K123375 · Siemens AG · LLZ · Nov 20, 2012 · Radiology
Device Facts
Record ID
K123375
Device Name
SYNGO.VIA
Applicant
Siemens AG
Product Code
LLZ · Radiology
Decision Date
Nov 20, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
syngo via is a software solution intended to be used for viewing, manipulation, communication, and storage of medical images. It can be used as a stand-alone device or together with a variety of cleared and unmodified syngo based software options. syngo via supports interpretation and evaluation of examinations within healthcare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments. The system is not intended for the displaying of digital mammography images for diagnosis in the U.S.
Device Story
syngo.via is a software-only PACS solution for viewing, manipulating, communicating, and storing medical images. Operated by clinicians in healthcare institutions (Radiology, Nuclear Medicine, Cardiology), it supports diagnosis and treatment planning. The system integrates with RIS via HL7/DICOM (IHE Year 5) to manage patient registration, scheduling, and worklists. It functions as a standalone device or with cleared syngo-based software options. The system provides a generic reader application for image display and post-processing. It does not include new imaging algorithms compared to the predicate. Healthcare providers use the output to assist in clinical decision-making and documentation. The system ensures data availability and security through a flexible storage hierarchy.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Software-only system; runs on common IT hardware meeting defined requirements. Backend: Windows 2008. Client: Windows XP, Vista, 7. Supports DICOM formatted images/objects. Connectivity: Networked via HL7/DICOM. Deployment: Installed on hardware; updates/upgrades via service contract.
Indications for Use
Indicated for healthcare professionals in Radiology, Nuclear Medicine, and Cardiology environments for viewing, manipulating, communicating, and storing medical images. Not indicated for digital mammography diagnosis in the U.S.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
syngo.x (K0925519)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary
# K123375
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 8807.92.
#### GENERAL INFORMATION I.
### Establishment:
- Address:
Siemens AG, Medical Solutions Henkestrasse 127 D-91052 Erlangen Germany
- . Registration Number: 3002808157
Contact Person:
Sabine Schroedel Regulatory Affairs Manager Telephone: +49 (9131) 84-8285 Fax: +49 (9131) 84-2792
### Device Name and Classification:
| • Trade Name: | syngo.via |
|-------------------------|---------------------------------------------|
| • Classification Name: | Picture Archiving and Communications System |
| • Classification Panel: | Radiology |
| • CFR Section: | 21 CFR §892.2050 |
| • Device Class: | Class II |
| • Product Code: | LLZ |
{1}------------------------------------------------
### II. SAFETY AND EFFECTIVENESS INFORMATION SUPPORTING THE SUB-STANTIAL EQUIVALENCE DETERMINATION
#### Device Description and Intended Use:
This premarket notification covers Siemens' enhanced PACS system syngo.via.
syngo.via is a software solution intended to be used for viewing, manipulation, communication, and storage of medical images. It can be used as a standalone device or together with a variety of cleared and unmodified syngo based software options. syngo via supports interpretation and evaluation of examinations within healthcare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments. The system is not intended for the displaying of digital mammography images for diagnosis in the U.S.
The system is a software only medical device. It defines minimum requirements to the hardware it runs on. The hardware itself is not seen as a medical device and not in the scope of this 510(k) submission.
It supports the physician in diagnosis and treatment planning. syngo.via also supports storage of Structured DICOM Reports.
In a comprehensive imaging suite syngo. Via integrates Radiology Information Systems (RIS) to enable customer specific workflows.
syngo.via allows to use a variety of advanced applications (clinical applications) designed for syngo.via just as the predicate device syngo.x (K0925519, cleared on August 27, 2009). These applications are medical devices on their own rights and filed separately. They are not part of this 510(k) submission and not part of the syngo. via medical device. syngo. via has a universal component called generic reader application which is part of this medical device and it allows no newly introduced imaging and post processing algorithms compared to the above mentioned predicate devices.
syngo.via is based on Windows. Due to special customer requirements and the clinical focus syngo.via can be configured in the same way as the predicate device with different combinations of syngo- or Windows -based software options and clinical applications which are intended to assist the physician in diagnosis and/or treatment planning. This includes commercially available post-processing software packages.
### syngo.via Data Management
... ensures all authorized personnel fast and continuous access to radiological data. It's main functionality ranges from availability of images with regard to data security, open interfaces, storage media and central system administration, to provide a flexible storage hierarchy.
#### Integration:
The Workflow Management enables by integration of any HL7- / DICOMcompatible RIS (IHE Year 5) to the syngo product family a consistent workflow -
{2}------------------------------------------------
from patient registration to requirement scheduling to a personal work list and supports therefore reporting, documentation or administrative tasks.
### Technological Characteristics:
syngo.via is a "software only"-system, which will be delivered on CD-ROM / DVD to be installed on common IT hardware. This hardware has to fulfil the defined requirements.
The backend communication and storage solution is based on Windows 2008 operating system. The client machines are based on Windows XP, Windows Vista and Windows 7. Any hardware platform, which complies to the specified minimum hardware and software requirements and with successful installation verification and validation activities can be supported.
The herewith described syngo. via supports DICOM formatted images and objects.
The syngo.via will be marketed as a software only solution for the end-user (with recommended hardware requirements). Any special needs such as integration in a specific environment and updates / upgrades will be covered by individual service contract and fulfilled by special trained service technicians.
syngo.via will be used for viewing, manipulation, communication, and storage of medical images. The predicate device syngo.x is also capable of viewing, manipulation, communication, and storage of medical images.
The difference between the syngo.via and the predicate device syngo.x are to give the subject device greater capabilities than the predicate device. syngo.via has similar technological characteristics as the predicate device and is similar to the functionalities of the predicate device, see table below:
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
November 20, 2012
Siemens AG Healthcare SY % Mr. Norbert Stuiber Responsible Third Party Official TÜV SÜD America, Inc. 1775 Old Hwy 8 NW, Ste 104 NEW BRIGHTON MN 55112-1891
Re: K123375
Trade/Device Name: syngo®.via Regulation Number: 21 CFR 2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: October 30, 2012 Received: November 1, 2012
Dear Mr. Stuiber:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - [Contact Person's Name]
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostics and Radiological Health at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
## Janine M. Morris -S
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### INDICATIONS FOR USE
510(k) Number (if known): K123375
Device Name: syngo®.via
#### Indications For Use:
syngo via is a software solution intended to be used for viewing, manipulation, communication, and storage of medical images.
It can be used as a stand-alone device or together with a variety of cleared and unmodified syngo based software options.
syngo via supports interpretation and evaluation of examinations within healthcare institutions, for example, in Radiology, Nuclear Medicine and Cardiology environments.
The system is not intended for the displaying of digital mammography images for diagnosis in the ഗ.ട.
AND / OR
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(Please do not write below this line - continue on another page if needed)
Concurrence of the CDRH, Office of In Vitto Diagnostic Devices (OIVD)
Димис Димив
D
Division Sign-Off Office of In Vitro Diaggostic Device Evaluation and Safety
510(k) K123325
Page 1 of 1
510(k) for syngo®.via
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.