swaddled using the Little Angels Swaddling Blanket by combining an overhead phototherapy source together with a light emitting pad type phototherapy source, each used as described above.
{2}------------------------------------------------
# 6. Comparison to Predicate Devices:
The following table summarizes the comparison of Bionix Development Corporation Little Angel Swaddling Blanket to the predicate device:
Image /page/2/Figure/2 description: The image is a table comparing the attributes of two products: BiliSoft Phototherapy System Disposable Pad Covers (Lumitex, Inc.) and Little Angels Swaddling Blanket (Bionix). The table lists attributes such as intended use, composition, light transmittance, material weight, single patient use, pocket for fiberoptic phototherapy source, material between infant and phototherapy source, skin contact duration, and whether the product swaddles the infant. For example, the light transmittance for BiliSoft is 59% while it is >90% for Little Angels.
Table 1 Comparison of Substantial Equivalence
## 7. Bench Testing:
Samples of the spun bond polypropylene material used in the construction of the Little Angels Swaddling Blanket from Bionix, along with the spun bond polypropylene material used in the manufacture of the BiliSoft disposable pad covers, were tested for light transmittance properties using a light meter as described in the section titled "Bench Testing" of this document. Fabric weights of the materials used in each device were also determined. The results of these testing are summarized in the following tables:
{3}------------------------------------------------
| Phototherapy Light<br>Source | Material<br>Tested | Baseline<br>(no material) | Material<br>Test #1 | Material<br>Test #2 | Material<br>Test<br>(Average) | %<br>Transmittance |
|------------------------------|--------------------------------------|---------------------------|---------------------|---------------------|-------------------------------|--------------------|
| GE Giraffe SPOT PT<br>LITE | Little Angels<br>Blanket<br>(Bionix) | 36.9<br>µw/cm2/nm | 34.3<br>µw/cm2/nm | 33.4<br>µw/cm2/nm | 33.9<br>µw/cm2/nm | 91.7 % |
| Drager PT 4000 | Little Angels<br>Blanket<br>(Bionix) | 30.4<br>µw/cm2/nm | 27.4<br>µw/cm2/nm | 27.3<br>µw/cm2/nm | 27.4<br>µw/cm2/nm | 90.0 % |
| GE Giraffe SPOT PT<br>LITE | Billisoft Cover<br>(Lumitex) | 35.3<br>µw/cm2/nm | 19.4<br>µw/cm2/nm | 22.4<br>µw/cm2/nm | 20.9<br>µw/cm2/nm | 59.2 % |
| Drager PT 4000 | Billisoft Cover<br>(Lumitex) | 31.2<br>µw/cm2/nm | 18.0<br>µw/cm2/nm | 18.7<br>µw/cm2/nm | 18.4<br>µw/cm2/nm | 58.8 % |
# Table 2 Bench Test Results Light Transmittance Testing
| Device | Manufacturer | Maximum Light<br>Transmittance<br>(%) | Fabric Weight |
|-------------------------------------------------------|---------------|---------------------------------------|---------------|
| BiliSoft Phototherapy System<br>Disposable Pad Covers | Lumitex, Inc. | 59.21 % | 118 g/m2 |
| Little Angels<br>Swaddling Blanket | Bionix | ≥90% | 15.29 g/m2 |
# Table 3 Bench Test Results Light Transmittance and Material Weight
{4}------------------------------------------------
## 8. Clinical Testing:
No clinical testing was performed on the Little Angels Swaddling Blanket from Bionix Development Corporation.
### 9. Conclusion:
The similarity of composition, performance, and features indicate that the Little Angels Swaddling Blanket from Bionix Development Corporation will perform as well as the legally marketed disposable covers used with the BiliSoft Phototherapy System by Lumitex, Inc. for the intended use as an adjunct to neonatal phototherapy systems used for the treatment of neonatal hyperbilirubinemia.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### --- -- March-7, 201-3 ---
James Huttner, M.S., Ph.D. Vice President Bionix Development Corporation 5154 Enterprise Boulevard TOLEDO OH 43612
Re: K123411
Trade/Device Name: Blanket, Neonatal Phototherapy Regulation Number: 21 CFR 880.5700 Regulation Name: Neonatal Phototherapy Unit Regulatory Class: II Product Code: PDH Dated: January 9, 2913 Received: January 23, 2013
Dear Dr. Huttner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Dr. Huttner
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to -registration and-listing (21-CFR --Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
for
S
with
Urowitz
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use Statement:
510(k) Number (if known): K123411
Device Name: Blanket, Neonatal Phototherapy
Indications for Use:
The Little Angels Swaddling Blanket developed and manufactured by Bionix Development Corp., Toledo, Ohio, is intended to be used as an adjunct to neonatal phototherapy systems used for the treatment of neonatal hyperbilirubinemia in the clinical or home setting.
It is intended to be used by or under the direction of a licensed physician.
X Prescription Use _ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Consyrrence of CDRH, Office of In Vitro Diagnostic Devices (OVID)
Jacqueline
S. Ryan
| DN: | c=US, o=U.S. Government, |
|-------|------------------------------------|
| | ou=HHS, ou=FDA, ou=People, |
| | 0.9.2342.19200300.100.1.1=20005702 |
| | 93, cn=Jacqueline S. Ryan |
| Date: | 2013.03.05 10:33:03 -05'00' |
(Division Sign-Off)
Division of Anestheslology, General Hospital
Infection Control, Dental Devices
Page 1 of 1
| 510(k) Number: | 123411 |
|----------------|--------|
|----------------|--------|
Bionix Development Corporation
、一意看照
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.